truelife B08CPN6CNR Pulse Digital Blood Pressure Monitor Instruction Manual
- June 6, 2024
- truelife
Table of Contents
- Product Layout
- Battery Installation
- INSTRUCTIONS FOR USE
- Product Layout
- Battery Installation
- Using the Adapter
- Preparing to Take a Reading
- Taking a Reading
- Table of Values
- Error indicator
- Maintenance
- Precautions
- Safety precautions
- Safety notices for batteries:
- Caring for your device
- Other information
- Declaration of Conformity:
- Manufacturer:
- Documents / Resources
PULSE
Product Layout
Battery Installation
INSTRUCTIONS FOR USE
Before first use, the consumer must first read the instructions and safety precautions of the product.
Product Layout
a. Cuff tube inlet
b. Reading memory button
c. On / Off button
d. SET button
e. USB power input
f. Display
g. USB cable
Battery Installation
Remove the battery cover on the back of the blood pressure monitor and insert 4 AA 1.5V alkaline batteries in the direction as shown by the arrow. Make sure the batteries are inserted in the correct direction of polarity as shown in the battery compartment. Do not use batteries after the expiry date.
The batteries must be replaced if the battery symbol appears on the display
when the pressure monitor is turned on. The symbol indicates low battery
power. At critically low battery power levels, the device will not take
readings. The device is not suitable for use with rechargeable batteries.
If you do not intend to use the device for an extended period of time, remove
the batteries.
Using the Adapter
Do not connect the adapter while the device is turned on. When connected to an adapter, the device does not draw any battery power. If you disconnect the adapter from the device whilst taking a reading, you must plug the adapter back in again and repeat the reading.
Preparing to Take a Reading
– Before taking a reading, rest for 510 minutes, avoid eating, drinking,
smoking, exercising, and bathing.
– Connect the cuff hose to the body of the blood pressure monitor.
– Take off your outerwear. Keep on only thin clothing or remove all clothing
from your arm.
– Put the sleeve of the cuff on your arm to a level above your elbow. Always
measure on the same arm (ideally the left one).
– As the pressure changes during the day (usually the highest in the morning),
take measurements at the same time each day.
– Repeated measurements accumulate blood in the arm and may cause measurement
errors. The interval between measurements should be at least 1 minute. You can
shorten the interval by raising your arm so that the blood drains.
– Any attempt to support your arm whilst taking a reading may increase the
pressure and distort the results. Any activity, including speaking, affects
the measurement results.
– The arm on which you take the reading should be kept so that the artery is
roughly level with the heart. The measurement results will be more accurate.
– Place the cuff on the left arm with the tubing pointing downwards and the
bottom of the cuff approximately 23 cm above the elbow.
– Make sure that the cuff is not tightened too tightly. You should be able to
insert two fingers between the cuff and the arm.
– Place your hand upside down on the table so that you feel comfortable with
the cuff roughly at heart level. Do not bend the tubing.
– If it is not possible to take a reading on the left arm, use the right arm
instead. However, the reading should be taken on one arm only.
Taking a Reading
– Press the START / STOP button. All the symbols appear on the display. Then
the display shows 0 and the device starts inflating the cuff. You can follow
the cuff pressure on the display.
– Once the required cuff pressure has been reached, the pump stops inflating
the cuff and the pressure gradually decreases.
– When the device detects a signal, a flashing heart symbol appears on the
display. The symbol flashes once each time the heart beats.
– When the measurement is complete, a beep sounds and the display shows
systolic and diastolic pressure and heart rate.
– The measurement results will remain on the display until the device is
turned off. If idle, the device will turn itself off within 3 minutes. If the
monitor does not measure the results correctly, the display will show the “E”
icon.
– If for any reason it is necessary to interrupt the measurement (eg due to
nausea), press the START / STOP button. The device immediately relieves the
cuff pressure automatically.
Table of Values
Blood pressure levels| SBP
mmHg| DBP
mmHg
---|---|---
Optimal| < 120| < 80
Normal| 120-129| 80-84
High-normal| 130-139| 85-89
Grade 1 hypertension| 140-159| 90-99
Grade 3 hypertension| 160-179| 100-109
Grade 3 hypertension| >180| > 110
Error indicator
a) Symbol E-1
You may have an incorrectly attached cuff Adjust it to the correct position
and repeat the measurement
b) Symbol E-4
The Cuff hose is improperly inserted Check tubing connection and adjust
or reconnect if necessary. Then repeat the measurement.
Monitor or arm have moved whilst taking a reading Keep your body and the
monitor motionless whilst taking a reading. Repeat measurement.
The systolic pressure is less than 30 mmHg Manually increases the pressure
for the desired result. Repeat the measurement.
c) The E-5
Pressure is greater than 300 mmHg The device is releasing air itself.
d) Low Battery Symbol
Replace discharged batteries with new ones and repeat reading.
e) Arm symbol in motion
The symbol appears when the hand is moving excessively.
f) Heart Symbol
The symbol appears when the heart beats irregularly.
Measurement Method | Oscillometric |
---|---|
Pressure Measurement Range | 0-300 mmHg |
Heart Rate Measurement Range | 40-180 bpm |
Pressure Measurement Accuracy | ± 3 mmHg |
Heart Rate Measurement Accuracy | ± 5% |
Memory | 60 readings |
Power Supply | 4x AA Battery DC 6V or AC Adapter |
AutoOff | After 3 Minutes |
Pressure Monitor Weight | 300 g (without cuff and batteries) |
Pressure Monitor Dimensions | 120 × 93 ×59 mm (without cuff) |
Lifetime | approx. 10,000 readings |
Operating Temperature | 5-40 ° C |
Operating Humidity | 10-90% RH |
Maintenance
– After use, store the blood pressure monitor in its packaging and protect it
from damage.
– Store the blood pressure monitor in a dry environment.
– Keep the blood pressure monitor clean. Clean it with a dry cloth. Do not use
abrasive or corrosive cleaners.
– Neither the blood pressure monitor nor any of the components should come
into contact with water.
– Make sure the blood pressure monitor is off before cleaning. For cleaning it
is possible to use distilled water with a 10 % bleach solution.
– Use a spray, moisten a soft cloth with bleach or detergent until it is
soaked.
– Squeeze excess moisture from the fabric to prevent the unit from dripping or
cuff getting wet.
– Wipe the entire surface of the cuff thoroughly. Make sure you clean the
inside and outside of the cuff properly. Be careful not to let any moisture
get into the blood pressure monitor.
– Thoroughly wipe all parts of the cuff with a dry cloth. Unfold the cuff and
allow it to air dry.
– Do not clean the blood pressure monitor or the cuff with diesel, thinner, or
petrol.
– Do not expose the device to extreme heat or cold, moisture, and direct
sunlight.
– If you will not be using the device for more than 3 months, remove the
batteries.
– Elem6 s.r.o. cannot be held responsible for any damage caused by
insufficient maintenance and non
-observance of the instructions in the manual.
Precautions
– Use only the original charger supplied as an optional accessory to the blood
pressure monitor.
– Do not use the charger in a high humidity environment. Never touch the
charger with wet hands or while standing in water.
– When supplying power, leave enough space around the charger for air
circulation. Do not cover the charger with paper or other objects that could
lead to poor cooling. Do not use the charger stored in the transport
packaging.
– Do not use the charger or device if it is obviously damaged. If the device
is damaged, do not repair it yourself!
– In case of excessive heating, switch off the appliance immediately and
disconnect it from the power supply.
– Charge the device only under supervision.
Warning: This product is intended to be used in households and similar applications.
Importer:
elem6 s.r.o., Braskovská 15, 16100 Prague 6, Czech Republic, Made in P.R.C.
Well Kang Limited, The Black Church, St. Mary’s Place, Dublin 7, D07 P4AX, Ireland
Xiamen Ants-Bro Technology Co., Ltd, 4F, 5th Building, Technology business establishing center, No. 289 Wengjiao Road, Haicang District, Xiamen City, Fujian Province, China
Date of last revision: 2020-04
| Manufacturer
---|---
| Before first use, read the instruction manual
| Batch number
| Type BF
| Manufacturer’s authorized representative in the EU
The consumer is required to learn the principles of safe use of the product before first use.
Precautions and notices
– Always consult your doctor. Self-diagnosis and self-medication on the basis
of readings taken at home are dangerous.
– Before using this device, please read these Safety Instructions.
– When using this product, please refer to the most up-to-date version of the
User Manual, available for download at: http://www.truelife.eu
– Misprints in the manual are reserved.
Safety precautions
– If the product package includes a charger use only the charger supplied with
your device. Use of another type of charger may result in a malfunction and/or
danger.
– Do not use the charger in the high moister environment. Never touch the
charger when your hands or feet are wet.
– Allow adequate ventilation around the charger when using it to operate the
device or to charge the battery. Do not cover the charger with paper or other
objects that will reduce cooling. Do not use the charger while it is inside a
carrying case.
– Connect the charger to a proper power source. The voltage requirements are
found on the product case and/or on the packaging.
– Do not use the charger if it is visibly damaged. In case of damage, do not
repair the equipment yourself!
– In case of device overheating, immediately disconnect the power source.
– Charge the device under supervision only.
– Packaging contains small pieces, that can be dangerous for children. Store
the product always out of reach of children. Bags or many of the pieces
containing may cause suffocation after swallowing.
– Do not use this device at the same time as using another medical electrical
device.
– Do not use the device on the injured arm or on an arm undergoing treatment.
– Do not apply the cuff to an arm that has a drip inserted or on which blood
transfusions occur.
– Do not use this device near high-frequency surgical equipment, an MRI or CT
scanner or in an oxygen-rich environment.
– In the case of pregnancy, including preeclampsia, or diagnosed arrhythmia or
arteriosclerosis, consult your doctor before use.
– Before using this device on an arm with arteriovenous (A-V) fistula, contact
your doctor.
– People with severe blood circulation problems or blood disorders should
consult their doctor before using this device, as the inflation of the cuff
can cause reduced blood flow, which can lead to the formation of bruises.
– If you are experiencing unusual complications whilst taking a reading,
remove the cuff.
– Do not use the device near a mobile phone or other devices that transmit
electromagnetic waves.
– Do not use the device in a moving vehicle (car, airplane).
– Only use the cuff that is intended for this device.
– Do not bend the cuff by force and or bend the air tube excessively.
– Do not compress the air tube.
– Do not inflate the cuff unless it is secured around your arm.
– Do not use this device for any purpose other than measuring blood pressure.
– Do not use this device for neonatal patients, children, or persons who
cannot express their consent. Do not inflate the cuff more than is needed.
Safety notices for batteries:
– Keep the battery out of the reach of children and animals.
– Do not dispose of worn or damaged batteries into the trash, fire, or into
heating devices, but hand them in at collection points for hazardous waste.
– Never insert mechanically damaged or bloated batteries into the device.
Using batteries in this state may cause an explosion.
Caring for your device
– Taking good care of your device will ensure trouble-free operation and
reduce the risk of damage.
– Keep your device away from excessive moisture and extreme temperatures and
avoid exposing your device to direct sunlight or strong ultraviolet light for
extended periods of time.
– Do not drop your device or subject it to severe shock.
– Do not subject your device to sudden and severe temperature changes. This
could cause moisture condensation inside the unit, which could damage your
device. In the event of moisture condensation, allow the device to dry out
completely before use.
– The screen surface can easily be scratched. Avoid touching it with sharp
objects.
– Never clean your device when it is powered on. Use a soft, lint-free cloth
to wipe the screen and the exterior of your device. Do not use paper towels to
clean the screen.
– Never attempt to disassemble, repair, or make any modifications to your
device. Disassembly, modification or any attempt at repair could cause damage
to your device and even bodily injury or property damage and will void any
warranty.
– Do not store or carry flammable liquids, gases, or explosive materials in
the same compartment as your device, its parts or accessories.
– To discourage theft, do not leave the device and accessories in plain view
in an unattended vehicle.
– Overheating may damage the device.
Other information
-
For households: This marking indicates that this product should not be disposed of with other household wastes throughout the EU. To prevent possible harm to the environment or human health from uncontrolled waste disposal, recycle it responsibly to promote the sustainable reuse of material resources. To return your used device, please use the return and collection systems or contact the retailer where the product was purchased. They can take this product for
environmentally safe recycling. In case of incorrect disposal of this kind of wastes, the user can be fined according to national laws. User information for disposal of electric and electronic devices in countries outside the EU: The symbol stated above (crossed recycle bin) is valid only in countries in the European Union. For correct disposal of electric and electronic devices ask your city hall or distributor of these devices. Everything is expressed by the symbol of crossed recycle bin on the product, cover, or printed materials. -
You can determine a warranty service at the place of purchase. In case of technical problems or queries contact your dealer who will inform you about the following procedure. Follow the rules for work with electric devices. The user is not allowed to disassembly the device either replace any of its components. To reduce the risk of electric shock, do not remove or open the cover. In case of incorrect setup and plugin of the device, you will expose yourself to the risk of electric shock.
Products are covered by a two-year warranty unless otherwise stated.
Declaration of Conformity:
Hereby, elem6 s.r.o., declares that all TrueLife products are in compliance
with the essential requirements and other relevant provisions of Directive
93/42/EEC.
Products are intended for sale without any restriction in Germany, the Czech
Republic, Slovak Republic, Poland, Hungary, and other EU countries. You can
download the Declaration of Conformity from the website: www.
truelife.EU/support/doc
Manufacturer:
Xiamen Ants-Bro Technology Co., Ltd.
address: 4F, 5th Building, Technology Business Establishing Center, 289
Wengjiao Road, Haicang District, Xiamen City, Fujian Province, 361026, P.R.
China
Misprints and changes in the manual are reserved.