etac 1a18702.2 Pressure Care Seat Instruction Manual

June 11, 2024
Etac

etac 1a18702.2 Pressure Care Seat

Product Information

Specifications

  • Brand: Star by Etac
  • Model Number: 1a18702.2
  • Manufacture Date: 2023-11-27
  • Reconditioning Instruction Language: English (en) and Danish (da)
  • Contact Information: Etac Supply Center Box 203 SE-334 33 Anderstorp
    Tel: +46 371 58 73 00
    Fax: +46 371 58 73 90
    Website: www.etac.com

Product Usage Instructions

  • StarLock, StabilAir, and Standard Air can be re-used.
  • Only official spare parts from Etac should be used.
  • The service life of the product is limited. This is crucial when prescribing the product for long-term use by a user.
  • Remove the cover before cleaning or disinfecting the cushion.

Handwash: Clean the cushion by hand with soap, bicarbonate, or vinegar. Rinse in cold water, then overinflate the cushion and air dry for 24 hours or dry in a drying cabinet on low heat.

StarLock: If water is in the bulb pump after machine washing, remove and drain the water. Reassemble the bulb pump.

  • Inspect the cushion for damage or puncture. Check for leaks by overinflating, closing the valve, placing it under water, and watching for air bubbles.
  • Mark damaged areas for repair following instructions on the manufacturer’s website.

Note: Do not spin-dry or tumble-dry the cushion.

FAQ

  • Q: Can I machine wash the entire product?
  • A: No, only the cover is machine washable. Follow handwashing instructions for the cushion.
  • Q: How should I disinfect the product?
  • A: Use a 70% alcohol-based disinfectant solution or a Chlorine-based solution as recommended in the manual.

Reconditioning instruction/Re-use

General

  • StarLock, StabilAir and Standard Air can be re-used.
  • Only official spare parts from Etac should be used.
  • Remember that the service life of the product is limited (see manual). This is particularly important when prescribing the product to a user who is expected to use it for an extended period of time.

Cleaning the Cover

Machine washing

  • Remove the cover, turn inside out and close the zip. Machine wash in warm water (maximum 60°C for the standard cover, maximum 80° for the incontinence cover). Do not bleach.
  • Rinse in cold water and tumble dry on low temperature or line dry.

Disinfection

  • Standard Cover:
    • The standard cover can only be washed (see ”Machine washing”). If disinfection is needed, we recommend to use an incontinence cover.
    • Incontinence cover:
  • Use a 70% alcohol-based disinfectant solution or a Chlorine-based solution, a maximum of 5,000 ppm. Rinse in cold water and tumble dry on low temperature or line dry.

. Cleaning the cushion

Before cleaning or disinfecting the cushion, remove the cover and wash separately (see ‘Cleaning the cover’).
Handwash
Inflate just until the cells contain air, then close the valve. Clean the cushion by hand with soap, bicarbonate or vinegar. Rinse in cold water. After rinsing, overinflate the cushion and leave it to air dry for 24 hours, or dry in a drying cabinet on low heat.
Machine wash (StarLock and Standard Air only)
Inflate just until the cells contain air, then close the valve. Machine wash at 60°/140°F for a maximum of 10 min. Use a mild detergent. Rinse in cold water. After rinsing, overinflate the cushion and leave it to air dry for 24 hours, or dry in a drying cabinet on low heat. N.B. The cushion must be filled with some air when washing in the machine. Note: Do not spin-dry or tumble- dry. Can be dried in a drying cabinet on low heat. StarLock – If there is water in the bulb pump after machine washing, remove the pump and drain the water. Mount the bulb pump again. Mesh laundry bags that are big enough to loosely fit around the cushion, can be used if additional protection is desired during machine washing.

Disinfection
Spot disinfection:

  1. Wipe down the affected areas with a 70% alcohol-based disinfectant solution or a chlorin-based solution, a maximum of 5000 ppm.
    Follow instructions on chemicals carefully.

  2. Rinse thoroughly.

Complete product

  1. Wipe down with a 70% alcohol-based disinfectant solution or a chlorin-based solution, a maximum of 5000 ppm. Follow instructions on chemicals carefully.
  2. Rinse thoroughly.

Drying
After rinsing, overinflate the cushion and leave it to air dry for 24 hours, or dry in a drying cabinet on low heat.

Visual inspection (see pictures)

Visual Inspection

  • Inspect the cushion for indications of damage or puncture.
  • If any air is lost over 24 hours, investigate the source of the leak.
  • If the cushion has no signs of damage or leaks, then the cushion can be reboxed and ready for reuse.

Check for leaks and damage

  1. Overinflate cushion.
  2. Close valve.
  3. Place the cushion under water and watch for a stream of air bubbles flowing from the cushion (this is the hole you need to patch).
  4. Mark the damaged area with a permanent marker or a toothpick (punctures).

Depending on the type of damage, see repair instructions on www.etac.com for action.

CONTACT

  • Etac Supply Center
  • Box 203
  • SE-334 33 Anderstorp
  • Tel +46 371 58 73 00
  • Fax +46 371 58 73 90
  • www.etac.com.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Etac User Manuals

Related Manuals