Citizen REF CT830 Ear Thermometer Instruction Manual

June 8, 2024
Citizen

INSTRUCTION MANUAL

DIGITAL EAR THERMOMETER
REF CT830

IDENTIFICATION OF PARTS

IDENTIFICATION OF PARTS

Warning

PLEASE READ THIS MANUAL BEFORE ANY ATTEMPT TO USE THIS PRODUCT IS MADE.

PRECAUTIONS
The ear thermometer must be inserted correctly to obtain correct temperature measurement. Be sure to consult a physician if you feel that your health is in poor condition.

DO NOT judge your health only on the presence or absence of a fever.

  1. Make sure that the probe lens is always clean.

PRECAUTIONS

  1. Very carefully and gently insert the probe into the ear.
  2. Avoid using the ear thermometer if there is inflammation on or in the ear.
  3. DO NOT use the ear thermometer if the main body is damaged (for example, the probe lens is broken). The continuous use of a damaged unit may cause injury, improper results, or serious danger. Contact a physician immediately if broken pieces are swallowed.

Protect the natural environment:

Please help protect natural environment by respecting national and/or local recycling regulations when disposing of the battery and the product at the end of their useful live.

WEEE

If you dispose this product, do not dispose of with general household waste. There are separate collection systems for used electronics products in accordance with legislation under the WEEE Directive (Directive 2002/96/EC), effective only within European Union.

Caution

  1. The body temperature measured in the ear may differ from measurements taken under the armpits, in the mouth, or in the rectum. Therefore, comparing them will not be of value. Take temperature readings periodically to determine the normal ear temperature measurement with this instrument, then use that reading as the basis for comparison with any measurement taken when you suspect a fever.
  2. Do NOT use this thermometer for measurements other than as specifi ed. This thermometer is designed for ear measurement. Room temperature and material surface temperature readings are approximate and are only for reference.
  3. Do NOT immerse the ear thermometer in water or any other liquid and do NOT allow the unit to come into contact with chemical solvents.
  4. Do NOT drop the ear thermometer and do NOT expose the ear thermometer to heavy shock. The ear thermometer is not shock-protected.
  5. Do NOT bend the ear thermometer.
  6. Do NOT disassemble the unit, do NOT make any modifi cations to it, and do NOT touch the circuitry.
  7. Do NOT store the unit under the direct sunlight, do NOT store it at high temperatures, and do NOT store it in areas with a high degree of humidity and/or dust.
  8. Do NOT boil the probe.
  9. Do NOT use the ear thermometer if it operates irregularly or shows signs of malfunction.
  10. Do NOT use thinner or benzene to clean the ear thermometer.
  11. When removing the ear thermometer from storage at low temperatures, place it in a warm area for 5 to 10 minutes before use.
  12. Keep the ear thermometer out of the reach of children.
  13. Wipe the ear thermometer clean before storing.

About Ear Thermometers

All objects radiate heat. To measure body temperature, this ear thermometer uses an infrared sensor incorporated in its probe to detect the heat radiated from the eardrum and the area around it. In general the ear canal (external auditory canal) of an infant is relatively straight.

As we grow older, it takes on some curvature, as shown in Fig. 2. It is important to point the probe lens of the ear thermometer toward the eardrum and its periphery inside the ear. To straighten the ear canal, gently pull on the outer ear in the direction of the rear of the head, as shown in Fig.

Each ear canal has a slightly different shape, be sure to check it before taking temperature readings. Insert the probe straight into the ear canal, as shown in Fig. 4.

About Ear Thermometers

Caution

  1. A person’s left and right ear canals are not necessarily symmetrical. This may cause different measurement results. Always take temperature readings in the same ear.
  2. A soiled probe lens will result in inaccurate measurement. Be sure to check that the probe lens is clean before use.
  3. Body temperature rises after vigorous exercise, a bath or shower, or eating. Wait at least 30 minutes before taking temperature readings.
  4. Do NOT use the ear thermometer when the ear canal is wet (after swimming, bathing, etc.).
  5. Let the ear thermometer stand at room temperature before use.
  6. It is important to know your normal temperature so you can compare it to the measurement result.

HOW TO OPERATE

FUNCTIONS
This thermometer has the following functions for care of users.

  • Ear Temperature: for human temperature.
  • Time and calendar: for memory.
  • Room Temperature: for monitoring the ambient temperature for a baby or patient.
  • Material surface temperature (NOT for body temperature): for temperature of a baby’s milk, etc.
  • Stop watch: for taking a patient’s pulse.
  • Fever alarm: an alarm when a user’s temperature is more than 37.5°C, indicative of a fever.
  • Memory: for recording up to 9 temperature readings.
  • Celsius / Fahrenheit option switch: for international use.

