BioFire SHIELD Control Kit COVID-19 Test

DO NOT DISCARD: Important product-specific information

For in vitro diagnostic use under Emergency Use Authorization (EUA) only.
Icons

Components not included

Materials required but not provided

  • External Control Vial
    Components Required

  • Pouch
    BioFire® COVID-19 Test
    Part No: 423744 423745
    Components Required

  • Hydration Injection Vial

  • Transfer Pipette
    Components Required

  • Sample Injection Vial
    Components Required

  • Sample Buffer Tube

  • Transport Medium, PBS or Saline
    Components Required

NOTE: FilmArray® Instrument should be powered on and ready for use prior to pouch preparation.
NOTE: Use clean gloves and other Personal Protective Equipment (PPE) when performing this procedure

STEP 1: Prepare Pouch

Refer to the BioFire® COVID-19 Test Upper Respiratory Quick Guide
STEP 1: Prepare Pouch
Prepare Pouch

STEP 2: Hydrate Pouch

Refer to the BioFire® COVID-19 Test Upper Respiratory Quick Guide
STEP 2: Hydrate Pouch
Hydrate Pouch

  • To prepare a Positive External Control, proceed to STEP 3a.
  • To prepare a Negative External Control, proceed to STEP 3b.

STEP 3a: Prepare COVID-19 Positive External Control

  1. a. Use the Transfer Pipette to draw the Transport Medium, PBS or Saline to the third line. Add to the Sample Injection Vial.
    Prepare COVID-19

  2. b. Remove rubber cap from Positive External Control Vial and place on a clean surface (a paper towel may be used).

  3. c. Add Sample Buffer to Positive External Control Vial

  4. Hold the Sample Buffer Tube with the tip facing up and firmly pinch at the textured plastic tab on the side of the tube until seal snaps.
    Prepare COVID-19
    NOTE: Do not touch tube tip.

  5. Dispense Sample Buffer into Positive External Control Vial using a slow, forceful squeeze, followed by a second squeeze.
    NOTE: Avoid generating excessive foam.

  6. d. Recap Positive External Control Vial and invert it 3 times to mix.

  7. e. Pour mixture of Sample Buffer and Control into the Sample Injection Vial.

  8. f. Dispose of Positive External Control Vial and change gloves.

  9. g. Tightly close lid of Sample Injection Vial and invert it 3 times, return it to the red well of Pouch Loading Station.
    Prepare COVID-19
    WARNING: Sample Buffer is harmful if swallowed and can cause serious eye damage and/or skin irritation.

  10. h. Proceed to STEP 4: Load Control Mix.

STEP 3b: Prepare COVID-19 Negative External Control

  1. i. Use the Transfer Pipette to draw the Transport Medium, PBS or Saline to the third line. Add to the Sample Injection Vial.

  2. j. Add Sample Buffer to Sample Injection Vial.

  3. Hold the Sample Buffer Tube with the tip facing up and firmly pinch at the textured plastic tab on the side of the tube until seal snaps.
    NOTE: Do not touch tube tip.

  4. Dispense Sample Buffer into Sample Injection Vial using a slow, forceful squeeze, followed by a second squeeze.
    NOTE: Avoid generating excessive foam.

  5. k. Tightly close lid of Sample Injection Vial and invert 3 times, return it to the red well of Pouch Loading Station.
    Prepare COVID-19
    WARNING: Sample Buffer is harmful if swallowed and can cause serious eye damage and/or skin irritation.

  6. l. Proceed to STEP 4: Load Control Mix.

STEP 4: Load Control Mix

Refer to the BioFire® COVID-19 Test Upper Respiratory Quick Guide
STEP 4: Load Sample Mix
Load Control Mix

STEP 5: Run Pouch

Refer to the BioFire® COVID-19 Test Upper Respiratory Quick Guide
STEP 5: Run Pouch
Run Pouch

NOTE: Select either Positive External Control v3.2 or Negative External Control v3.2.

STEP 6: Review Results

Run Information – Displays information about the pouch, protocol, run status, operator, serial number, instrument and lot number

  1. Internal Controls:
    • If ‘Passed’, results are valid.
    • If ‘Failed’ or ‘Invalid’, RETEST CONTROL once.
  2.   Run Status:
    • If ‘Completed’, run is complete.
    • If ‘Aborted’, or any other error message, RETEST CONTROL once.
  3.   External Control Results:
    • If ‘Passed’, results are valid.
    • If ‘Failed’,
    • Positive External Control: RETEST EXTERNAL CONTROL once.
    • Negative External Control: Decontaminate the area; and RETEST EXTERNAL CONTROL once.
    •  If ‘Invalid’, RETEST EXTERNAL CONTROL once.

NOTE: Refer to Instructions for Use for reporting information. If repeated error messages are obtained, contact BioFire Defense Technical Support.

Conditions of Authorization

The BIOFIRE SHIELD Control Kit for the BioFire COVID-19 Test has not been FDA cleared or approved but has been authorized for emergency use by FDA under an EUA for use by authorized laboratories.

The emergency use of the BIOFIRE SHIELD Control Kit for the BioFire COVID-19 Test is only authorized for the duration of the declaration that circumstances exist justifying the authorization of emergency use of in vitro diagnostics for detection and/or diagnosis of COVID-19 under Section 564(b)(1) of the Federal Food, Drug and Cosmetic Act, 21 U.S.C. § 360bbb-3(b)(1), unless the declaration is terminated or authorization is revoked sooner.