TROTEC BT22WP Smartphone Controllable Thermometer
Product Information
Specifications:
- Model: BT22WP
- Type: Smartphone-controllable thermometer
- Compatible with Trotec MultiMeasure Mobile app
- Measures air temperature, non-aggressive gases, andliquids
- Data logging and display capabilities
- Approved accessories and spare parts by Trotec
Product Usage Instructions
Safety Precautions:
Read the manual before use and store it near the device.
Intended Use:
Use with a compatible terminal device and Trotec MultiMeasure Mobile app for air temperature and non-aggressive gas/liquid measurements.
Foreseeable Misuse:
- Avoid use in explosive atmospheres or near medical equipment.
- Maintain a 20cm distance from pacemakers.
- Avoid interference with alarm systems and automatic doors.
Operation:
- Ensure the device is connected to a compatible terminal with the app installed.
- Select the type of measurement (air temperature, gas, or liquid).
- Follow on-screen instructions for data logging and display options.
Maintenance:
Regularly clean the device using a soft, dry cloth. Do not expose to liquids or extreme temperatures.
Errors and Faults:
If the device malfunctions, refer to the troubleshooting section in the manual or contact customer support.
Frequently Asked Questions
- Q: Can I use the device for measuring aggressive liquids?
- A: No, only use the device for non-aggressive gases and liquids as stated in the manual.
- Q: How far should I keep the device from pacemakers?
- A: Maintain a minimum distance of 20cm between the device and pacemakers.
- Q: What should I do if the device interferes with other electronic equipment?
- A: Ensure no other devices are malfunctioning during use and avoid using near-sensitive equipment.
Notes regarding the instructions
Symbols
You can download the current version of the instructions and the EU declaration of conformity via the following link
Safety
Read this manual carefully before starting or using the device. Always store the manual in the immediate vicinity of the device or its site of use
Warning
Read all safety warnings and all instructions. Failure to follow the warnings and instructions may result in electric shock, fire and/or serious injury.
Save all warnings and instructions for future reference.
- Do not use the device in potentially explosive rooms or areas and do not install it there.
- Do not use the device in aggressive atmosphere.
- Protect the device from permanent direct sunlight.
- Do not remove any safety signs, stickers or labels from the device. Keep all safety signs, stickers and labels in legible condition.
- Do not open the device.
- Never charge batteries that cannot be recharged.
- Different types of batteries and new and used batteries must not be used together.
- Insert the batteries into the battery compartment according to the correct polarity.
- Remove discharged batteries from the device. Batteries contain materials hazardous to the environment. Dispose of the batteries according to the national regulations.
- Remove the batteries from the device if you will not be using the device for a longer period of time.
- Never short-circuit the supply terminal in the battery compartment!
- Do not swallow batteries! If a battery is swallowed, it can cause severe internal burns within 2 hours! These burns can lead to death!
- If you think batteries might have been swallowed or otherwise entered the body, seek medical attention immediately!
- Keep new and used batteries and an open battery compartment away from children.
- Observe the storage and operating conditions (see Technical data).
Intended use
Use this device in combination with a terminal device that is compatible with the installed Trotec MultiMeasure Mobile app. Only use the device to measure the air temperature and the temperature of non-aggressive gases and non- aggressive liquids. The Trotec MultiMeasure Mobile app on the terminal device is used for both operation and the evaluation of the measured values. Data logged by the device can be displayed, saved or transmitted either numerically or in form of a chart. To use the device for its intended use, only use accessories and spare parts which have been approved by Trotec.
Foreseeable misuse
Do not use the device in potentially explosive atmospheres, for measurements in liquids or on live parts. Radio waves may interfere with the operation of medical equipment and cause malfunctions. Do not use the device near medical equipment or within medical institutions. Persons with pacemakers must observe a minimum distance of 20 cm between the pacemaker and the device. Also do not use the device near automatically controlled systems such as alarm systems and automatic doors. Radio waves may interfere with the operation of such equipment and cause malfunctions. Make sure that no other devices are malfunctioning during the use of your device. Any unauthorised changes, modifications or alterations to the device are forbidden.
Personnel qualifications
People who use this device must:
- have read and understood the instructions, especially the Safety chapter.
Safety signs and labels on the device
Note
Do not remove any safety signs, stickers or labels from the device. Keep all safety signs, stickers and labels in legible condition.
The following safety signs and labels are attached to the device:
Residual risks
-
Warning of electrical voltage
There is a risk of a short-circuit due to liquids penetrating the housing! Do not immerse the device and the accessories in water. Make sure that no water or other liquids can enter the housing. Warning of electrical voltage Work on the electrical components must only be carried out by an authorised specialist company! -
Warning
Magnetic field!
The magnet attachment can affect pacemakers and implanted defibrillators! Always keep a minimum distance of 20 cm between the device and pacemakers or implanted defibrillators. Persons with pacemakers or implanted defibrillators must not carry the device in their breast pocket. -
Warning
Risk of damages or data loss due to magnetic field! Do not store, carry or use the device in the vicinity of data storage media or electronic devices such as hard drives, television units, gas meters or credit cards! There is a risk of data loss or damages. If possible, keep the highest safety distance possible (at least 1 m). -
Warning
Risk of suffocation!
Do not leave the packaging lying around. Children may use it as a dangerous toy. -
Warning
The device is not a toy and does not belong in the hands of children. -
Warning
Dangers can occur at the device when it is used by untrained people in an unprofessional or improper way! Observe the personnel qualifications! -
Caution
Keep a sufficient distance from heat sources. -
Note
To prevent damages to the device, do not expose it to extreme temperatures, extreme humidity or moisture. -
Note
Do not use abrasive cleaners or solvents to clean the device.
Information about the device
Device description
Used in combination with Trotec’s MultiMeasure Mobile app the thermometer permits air temperature measurements. In case of individual measurements, the measurement value display can be refreshed both via the app and by a brief actuation of the measurement button at the measuring device. Apart from the hold function, the measuring device can indicate minimum, maximum and average values and carry out series measurements. In the app you can specify MAX and MIN alarm thresholds for all \parameters measured with the device. The measurement results can be displayed and saved on the terminal device either numerically or in form of a chart. Then, the measurement data can be sent in PDF or Excel format. The app also includes a report generation function, an organiser function, one for customer management and further analysis options. Moreover, it is possible to share measurements and project data with colleagues in another subsidiary. If MultiMeasure Studio Professional is installed on a PC, you can even use report templates and ready-made text blocks for various fields of application to turn the data into professional reports.
Device depiction
No. | Designation |
---|---|
1 | Measuring sensor |
2 | LED |
3 | On / off / measurement button |
4 | Battery compartment with cover |
5 | Lock |
Technical data
Parameter | Value |
---|---|
Model | BT22WP |
Temperature
Measuring range| -50 to 150 °C
Accuracy| ±1 °C
Measuring range resolution| 0.1 °C
General technical data
Bluetooth standard| Bluetooth 4.0, Low Energy
Bluetooth frequency range| 2.4 GHz
Bluetooth max. transmission power| 10 dBm
Radio range| approx. 10 m
(depending on measuring environment)
Operating temperature| -20 to 50 °C, < 80 % RH
Storage temperature| -10 to 60 °C, 10 to 90 % RH
Power supply| 3 x 1.5 V battery AAA
Device switch-off| after approx. 3 minutes without active Bluetooth connection
Type of protection| IP40
Weight| 90 g (incl. batteries)
Dimensions
(length x width x height)
| 220 mm × 38 mm × 23 mm
Scope of delivery
- 1 x Device BT22WP
- 3 x 1.5 V battery AAA
- 1 x Wrist strap
- 1 x Quick guide
Transport and storage
Note
If you store or transport the device improperly, the device may be damaged.
Note the information regarding transport and storage of the device.
Warning
Risk of damages or data loss due to magnetic field! Do not store, carry or use the device in the vicinity of data storage media or electronic devices such as hard drives, television units, gas meters or credit cards! There is a risk of data loss or damages. If possible, keep the highest safety distance possible (at least 1 m).
Transport
When transporting the device, ensure dry conditions and and protect the device from external influences e.g. by using a suitable bag.
Storage
When the device is not being used, observe the following storage conditions:
- dry and protected from frost and heat
- protected from dust and direct sunlight
- the storage temperature complies with the values specified in the Technical data
- Batteries are removed from the device
Operation
Inserting the batteries
Note
Make sure that the surface of the device is dry and the device is switched off.
- Unlock the battery compartment by turning the lock (5) in a way that the arrow points towards the opened padlock icon.
- Remove the cover from the battery compartment (4).
- Insert the batteries (3 batteries of type AAA) into the battery compartment with correct polarity.
- Put the cover back onto the battery compartment.
- Lock the battery compartment by turning the lock (5) in a way that the arrow points towards the closed padlock icon.
MultiMeasure Mobile app
Install the Trotec MultiMeasure Mobile app on the terminal device you want to use in combination with the device.
Info
Some of the app’s functions require access to your location and an active Internet connection. The app is available for download in the Google Play Store as well as in Apple’s app store and via the following link:
Start-up
Info
Allow for an acclimatization period of about 10 minutes in the respective measuring environment prior to the appSensor’s measuring operation.
Connecting the appSensor
Info
The app can simultaneously be connected to several different appSensors or
appSensors of the same type and also record several measurements at the same
time.
Proceed as follows to connect the appSensor to the terminal device:
- The Trotec MultiMeasure Mobile app is installed.
- The Bluetooth function on your terminal device is activated.
-
Start the Trotec MultiMeasure Mobile app on the terminal device.
-
Briefly actuate the On / off / measurement button (3) three times to switch on the appSensor.
- The LED (2) flashes yellow.
-
Press the Sensors button (6) on the terminal device.
- The sensors overview opens.
-
Press the Refresh button (7).
- If the scanning mode was not active before, the colour of the Refresh button (7) will change from grey to black.
The terminal device now scans the surroundings for all available appSensors.
- If the scanning mode was not active before, the colour of the Refresh button (7) will change from grey to black.
-
Press the Connect button (8) to connect the desired sensor to the terminal device.
- The LED (2) flashes green.
- The appSensor is connected to the terminal device and starts measuring.
- The on-screen display changes to the continuous measurement.
No. | Designation | Meaning |
---|---|---|
6 | Sensors button | Opens the sensors overview. |
7 | Refresh button | Refreshes the list of sensors near the terminal device. |
8 | Connect button | Connects the displayed sensor to the terminal device. |
Continuous measurement
Info
Note that moving from a cold area to a warm area can lead to condensation forming on the device’s circuit board. This physical and unavoidable effect can falsify the measurement. In this case, the app will either display incorrect measured values or none at all. Wait a few minutes until the device has become adjusted to the changed conditions before carrying out a measurement. When the appSensor has been successfully connected to the terminal device, a continuous measurement is started and indicated. The refresh rate is 1 second. The 12 most recently measured values are displayed graphically (9) in chronological sequence. The currently determined and calculated measured values are displayed numerically (10).
No. | Designation | Meaning |
---|---|---|
9 | Graphic display | Indicates the temperature as measured over the course of |
time.
10| Numeric temperature display| Indicates minimum, maximum and average value
as well as the current value.
11| Menu button| Opens the menu to adjust the settings of the current
measurement.
Info
The indicated measured values will not be saved automatically.
Info
By tapping on the graphic display (9) you can switch over to the numeric display and vice versa.
Measurement settings
Proceed as follows to adjust the settings for the measurement:
- Press the Menu button (11) or the free area below the measured value display.
- The context menu opens.
- Adjust the settings as required.
No. | Designation | Meaning |
---|---|---|
12 | Reset min / max / Ø button | Deletes the determined values. |
13 | X/T measurement |
button
| Switches between continuous measurement and individual measurement.
14| Disconnect sensor
button
| Disconnects the connected appSensor from the terminal device.
15| Sensor settings
button
| Opens the context menu for the connected appSensor.
16| Start recording
button
| Starts a recording of the determined measured values for later evaluation.
Individual value measurement
Proceed as follows to select the individual value measurement as measuring mode:
- Press the Menu button (11) to open the context menu for sensors.
- Press the X/T measurement button (13) to switch from continuous measurement to individual value measurement.
- Individual value measurement has been selected as measuring mode.
- Return to the screen displaying the measured values.
- The first measured value is automatically determined and displayed.
No. | Designation | Meaning |
---|---|---|
17 | Individual value indication | Indicates the value for temperature |
determined at the time of the measurement.
18| Numeric temperature display| Indicates minimum, maximum and average value
as well as the current value.
19| Refresh measured value button| Performs an individual value measurement
and refreshes the displays (17) and (18).
Refreshing the measured value
Proceed as follows to refresh the measured values in the individual value measurement mode:
- Press the Refresh measured value button (19) on the terminal device.
- The appSensor determines the current measured value which is then displayed on the terminal device.
- You can also press the On / off / measurement button (3) on the appSensor.
- The appSensor determines the current measured value which is then displayed on the terminal device.
Recording measured values
Proceed as follows to record measured values for later evaluation:
- Press the Menu button (11) or the free area below the measured value display.
- The context menu for sensors opens.
- Press the Start recording button (16).
- The REC button (20) replaces the Menu button (11).
- If you perform a continuous measurement, the measuredvalues determined from then on will be recorded.
- If you perform individual value measurements, repeatedly press the On / off / measurement button (3) on the appSensor or the Refresh measured value button (19) on the terminal device until you have logged all the required measured values.
No. | Designation | Meaning |
---|---|---|
20 | REC button | Opens the sensor settings menu. |
21 | Stop recording |
button
| Stops the current recording of measured values.
Opens the submenu for saving recordings.
Stopping a recording
Proceed as follows to stop recording the measured values:
- Press the REC button (20).
- The context menu for sensors opens.
- Press the Stop recording button (21).
- The context menu for saving the recording opens.
- You can optionally save, discard or resume the measurement.
Saving a recording
Proceed as follows to save the recorded measured values:
- Press the Save button (22) to save the recorded measured values on the terminal device.
- The input mask for logging the recorded data opens.
- Enter all the data relevant for an unambiguous assignment, then save the recording.
- The recording will be saved on the terminal device.\
No. | Designation | Meaning |
---|---|---|
22 | Save button | Stops the current recording of measured values. |
Opens the input mask for logging recording data.
23| Discard button| Stops the current recording of measured values.
Discards the recorded measured values.
24| Continue button| Resumes the recording of the measured values without saving.
Analysing measurements
Proceed as follows to call up the saved measurements:
- Press the Measurements button (25).
- An overview of already saved measurements will bedisplayed.
- Press the Display measurement button (27) for the desired measurement to be indicated.
- A context menu for the selected measurement opens.
No. | Designation | Meaning |
---|---|---|
25 | Measurements |
button
| Opens the overview of saved measurements.
26| Indication of the date of the measurement| Indicates the date on which
the measurement was recorded.
27| Display measurement button| Opens the context menu for the selected
measurement.
28| Indication of the number of measured values| Indicates the number of
individual measured values constituting the saved measurement.
The following functions can be called up in the context menu of the selected measurement:
No. | Designation | Meaning |
---|---|---|
29 | Basic data button | Opens an overview of the data saved for the |
measurement.
30| Evaluations button| Opens an overview of the evaluations generated for
the measurement (graphics and tables).
31| Evaluation parameters button| Opens a menu to select and deselect
individual evaluation parameters.
32| Values button| Opens a tabular overview of all the values logged for
the measurement.
33| Generate table
button
| Creates a table containing the logged values of the measurement and saves it
as a *.CSV file.
34| Generate graphic
button
| Creates a graphic representation of the logged values and saves it as a
*.PNG file.