TIME AND CALENDAR SETTINGS

  1. Press and hold the ON/MEM button for about 5 sec and the SET icon starts to fl ash, indicating the setting mode (Fig.5).

TIME AND CALENDAR SETTINGS

2. Press the SCAN button for the 12/24-hour setting mode. The 24 icon starts to fl ash, indicating the 24-hour display. Press the SCAN button again and the 24 icon will disappear from the screen, indicating the 12-hour display. When this display is used, afternoon times will be indicated with the PM icon. Press the SCAN button to toggle between the 12 and 24-hour displays.

3. Press the ON/MEM button to cycle the fl ashing icons in the
sequence below to select the respective setting.

hour-minute-year-month-day-leave the setting screen

While the setting section is fl ashing, press SCAN button to change the numbers. If no buttons are pressed in the setting mode for 8 seconds, the display automatically returns to the EAR TEMPERATURE mode.

EAR TEMPERATURE

  1. Make sure that the probe lens is clean.
  2. Press ON/MEM button and after the all indicators show up on the display (Fig.6), the ready message with icon displays, accompanied by two short beeps (Fig.7).

EAR TEMPERATURE

3. Gently pull the ear back to straighten the ear canal (Fig.3) and gently position the probe into the ear canal, aiming towards the membrane of the eardrum to obtain an accurate reading.

4. After the probe is positioned in the ear, press and hold the SCAN button until you hear one beep. The measurement is now complete. Remove the probe and read the temperature on the display (Fig.8).

Ear temperature mark

NOTE: The time and calendar settings should be done before measurement readings are taken. The thermometer always shuts off automatically in about one minute and the stand-by mode display (room temperature and date/time) will be shown. Without the time and calendar settings, the display will not show any indications after the thermometer shuts off. Pressing the buttons will NOT shut off the thermometer.

ROOM TEMPERATURE = STAND-BY MODE

  1. Set the TIME AND CALENDAR FIRST. Then the stand-by mode display will appear after the automatic shut-off.
  2. When in stand-by mode, the room temperature will always appear on the display with the “ Room temperature mark ” icon (Fig.9).

Room temperature mark

3. The room temperature is renewed every minute. After measuring body or material surface temperature, the sensor may be warm which may affect the room temperature reading. Therefore during the fi rst 15 minutes after every use, the room temperature on the display may be inaccurate.

MATERIAL SURFACE TEMPERATURE / STOP WATCH

  1. Make sure that the probe lens is clean.
  2. Press ON/MEM button to enter the EAR TEMPERATURE mode.
  3. MATERIAL SURFACE TEMPERATURE:
    Press and hold the ON/MEM button and then press the SCAN button once to leave the EAR EMPERATURE mode and enter the MATERIAL SURFACE TEMPERATURE scan mode with ” Material surface temperature mark “icon (Fig.10).

Material surface temperature mark

3.1 Aim the probe toward the material and press the SCAN button, and the material surface temperature appears on the display. To measure the temperature of another material, repeat the procedure. If the SCAN button is pressed and held, the measurement reading will be continuously updated.
3.2 Applications include temperature measurements for water, milk, cloth, etc.

Press and hold the ON/MEM button and then press the SCAN button once to leave MATERIAL SURFACE TEMPERATURE mode and enter the STOPWATCH mode with the ” Stop watch mark ” icon (Fig.11).

NOTE:

  • Do NOT submerge the probe tip in liquid.

  • This mode shows the surface temperature of material only and is NOT for taking human body temperature.

4. STOP WATCH:

4.1 This mode counts from 0.01sec to 3 minutes.
4.2 To start and stop the stopwatch, press the SCAN button. To reset the counter, press the ON/MEM button.

Press and hold the ON/MEM button and then press the SCAN button once to leave STOP WATCH mode and return to the EAR TEMPERATURE mode.

FEVER ALARM
If the thermometer detects a temperature over 37.5°C in the EAR TEMPERATURE mode, there will be a long beep followed by three short beeps to warn the user of a fever.

MEMORY
There are a total of 9 memory slots for recording EAR TEMPERATURE measurements. Each memory slot also records the date and time of the measurement. Memory slots can be cycled by pressing the ON/MEM button during the EAR TEMPERATURE mode (Fig.12).

NOTE: Even when a Memory slot is being displayed, users can measure ear temperature.

CELSIUS/FAHRENHEIT OPTION SWITCH

  1. Make sure the thermometer is in stand-by mode and indicating the room temperature, date, and time.
  2. Press and hold the SCAN button, then immediately press the ON/MEM button. After 3 seconds, the °C icon will switch to the °F icon. (use same procedure to switch back to °C).

AUTO POWER OFF FUNCTION

If this unit is left unattended for about 1 minute, it will automatically power down from the current mode and enter the stand-by mode, showing room temperature, time, and date. The display alternates between time and date in a pattern that repeats every minute as follows: time 11 sec / date 4 sec / time 11 sec / date 4 sec / time 30 sec.

