MSDR-D-KERN Digital Refractometer User Manual

June 6, 2024
KERN

MSDR-D-KERN Digital Refractometer

MSDR-D-KERN Digital Refractometer
product

Introduction

Panel DescriptionsMSDR-D-KERN Digital Refractometer
fig-1MSDR-D-KERN Digital Refractometer
fig-2

The Packing Accessories

  • Packaging x1 The instructions x1 Dropper x1
  • Screwdriver x1 USB Line x1 Charger x1

Display and Buttons

Display Areas and ButtonsMSDR-D-KERN Digital Refractometer
fig-3

Preparations before Operating

Install the Battery

Use the screwdriver to open the battery hatch.

USB Connection
Connect to USB for charging/connecting to the computer.

Booting and Measurement

BootingMSDR-D-KERN Digital Refractometer fig-7

Note :

  1. When used outdoors,please avoid strong light so as not to affect the measurement accuracy.
  2. Before dripping into the sample liquid, please clean the sample plate and Prism with soft clean cloth or soft paper.
  3. Please keep the instrument in a stable status to measure.
  4. Please ensure instrument, environment and sample are in the same temperature level before measuring.

Measurement

After tum on,clear the distilled water and dry the sample plate, drip 0.3~0.4ml sample then close the cover to measure.MSDR-D-KERN Digital
Refractometer fig-8

MSDR-D-KERN Digital Refractometer fig-9

If press the “Read” button for 2 seconds, the instrument would make the automatic measurements upon programmed times (default 15times), the final value is the average of 15 times’ measurements, After measurements, it will return to the normal state
and the scales display area would display the average of the automated measurements

MSDR-D-KERN Digital Refractometer fig-10

The Calibration

The meter only supports pure water calibration. The calibration method is as follows: drop 0.4ml pure water then close the cover to measure.MSDR-D-KERN
Digital Refractometer fig-11

Press “CAL” button once again during the ‘CAL’ flashing to start to calibrate. When the calibration is completed showing as follows. If no operations for 10seconds the instrument would return back to booting status.

If fail to complete the calibration, the multi-function display area would show an error code.MSDR-D-KERN Digital Refractometer
fig-13

Scales converting and temperature systems converting

Scales ConvertingMSDR-D-KERN Digital Refractometer
fig-14

Temperature System ConvertingMSDR-D-KERN Digital Refractometer
fig-15

If exceed the temperature limitations, the signs “HHH” or “LLL” would show.MSDR-D-KERN Digital Refractometer fig-16

Turn Off

  1. After turn on if without any operations for 3 minutes, the instrument would be automatically turned off.
  2. Press “on/ off ” button for 2-3 seconds, the instrument would be turned off.

 Maintenance and Preservation

  1. Please clean and wash the sample plate with distilled water and dry it with soft cleaning cloth or paper towel after finishing the measuring one kind sample.
  2. Never left the remains and residuals of samples in the sample plate for long time.
  3. After finishing measurements of the corrosive liquid, please clean the sample plate as quick as possible to avoid irreparable damage of the prism and metal surface of the plate.
  4. Please use soft cleaning cloth or paper towel to clean the sample plate to avoid scribing the prism’s glass.
  5. When the dropper and dust-free cloth are not used, please clean it with clean water and put it in the packing box after drying.
  6. If no using the instrument for a long time, please remove the battery, and preserved in a cool and dry environment.

Appendix

Performance:MSDR-D-KERN Digital Refractometer
fig-17

The Error Codes TableMSDR-D-KERN Digital Refractometer
fig-18