Info
If you have saved a previous measurement with certain parameters and then realize, that some parameters are missing, you can subsequently edit them via the menu item Evaluation parameters. They will not be added to the already saved measurement, to be sure, but if you save the measurement again with a different name, these parameters will be added to the initial measurement.
Generating a report
The reports generated in the MultiMeasure Mobile app are short reports providing a fast and simple documentation. Proceed as follows to generate a new report:
- Press the Reports button (35).
- The reports overview opens.
- Press the New report button (36) to create a new report.
- An input mask for entering all the relevant information opens.
- Enter the information via the input mask and save the data.
No. | Designation | Meaning |
---|---|---|
35 | Reports button | Opens the overview of saved reports. |
36 | New report button | Creates a new report and opens an input mask. |
Info
The customer can acknowledge the report directly in the integrated signature field.
Calling up a report
Proceed as follows to call up a created report:
- Press the Reports button (35).
- The reports overview opens.
- Press the corresponding button (37) to display the desired report.
- An input mask opens in which you can view and edit all the information.
No. | Designation | Meaning |
---|---|---|
37 | Display reports |
button
| Opens the selected report.
Creating a new customer
Proceed as follows to create a new customer:
- Press the Customers button (38).
- The customers overview opens.
- Press the New customer button (39) to create a new customer.
- An input mask for entering all the relevant information opens.
- Enter the information via the input mask and save the data.
- Alternatively, you can also import existing contacts from the phone book of the terminal device
Info
You can perform a new measurement directly from the input mask.
Calling up customers
Proceed as follows to call up an already created customer:
- Press the Customers button (38).
- The customers overview opens.
- Press the corresponding button (40) to display the desired customer’s details.
- An input mask opens in which you can view and edit all the information for the selected customer as well as directly start a new measurement.
- The New customer button (39) changes. In this menu it can be used to delete the selected customer data record.
App settings
Proceed as follows to make settings in the Trotec MultiMeasure
Mobile app:
- Press the settings button (41).
- The settings menu opens.
- Adjust the settings as required.
appSensor settings
Proceed as follows to adjust the settings for the appSensor:
- Press the Sensors button (6).
- A list of connected and available sensors will be displayed.
- Select the line with the appSensor the settings of which you want do adjust and swipe right at the yellow marking.
- Confirm your input.
- The sensor menu opens.
- Alternatively, you can press the Sensors button (6).
- Press the Menu button (11).
- The context menu opens.
- Press the Sensor settings button (15).
- The sensor menu opens
Disconnecting an appSensor
Proceed as follows to disconnect an appSensor from the terminal device:
- Press the SENSORS button (6).
- A list of connected and available sensors will be displayed.
- Select the line with the appSensor to be disconnected and swipe left at the red marking.
- Confirm your input.
- The appSensor is now disconnected from the terminal device and can be switched off.
- Alternatively, you can press the Menu button (11).
- The context menu opens.
- Press the Disconnect sensor button (14).
- Confirm your input.
- The appSensor is now disconnected from the terminal device and can be switched off.
Switching off an appSensor
Info
Always terminate the connection between appSensor and app before you switch off the appSensor.
Proceed as follows to switch off an appSensor:
- Press and hold the On / off / measurement button (3) for approx. 3 seconds.
- The LED (2) on the appSensor goes out.
- The appSensor is switched off.
- You may now exit the Trotec MultiMeasure Mobile app on the terminal device.
Maintenance and repair
Battery change
A battery change is required when the LED at the device flashes red or the device can no longer be switched on. See chapter Operation.
Cleaning
Clean the device with a soft, damp and lint-free cloth. Make sure that no moisture enters the housing. Do not use any sprays, solvents, alcohol-based cleaning agents or abrasive cleaners, but only clean water to moisten the cloth.
Repair
Do not modify the device or install any spare parts. For repairs or device testing, contact the manufacturer.
Errors and faults
The device has been checked for proper functioning several times during production. If malfunctions occur nonetheless, check the device according to the following list.
Bluetooth connection is terminated or interrupted
- Check whether the LED at the appSensor flashes green. If so, briefly switch it off completely, then turn it back on. Establish a new connection to the terminal device.
