nuwave LumAir LA100 Pathogen Detection System User Guide

June 6, 2024
nuwave

nuwave LumAir LA100 Pathogen Detection System - logo LumAir LA100
Pathogen Detection System
Cartridge Loading/Unloading Guide

This guide is intended to provide instructions on how to insert or remove a cartridge from an installed LA100 Pathogen Detection System.

nuwave LumAir LA100 Pathogen Detection System

Note: Please see the cartridge data sheet for detailed information on the application, use, and disposal of cartridges

Introduction

This guide is intended to provide instructions on how to insert or remove a cartridge from an installed LumAir LA100 Pathogen Detection System.

nuwave LumAir LA100 Pathogen Detection System - figure
1

Package Contents

nuwave LumAir LA100 Pathogen Detection System - figure
2| nuwave LumAir LA100
Pathogen Detection System - figure 3
---|---
Cartridge (in Foil Bag)| Sterile Gloves
nuwave LumAir LA100 Pathogen Detection System - figure
4| nuwave LumAir LA100
Pathogen Detection System - figure 5
Bio-Hazard Safety Disposal Bag| Cartridge Removal Tab

Step 1: Inserting a New Cartridge (visual steps)

1. Put on sterile gloves using sterile technique| 2. Remove Cartridge from foil pack & zip lock bag
---|---
nuwave LumAir LA100 Pathogen Detection System - figure
6| nuwave LumAir LA100
Pathogen Detection System - figure 7
3. Peel back sticker to remove cartridge pull tab| 4. Once the pull tab is removed, re-seal with sticker
nuwave LumAir LA100 Pathogen Detection System - figure
8| nuwave LumAir LA100
Pathogen Detection System - figure 9
5. Insert cartridge fully into the unit| 6. LED status changes to green Active
nuwave LumAir LA100 Pathogen Detection System - figure
10| nuwave LumAir LA100
Pathogen Detection System - figure 11

Step 2 – Removing the Cartridge

1. Put on sterile gloves using sterile technique| 2. Push button, turning the LED status light to blue
---|---
nuwave LumAir LA100 Pathogen Detection System - figure
12| nuwave LumAir LA100
Pathogen Detection System - figure 13
3. Peel back the sticker fully| 4. Insert the locking slide (included in cartridge box)
nuwave LumAir LA100 Pathogen Detection System - figure
14| nuwave LumAir LA100
Pathogen Detection System - figure 15
5. Ensure locking slide is completely inserted| 6. Remove cartridge and place in yellow Biohazard bag – Seal bag
nuwave LumAir LA100 Pathogen Detection System - figure
16| nuwave LumAir LA100
Pathogen Detection System - figure 17

For questions not covered in this guide:
Email:[email protected]
Tel: (+353) 1 254 4188 (International)

Documents / Resources

| nuwave LumAir LA100 Pathogen Detection System [pdf] User Guide
nuwave, LumAir, LA100, Pathogen, Detection, System, Cartridge Loading, Unloading Guide
---|---

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

nuwave User Manuals

Related Manuals