GASTEC 154 Cyclohexanone Detector Tube Instruction Manual
- June 6, 2024
- GASTEC
Table of Contents
- FOR SAFE OPERATION :
- APPLICATION OF THE TUBE :
- SPECIFICATION :
- CORRECTION FOR TEMPERATURE, HUMIDITY AND PRESSURE :
- MEASUREMENT PROCEDURE:
- INTERFERENCES :
- APPLICATION FOR OTHER SUBSTANCES :
- CORRECTION FACTOR :
- DANGEROUS AND HAZARDOUS PROPERTIES :
- INSTRUCTIONS ON DISPOSAL :
- WARRANTY:
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
GASTEC 154 Cyclohexanone Detector Tube Instruction Manual
FOR SAFE OPERATION :
Carefully read this manual and the instruction manual of your Gastec Gas Sampling Pump.
WARNING:
- Use only Gastec detector tubes in a Gastec Pump.
- Do not interchange or use non-Gastec parts or components in Gastec’s detector tube and pump system.
- The use of non-Gastec parts or components in Gastec’s detector tube and pump system, or use of a non-Gastec detector tube with a Gastec pump, or use of a Gastec detector tube with a non-Gastec pump may result in property damage, serious bodily injury, and death; voids all warranties; and voids all performance and data accuracy guarantees.
CAUTION : If you do not observe the following precautions, you may suffer injuries or damage the product.
- When breaking the tube ends, keep away from eyes.
- Do not touch the broken glass tubes, pieces or reagent with bare hand(s).
- The sampling time represents the time necessary to draw the air sample through the tube.
The tube must be positioned in the desired sampling area for the entire sampling time or until the flow finish indicator indicates the end of the sample.
NOTES : For maintaining performance and reliability of the test results, observe the following
- Use Gastec Gas Sampling Pump together with Gastec Detector Tubes only for the purposes specified in the instruction manual of the detector tube.
- Use this tube within the temperature range of 0 40°C (32 104°F ).
- Use this tube within the relative humidity range of 0 – 90%.
- This tube may be interfered with by the coexisting gases. Please refer to the table “INTERFERENCES” below.
- Shelf life and storage condition of the tube are marked on the label of the box of tube.
APPLICATION OF THE TUBE :
Use this tube for detecting Cyclohexanone in the air or in industrial areas and for determining the environmental atmospheric condition.
SPECIFICATION :
(Because of Gastec’s commitment to continued improvement, specifications are subject to change without notice.)
Coefficient of Variation: 10% (for 2 to 10 ppm), 5% (for 10 to 30 ppm)
Shelf Life : Please refer to the validity date printed on the box of tubes.
Store the tubes at 10°C (50°F} or below in a refrigerator.
CORRECTION FOR TEMPERATURE, HUMIDITY AND PRESSURE :
Temperature: Correct for temperature by the table below:
MEASUREMENT PROCEDURE:
-
For checking the leakage of the pump, insert a fresh sealed detector tube into the pump.
Follow instructions provided with the pump operating manual. -
Break tips off a fresh detector tube with the tube tip breaker of the pump.
-
Insert the tube into the pump inlet with arrow ( G► ) on the tube pointing toward pump.
-
Make certain the pump handle is all the way in. Align the guide mark on the pump body with the guide mark on the handle.
-
Pull the handle all the way out until it locks at one pump stroke (100 ml). Wait three minutes and confirm the completion of the sampling. Repeat the above sampling procedure three more times.
-
For measurements higher than 30 ppm, prepare a fresh tube and perform two pump strokes.
-
Read concentration level at t11e interface where the stained reagent meets the unstained reagent.
-
If temperature correction is necessary, obtain the true concentration by using the temperature correction factor. Afterwards multiply the correction factor of pump stroke if necessary.
-
If pressure correction is necessary, use the pressure correction formula.
INTERFERENCES :
This table of interference gases primarily expresses the interference of each coexisting gas in the gas concentration range, that is equivalent to the gas concentration. Therefore, the test result may give a positive result from other substances not listed in the table. If more information is needed, please contact us or your Gastec representatives.
APPLICATION FOR OTHER SUBSTANCES :
Tube 154 can also be used for the other substances as below:
CORRECTION FACTOR :
Detector tubes are primarily designed to measure specific gases. But it is also possible to measure other substances of similar chemical properties with the aid of a correction factor or chart. Therefore, please make use of the correction factor/chart measuring ranges as a reference. For a more precise factor please contact your Gastec representatives.
DANGEROUS AND HAZARDOUS PROPERTIES :
Threshold Limit Value-Time Weighted Average by ACGIH (2020): 20 ppm
Threshold Limit Value-Short Term Exposure Limit by ACGIH (2020): 50 ppm
Explosive Range : 1.1 9.4 %
INSTRUCTIONS ON DISPOSAL :
The reagent of the tube does not use toxic substances. When disposing the tube regardless of whether it has been used or not, follow the rules and regulations of your local government.
WARRANTY:
If you have any questions regarding gas detection and quality of the tubes, please feel free to contact your Gastec representatives.
Manufacturer : Gastec Corporation
8-8-6 Fukayanaka, Ayase-City, Kanagawa 252-·1195, Japan
https://www.gastec.co.jp/
Telephone +81-467-79-3910 Facsimile +81-467-79-3979
IM00154E4
Printed in Japan
20UZ