Winsen MH-441D-LOW NDIR Infrared CO2 Gas Sensor User Manual
- June 8, 2024
- Winsen
Table of Contents
NDIR Infrared CO2 Gas Sensor
(Model:MH-441D-LOW)
Manual
Version: 1.0
Valid from: 2020.09.14
Statement
This manual copyright belongs to Zhengzhou Winsen Electronics Technology Co.,
LTD. Without the written permission, any part of this manual shall not be
copied, translated, stored in database or retrieval system, also can’t spread
through electronic, copying, record ways.
Thanks for purchasing our product. In order to let customers use it better and
reduce the faults caused by misuse, please read the manual carefully and
operate it correctly in accordance with the instructions. If users disobey the
terms or remove, disassemble, change the components inside of the sensor, we
shall not be responsible for the loss.
The specific such as color, appearance, sizes &etc, please in kind prevail.
We are devoting ourselves to products development and technical innovation, so
we reserve the right to improve the products without notice. Please confirm it
is the valid version before using this manual. At the same time, users’
comments on optimized using way are welcome.
Please keep the manual properly, in order to get help if you have questions
during the usage in the future.
Introduction
MH-441D infrared gas sensor is a miniature universal intelligent sensor, which
adopts NDIR theory to detect concentration of CH4 in air and has good
selectivity, stable performance, long life, also is independent of Oxygen. The
inside temperature sensor could be used for temperature compensation. This
miniature infrared gas sensor is developed by the tight integration of mature
infrared absorbing gas detection technology, micro machine workout and
superior circuit design.
It is convenient in use and also instead of catalytic component directly,
widely used in various occasions with flammable and explosion hazard gas.
Features
- High sensitivity, high resolution, low power consumption, Fast response time
- Output method: UART, analog voltage signal
- Temperature compensation, excellent linear output
- Long lifespan, Excellent stability, Anti-poisons, anti-vapor interference
Applications
Widely used for HVAC refrigeration, industrial-process control and safety protection, agriculture and animal husbandry.
Main Parameters
Table 1
Model | MH-411D-LOW |
---|---|
Detection Gas | CO2 |
Working Voltage | 3.6-5V DC(Need to be powered by safety barrier) |
Average Current | <30 mA |
Detection range | 0-10%vol Optional(refer Table 2.) |
Interface Level | 3.0V |
Output Signal | UART |
0.4-2V(Need to be powered by safety barrier)
Warm-up time| 3 min
Response time| T90<30s
Working Temperature| -20°C —60°C
Working Humidity| 0″95%RH(no condensation)
Sizes| 020 mmx22.2 mm
Weight| 35g
Lifetime| >5 years
Defense Grade| IP54
Power, communication terminal Intrinsic safety| Ui=7.5VDC, Ii=265mA,
Pi=0.5W, Ci=10 μF, Li=0mH
Figure 1: Sensor Structure diagram
Table 2.Measuring Range and Resolution
Target Gas| Measuring Range| Resolution| Decimal places|
Remark
---|---|---|---|---
CH4| 0-5%VOL| 0.01% VOL| 2| Temperature compensation
0-10%VOL| 2
0100%LEL| 1%LEL| None
Structure Size
Pin definition
Table3.Pin definition
Name of Pin | Explanation |
---|---|
Pin 2 | V+ Voltage input |
Pin 1 | GND |
Pin 4 | Vout (0.4-2 V) |
Pin 3 | UART ( RXD ) 0-3.0 V data input |
Pin 5 | UART ( TXD ) 0-3.0 V data output |
Analog voltage output
The output of Vout is proportional to the gas concentration, 0.4-2.0V output
stands for 0 to full range.
Connection: Vin –5V,GND- Power Ground, Vout-ADC input.
After warm-up, If self-checking detect a fault, output voltage is 0V.
Digital Output
Please referMH-410D communication protocol.
Intrinsically safe explosion-proof
-
This product meets the standards of GB3836.1-2010 “Explosive Atmosphere Part 1: General Requirements for Equipment” and GB3836.4-2010 “Explosive Atmosphere Part 4: Equipment Protected by Intrinsically Safe “i”” standards ” ; the explosion-proof mark is Exib Ⅱ B T4 Gb, it is suitable for zone 1 and zone 2, contains Class IIA, T1-T3 explosive environment formed by the flammable gas, mixture of steam and air; it has passed the inspection by the National Quality Inspection Center for Explosion-proof Electrical Products and obtained the explosion-proof certificate.
When using, please note the following: -
The intrinsically safe power supply must be used to power the sensor, otherwise the explosion-proof performance will be affected.
-
It is forbidden to replace the sensor in dangerous places.
-
It is forbidden to disassemble or replace the sensor element to avoid affecting the explosion-proof performance.
-
It is not allowed to replace components or structures, so as not to affect the explosion-proof performance.
-
The installation and wiring of the safety barrier must be carried out in accordance with the safety barrier instruction manual, and the safety barrier must obtain an explosion-proof certificate.
Connection diagram of intrinsically safe explosion-proof system
The on-site installation must comply with the relevant regulations of the
GB3836.15 — 2000 “Electrical Equipment for Explosive Gas Environment Part 15:
Electrical Installation in Hazardous Locations (Except Coal and
Mines).
The distribution parameters of the connecting cable between the safety barrier
and the sensor should meet:
Cc ≤Co-Ci Lc ≤Lo-Li Ui ≥Uo Ii ≥Io Pi ≥Po
Note:
Uo: Maximum output voltage of safety barrier;
Io: Maximum output current of safety barrier
Po: Maximum output power of safety barrier
Co: Maximum external capacitance of safety barrier
Lo: the maximum external inductance of the safety barrier (see the safety
barrier instructions for the above parameters book)
Cc: Maximum allowable distributed capacitance of connecting cable
Ui: sensor maximum input voltage
Ii: Maximum sensor input current
Pi: sensor maximum input power
Ci: Maximum internal capacitance of the sensor
Li: Maximum internal inductance of the sensor
Lc: Maximum allowable distributed inductance of connecting cable
Note:
- The sensor should be calibrated regularly, and the recommended calibration period is 6 months.
- Do not use the sensor for a long time in an environment with high dust density.
- Please use the sensor within the power supply range of the sensor.
- It is forbidden to cut or weld the sensor pins.
Zhengzhou Winsen Electronics Technology Co., Ltd
Add: No.299, Jinsuo Road, National Hi-Tech Zone,
Zhengzhou 450001 China
Tel: +86-371-67169097/67169670
Fax: +86-371-60932988
E-mail: sales@winsensor.com
Website: www.winsen-sensor.com
Tel: 86-371-67169097/67169670
Fax: 86-371-60932988
Email: sales@winsensor.com
Leading gas sensing solutions supplier in China!
Zhengzhou Winsen Electronics Technology Co., Ltd
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>