- Check the battery voltage and insert new or freshly charged batteries, if required.
- Does the distance between appSensor and terminal device exceed the appSensor’s radio range (see chapter Technical data) or are there any solid building parts (walls, pillars etc.) situated between appSensor and terminal device? Shorten the distance between the two devices and ensure a direct line of sight.
The sensor cannot be connected to the terminal device although it is displayed there.
- Check the Bluetooth settings of your terminal device. A possible reason for this could be special, manufacturerspecific settings relating to an improved location accuracy.
Enable these settings, then try to establish a connection to the sensor again. Further information and assistance regarding the used sensor type will be provided in the MultiMeasure Mobile app via the menu item Settings => Help. Selecting the menu item Help opens a link to the app’s help page. You can open a drop-down menu with numerous support entries from the Table of contents. Optionally, you can also scroll through the entire help page and thoroughly acquaint yourself with the individual help topics.
Disposal
Always dispose of packing materials in an environmentally friendly manner and in accordance with the applicable local disposal regulations. The icon with the crossed-out waste bin on waste electrical or electronic equipment is taken from Directive 2012/19/EU. It states that this device must not be disposed of with the household waste at the end of its life. You will find\ collection points for free return of waste electrical and electronic equipment in your vicinity. The addresses can be obtained from your municipality or local administration. You can also find out about other return options that apply for many EU countries on the website https://hub.trotec.com/?id=45090. Otherwise, please contact an official recycling centre for electronic and electrical equipment authorised for your country. The separate collection of waste electrical and electronic equipment aims to enable the re-use, recycling and other forms of recovery of waste equipment as well as to prevent negative effects for the environment and human health caused by the disposal of hazardous substances potentially contained in the equipment. In the European Union, batteries and accumulators must not be treated as domestic waste, but must be disposed of professionally in accordance with Directive 2006/66/EC of the European Parliament and of the Council of 6 September 2006 on batteries and accumulators. Please dispose of batteries and accumulators according to the relevant legal requirements.
Only for United Kingdom
According to Waste Electrical and Electronic Equipment Regulations 2013 (SI 2013/3113) (as amended) and the Waste Batteries and Accumulators Regulations 2009 (SI 2009/890) (as amended), devices that are no longer usable must be collected separately and disposed of in an environmentally friendly manner.
Declaration of conformity
We – Trotec GmbH – declare in sole responsibility that the product designated below was developed, constructed and produced in compliance with the requirements of the EU Radio Equipment Directive in the version 2014/53/EU.
- Product model / Product: BT22WP
- Product type: smartphone-controllable
- thermometer Year of manufacture as of: 2022
Relevant EU directives:
- 2011/65/EU
- 2015/863/EU
Applied harmonised standards:
- EN 300 328 V2.2.2
- EN 61010-1:2010/A1:2019-02
Applied national standards and technical specifications:
-
Regulation (EC) 1907/2006
-
EN 301 489-1 Draft Version 2.2.0:2017-03
-
EN 301 489-17 Draft Version 3.2.0:2017-03
EN 50663:2017-10 -
EN 61010-1:2010/A1:2019-02
-
EN 61010-1:2010
-
EN 61326-1:2013
-
EN 61326-2-2:2013
-
EN 62479:2010
-
IEC 62321-3-1:2013
-
IEC 62321-6:2015
-
IEC 62321-7-1:2015
-
IEC 62321-7-2:2017
-
IEC 62321-8:2017
Manufacturer and name of the authorised representative of the technical documentation:
- Trotec GmbH
- Grebbener Straße 7, D-52525 Heinsberg
- Phone: +49 2452 962-400
- E-mail: info@trotec.de
Place and date of issue: Heinsberg, 17.02.2023
Joachim Ludwig, Managing Director
Contact
- Trotec GmbH
- Grebbener Str. 7
- D-52525 Heinsberg
- +49 2452 962-400
- +49 2452 962-200
- info@trotec.com
- www.trotec.com