GASTEC 146 Isopropyl Acetate Detector Tube Instructions
- June 5, 2024
- GASTEC
Table of Contents
GASTEC 146 Isopropyl Acetate Detector Tube
FOR SAFE OPERATION
Carefully read this manual and the instruction manual of your Gastec Gas Sampling Pump.
WARNING:
- Use only Gastec detector tubes in a Gastec Pump.
- Do not interchange or use non-Gastec parts or components in Gastec’s detector tube and pump system.
- The use of non-Gastec parts or components in Gastec’s detector tube and pump system or use of a non-Gastec detector tube with a Gastec pump or use of a Gastec detector tube with a non-Gastec pump may result in property damage, serious bodily injury, and death; voids all warranties; and voids all performance and data accuracy guaranties.
CAUTION: If you do not observe the following precautions, you may suffer injuries or damage to the product.
- When breaking the tube ends, keep away from eyes.
- Do not touch the broken glass tubes, pieces and reagent with bare hand(s).
- The sampling time represents the time necessary to draw the air sample through the tube. The tube must be positioned in the desired sampling area for the entire sampling time or until the flow finish indicator indicates the end of the sample.
NOTES:
For maintaining performance and reliability of the test results, observe the following.
- Use Gastec Gas Sampling Pump together with Gastec Detector Tubes only for the purposes specified in the instruction manual of the detector tube.
- Use this tube within the temperature range of 0 40°C (32 104°F).
- Use this tube within the relative humidity range of 0 90%. (When the dehumidifying layer is broken through the detecting reagent changes colour to orange.)
- This tube may be interfered with by the coexisting gases. Please refer to the table “INTERFERENCES” below.
- Shelf life and storage condition of the tube are marked on the label of the box of tube.
APPLICATION OF THE TUBE :
Use this tube for the detection of isopropyl acetate in air or the industrial areas and environmental atmospheric condition.
SPECIFICATION
Measuring Range | 10-500 ppm |
---|---|
Number of Pump Strokes | 2 |
Correction Factor | 1 |
sampling Time | 1.5 minutes per pump stroke |
Detecting Limit | 4 ppm (n=2) |
Colour Chanqe | Yellow -► Blackish brown |
Reaction Principle | CH3CO2CH(CH3)2 + Cr6 + **** + H2SO4 -► Cr3+ |
(Because of Gastec’s commitment to continuous improvement, specifications are subject to change without notice)
Coefficient of Variation: 10% (for 10 to 500 ppm)
Shelf Life: Please refer to the validity date printed on the tube box.
Store the tubes in the cool and dark place.
CORRECTION FOR TEMPERATURE, HUMIDITY AND PRESSURE
Temperature: Correct for temperature by the table below :
If the value of the Tube Reading multiplied by the Temperature Correction Factor is less than 1 0 ppm a measurement cannot be performed.
Humidity : No correction is required between 0 90% R.H.
Pressure : To correct for pressure, use the formula below
Tube Reading* (ppm) x 1013 (hPa) Atmospheric Pressure (hPa) This value is after other correction(s), if any are applied.
MEASUREMENT PROCEDURE
- For checking the leakage of the pump, insert a fresh sealed detector tube into the pump. Follow instructions provided with the pump operating manual.
- Break tips off a fresh detector tube with the tube tip breaker of the pump.
- Insert the tube into the pump inlet with arrow ( G► ) on the tube pointing toward pump.
- Make certain the pump handle is all the way in. Align the guide mark on the pump body with the guide mark on the handle.
- Pull handle all the way out until it locks at one pump stroke (‘100 ml). Wait 1.5 minutes and confirm the completion of sampling. Repeat the above sampling procedure one more time.
- Read concentration level at the interface where the stained reagent meets the unstained reagent.
- If temperature correction is necessary, obtain the true concentration by using the temperature correction factor.
- If pressure correction is necessary, use the pressure correction formula.
INTERFERENCES
This table of interference gases primarily expresses the interference of each coexisting gas in the gas concentration range, that is equivalent to the gas concentration. Therefore, the test result may be given positive results by the other substances not listed in the table. For more information is needed, please contact us or Gastec representatives. ****
DANGEROUS AND HAZARDOUS PROPERTIES :
Threshold Limit Value-Time Weighted Average by ACGIH (2020) : 100 ppm
Threshold Limit Value-Short Tenn Exposure Limit by ACGIH (2020): 150 ppm
INSTRUCTIONS ON DISPOSAL
The reagent of the tube uses a small amount of hexavalent chromium. When disposing the tube regardless of whether it has been used or not, follow the rules and regulations of your local government.
WARRANTY
If you have any questions regarding gas detection and quality of the tubes,
please feel free to contact your Gastec representatives.
Manufacturer : Gastec Corporation
8-8-6 Fukayanaka, Ayase-City, l<anagawa 252-1195, Japan
https://www.gastec.co.jp/
Telephone +81-467-79-3910 Facsimile +81-467-79-3979