OMRON Wrist Blood Pressure Monitor Instruction Manual
- June 7, 2024
- Omron
Table of Contents
Wrist Blood Pressure Monitor
Wrist Blood Pressure Monitor
HEM-6181 Instruction Manual
2829552-3B
Table of Contents
Introduction Important Safety Information 1. Know Your Monitor 2. Preparation
3. Using Your Monitor 4. Using Memory Function 5. Other Settings 6. Error
Messages and Troubleshooting 7. Maintenance 8. Specifications 9. Guidance and
Manufacturer’s Declaration
Introduction
Thank you for purchasing the OMRON HEM-6181 Wrist Blood Pressure Monitor. This
blood pressure monitor is to be worn on your wrist. This new blood pressure
monitor uses the oscillometric method of blood pressure measurement. When the
cuff inflates, this monitor senses the pressure pulsations of the artery
underneath the cuff. The pulses are called oscillometric pulses. The
electronic pressure sensor displays a digital reading of blood pressure.
Safety Instructions
This instruction manual provides you with important information about the
OMRON HEM-6181 Wrist Blood Pressure Monitor. To ensure the safe and proper use
of this monitor, READ and UNDERSTAND all of the safety and operating
instructions. If you do not understand these instructions or have any
questions, contact your local OMRON representative before attempting to use
this monitor. For specific information about your own blood pressure, consult
with your physician.
Intended Use
This device is a digital monitor intended for use in measuring blood pressure
and pulse rate in adult patient population with wrist circumference ranging
from 13.5 cm to 21.5 cm. This monitor detects the appearance of irregular
heartbeats during measurement and gives a warning signal with readings. It is
mainly designed for general household use.
Receiving and Inspection
Remove this monitor from the packaging and inspect for damage. If this monitor
is damaged, DO NOT USE and consult with your local OMRON representative.
Important Safety Information
Please read the Important Safety Information in this instruction manual before
using this monitor. Please follow this instruction manual thoroughly for your
safety. Please keep for future reference. For specific information about your
own blood pressure, CONSULT WITH YOUR PHYSICIAN.
Indicates a potentially hazardous situation
Warning which, if not avoided, could result in death or
serious injury.
· DO NOT use this monitor on infants, toddlers, children or persons who cannot
express themselves.
· DO NOT adjust medication based on readings from this blood pressure monitor.
Take medication as prescribed by your physician. ONLY a physician is qualified
to diagnose and treat high blood pressure.
· DO NOT use this monitor on an injured wrist or a wrist under medical
treatment.
· DO NOT apply the wrist cuff on your wrist while on an intravenous drip or
blood transfusion.
· DO NOT use this monitor in areas containing high frequency (HF) surgical
equipment, magnetic resonance imaging (MRI) equipment, computerized tomography
(CT) scanners. This may result in incorrect operation of the monitor and/or
cause an inaccurate reading.
· DO NOT use this monitor in oxygen rich environments or near flammable gas.
· Consult with your physician before using this monitor if you have common
arrhythmias such as atrial or ventricular premature beats or atrial
fibrillation; arterial sclerosis; poor perfusion; diabetes; pregnancy; pre-
eclampsia or renal disease. NOTE that any of these conditions in addition to
patient motion, trembling, or shivering may affect the measurement reading.
· NEVER diagnose or treat yourself based on your readings. ALWAYS consult with
your physician.
· This product contains small parts that may cause a choking hazard if
swallowed by infants, toddlers or children.
Battery Handling and Usage
· Keep batteries out of the reach of infants, toddlers or children.
Indicates a potentially hazardous situation
Caution
which, if not avoided, may result in minor or moderate injury to the user or patient, or cause
damage to the equipment or other property.
· Stop using this monitor and consult with your physician if you experience
skin irritation or discomfort.
· Consult with your physician before using this monitor on a wrist where
intravascular access or therapy, or an arteriovenous (A-V) shunt, is present
because of temporary interference to blood flow which could result in injury.
· Consult with your physician before using this monitor if you have had a
mastectomy.
· Consult with your physician before using this monitor if you have severe
blood flow problems or blood disorders as cuff inflation can cause bruising.
· DO NOT take measurements more often than necessary because bruising, due to
blood flow interference, may occur.
· ONLY inflate the wrist cuff when it is applied on your wrist. · Remove the
wrist cuff if it does not start deflating during a
measurement. · DO NOT use this monitor for any purpose other than
measuring blood pressure. · During measurement, make sure that no mobile
device or any
other electrical device that emits electromagnetic fields is within 30 cm of
this monitor. This may result in incorrect operation of the monitor and/or
cause an inaccurate reading. · DO NOT disassemble or attempt to repair this
monitor or other components. This may cause an inaccurate reading. · DO NOT
use in a location where there is moisture or a risk of water splashing this
monitor. This may damage this monitor. · DO NOT use this monitor in a moving
vehicle such as in a car. · DO NOT drop or subject this monitor to strong
shocks or vibrations. · DO NOT use this monitor in places with high or low
humidity or high or low temperatures. Refer to chapter 8. · Ensure this
monitor is not impairing blood circulation by observing the wrist while
measurement is occurring. · DO NOT use this monitor in high-use environments
such as medical clinics or physician offices. · DO NOT use this monitor with
other medical electrical (ME) equipment simultaneously. This may result in
incorrect operation of the monitor and/or cause an inaccurate reading. · Avoid
bathing, drinking alcohol or caffeine, smoking, exercising and eating for at
least 30 minutes before taking a measurement. · Rest for at least 5 minutes
before taking a measurement. · Remove tight-fitting, thick clothing and any
accessories from your wrist while taking a measurement. · Remain still and DO
NOT talk while taking a measurement. · ONLY use this monitor on persons whose
wrist circumference is within the specified range of the cuff. · Ensure that
this monitor has acclimated to room temperature before taking a measurement.
Taking a measurement after an extreme temperature change could lead to an
inaccurate reading. OMRON recommends waiting for approximately 2 hours for the
monitor to warm up or cool down when the monitor is used in an environment
within the temperature specified as operating conditions after it is stored
either at the maximum or at the minimum storage temperature. For additional
information of operating and storage/transport temperature, refer to chapter
8. · DO NOT crease the wrist cuff excessively. · Read and follow the “Correct
Disposal of This Product” in chapter 9 when disposing of the device and any
used accessories or optional parts. · DO NOT use this monitor after the
durable period has ended. Refer to chapter 8.
Battery Handling and Usage
· DO NOT insert batteries with their polarities incorrectly aligned.
· ONLY use 2 “AAA” alkaline batteries with this monitor. DO NOT use other
types of batteries. DO NOT use new and used batteries together. DO NOT use
different brands of batteries together.
· Remove batteries if this monitor will not be used for a long period of time.
· If battery fluid should get in your eyes, immediately rinse with plenty of
clean water. Consult with your physician immediately.
· If battery fluid should get on your skin, wash your skin immediately with
plenty of clean, lukewarm water. If irritation, injury or pain persists,
consult with your physician.
· DO NOT use batteries after their expiration date. · Periodically check
batteries to ensure they are in good
working condition. · ONLY use batteries specified for this monitor. Use of
unsupported batteries may damage and/or may be hazardous to this monitor.
1. Know Your Monitor
1.1 Contents
Monitor, 2 “AAA” alkaline batteries, storage case, instruction manual
1.2 Monitor d
a
b
e
f c
a. Wrist cuff (Wrist circumference 13.5 cm to 21.5 cm)
b. Display c. Battery compartment d. [Memory] button e. [Morning average]
button f. [START/STOP] button
1.3 Display and Symbols
A
H
I
B
J
K C
L
D
M
E
N
F
O
G
A Date and time display
B
Memory symbol
Appears when viewing readings stored in the memory.
C Systolic blood pressure reading
D Diastolic blood pressure reading
Pulse display / Memory number
Pulse rate appears after the measurement. E When pressing the button, the
memory number appears
for approximately one second before the pulse rate appears on the display.
F
Deflation symbol
Appears during cuff deflation.
G
Heartbeat symbol
Flashes while taking a measurement.
Morning average symbol
Appears when viewing the weekly average readings
H
taken in the morning.
Weekly averages for the measurement taken in the
morning within 4 weeks appear on the display.
Average value symbol
I
Appears when viewing the average readings of the
latest 2 or 3 readings taken within 10 minutes span.
Battery symbol (low)
Flashes when batteries are low. J
Battery symbol (depleted)
Appears when batteries are depleted.
Hypertension symbol
Appears if your systolic blood pressure is 135 mmHg or
above and/or the diastolic blood pressure is 85 mmHg
K
or above.
Morning hypertension symbol
Appears when the morning weekly average is 135/85 mmHg or above.
Positioning indicator
Your monitor has a built-in advanced positioning sensor that is used as an aid
in determining if the monitor is at the correct height.
L
The ” ” symbol appears when your monitor is at the
correct position relative to your heart during a
measurement. When your monitor is higher or lower
position relative to your heart, the heart position locator
bar will appear up above, or below the heart symbol to
guide correct wrist positioning.
Irregular heartbeat symbol
Appears along with
Normal Heartbeat
readings when an
Pulse
irregular rhythm is detected 2 or more
Blood pressure
times during a
measurement.
Irregular Heartbeat
M
An irregular heartbeat
rhythm is defined as a
Pulse
Short Long
rhythm that is 25% Blood pressure
less or 25% more
than the average rhythm detected while your monitor is
measuring blood pressure. If it continues to appear, we
recommend you to consult with and follow the
directions of your physician.
Movement error symbol
Appears along with readings when your body is moving
N
during a measurement. If it appears, remove the wrist
cuff, and wait for 2 to 3 minutes. Then, remain still and
take another measurement.
Cuff wrap guide symbol (OK)
Appears if the wrist cuff is wrapped around the wrist correctly while taking a
measurement. It also appears when viewing past readings. O
Cuff wrap guide symbol (loose)
Appears if the wrist cuff is not wrapped around the wrist correctly while
taking a measurement. It also appears when viewing past readings.
2013 ESH/ESC* Guidelines for the management of arterial hypertension
Definitions of hypertension by office and home blood pressure levels
Systolic Blood Pressure
Diastolic Blood Pressure
Office 140 mmHg 90 mmHg
Home 135 mmHg 85 mmHg
These ranges are from statistical values for blood pressure.
- European Society of Hypertension (ESH) and European Society of Cardiology (ESC).
Warning
· NEVER diagnose or treat yourself based on your readings. ALWAYS consult with
your physician.
2. Preparation 2.1 Installing Batteries 1. Push down the hook of the battery
cover and pull downward.
2. Insert 2 “AAA” alkaline batteries as indicated in the battery compartment.
3. Close the battery cover.
2
1
Note · When the ” ” symbol flashes on the display, it is recommended
you replace batteries. · To replace batteries, turn your monitor off and
remove all batteries.
Then replace with 2 new alkaline batteries at the same time. · As soon as
batteries are correctly installed, the year flashes on the
date and time display to set the date and time. · Replacing batteries will not
delete previous readings. · The supplied batteries may have a shorter life
than new batteries. · Disposal of used batteries should be carried out in
accordance
with local regulations.
2.2 Setting the Date and Time
Note · Set your monitor to the correct date and time before taking a
measurement for the first time. · As soon as batteries are correctly
installed, the year flashes on the
display of your monitor.
1. Press the or
Press and hold the Press and hold the the years.
button once to adjust the year.
button to quickly advance the years. button to quickly go backwards through
2. Press the [START/STOP] button to confirm the year and the month will flash. Repeat the same steps to adjust the month, day, hour, and minute.
Year
Month
Day
Hour
Minute
3. Press the [START/STOP] button to store the setting.
It automatically turns off after 3 seconds.
Note · If you need to reset the date and time, replace batteries or press
and hold the button until the year flashes, and set the date and time as above
steps.
2.3 Blood Pressure Measurement Tips
To help ensure an accurate reading, follow these directions: · Stress raises
blood pressure. Avoid taking measurements during
stressful times. · Measurements should be taken in a quiet place. · It is
important to take measurements at same times each day.
Taking measurements in the morning and in the evening is recommended. ·
Remember to have a record of your blood pressure and pulse readings for your
physician. A single measurement does not provide an accurate indication of
your true blood pressure. You need to take and record several readings over a
period of time.
Caution · Avoid bathing, drinking alcohol or caffeine, smoking, exercising
and eating for at least 30 minutes before taking a measurement. · Rest for at
least 5 minutes before taking a measurement.
2.4 Applying the Wrist Cuff
Note · Following steps are for applying the wrist cuff to your left wrist.
When you take a measurement on your right wrist, apply the wrist cuff using
the same steps as on your left wrist. · The blood pressure can differ between
the right wrist and the left wrist, and the measured blood pressure values can
be different. OMRON recommends to always use the same wrist for measurement.
If the values between both wrists differ substantially, please check with your
physician as to which wrist to use for your measurements.
Caution · Remove tight-fitting, thick clothing and any accessories from your
wrist while taking a measurement.
1. Apply the wrist cuff to your left wrist.
Roll up sleeve. Make sure your sleeve is not rolled up too tightly on your
arm. This may constrict the flow of blood in your arm.
2. Put your wrist through the cuff loop.
Your palm and monitor display should both face upward.
3. Position the wrist cuff leaving a clearance of 1 to 2 cm between the cuff and the bottom of your palm.
Wrist cuff 1 to 2 cm
4. Wrap the wrist cuff firmly around your wrist. Do not apply over clothing.
Check to make sure that there is no gap between your wrist and the cuff.
Note · Make sure the wrist cuff does not
cover the protruding part of the wrist bone on the outside of your wrist.
2.5 Sitting Correctly
To take a measurement, you need to be relaxed and comfortably seated in a room
with a comfortable temperature. Place your elbow on the table to support your
arm.
· Sit in a chair with your legs uncrossed and your feet flat on the floor.
· Sit with your back and arm being supported.
· Your monitor must be approximately the same level as your heart. If the
monitor is too high above your heart, your blood pressure will be artificially
low. If the monitor is too low below your heart, your blood pressure will be
artificially high.
· Relax your wrist and hand. Do not bend your wrist back, clench your fist, or
bend your wrist forward.
3. Using Your Monitor
3.1 Taking a Measurement
Note · To stop the measurement, press the [START/STOP] button once
to deflate the wrist cuff.
Caution · DO NOT use this monitor with other medical electrical (ME)
equipment simultaneously. This may result in incorrect operation of the
monitor and/or cause an inaccurate reading. · Remain still and DO NOT talk
while taking a measurement.
1. Press the [START/STOP] button.
All symbols appear on the display before starting the measurement.
2. Check the positioning indicator.
Your monitor has a built-in advanced positioning sensor that is used as an aid
in determining if the monitor is at the correct height. It has been designed
to work with most people so that when your wrist is at the correct position
relative to your heart,
the ” ” symbol will appear. When your wrist is higher or lower position
relative to your heart, a locator bar will appear according to the wrist
position. Due to difference in individual size and physique, this feature may
not be helpful in all cases and you may wish to disable this feature. If you
feel the position of the wrist according to positioning indicator’s guidance
does NOT match your heart level, disable this feature and follow your
judgement. To disable it, refer to section 5.1.
Monitor height Wrist is too high
Positioning indicator Any locator bar (-) above the ” ” symbol appears.
Proper position
The ” ” symbol appears.
Wrist is too low
Any locator bar (-) below the ” ” symbol appears.
Note · After 5 seconds from pressing the [START/STOP] button,
even if your monitor is not positioned properly, your monitor will start the
measurement and the wrist cuff will start to inflate.
3. Remain still and do not move or talk until the entire measurement process
is completed.
As the cuff inflates, your monitor automatically determines your ideal
inflation level. This monitor detects your blood pressure and pulse rate
during inflation.
The ” ” symbol flashes at every heartbeat.
Note · The ” ” symbol appears if the wrist cuff is wrapped
around the wrist correctly. · When the ” ” symbol is displayed, the wrist cuff
is not
applied correctly. Press the [START/STOP] button to turn your monitor off,
then apply the cuff correctly.
After your monitor has detected your blood pressure and pulse rate, the cuff
automatically deflates. Your blood pressure and pulse rate are displayed. If
either the systolic or the diastolic reading is high (refer to section 1.3),
the ” ” symbol appears.
START
INFLATING COMPLETED
4. Press the [START/STOP] button to turn your monitor off.
Note · Your monitor will automatically turn off after 2 minutes. · Wait 2-3
minutes between measurements. The wait time
allows the arteries to decompress and return to their premeasurement form. You
may need to increase the wait time depending on your individual physiological
characteristics.
4. Using Memory Function
Your monitor automatically stores up to 60 readings.
4.1 Viewing the Readings Stored in Memory
1. Press the button.
Alternating date/time display
The Memory number appears for one second before the pulse rate is displayed.
The most recent reading set is numbered “1”.
Note · If the positioning indicator is enabled, the positioning
indicator is displayed with the readings. · If the reading is high (refer to
section 1.3),
the ” ” symbol appears. · If there are no readings stored in the
memory, the screen to the right is displayed. · If the memory is full, the
monitor will delete
the oldest readings.
2. Press the button repeatedly to scroll through the previous readings stored
in the memory.
4.2 Viewing the Average Reading
Your monitor can calculate an average reading based on the latest 2 or 3
readings taken within 10 minutes span.
1. When your monitor is off, press and hold the button for more than 2
seconds.
Note · If there are only 2 readings in the memory
for that period, the average will be based on these 2 readings. · If the
average reading is high (refer to section 1.3), the ” ” symbol appears.
4.3 Viewing Morning Weekly Averages
Your monitor calculates and displays weekly averages for the measurements
taken in the morning for 4 weeks.
1. Press the button once.
The morning weekly average for the current week “THIS WEEK” appears on the
display. If the morning weekly average is high (refer to section 1.3), the ” ”
symbol appears.
2. Press the button repeatedly to display the average for the previous weeks.
Your monitor displays “-1 WEEK” for the previous week’s average, to “-3 WEEK”
for the oldest set of averages.
3. Press the [START/STOP] button to turn off your monitor.
Morning Weekly Average Calculation This is the average for the measurements taken during the morning (4:00am – 9:59am) between Sunday and Saturday. Only the first 3 measurements taken in the morning between 4:00am – 9:59am will be used to calculate the morning average for each day.
Sunday 4:00
Monday
Measurements of the morning Tuesday Wednesday Thursday
Friday
Saturday
1
2
The first 3 measurements
3
4
9:59
5. Other Settings
5.1 Disabling/Enabling Positioning Indicator
The positioning indicator is enabled as default.
1. When your monitor is off, press and hold the for more than 10 seconds.
button
Note · Your monitor will automatically turn off after 3 seconds from
the completion. · To enable the positioning indicator, follow the same steps
as
above.
5.2 Deleting All Readings Stored in Memory
1. Press the button, then the ” ” symbol appears.
2. While holding the button down, press
and hold the [START/ STOP] button for more than 2 seconds.
6. Error Messages and Troubleshooting
In case of any of the below problems occur during measurement, first check
that no other electrical device is within 30cm. If the problem persists,
please refer to the table below.
Display/ Problem
Possible Cause
Solution
The wrist cuff is not applied correctly. appears or the wrist cuff does not inflate. Air is leaking from the wrist cuff.
Apply the wrist cuff correctly, then take another measurement. Refer to
section 2.4.
Contact local OMRON representative.
appears
The wrist cuff is overinflated exceeding 300 mmHg.
Do not touch the wrist cuff while taking a measurement.
appears
You move or talk during a measurement. Vibrations disrupt a measurement.
Remain still and do not talk during a measurement.
appears /
appears
appears
appears
appears
The pulse rate is not detected correctly.
Apply the wrist cuff correctly, then take another measurement. Refer to section 2.4. Remain still and sit correctly during a measurement.
If the ” ” symbol continues to appear, we recommend you to consult with your physician.
Your wrist is moved up and down during a measurement.
Do not move your wrist, then take another measurement. Refer to chapter 3.
The monitor is malfunctioned.
Press the [START/ STOP] button again. If “Er” still appears, contact local OMRON representative.
heart symbol
The heart symbol in the
positioning
The positioning indicator is disabled.
Enable the positioning indicator. Refer to section 5.1.
indicator does
not appear and
a locator bar
does not
appear.
flashes
Batteries are low.
Replacing all batteries with 2 new alkaline batteries is recommended. Refer to section 2.1.
appears or the monitor is turned off Batteries are depleted.
unexpectedly during a
measurement
Immediately replace all batteries with 2 new alkaline batteries. Refer to section 2.1.
No power. Nothing appears on the display of the monitor.
Batteries are completely depleted.
Battery polarities are not properly aligned.
Check the battery installation for proper placement. Refer to section 2.1.
Blood pressure varies constantly. Many factors
Readings
including stress, time of day, and/or how you
appear too high apply the wrist cuff, may affect your blood
or too low.
pressure. Review sections 2.3 – 2.5 and chapter
Press the [START/STOP] button to turn the
monitor off, then press it again to take a
Any other problems occur.
measurement. If the problem continues, remove all batteries and wait for 30 seconds. Then reinstall batteries.
If the problem still persists, contact local OMRON
representative.
Note · All readings will be deleted. You cannot partially delete the
readings stored in the memory.
7. Maintenance
7.1 Maintenance
To protect your monitor from damage, please follow the directions below: ·
Changes or modifications not approved by the manufacturer will
void the user warranty.
Caution · DO NOT disassemble or attempt to repair this monitor or other
components. This may cause an inaccurate reading.
7.2 Storage
Keep your monitor in the storage case when not in use. · Store your monitor in
a clean, safe location. Do not store your monitor: · If your monitor is wet. ·
In locations exposed to extreme temperatures, humidity, direct
sunlight, dust or corrosive vapors such as bleach. · In locations exposed to
vibrations or shocks.
7.3 Cleaning
· Do not use any abrasive or volatile cleaners. · Use a soft dry cloth or a
soft cloth moistened with neutral soap to
clean your monitor and wrist cuff, and then wipe them with a dry cloth. · Do
not wash or immerse your monitor and wrist cuff in water. · Do not use
gasoline, thinners or similar solvents to clean your monitor and wrist cuff.
7.4 Calibration and Service
· The accuracy of this blood pressure monitor has been carefully tested and is
designed for a long service life.
· It is generally recommended to have the unit inspected every two years to
ensure correct functioning and accuracy. Please consult your authorised OMRON
dealer or local OMRON representative at the address given on the packaging or
attached literature.
8. Specifications
Product description Model Display Cuff pressure range Blood pressure
measurement range Pulse measurement range
Accuracy
Inflation Deflation Measurement method Operation mode IP classification Rating
Power source
Battery life
Durable period (Service life)
Operating conditions
Storage / Transport conditions
Weight
Dimensions
Measurable wrist circumference Memory
Contents
Protection against electric shock Cuff Material Applied part Maximum
temperature of the applied part
Wrist Blood Pressure Monitor HEM-6181 LCD digital display 0 to 299 mmHg SYS:
60 to 260 mmHg DIA: 40 to 215 mmHg
40 to 180 beats / min.
Pressure: ±3 mmHg Pulse: ±5% of display reading Automatic by electric pump
Automatic rapid deflation Oscillometric method Continuous operation IP 22 DC3
V 3.0 W 2 “AAA” alkaline batteries 1.5V Approximately 300 measurements (using
new alkaline batteries)
5 years
+10°C to +40°C / 15 to 90% RH (noncondensing) / 800 to 1060 hPa -20°C to +60°C
/ 10 to 90% RH (noncondensing) Approximately 86 g not including batteries
Approximately 93 mm (w) × 62 mm (h) × 20 mm (l) (not including the wrist cuff)
13.5 to 21.5 cm
Stores up to 60 readings Monitor, storage case, 2 “AAA” alkaline batteries,
instruction manual
Internally powered ME equipment
Nylon and polyester Type BF (wrist cuff)
Lower than +48°C
Note · These specifications are subject to change without notice. · In the
clinical validation study, K5 was used on 85 subjects for
determination of diastolic blood pressure. · This monitor is clinically
investigated according to the
requirements of ISO 81060-2:2013. · IP classification is degrees of protection
provided by enclosures in
accordance with IEC 60529. This monitor is protected against solid foreign
objects of 12.5 mm diameter and greater such as a finger, and against oblique
falling water drops which may cause issues during a normal operation. · This
device has not been validated for use on pregnant patients.
9. Guidance and Manufacturer’s Declaration
· This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
· This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and Part
3: Supplementary requirements for electromechanical blood pressure measuring
systems.
· This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan. The Core component for OMRON blood pressure
monitors, which is the Pressure Sensor, is produced in Japan.
Symbols description
Applied part – Type BF Degree of protection against electric shock (leakage current)
IP XX
Ingress protection degree provided by IEC 60529
CE Marking
Serial number
Temperature limitation
Humidity limitation
Atmospheric pressure limitation
Need for the user to consult this instruction manual.
Indicates the correct positioning for the monitor on the wrist Measurable
wrist circumference
Battery
Direct current
Important information regarding Electro Magnetic Compatibility (EMC)
HEM-6181 manufactured by OMRON HEALTHCARE Co., Ltd. conforms to
EN60601-1-2:2015 Electro Magnetic Compatibility (EMC) standard. Further
documentation in accordance with this EMC standard is available at http://www
.omronhealthcare-ap.com/emc-information. Refer to the EMC information for
HEM-6181 on the website.
Correct Disposal of This Product (Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it should
not be disposed of, with other household wastes at the end of its working
life. To prevent possible harm to the environment or human health from
uncontrolled waste disposal, please separate this product from other types of
wastes and recycle it responsibly to promote the sustainable reuse of material
resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they
can return this item for environmentally safe recycling. Business users should
contact their supplier and check the terms and conditions of the purchase
contract. This product should not be mixed with other commercial waste for
disposal.
Manufacturer
OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, KYOTO, 617-0002 JAPAN
EU-representative OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS www.omron-healthcare.com
Asia Pacific HQ
OMRON HEALTHCARE SINGAPORE PTE LTD. 438A Alexandra Road, #05-05/08 Alexandra Technopark, Singapore 119967 www.omronhealthcare-ap.com
Production Facility
OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD. Binh Duong Province, Vietnam
Made in Vietnam
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>