BATTERY REPLACEMENT

  1. Replace the battery when the display behaves as follows:
    The stand-by mode shows NOTHING on the display except the time and date. When the ON/MEM button is pressed, NOTHING is displayed except the battery mark blinking 12 times followed by OFF for 1 second and then the display returns to NOTHING except the time and date.

BATTERY REPLACEMENT

2. Insert a pointed object such as a pen into the battery cover access hole (Fig.13) while using your thumb to push the battery cover out.

3. Hold the device and pry the battery out with a small screwdriver (Fig.14).

4. Insert a new battery (CR2032) under the metal hook and press down on the battery until you hear a click. The positive (+) side should be up and the negative (-) side should be down.

5. Replace the battery cover.

Caution

  1. Remove and then reinstall the battery if the Er message is displayed, if the display is abnormal, or if nothing is displayed when the ON/MEM button is pressed.
  2. To protect the environment, dispose of the used battery according to local regulations.
  3. Remove the battery if the ear thermometer will not be used for a long time.
  4. Keep the battery away from children.

CARE AND CLEANING, STORAGE

The probe is the most delicate part of the thermometer. Use caution when cleaning the probe lens to avoid damage.

Holding the thermometer too long may cause the probe to make a higher ambient temperature reading. This could make the body temperature measurement lower than usual.

  • After each use, clean the probe and probe lens with a cotton swab or soft cloth moistened with 70% alcohol.
  • Let the lens fully dry for at least 30 minutes.
  • Keep the unit dry and away from where it might be exposed to moisture, liquids, direct sunlight, high temperature, high humidity, or excessive dust.
  • Storing the thermometer at 10 to 40°C (50 to 104°F) allows you to measure body temperature at any time. Storing the thermometer below or above this temperature range may cause it to display Er 3. In such a case, leave the thermometer at 10 to 40°C (50 to 104°F) for at least 30 minutes before taking a measurement reading.

DESCRIPTION OF DISPLAY MARK

DESCRIPTION OF DISPLAY MARK

Questions and Answers

Q1: Why is temperature measured in the ear different from measurements taken under the armpit, in the mouth, or in the rectum?
A1: Temperatures of different parts of body may vary. Compare the measurement in question with a normal temperature previously taken in the same ear

Q2: The temperature differs each time I take a measurement. Why? What can I do?
A2: Your method for straightening the ear canal may be inconsistent, the way you insert the probe may be not consistent, or you may be taking a second measurement too soon. Pull the ear rearward to straighten ear canal, then very gently insert the probe into the ear canal for measurement. Once a measurement is taken, the temperature in the ear will be affected. Therefore, wait at least 1 minute before taking another measurement.

Q3: Can I measure temperature without the probe cover?
A3: This instrument is designed to measure temperature without the probe cover. Do NOT attach the probe cover or any film or covering.

Q4: Is it hygienically safe to use the ear thermometer without the probe cover?
A4: The probe and probe lens should be kept clean by using a cotton swab or soft cloth moistened with 70% alcohol. Therefore, the thermometer can be used hygienically without the probe cover.

Q5: Can I use the ear thermometer in cold/hot places?
A5: This instrument can be used in room temperature from 10 to 40°C (50 to 104°F). Error messages (Er 2 or Er 3) may be displayed if the temperature during storage greatly differs from that of the room temperature where you take a measurement reading. In such a case, let the ear thermometer stand for a while at room temperature before measurement.

Q6: Can I use the ear thermometer on an infant? Their ear canals are very small.
A6: Do NOT attempt to force the probe into the ear canal. Use caution and pay close attention while very gently inserting the probe just into the entrance of ear canal.

Q7: Can I use the ear thermometer to measure temperature under the armpit or in the mouth?
A7: No. This instrument is designed to measure temperature only in the ear.

Q8: Can I use the ear thermometer on animals?
A8: No. This instrument is designed to measure temperature in the ears of humans.

SPECIFICATIONSSPECIFICATIONS

Attention: consult accompanying documents. Specifications may be changed without notice in the event of improvements being made.

Reference standard: Complies with ASTM E1965-98, EN12470-5: 2003
Clinical thermometers-Part 5: Performance of infra-red ear thermometers
(with maximum device), IEC60601-1-2(EMC), EN60601-1(Safety) standards.
Manufacturer:
Name: Radiant Innovation Inc.
Address: 1F, No.3, Industrial East 9th Road, Science-Based Industrial Park,
HsinChu, Taiwan 300.
European Representative:
Name: Medical Technology Promedt Consulting GmbH
Address: Altenhofstrasse 80, D-66386 St. Ingbert, Germany

Citizen-REF-CT830-Manual Download

Questions about your Manual ? Post in the comments!

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals