INVACARE IRC1001 Envoy Jr Aerosol Compressor
SPECIAL NOTES
DANGER WARNING/CAUTION notices as used in this manual apply to hazards or unsafe practices which could result in personal injury or property damage.
NOTICE
THE INFORMATION CONTAINED IN THIS DOCU. MENT IS SUBJECT TO CHANGE WITHOUT
NOTICE.
THIS AEROSOL COMPRESSOR IS INTENDED FOR HOUSEHOLD USE.
WARNING
DO NOT INSTALL, MAINTAIN OR OPERATE THIS EQUIPMENT WITHOUT FIRST READING
AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE
DANGERS/WARNINGS/ CAUTIONS AND INSTRUCTIONS, CON- TACT YOUR HOME HEALTHCARE
DEALER BEFOREATTEMPTING TO USE THIS EQUIP. MENT OTHERWISE INJURY OR DAMAGE MAY
RESULT.
SAVE THESE INSTRUCTIONS
IMPORTANT SAFEGUARDS
READ ALL INSTRUCTIONS BEFORE USING
Caution: Federal law restricts this device to sale by or on order of a
physician, or any other practitioner licensed by the law of the State in which
he practices to use or order the use of this device.
When using electrical products, especially when children are present, basic
safety precautions should always be followed, including the following
DANGER
TO REDUCE THE RISK OF ELECTROCUTION
- ALWAYS unplug this product IMMEDIATELY after using.
- DO NOT use while bathing.
- DO NOT place or store product where it can fall or be pulled into a tub or sink.
- DO NOT reach for product that has fallen into water.
- UNPLUG IMMEDIATELY. DO NOT place or drop into water or other liquid.
WARNING
TO REDUCE THE RISK OF BURNS, ELECTROCUTION, FIRE OR INJURY TO PERSONS
- This product should NEVER be left unattended when plugged in.
- Close supervision is necessary when this product is used by, on, or near CHILDREN AND PHYSICALLY CHALLENGED Individuals.
- Use this product only for its intended use as described in this manual. Use this product only under physician’s direction.
DO NOT use attachments not recommended by the manufacturer.
WARNING
- NEVER operate this product with a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water; call Qualified Service Personnel for examination and repair. Keep the cord away from HEATED or HOT surfaces.
- NEVER drop or insert any object into any opening
- NEVER use while sleeping or drowsy.
- NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air openings may be blocked. Keep the air openings free from lint, hair and the like. Blocked openings may cause the unit to shut down. Contact dealer immediately.
- DO NOT use outdoors or operate where aerosol (spray) products are being used or where oxygen is being administered in a closed environment such as an oxygen tent.
This unit is oil-less. DO NOT lubricate.
CAUTION
Risk of electric shock. Do not disassemble. Refer servicing to qualified
service personnel.
SPECIFICTAION AND SUPPLEMENTARY INFORMATION
STANDARD SPECIFICATIONS
SPECIFICATIONS AND SUPPLEMENTAL INFORMATION
ELECTRICAL This appliance has a polarized plug (one blade is wider than the other). As a safety feature, this plug will fit in a polarized outlet only one way. If the plug does not fit fully in the outlet, reverse the plug. If it still does not fit, contact a qualified electrician. Do not attempt to defeat this safety feature.
Compressor:
If the unit shuts down frequently, you may have a low voltage situation. Low
voltage can also be suspected when:
- The motor does not get up to full power or speed.
- Home fuses or circuit breakers activate when starting compressor.
- Lights dim or remain dim when compressor is started.
- Other motor operated appliances fail to operate properly or too many motor operated appliances are on same circuit.
The motor of this compressor has a thermal overload protector. If the motor should overheat, the overload protector will shut off the motor. If the unit shuts down and cannot be restarted, the motor must be replaced. Call your dealer immediately.
Extension Cords:
Make certain extension cord is in good condition. Make certain extension cord
wire is NOT LESS than 18 gauge for 25 feet, 16 gauge for 50 feet and 14 gauge
for 100 feet.
MAINTENANCE
Only authorized or factory-trained personnel should perform preventative
maintenance.
FEATURES
FIGURE 1- FEATURES
HANDLING PROCEDURES
SHIPPING AND TRANSPORTATION INSTRUCTIONS
The shipping container has been designed to assure protection of this device. If this device is to be reshipped by common carrier, it should be packed in the same carton. Extra cartons are available from Invacare.
UNPACKING
- Check for any obvious damage to the carton or its contents. If damage is evident, do NOT use. Notify Carrier/Invacare or your local Invacare Dealer.
- Remove all the loose packing from the carton.
- 3. Carefully remove all the components from the carton.
NOTE Retain carton and packing materials for use in shipping or storing the unit until use is required.
INSPECTION
- Examine exterior for nicks, dents, scratches or other damage.
Inspect all components.
STORAGE
- Store the repackaged device in a dry area
- DO NOT place other objects on top of the repackaged device.
OPERATION
Your disposable nebulizer package includes a nebulizer mouthpiece, nebulizer,
and air tubing. These items are sold as a package only.
NOTE: Before initial operation, nebulizer should be deaden. Follow the
recommended Cleaning instructions in this Owner’s Manual.
OPERATION
WARNING
DO NOT run unit without filter. Avoid operating this product in a dusty
environment which could cause premature compressor wear. SETUP
-
Place the device on a table so that the controls can be reached when seated.
-
Secure one end of the nebulizer tubing to the compressor air outlet (FIGURE I).
-
Connect the other end to the nebulizer.
-
FIGURE I- LOCATION OF COMPRESSOR AIR OUTLET
WARNING
In the case of high humidity, condensation may form inside the internal tubing. Run the unit for two (2) minutes before connecting the nebulizer to the tubing. -
Add liquid medication to the nebulizer according to your physician’s prescription (FIGURE 2). WARNING Before plugging in the line cord, make sure that the power switch is in the OFF (O) position.
-
Unwrap the line cord.
-
Plug into a l I5VI60Hz wall outlet.
-
Insert the nebulizer mouthpiece into the top of the nebulizer.
NOTE: Before attempting to start a cold unit, let the unit reach room temperature in a heated area. -
Press the power switch ON to start the compressor.
-
Place mouthpiece between your teeth and close your mouth (FIGURE 3).
-
T0. As aerosol begins to flow, inhale deeply, slowly, and hold breath for 10 seconds. Then exhale slowly through the mouthpiece. NOTE: Before storing supplies between treatments, clean and dry nebulizer. (Refer to Maintenance Section in this manual). Tighten caps on medication jars. NOTE: In the event a substantial increase in back pressure is achieved, ie. kinked tubing, an audible “pop” or air leak may be heard. A relief valve has been included to reduce possible damage to the compressor; which doses once the back pressure has decreased.
MAINTENANCE
WARNING Electrical shock hazard. DO NOT remove outer case of this unit. All disassembly and maintenance of this unit MUST BE done by a qualified Invacare Dealer.
CLEANING THE OUTER CASE
WARNING UNPLUG unit before cleaning. DO NOT submerge in water to clean. Clean outer case with damp cloth.
CLEANING THE NEBULIZER
NOTE: With proper maintenance your nebulizer should last 30 days. Since the nebulizer is disposable, it is recommended that an extra nebulizer be kept at all times.
NOTE: Clean the Nebulizer after every use and at least once a day.
NOTE: Your physician and/or a Invacare Dealer may specify certain cleaning procedures so, follow their recommendations. If not, clean according to the following instructions:
-
Remove the nebulizer and mouthpiece from the tubing.
-
Disassemble nebulizer chamber by turning counterclockwise and separating8
-
Fill two (2) plastic containers or bowls:
A. Washing: Hot water/detergent solution.
B. Soaking: Hot water/vinegar solution (one [) part vinegar to three 3] parts water). -
Thoroughly clean the nebulizer and mouthpiece in hot water detergent solution.
-
Remove from the water/detergent solution and rinse which clear hot tap water.
-
Soak in hot water/vinegar solution for 30 minutes.
-
Rinse with hot tap water again and air dry thoroughly.
NOTE fusing medical disinfectant cleaners, follow manufacturer’s instructions carefully. -
Keep the outer surface of the tubing dust-free by wiping regularly.
NOTE: The nebulizer air tubing does not have to be washed internally because only filtered air passes through the tubing
CLEANING OR REPLACING THE FOAM INTAKE FILTER
NOTE: Clean the filter once a month or sooner if fliter discolors.
- Unplug unit.
- Remove filter cover to gain access to filter.
- Remove filter and clean with soap and water. Let dry thoroughly before use.
- Replace filter if it becomes clogged, torn or has a worn appearance.
NOTE: Regular filter cleaning/replacing is necessary to help ensure proper compressor performance.
WARNING
DO NOT run unit without filter
ACCESSORIES
NOTE: The nebulizer and accessories are not manufactured by Invacare. Invacare has maintained the integrity of the nebulizer manufacturers original sealed packaging and labeling.
Intake Filter/ Filter Cover Owner’s Manual
MS2100 Disposable Nebulizer (5 ml Capacity), w/Mouthpiece, T-Piece, and 7 ft.
tubing
MS2200 Disposable Side stream Nebulizer (10 ml Capacity), w/Mouthpiece and 7
ft. Tubing.
MS2400 Durable Side stream Nebulizer (10 ml Capacity), w/Mouthpiece and 7 ft.
Tubing. One (1) Year Limited Warranty. IRCI198 TruZone° PFM Peak Flow Meter
NOTE: See Invacare Dealer for additional Nebulizers.
WARRANTY
LIMITED WARRANTY
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH
FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4,1975.
This warranty is extended only to the original purchaser of our products. This
warranty gives you specific legal rights and you may also have other legal
rights which vary from state to state.
Invacare warrants the compressor of the Invacare Aerosol Com- pressor to be free from defects in materials and workmanship for a period of three (3) years from date of purchase. The nebulizer chamber, nebulizer air tubing, and any other accessories are excluded in this warranty. If within such warranty period any such product shall be proven to Invacare’s satisfaction to be defective, such product shall be repaired or replaced, at Invacare’s option. This warranty does not include normal wear and tear or shipping charges incurred in replacement part installation or repair of such product. Invacare’s sole obligation and your exclusive remedy under this warranty shall be limited to such repair or replacement.
For warranty service, please contact Invacare’s Service Department during normal working hours at the Toll Free Number listed.
Upon receiving notice of an alleged defect in a product, Invacare will issue a serialized return authorization. It shall then be the responsibility of the purchaser to return, at purchaser’s cost, to either Invacare’s plant or service center as specified by Invacare in advance. Defective component part(s) must be returned for warranty inspection using the serial number as identification within thirty (30) days of return authorization date. DO NOT return products to our factory without prior consent. C.O.D shipments will be refused-please prepay shipping charges.
Limitations and Exclusions: The foregoing warranty shall not apply to products subjected to negligence, accident, improper operation, maintenance or storage, commercial or institutional use, products modified without Invacare’s express written consent (including, but not limited to, modification through the use of unauthorized parts or attachments) or to products damaged by reason of repairs made to any component without the specific consent of Invacare, or to a product damaged by circumstances beyond Invacare’s control.
The foregoing warranty is exclusive and In lieu of all other express warranties. Implied warranties, if any, including the implied warranties of merchantability and fitness for a particular purpose, shall not extend beyond the duration of the expressed warranty provided herein and the remedy for violations of any implied warranty shall be limited to repair or replacement of the defective product pursuant to the terms contained herein. Invacare shall not be liable for any consequential or incidental damages whatsoever.
Some states do not allow the exclusion or limitation of incidental or consequential damage, or limitation on how long an implied warranty lasts, so the above exclusion and limitation may not apply to you.
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE/PROVINCIAL LAWS AND REQUIREMENTS.