Corsano CardioWatch 287-1B Bracelet and Trials App Instruction Manual
- June 3, 2024
- Corsano
Table of Contents
CardioWatch 287-1B Bracelet and Trials App
EN
CardioWatch 287-1B
Bracelet & Corsano Trials App Instruction Manual
As of 2022-07-27 Revision status: 10
1
1 Table of Contents
2 INTRODUCTION ……………………………………………………………………………………………………….3
3 SAFETY INSTRUCTIONS ……………………………………………………………………………………………..3 3.1 Intended
Use……………………………………………………………………………………………………………….3 3.2 Receiving and Inspection
……………………………………………………………………………………………….3 3.3 Data Transmission
………………………………………………………………………………………………………..4 3.4 Handling and Usage
………………………………………………………………………………………………………5 3.5 Charger Handling and Usage
…………………………………………………………………………………………..5 3.6
Warnings…………………………………………………………………………………………………………………….6 3.7 Residual
risks……………………………………………………………………………………………………………….6 3.8 Clinical
benefit……………………………………………………………………………………………………………..6 3.9 Cleaning and service
life…………………………………………………………………………………………………6
4 SYMBOLS………………………………………………………………………………………………………………..8
5 CONTENTS / PRODUCT INCLUDES………………………………………………………………………………..9
6 KNOW YOUR BRACELET ………………………………………………………………………………………….. 10 6.1 Front of the
bracelet……………………………………………………………………………………………………10 6.2 Back and bottom of the
bracelet ……………………………………………………………………………………10 6.3 Charging the
bracelet…………………………………………………………………………………………………..11
7 USING YOUR BRACELET WITH A SMART DEVICE ………………………………………………………….. 13 7.1
Download and install the free “Corsano Trials” app onto your smart device.
………………………….13 7.2 First time use……………………………………………………………………………………………………………..13
7.3 Pairing Your Bracelet with a Smart Device ……………………………………………………………………….18
7.4 Troubleshooting the Bluetooth Connection ……………………………………………………………………..21
7.5 Troubleshooting the Cloud Connection……………………………………………………………………………22
8 NON-MEDICAL INFORMATION …………………………………………………………………………………. 23
9 HEALTH APP AND GOOGLE FIT INTEGRATION ……………………………………………………………… 28
10 CLINICAL PERFORMANCE ……………………………………………………………………………………… 29
11 SPECIFICATION …………………………………………………………………………………………………… 30
12 ELECTRICAL SAFETY AND ELECTROMAGNETIC COMPATIBILITY ……………………………………. 31
13 DISPOSAL OF THE DEVICE …………………………………………………………………………………….. 35
14 CORSANO CONTACT INFORMATION ………………………………………………………………………. 36
As of 2022-07-27 Revision status: 10
2
2 INTRODUCTION
Thank you for purchasing the Corsano CardioWatch 287-1B. The CardioWatch
bracelet records raw PPG and ACC data, which is transferred to the application
and cloud.
Irregular heartbeats (e.g. extrasystoles) and atrial fibrillation can only be
diagnosed in accordance with the guidelines with an ECG of the thoracic wall,
generally carried out by cardiologists.
If you are feeling unwell or experience other troubling symptoms, please seek
medical advice immediately.
3 SAFETY INSTRUCTIONS
This instruction manual provides you with important information about the
Corsano CardioWatch 287-1B Bracelet. To ensure the safe and proper use of this
bracelet, READ and UNDERSTAND all of the safety and operating instructions. If
you do not understand these instructions or have any questions, contact
support@corsano.com before attempting to use this bracelet. For specific
information about your own heartbeats, consult with your physician.
3.1 Intended Use
This device is a digital bracelet intended for use in measuring pulse rate and
activity in adult patient population. Environments of Use: Hospital and Home
Patient Population: Adult
3.2 Receiving and Inspection
Remove this bracelet and other components from the packaging and inspect for damage. If this bracelet or any other components is damaged, DO NOT USE and contact support@corsano.com
Read the Important Safety Information in this instruction manual before using this bracelet. Follow this instruction manual thoroughly for your safety.
Keep for future reference. For specific information about your own heartbeats, CONSULT WITH YOUR PHYSICIAN.
· DO NOT use this bracelet on infants, toddlers, children or persons who cannot express themselves.
· DO NOT adjust medication based on readings from this bracelet. Take medication as
As of 2022-07-27 Revision status: 10
3
prescribed by your physician. ONLY a physician is qualified to diagnose and
treat high or irregular heartbeats.
· DO NOT use this bracelet on an injured arm or an arm under medical
treatment.
· DO NOT use this bracelet in areas containing high frequency (HF) surgical
equipment, magnetic resonance imaging (MRI) equipment, computerized tomography
(CT) scanners. This may result in incorrect operation of the bracelet and/or
cause an inaccurate reading.
· DO NOT take recordings in close vicinity to strong electromagnetic fields
(e.g. electromagnetic anti-theft systems, metal detectors).
· DO NOT use this bracelet in oxygen rich environments or near flammable gas.
· Consult with your physician before using this bracelet if you have common
arrhythmias such as atrial or ventricular premature beats or atrial
fibrillation; arterial sclerosis; poor perfusion; diabetes; pregnancy; pre-
eclampsia or renal disease. NOTE that any of these conditions in addition to
patient motion, trembling, or shivering may affect the measurement reading.
· NEVER diagnose or treat yourself based on your readings. ALWAYS consult with
your physician.
· To help avoid strangulation, keep the charger cable away from infants,
toddlers or children.
· This product contains small parts that may cause a choking hazard if
swallowed by infants, toddlers or children.
3.3 Data Transmission
This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT use this
product in locations where RF is restricted, such as on an aircraft or in
hospitals. Turn off the Bluetooth® feature in this bracelet and remove
batteries and/or unplug the charger when in RF restricted areas. For further
information on potential restrictions refer to documentation on the Bluetooth
usage by the FCC.
As of 2022-07-27 Revision status: 10
4
3.4 Handling and Usage
· Stop using this bracelet and consult with your physician if you experience
skin irritation or discomfort.
· Consult with your physician before using this bracelet on an arm where
intravascular access or therapy, or an arterio-venous (A-V) shunt, is present
because of temporary interference to blood flow, which could result in injury.
· Consult with your physician before using this bracelet if you have severe
blood flow problems or blood disorders.
· DO NOT use this bracelet for any purpose other than measuring heartbeats.
· During measurement, make sure that no mobile device or any other electrical
device that emits electromagnetic fields is within 12 inches (30 cm) of this
bracelet. This may result in incorrect operation of the bracelet and/or cause
an inaccurate reading.
· DO NOT disassemble or attempt to repair this bracelet or other components.
This may cause an inaccurate reading.
· DO NOT drop or subject this bracelet to strong shocks or vibrations.
· DO NOT use this bracelet with other medical electrical (ME) equipment
simultaneously. This may result in incorrect operation of the bracelet and/or
cause an inaccurate reading.
· Ensure that this bracelet has acclimated to room temperature before taking a
measurement. Taking a measurement after an extreme temperature change could
lead to an inaccurate reading.
· Ensure the bracelet is well adjusted on the wrist to have the best
performance of the heart rate sensor, and not too tight to avoid skin injuries
3.5 Charger Handling and Usage
· USE the charger cable with a CE marked adapter with the following characteristics: o Input voltage: 100/240 V – 50hz 60hz o Output voltage: DC 5V (+-5%) o Maximum current: 500 mA
· DO NOT use the charger if this bracelet or the charger cable is damaged. If this bracelet or the cable is damaged, unplug the charger immediately.
· Plug the charger into the appropriate USB outlet. DO NOT use in a multi- outlet plug.
· NEVER plug in or unplug the charger from the electric outlet with wet hands.
· DO NOT disassemble or attempt to repair the charger.
As of 2022-07-27 Revision status: 10
5
· Fully insert the USB plug at the end of the charger into the USB outlet.
· When unplugging the charger from the outlet, be sure to safely pull from the
USB outlet. DO NOT pull from the charger cable.
· When handling the charger cable:
DO NOT damage it. DO NOT break it. DO NOT tamper with it. DO NOT forcibly bend
or pull it. DO NOT twist it. DO NOT use it if it is gathered in a bundle. DO
NOT pinch it. DO NOT place it under heavy objects.
· Wipe any dust off of the charger.
· Unplug the charger when not in use.
· Unplug the charger before cleaning this bracelet.
3.6 Warnings
Regardless of the measurement taken using this device, you should immediately
consult your practitioner when you experience symptoms that could indicate a
disease, such as chest pain, pressure, tightness, etc.
You may be experiencing a cardiac arrhythmia or other disease even in the
absence of a notification from the APP. You should notify your practitioner at
any changes of your health condition.
If a serious incident occurs in relation to the device, it must be reported to
the manufacturer and the competent authority.
3.7 Residual risks
In rare cases, the device may detect arrhythmia while you experienced no cardiac problems. You should contact your practitioner to get a diagnostics confirmation.
3.8 Clinical benefit
The CardioWatch 287-1B provides a non-invasive and comfortable solution to
continuously and accurately monitor vital signs and enables off-line analysis
and screening of cardiac arrhythmia (AFib for instance) by third-party medial
applications.
3.9 Cleaning and service life
Use a lint-free cloth moistened with warm water to clean the housing and casing of your device.
As of 2022-07-27 Revision status: 10
6
Use warm water and hypoallergenic soap to clean the strap. Dry the wrist band
with a soft cloth. It is not necessary to sterilize the device.
The Smartwatch is an electronic device with rechargeable battery. The expected
service life is 5 years.
As of 2022-07-27 Revision status: 10
7
4 SYMBOLS
These instructions for use contain the following symbols (color and size may
vary):
Symbol
1912
Meaning
This stand-alone software is a medical device classified as risk category IIa,
in accordance with rule 10 of EU Directive 93/42/EEC, last amended by
2007/47/EC of the European Parliament and Council of 5 September 2007.
Indicates the medical device manufacturer
Warning Indicates the need for the user to consult the instructions for use
for important information such as warnings and cautions. A warning is always
related to safety. Note Indicates the need for the user to consult the
instructions for use
Applied Part TYPE BF Applied Part (IEC 60417-5333)
Indicates the manufacturer’s catalogue number so the medical device can be
identified
Indicates the manufacturer’s serial number so that a specific medical device
can be identified
CE marking indicates that a product complies with applicable European Union
regulations
FCC marking indicates the electronic device, which sold in the United States,
is certified and the electromagnetic interference from the device is under the
limits that are approved by Federal Communications Commission
Indicates a product should not be disposed of in a landfill; the black bar
indicates that the equipment was manufactured after 2005
As of 2022-07-27 Revision status: 10
8
5 CONTENTS / PRODUCT INCLUDES
Figure 1 -Bracelet (CS-287-1B) Figure 2 – Charger (CS-287CH-1) Figure 3 – Instruction Manual (CS-287IFUEN-1)
As of 2022-07-27 Revision status: 10
9
6 KNOW YOUR BRACELET
6.1 Front of the bracelet
(A) Ambient Temperature Sensor (B) LEDs (C) Pusher button
Figure 4 Front view of Bracelet
6.2 Back and bottom of the bracelet
(D) Temperature Sensors (E) PPG Sensor (F) Magnets (G) Charge Contacts
Figure 5 Back view of Bracelet
As of 2022-07-27 Revision status: 10
10
6.3 Charging the bracelet
Figure 6 Charging view
Attach the charger cable to the backside of the bracelet. The magnets will
pull the charger head to the bracelet.
Figure 7 Charging view charger attached
The Magnets will click the charger into position. The LEDS will light up to
indicate that charging has started. While charging, the Bracelet will not
perform any measurement.
As of 2022-07-27 Revision status: 10
11
Figure 8 Charging view charger attached #2
The polarity of the magnets in the bracelet and the charger will ensure that
the charger contacts will align.
Figure 9 Bracelet side view
LED
Pattern
State
Green (B1)
Flashing
Bracelet charging
Green (B1) Orange (B2) Orange (B2) Orange (B2)
OFF Flashing for 5 sec ON for 5 seconds ON for 10 seconds
Bracelet fully charged or not on charger After the pusher is pressed, when the Bracelet is looking for the Bluetooth Low Energy connection After the pusher is pressed, when the Bracelet is connected to a smartphone After a successful pairing with the APP.
When the bracelet is close to the end of its battery autonomy, the user gets a notification through the mobile APP (20% remaining).
As of 2022-07-27 Revision status: 10
12
7 USING YOUR BRACELET WITH A SMART DEVICE
7.1 Download and install the free “Corsano Trials” app onto your smart device.
7.2 First time use Select Sign Up (2) to create your account. If you have an
account, select Sign In (1)
1
2
As of 2022-07-27 Revision status: 10
13
Sign Up
Enter your Email Address, First Name and Last Name. Agree to Terms of Service and Privacy Policy. Press the Sign Up button
As of 2022-07-27 Revision status: 10
14
You will receive an email with a link to the confirmation page. Please fill your password and confirm password. When ready, press the Continue button.
As of 2022-07-27 Revision status: 10
15
A password should have at least 8 characters, at least one letter a, b, c…, at
least one number, and both Upper and Lower case characters.
Password must NOT contain only numbers or consecutive characters
As of 2022-07-27 Revision status: 10
16
After you have signed up, you can Sign-In:
As of 2022-07-27 Revision status: 10
17
7.3 Pairing Your Bracelet with a Smart Device
Upon first sign-in, user will be asked to pair a bracelet, follow the
instructions. First, select “287-1B” in the list:
As of 2022-07-27 Revision status: 10
18
Press the pusher, the orange LED will flash, the bracelet is waiting for pairing.
Press on button in the app to start pairing.
After button is pressed app will look for the bracelet
As of 2022-07-27 Revision status: 10
19
Once the device is found, confirm by pressing on the pusher. On successful pairing, the bracelet will light the orange LED up for 10 seconds and then turn it OFF.
The app will display the Serial Number found. Please double check the S/N on the device and click Confirm. The devices is now paired to the app.
Figure 10 APP Pairing, complete
When your bracelet is connected successfully to your smart device, the green “V” symbol appears under Settings:
As of 2022-07-27 Revision status: 10
20
7.4 Troubleshooting the Bluetooth Connection
If the connection between the bracelet and app is lost, a red “X” will appear:
Click on the button Please Reconnect and follow instructions: 1. Make sure
your phone is nearby 2. Check if watch is charged 3. Check if GPS is on
(Android only) 4. Force quit the app on your phone 5. Turn your phone’s
Bluetooth off and on again 6. Re-open the app
If these steps did not reconnect, please proceed: 1. Shut down your phone (do
not do restart; completely shut the phone off and turn it back on). This will
fully reset the Bluetooth system in the phone 2. Turn your phone on again 3.
Re-open the app
If none of the above did not resolve, you will need to re-pair your bracelet:
- Go to watch settings: Remove/Clean old pairing 2. Go to Bluetooth settings, find 287, 286, 284 > Click Forget Device/Unpair 3. Force quit the app on your phone 4. Re-open the app 5. Press the (+) inside the watch icon in upper right corner 6. Follow pairing instructions
As of 2022-07-27 Revision status: 10
21
7.5 Troubleshooting the Cloud Connection
If the connection between the app and the cloud is lost, a red “X” will
appear:
Click on the button Please Reconnect and follow instructions: 1. Make sure
your phone is connected to internet 2. Force quit the app on your phone 3.
Turn your phone’s Airplane Mode off and on again 4. Re-open the app
If these steps did not reconnect, please proceed: 1. Shut down your phone (do
not do restart; completely shut the phone off and turn it back on). This will
fully reset your phone 2. Turn your phone on again 3. Re-open the app
If none of the above did not resolve, you will need to to logout and login
again to the cloud:
1. Go to profile settings and log out 2. Force quit the app on your phone 3.
Re-open the app 4. Enter user and password 5. Follow instructions
As of 2022-07-27 Revision status: 10
22
8 RECORD PATIENTS
Click on “New Patient”
Click on “Start Measurement”
During the measurement, please do not remove the app from multitasking. The screen of the phone can be locked. Please keep an internet connection on the phone while measuring. Please keep the phone close to the bracelet (few meters maximum) when measuring.
As of 2022-07-27 Revision status: 10
23
When you want to stop the
measurement, click on “Stop Measurement”
The app will stop the measurement. At this stagethe note is still editable.
Click on “Done”
You can start these steps again to record other patients. Confirm the stopping of the measurement by clicking on “Yes”
As of 2022-07-27 Revision status: 10
24
9 NON-MEDICAL INFORMATION
The Corsano Trials app provides non-medical information like Activity:
As of 2022-07-27 Revision status: 10
25
Sleep:
As of 2022-07-27 Revision status: 10
26
Heart Rate:
As of 2022-07-27 Revision status: 10
27
10 HEALTH APP AND GOOGLE FIT INTEGRATION
The Corsano Trials app is integrated with the Apple Health and Google Fit
apps. Information about your activities and vitals helps us provide you with a
monthly (premium) personal report. You have the option of turning off the
integration within the Apple Health and Google Fit apps; however, doing so
will limit the information available in your personal report. The Corsano
Trials app shares the following pieces of information with the Apple Health
and Google Fit app:
· Heart Rate · Blood pressure · Height · Weight
The Corsano Trials app collects the following pieces of information from the
Apple Health and Google Fit apps:
· Active Energy · Blood Glucose · Diastolic Blood Pressure · Flights Climbed ·
Heart Rate · Height · Oxygen Saturation · Resting Energy · Sleep Analysis ·
Steps · Systolic Blood Pressure · Walking + Running Distance · Weight ·
Workouts
As of 2022-07-27 Revision status: 10
28
11 CLINICAL PERFORMANCE
Heart Rate
RR Interval
Heart Rate Variability Respiration Rate
Sleep Stages
Sleep Score
Definition
Unit
Number of beats of the heart per minute
Elapsed time between two consecutive heart beats
beats/minute (bpm)
msec
Range 30-220
Acquisition Time
5-10 sec
300-2000 ms 5-10 sec
Beat to beat (RR interval) variations Number of breaths (inhalation –
exhalation cycles) per minute Detection of specific sleep stages & sleep HR
Sleep performance and sleep consistency with equal weight
msec
0-200 ms
breaths/minute 5-45 brpm (brpm)
awake, light sleep, deep sleep, REM %
sleep stage 0-100%
5-10 sec
20-30 sec
upon end of the entire sleep event 10 sec
Update interval 1 sec 1 sec
1 sec 1 sec
1 min
1 sec
Accuracy *
HR is ±4 bpm MAD, ±5% MARD (at rest) RR Interval ±50 ms MAD, ±5% MARD (at
rest) HRV ±10 ms MAD, ±5% MARD (at rest) Respiration Rate ±1 bprm MAD, ±5%
MARD (at rest) Sleep Stage ±10 % MAD
Sleep Score ±5 % MAD
Remarks: * MAD=Mean absolute difference & MARD=Mean absolute relative difference under motion
As of 2022-07-27 Revision status: 10
29
12 SPECIFICATION
Minimum requirements for mobile device Operating Systems: · iOS 12.2 or higher
· Android 8.0 or higher
PPG Sensor Characteristics PPG……………………………………………………………………………………..Single Channel
PPG LEDs number ………………………………………………………………..2 PPG LEDs Peak wavelength
………………………………………………….525 nm PPG LEDs max current ………………………………………………………….30 mA
PPG sampling resolution………………………………………………………..19 bits
Motion Sensor Characteristics Type ………………………………………………………………………………………….. 3-axis
Acquisition noise…………………………………………………………………………. 1.3 mg RMS Sensor range
…………………………………………………………………………….. ±16 g full scale
Data Acquisition PPG sampling rate …………………………………………………………………….. 25 Hz Motion
sampling rate …………………………………………………………………… 100 Hz Flash Memory Size
……………………………………………………………………… 64 Mbit Recording
………………………………………………………………………………….. Continuous
Power Requirements Average current …………………………………………………………………….1 mA Max current
consumption ………………………………………………………. 45 mA Battery
type…………………………………………………………………………..Rechargeable Technology
…………………………………………………………………………..Lithium Polymer Battery capacity (Bracelet)
……………………………………………………..93 mAh Autonomy (Bracelet)………………………………………………………………up to
1 week
Dimensions Length x Width x Height ……………………………………………..25mm x 37mm x 9mm
Environmental Specifications Ingress Protection……………………………………………………………… IP66
Operational Temperature …………………………………………………… +10 to +40 degrees C Transport
and storage Temperature ……………………………………. -20 to +60 degrees C Operational
Humidity………………………………………………………….. 20% to 80% Transport and storage
Humidity………………………………………….. 20% to 90%
Interface Wireless Communication………………………………………………………..BLE 5.0 Display LEDs
………………………………………………………………………..1 green, 1 orange User
action……………………………………………………………………………pusher button
PPG sensor are exposed on the back of the Corsano bracelet. The PPG sensor
makes contact with the user’s skin.
As of 2022-07-27 Revision status: 10
30
13 ELECTRICAL SAFETY AND ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacturer’s declaration – electromagnetic emissions
The Corsano Bracelet is intended for use in the electromagnetic environment specified below. The customer or the user of Corsano Bracelet should assure that it is used in such an environment.
Emissions test Compliance RF emissions Group 1 CISPR 11
RF emissions Class B
CISPR 11
Harmonic
Class A
emissions
IEC 61000-3-2
Voltage
fluctuations /
Complies
flicker emissions
IEC 61000-3-3
Electromagnetic environment – guidance
The Corsano Bracelet uses RF energy only for its internal function. Therefore,
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The Corsano Bracelet is suitable for use in all establishments other than
domestic and those directly connected to the public low- voltage power supply
network that supplies buildings used for domestic purposes.
IEC 60417-5333
TYPE BF APPLIED PART
As of 2022-07-27 Revision status: 10
31
Guidance and manufacturer’s declaration–electromagnetic immunity
Corsano bracelet is intended for use in the electromagnetic environment
specified below. The customer or the user of Corsano bracelet should assure
that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
Electrostatic Discharge (ESD) IEC 61000-4-2
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burs t IEC 610004-5
Surge IEC 61000-4-6
Power frequency (50/60Hz) magnetic field IEC 61000-4-8
±2 kV for power supply lines ±1 kV for input/outpu t lines
±1 kV differential mode ±2 kV common mode
3 A/m
±2 kV for power supply lines ±1 kV for input/output lines
Mains power quality should be that of a typical commercial or hospital environment.
±1 kV differential mode ±2 kV common mode
3 A/m
Mains power quality should be that of a typical commercial or hospital
environment.
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
As of 2022-07-27 Revision status: 10
32
Guidance and manufacturer’s declaration–electromagnetic immunity
Corsano bracelet is intended for use in the electromagnetic environment specified below. The customer or the user of Corsano bracelet should assure that it is used in such an environment.
Immunity test IEC 60601 Compliance
test
level
Electromagnetic environment – guidance
Conducted
level
3 Vrms
NA
Portable and mobile RF communications equipment should be used no closer to any part of Corsano bracelet,
including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance (m)
Conducted RF 150 kHz to
IEC 61000-4-6 80
NA
MHz
d = 1.2 P d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 800 MHz to 2.5 GHz
Radiated RF 3 V/m
NA
IEC 61000-4-3 80 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a, should be less than the compliance level in each frequency range b. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1–At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2–These
guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people
a Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast, and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which Corsano bracelet is used
exceeds the applicable RF compliance level above, Corsano bracelet should be
observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating
Corsano Bracelet.
As of 2022-07-27 Revision status: 10
33
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications
equipment and Corsano
Bracelet
Corsano Bracelet is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the
Corsano bracelet can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Corsano bracelet as recommended below, according to the
maximum output power of the communications equipment.
Separation distance according to frequency of
Rated maximum output power of
transmitter W
150 kHz to 80 MHz
d = 1.2 P
transmitter (m)
80 MHz to 800
800 MHz to 2.5
MHz
GHz
d = 1.2 P
d = 2.3 P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1–At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2–These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
As of 2022-07-27 Revision status: 10
34
14 DISPOSAL OF THE DEVICE
Once your bracelet has reached its end of life it has to be properly recycled
so that the material can be reused and will not end up in the environment.
Preferably bring your device to a recycling service for Waste Electrical and
Electronic Equipment.
As of 2022-07-27 Revision status: 10
35
15 CORSANO CONTACT INFORMATION
Corsano Health B.V. Wilhelmina van Pruisenweg 35 2595 AN The Hague The
Netherlands
www.corsano.com
As of 2022-07-27 Revision status: 10
36
References