BioFire DFA2-ASY-0004 Global Fever Panel User Guide

June 6, 2024
BioFire

BioFire DFA2-ASY-0004 Global Fever Panel User Guide

For In Vitro Diagnostic Use.

NOTE: Components from this kit should not be stored or used with any other kit.

Components

To avoid contamination, always wear clean gloves and work behind a protective shield.

Step 1: Prepare Pouch

Prepare Pouch

a. Insert pouch into Pouch Loading Station.
b. Remove clear cap from Sample Injection Vial and place into red well.
c. Place Hydration Injection Vial into blue well.

Step 2: Hydrate Pouch

a. Unscrew Hydration Injection Vial from cover.
b. Remove Hydration Injection Vial, leaving blue cover in Pouch Loading Station.
c. Insert Hydration Injection Vial into hydration port.
d. Push down to puncture seal and wait as Hydration Solution is drawn into the pouch.

NOTE: Verify the pouch has been hydrated.

Step 3: Prepare Sample Mix

Prepare Sample Mix

NOTE: Gently invert whole blood sample container until thoroughly mixed.

a. Use the Transfer Pipette to draw sample to the 2 nd line. Add the sample to Sample Injection Vial.
b. Hold the Sample Buffer Tube with the tip facing up and firmly pinch at textured plastic tab on side of tube until the seal snaps.

NOTE: Do not touch the tip of the tube.

c. Dispense Sample Buffer into Sample Injection Vial using a slow, forceful squeeze, followed by a 2 nd squeeze.

NOTE: Avoid generating excessive foam.

d. Tightly close lid and invert the Sample Injection Vial 3 times.
e. Return Sample Injection Vial to red well of Pouch Loading Station.

WARNING: Sample Buffer is harmful if swallowed and can cause serious eye damage and/or skin irritation.

Step 4: Load Sample Mix

Load Sample Mix

a. Unscrew Sample Injection Vial from red cover.
b. Wait for 5 seconds, then remove Sample Injection Vial, leaving red cover in Pouch Loading Station.

NOTE: Waiting 5 seconds decreases the contamination risk.

c. Insert Sample Injection Vial into pouch sample port.
d. Push down to puncture seal, then wait as sample mix is drawn into the pouch.

Step 5: Run Pouch

Run Pouch

a. Screw vials back into covers in Pouch Loading Station before disposing of them in a biohazard container.
b. Remove pouch from Pouch Loading Station and load into the instrument.

Optional added operator protection: Before removal from biosafety cabinet, run a bleach wipe, a paper towel with 10% bleach (one part bleach to nine parts water) across the top of the pouch from the pouch hydration port to the pouch sample port, and follow with a water wipe. This reduces the potential for contact with small amounts of sample mixed with sample buffer that may be retained at the sample injection port.

c. Follow instructions on computer for starting a test.

NOTE: Select the GF Blood protocol.

Step 6: Review Report

Run Information – Displays information about the sample, protocol, pouch, instrument, run status, operator, and results.

Review Report

  1. Internal Controls:

    • If ‘Passed’, results are valid.
    • If ‘Failed’ or ‘Invalid’, RETEST SAMPLE.
  2. Run Status:

    • If ‘Completed’, run is complete.
    • If ‘Incomplete’, ‘Aborted’, or any other error message, RETEST SAMPLE.
  3. Results Banner: Displays the test results.

    • Follow the instructions listed in the banner.
      Result Summary: Displays the test results for each organism.
  4. ‘Detected’: Lists names of any detected organisms.

  5. ‘Not Detected’: Lists names of organisms that were not detected.

NOTE: Refer to the BioFire Global Fever Panel Instructions for Use for reporting information. If repeated error messages are obtained, contact BioFire Defense Technical Support.

BioFire Defense, LLC

79 West 4500 South, Suite 14, Salt Lake City, Utah 84107 USA
801-262-3592
www.BioFireDefense.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals