Honeywell Handheld PID Monitors User Guide
- June 7, 2024
- Honeywell
Table of Contents
THE POWER OF CONNECTED
Handheld PID Monitors
Quick Start Guide
User Interface
Functions are controlled via the [MODE] and [Y/+] keys. The display shows information such as monitored threats, real-time readings and measurement units, alarm type (when in alarm, including cal. overdue), battery status, datalog (if on), and radio and connection quality (if available).
Charging The Instrument
Always fully charge the battery before use. Contacts on the bottom of the instrument meet the Travel Charger’s or Charging Cradle’s contact pins, transferring power. Make sure the charger is firmly attached. Then connect the AC Adapter’s plug to the charger, and plug its transformer into an AC outlet. While charging, the LED on the cradle glows red. When the battery is fully charged, the LED glows green. (Consult the User’s Guide for more details on using the Charging Cradle.)
Download The User’s Guide
The free Handheld PID Monitors User’s Guide is the complete manual for
operating and maintaining all Handheld PID Monitors. It must be carefully read
by all individuals who have or will have the responsibility of using,
maintaining, or servicing the product.
Download the Handheld PID Monitors User’s Guide here:
https://www.raesystems.com/customercare/resource-center/manuals
Or, use your smartphone’s camera:
WARNINGS
Read Before Operating
The User’s Guide must be carefully read by all individuals who have or will
have the responsibility of using, maintaining, or servicing this product. The
product will perform as designed only if it is used, maintained, and serviced
in accordance with the manufacturer’s instructions.
Note: When the battery’s charge falls below a preset voltage, the
instrument beeps and flashes once every minute, and the “empty battery” icon
blinks on and off once per second. You should turn off the instrument within
10 minutes and recharge the battery or replace it with a fresh one with a full
charge.
Turning The Instrument On/Off
With the instrument turned off, press and hold the [MODE] key until the
audible alarm stops, and then release. During startup, it tests the battery,
buzzer, vibration alarm, LEDs, and internal functions. Then the main reading
screen is shown.
Turning The Instrument Off
Press and hold [MODE] for the 5-second countdown to shutoff When the countdown
ends and the screen displays “Unit Off,” release your finger from the [MODE]
key. The instrument is now off.
Hygiene & Search Modes
The instrument has two main operating modes:
Hygiene Mode: After a warm-up, the instrument samples continuously.
Search Mode: After the warm-up, the pump shuts off, and “Ready” is displayed.
This is ideal for headspace sampling.
Compound-Specific Measurement
The UltraRAE 3000+ can perform the compound-specific measurement in addition
to general VOC measurement. This requires using an RAE-Sep™ separation tube
(butadiene or benzene) and having the UltraRAE 3000+ in Tube Mode, operating
with a 9.8eV lamp.
The following are the 3 steps to make specific measurements:
-
Start the instrument and measure in TVOC.
-
If the reading is 0ppm, there is no need to enter a specific measurement mode.
-
If the reading is higher than acceptable values, take an unopened RAE-SepTM tube and break off both ends.
-
Insert the tube following the arrow.
-
Trigger the specific measurement mode.
Recommended Procedures
Important! Test the alarms prior to use. If any do not work, refer to the
User’s Guide for instructions. Do not use the instrument without functioning
alarms.
Check the pump: Start the instrument and block the inlet to make sure the
alarm is triggered and sounds and lights are functional. If the gas inlet is
blocked but the pump does not shut down, or the pump shuts down too easily
with a slight blockage, the pump threshold value may be set too high or low.
Refer to the User’s Guide for steps to correct this.
Check whether the instrument responds to moisture: Cup your hand over the
inlet for 10 to 20 seconds (do not block flow). If the instrument reads >2 ppm
or if the ppbRAE 3000+ reads >500 ppb, then the sensor needs
cleaning.
Always Use An External Filter
Before using the instrument, attach an external filter to the sampling probe. This prevents moisture, debris, and other potentially clogging materials from entering the probe, preventing premature pump failure and keeping the sensor and lamp cleaner.
Download The Device Configurator Mobile App
The Honeywell Device Configurator Mobile App provides the experience of
managing safety (gas monitor) instruments on a mobile platform. It simplifies
interaction with both portable and transportable gas monitors when they are in
close proximity. It also provides the flexibility and convenience to access
information stored on instruments without having to move them.
Download the free Device Configurator Mobility App
https://play.google.com/store/apps/details?id=com.honeywell.his.ha.dc
|
https://itunes.apple.com/cn/app/device-
configurator/id1194056427?l=en&mt=8
---|---
|
Rev. A P/N: 059-4027-000
November 2018