Posey 2790 Series Twice As Tough Cuffs Instructions

July 13, 2024
Posey

Posey 2790 Series Twice As Tough Cuffs

Specifications

  • Product Name: Posey Twice-As-Tough Cuffs Limb Holder
  • Models Available: 2792 D-Ring Cuff (Regular/Blue), 2793 D-Ring Cuff (Large/Red), 2798 Lock Cuff (Regular/Blue), 2799 Lock Cuff (Large/Red)
  • Key Included: 1074 Posey Universal Key

Product Usage Instructions
Applying Limb Restraints

  1. Follow the steps to apply the device as instructed in the manual. Ensure the straps are securely attached to the frame.
  2. Make sure the locks click shut properly. Test the lock before leaving the patient’s side.

Applying the Cuffs

  1. Wrap the neoprene piece around the wrist/ankle as directed, ensuring proper alignment of the buckle and strap.
  2. Secure the hook-and-loop closure firmly and check for proper adherence.
  3. Adjust the strap length according to the patient’s comfort and safety.

Removing the Cuffs
To remove the cuffs, release them from the patient and detach the anchor strap from the bed frame.

Unlocking the Cuffs
Insert the Posey Key into the lock and turn counterclockwise to unlock. Avoid turning clockwise to prevent jamming.

FAQ
Q: Can the limb restraints be used for purposes other than on a bed or stretcher?
A: No, the padded limb restraints are designed for bed and stretcher use only.

Posey® Twice-As-Tough® Cuffs 2792, 2793, 2798, 2799 Application Instructions for Wrist or Ankle

DESCRIPTION OF PRODUCT: Padded limb restraints with double-security closure. For bed and stretcher use only.

INTENDED USE:

  • To protect patients assessed to be at risk of disrupting life-saving treatments (e.g., chronic tube pulling) or in danger of injury to themselves or to others by limiting limb movement in a bed or stretcher.
  • Follow your hospital’s restraint policies and procedures that are compliant with USA CMS guidelines and state laws, or other governing agencies outside the USA.

Posey-2790-Series-Twice-As-Tough-Cuffs- \(4\)

CONTRAINDICATIONS
None

APPLICATION INSTRUCTIONS
The order of either application or removal of the straps to the bed OR the cuff to the patient is to be determined by the clinician.

CAUTION
Posey Keys MUST be readily available to staff at all times. In an emergency, it is difficult to cut the cuffs with scissors to free the patient. Follow these steps to apply device (repeat steps 1-2 on each side):

  1. Bring the ends of the strap down through the inside Fig. 1 of the side rails so they do not interfere when the side rails are raised. Thread the end of the strap over the top, around the frame, at least once, and through the lock.
    Make sure that the straps are attached to the frame that moves with the patient (Fig. 1) (do not attach to side rail or head/footboard), out of the patient’s reach.

  2. Pull the strap(s) snug and close the lock. Check that the lock “clicks” shut. If a lock is not completely closed, it can pop open. Before leaving the patient’s side, test the lock by trying to open it without the key.

APPLYING THE CUFFS: (Repeat steps 1-3 for each side)

  1. Wrap the neoprene piece around wrist/ankle so the buckle and connecting strap is on the ulnar side of the wrist or lateral malleolus of the ankle. Attach the black hook-and-loop pieces together and pull back the black pull tab on the blue or red fuzzy piece, followed by the blue or red hook-and-loop pieces. The fuzzy piece should be sandwiched between the two pieces of hook (Fig. 2). Be sure to overlap at least one inch (3 cm).
  2.  Press the hook-and-loop closure together firmly and make sure it adheres securely. Slide ONE finger (flat) between the cuff and the inside of the patient’s wrist/ankle to ensure proper fit (Fig. 3).
  3. 2792/2793: Pass the end of the limb strap over the top of the cuff and through the two D-rings on the cuff. Bring
    the strap back over the first ring and through the second
    D-ring on the cuff (Fig. 4). Slide one finger (flat) between the cuff and limb strap. 2798/2799: Pass the end of the short strap through the lock and adjust it to the desired length. Check that the lock “clicks” shut. If a lock is not completely closed, it can pop open. Before leaving the patient’s side, test the lock by trying to open it without a key.

ADJUSTING STRAP LENGTH:
Adjust the connecting strap(s) length from the frame to allow desired freedom of movement without compromising patient or caregiver safety. Ensure the remaining strap end(s) are secured and out of the reach of the patient.

POSEY TWICE-AS-TOUGH CUFFS LIMB HOLDER

Posey-2790-Series-Twice-As-Tough-Cuffs- \(5\)

  • TO REMOVE: Release the cuff from the patient. Release the anchor strap from the bed frame.
  • TO UNLOCK THE CUFFS: Insert the Posey Key into the lock and turn counterclockwise. Turning clockwise may cause jamming (Fig. 5).
  • NOTE : Universal Posey Keys fit all Posey locks.

PRECAUTIONS

  • A clinical assessment and decision are required when used with monitoring lines or if patient has a wound or dislocated/fractured limb.
  • Check the patient regularly to ensure that circulation is not impaired. Serious injury may occur if the cuffs restricts circulation when the limb holder is applied.

WARNING: Additional or different body or limb restraints may be needed (visit tidiproducts.com):

  • Some patients may require additional interventions in conjunction with a restraint in order to prevent injury to self or others.
    • If the patient pulls violently against the bed straps.
    • To reduce the risk of the patient getting access to the line/wound/tube site.
    • To prevent the patient from flailing or bucking up and down and causing self-injury.
  • It is a clinical decision to decide when a restraint is no longer deemed safe.

BED SAFETY: Refer to the Food and Drug Administration (FDA) , or other governing agencies outside the USA, for the most recent Hospital Bed Safety Guidelines and the bed manufacturer’s instructions for use.

Report any serious incident related to device use to TIDI Products and the Member State Competent Authority. | Translations of this document in French, Spanish, Dutch, German, Italian, and Portuguese are located at www.tidiproducts.com/IFU

ADDITIONAL SAFETY AND LAUNDERING INSTRUCTIONS ON OTHER SIDE.

  • TIDI Products, LLC • 570 Enterprise Drive, Neenah, WI 54956 USA
    Phone: 1.800.521.1314 • International: +1.920.751.4036 • www.tidiproducts.com

  • © 2023 TIDI Products, LLC. All rights reserved. Posey is a registered trademark of TIDI Products, LLC.

  • MDSS GmbH Schiffgraben 41
    30175 Hannover, Germany

  • UK Responsible Person
    Emergo Consulting (UK) Limited c/o Cr360 – UL International Compass House, Vision Park Histon Cambridge CB249BZ
    United Kingdom

Safety Information for the use of Posey® Restraining Products

  • For RX PRODUCTS ONLY. Federal law (USA) restricts RX devices to sale by  or on order of a physician. For use in a licensed healthcare facility only.
  • STAFF TRAINING: TIDI offers in-service training aids at no charge. Contact TIDI online at www.tidiproducts.com or call toll-free at 1.800.521.1314.
  • BEFORE APPLYING ANY RESTRAINT: Follow your hospital’s restraint policies and procedures that are compliant with USA CMS guidelines and state laws or other governing agencies outside the USA.

ADDITIONAL WARNINGS

  1. ALWAYS monitor patient per facility policy. Improper application or use of any restraint may result in serious injury or death. Be aware that constant monitoring may be required for:
    • Aggressive or agitated patients; and
    • Patients in a supine position; ensure that being in this position would not cause any complications for those who are at risk of vomiting while being restrained.
  2.  NEVER alter or repair this product. ALWAYS Inspect before each use: Check for broken stitches or parts; torn, cut or frayed material; or locks, buckles, or hook-and-loop fasteners that do not hold securely. DO NOT use soiled or damaged products. Doing so may result in serious injury or death. Dispose of damaged products per facility policy for BIOHAZARDOUS material.
  3. ALWAYS secure straps to a part of the frame that moves with the patient (not foot/head board or side rails), out of the patient’s reach, using quick-release ties (see drawing below) or buckles. These allow easy release in the event of an accident or fire.

Test to make sure straps cannot tighten, loosen, or slip and create excess slack. If this occurs, the patient may slide off the chair or bed, increasing the risk of serious injury or suffocation. Restraint release is an important part of facility fire and disaster drills. Straps can be cut with scissors in an emergency.

  1. NEVER secure restraint strap to side rail or head/footboard.
  2. NEVER use Posey products on toilets, or on any chair or furniture that does not allow proper application as directed in the Application Instructions. DO NOT use at home.
  3.  NEVER expose this product to open flame, fire, smoking materials, or high heat sources. Some products may melt or ignite and burn. The facility smoking/no smoking policy should be strictly enforced.
  4. NEVER use a Posey product as a seat belt in a moving vehicle. Posey products are not designed to withstand the force of a crash or sudden stop.
  5. Test Zippers or hook-and-loop fasteners before each use. DISCARD device if it does not fasten securely.

LAUNDERING INSTRUCTIONS (if applicable)

  • Fasten all buckles and locks to reduce risk of damage during wash and dry cycles. DO NOT put buckles or locks through extractors. For maximum life, launder in a laundry bag.

  • Stainless Steel Locks: Posey recommends that locks be treated with a dryfilm lubricant (silicone) after each cleaning. For best results, spray a small amount into the lock. Work the action of the lock with a key several times to ensure that all the surfaces are well lubricated.

  • Before laundering, zip up and turn the product inside out to protect zipper.

  • Hook-and-loop fasteners may collect lint after repeated use or laundering, reducing grip strength. Fasten the “hook” to the “loop” before laundering to help prevent lint buildup. As needed, use a stiff-bristle brush to remove lint from the “hook” side.

  • These products, other than foam products, can be machine washed under CDC guidelines for material soiled with blood or bodily fluid

  • For non-contaminated material, use lower temperature wash and dry cycles to extend product life.

  • For foam products:

  • Make sure products are completely dry before use.

STORAGE AND HANDLING:

  • This device is designed for use in normal indoor environments.
  • This device may be stored in ambient warehouse temperatures at normal humidity levels. Avoid excess moisture or high humidity that may damage product materials.

HOW TO TIE THE POSEY QUICK-RELEASE TIE

Posey-2790-Series-Twice-As-Tough-Cuffs- \(1\)

  1. Wrap strap once around the frame that moves with the patient (do not attach to side rail or head/footboard), leaving at least an 8” (20 cm) tail. Fold the loose end in half to create a loop and cross it over the other end.
  2. Insert the folded strap where the straps cross over each other, as if tying a shoelace. Pull on the loop to tighten.
  3.  Fold the loose end in half to create a second loop.
  4. Insert the second loop into the first loop.
  5.  Pull on the loop to tighten. Test to make sure strap is secure and will not slide in any direction.
  6. Repeat on other side. Practice quick-release ties to ensure the knot releases with one pull on the loose end of the strap.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals