accucold ARS1PV Medical Refrigerator User Manual
- June 5, 2024
- accucold
Table of Contents
ARS1PV Medical Refrigerator
User Manual
Models
ARS1PV | ARS1ML |
---|---|
ARG1PV | ARG1ML |
ARS3PV | ARS3ML |
ARG3PV | ARG3ML |
ARS6PV | ARS6ML |
ARG6PV | ARG6ML |
ARS8PV | ARS8ML |
ARG8PV | ARG8ML |
ARS12PV | ARS12ML |
ARG12PV | ARG12ML |
ARS15PV | ARS15ML |
ARG15PV | ARG15ML |
ARG31PVBIADA | ARG31MLBIADA |
ARS32PVBIADA | ARS32MLBIADA |
ARG61PVBIADA | ARG61MLBIADA |
ARS62PVBIADA | ARS62MLBIADA |
BEFORE USE, PLEASE READ AND FOLLOW ALL SAFETY RULES AND OPERATING INSTRUCTIONS
Write Model and Serial Numbers here:
Model: _____
Serial No.: __
Accucold Division of Felix Storch, Inc.
An ISO 9001:2015 registered company
770 Garrison Avenue
Bronx, NY 10474
www.accucold.com
EQUIPMENT SAFETY
Your safety and the safety of others are very important.
We have provided many important safety messages in this manual and on your
unit. Always read and obey all safety messages.
This is the Safety Alert Symbol. The symbol alerts you to potential hazards that can kill or injure you and others. All safety messages will follow the Safety Alert Symbol and either the words “DANGER” or “WARNING”.
DANGER
DANGER means that failure to heed this safety statement may result in
severe personal injury or death.
WARNING
WARNING means that failure to heed this safety statement may result in
extensive product damage, serious personal injury, or death.
All safety messages will alert you about the potential hazard, tell you how to
reduce the chance of injury, and let you know what can happen if the
instructions are not followed. Before the equipment is used, it must be
properly positioned and installed as described in this manual, so read the
manual carefully. To reduce the risk of fire, electrical shock, or injury
when using this equipment, follow basic precautions, including the following:
DANGER
-
Plug into a grounded 3-prong outlet, do not remove the grounding prong, do not use an adapter, and do not use an extension cord.
-
Replace all panels before operating.
-
It is recommended that a separate circuit serving only your unit be provided. Use receptacles that cannot be turned off by a switch or pull chain.
-
Never clean the equipment parts with flammable fluids. These fumes can create a fire hazard or explosion. And do not store or use gasoline or other flammable vapors and liquids in the vicinity of this or any other pieces of equipment. The fumes can create a fire hazard or explosion.
-
Before proceeding with cleaning and maintenance operations, make sure the power line of the unit is disconnected.
-
Do not connect or disconnect the electric plug when your hands are wet.
-
Unplug the unit or disconnect power before cleaning or servicing. Failure to do so can result in electrical shock or death.
-
Do not attempt to repair or replace any part of your unit unless it is specifically recommended in this manual. All other servicing should be referred to a qualified technician.
-
This unit is CFC- and HFC-free and contains a small quantity of Isobutane (R600a), which is environmentally friendly but flammable. It does not damage the ozone layer, nor does it increase the greenhouse effect. care must be taken during transportation and setting up of the unit that no parts of the cooling system are damaged. leaking coolant can ignite and may damage the eyes.
-
In the event of any damage:
– Avoid open flames and anything that creates a spark,
– Disconnect from the electrical power line,
– Air the room in which the unit is isolated for several minutes, and
– Contact the Service Department for advice. -
The more coolant there is in a unit, the larger the room it should be installed in. In the event of a leakage, if the unit is in a small room, there is the danger of combustible gases building up. For every ounce of coolant, at least 325 cubic feet of room space is required. The amount of coolant in the unit is stated on the data plate inside the unit. It is hazardous for anyone other than an Authorized Service Person to carry out servicing or repairs to this piece of equipment.
-
Take serious care when handling, moving, and using the unit to avoid either damaging the refrigerant tubing or increasing the risk of a leak.
-
Replacing component parts and servicing shall be done by factory authorized service personnel so as to minimize the risk of possible ignition due to incorrect parts or improper service.
-
WARNING FOLLOW WARNING CALLOUTS BELOW ONLY WHEN APPLICABLE TO YOUR MODEL
-
Use two or more people to move and install the unit. Failure to do so can result in back or another injury.
-
To ensure proper ventilation for your unit, the front of the unit must be completely unobstructed.
Choose a well-ventilated area with temperatures above 60°F (16°C) and below 90°F (32°C). [For optimal performance, install the unit where the ambient temperature is between72º and 78ºF (23º26ºC).] This unit must be installed in an area protected from the elements, such as wind, rain, water spray or drips. -
The unit should not be located next to ovens, grills, or other sources of high heat.
-
The unit must be installed with all electrical, water, and drain connections in accordance with state and local codes. A standard electrical supply (115 V AC only, 60 Hz), properly grounded in accordance with the National Electrical Code and local codes and ordinances, is required.
-
Do not kink or pinch the power supply cord of the unit.
-
The size of the fuse (or circuit breaker) should be 15 amperes.
-
It is important that the equipment be leveled in order to work properly. You may need to make several adjustments to level it.
-
All installations must be in accordance with local plumbing code requirements.
-
Make certain that the pipes are not pinched, kinked or damaged during installation.
-
Check for leaks after connection.
-
Never allow children to operate, play with or crawl inside the unit.
-
Do not use solvent-based cleaning agents or abrasives on the interior. These cleaners may damage or discolor the interior.
-
Use this equipment only for its intended purpose as described in this Instruction Manual.
-
Keep fingers out of the “pinch point” areas. Clearances between the door and cabinet are necessarily small. Be careful closing the door when children are in the area.
DANGER
Child entrapment and suffocation are not problems of the past. Junked or
abandoned appliances are still dangerous, even if they will “just sit in the
garage a few days”.
Before discarding your old refrigerator:
- Take off the doors
- Leave the shelves in place so that children may not easily climb inside.
SAVE THESE INSTRUCTIONS
INSTALLATION INSTRUCTIONS
Before using your equipment
-
Remove the exterior and interior packing.
CAUTION: After unpacking you MUST allow this unit to stand upright for at least 2 hours to allow the lubricant and refrigerant to drain back into the compressor and stabilize. Failure to do so may adversely affect performance and the lifetime of this unit. -
Clean the interior surface with lukewarm water using a soft cloth.
Installation of your equipment
-
Units with BI in their suffix are designed to be built-in or recessed installation. The rest of the units listed in this manual are designed for freestanding use. All models are for indoor use only.
-
WARNING: Do not store or install the unit outdoors.
-
CAUTION: This equipment is designed for the storage of medicine or other medical products. Do not store beverages or perishable food in this unit.
-
Place the refrigerator on a floor that is strong enough to support it when it is fully loaded. To level the unit, adjust the front leveling legs.
-
For free-standing installation, allow at least 5 inches (127mm) of space between the back, top, and sides of the unit. This allows the proper air circulation to cool the compressor and condenser for energy saving. Even for recessed or built-in installation (units with “BI” in the model number), it is a must to keep ¼” (6.35mm) of space on each side and 2” (51mm) on the top and at the rear. The air vent at the front of the unit must never be covered or blocked in any way.
NOTE: It is recommended that you do not install the unit near an oven, radiator, or another heating source. Direct sunlight may affect the acrylic coating and heat sources may increase electrical consumption. Don’t install in a location where the temperature will fall below 60°F (16°C). For best performance, do not install the unit behind a cabinet door or block the base grille. -
Avoid locating the unit in moist areas.
-
Plug the unit into an exclusive, properly grounded wall outlet. Do not under any circumstances cut or remove the third (ground) prong from the power cord. Any questions concerning power and/or grounding should be directed toward a certified electrician or an authorized service center.
Electrical connection
WARNING
Improper use of the grounded plug can result in the risk of electrical shock.
If the power cord is damaged, have it replaced by a qualified electrician or
an authorized service center.
This unit should be properly grounded for your safety. The power cord of this
unit is equipped with a three-prong plug which mates with standard three-prong
wall outlets to minimize the possibility of electrical shock.
Do not under any circumstances cut or remove the third (ground) prong from the
power cord supplied. For personal safety, this equipment must be properly
grounded.
This unit requires a standard 115/120 Volt AC ~ 60Hz three-prong grounded
electrical outlet. Have the wall outlet and circuit checked by a qualified
electrician to make sure the outlet is properly grounded? When a standard
2-prong wall outlet is encountered, it is your responsibility and obligation
to have it replaced with a properly grounded 3- prong wall outlet.
To prevent accidental injury, the cord should be secured behind the unit and
not left exposed or dangling.
The unit should always be plugged into its own individual electrical outlet
which has a voltage rating that matches the rating label on the unit. This
provides the best performance and also prevents overloading house wiring
circuits that could cause a fire hazard from overheating. Never unplug the
unit by pulling on the power cord. Always grip the plug firmly and pull it
straight out from the receptacle. Repair or replace immediately all power
cords that have become frayed or otherwise damaged. Do not use a cord that
shows cracks or abrasion damage along its length or at either end.
When moving the equipment, be careful not to damage the power cord.
Electrical connection
Because of potential safety hazards under certain conditions, it is strongly
recommended that you do not use an extension cord with this equipment.
However, if you must use an extension cord it is absolutely necessary that it
be a UL/CUL-Listed, 3-wire grounding type equipment extension cord having a
grounding type plug and outlet and that the electrical rating of the cord is
115 volts and at least 10 amperes.
Reversing the door
Unless ordered with option “LHD,” your unit has shipped with a right-hand
self-closing door swing. If you wish to reverse the door to open from the
opposite direction, follow the steps listed below.
NOTE: Reversing the door will undo the self-closing function. Doors on
units with the “DL2B” data logger suffix are not reversible.
- Carefully lay the unit on its back and remove the two front leveling legs
- Remove the two hex head screws that connect the door hinges to the body on the bottom of the unit
- Remove the two flathead screws that connect the hinge assembly to the bottom of the door
- Remove the white hinge from the square rod
- Slide the door towards the bottom until it is free of the top hinge
- Unscrew the top hinge pin from the unit and reinstall on the left side
- Flip the door so the handle is now on the right side
- Push the door onto the top hinge piece.
- Reinstall the hinge assembly on the left side bottom of the door using the flathead screws
- Push the white hinge onto the square rod at a position of 9 o’clock
- Turn the white hinge clockwise until the holes line up with screws located on the bottom of the unit
- Replace the hex head screws and front leveling legs
NOTE: Once the door has been reversed, the Pharma-Vac or Med-Lab sticker
will be upside down.
You can remove and reapply to the preferred position. If you would like a
replacement sticker, contact our parts department at
parts@summitappliance.com.
CAUTION: After setting the unit upright you MUST allow this unit to stand upright for at least 2 hours to allow the lubricant and refrigerant to drain back into the compressor and stabilize. Failure to do so may adversely affect performance and the lifetime of this unit.
LOCATION OF PARTS
1. Keyed lock | 7. Handle |
---|---|
2. Microprocessor control panel | 8. Leveling legs |
3. Reset the switch | 9. Kickplate |
4. Light switch | 10. Temperature sensor in a glycol bottle |
5. Internal light | 11. Dry contacts (located in the rear) |
6. Internal fan | 12. Access port |
OPERATION – QUICK SETUP GUIDE
This section explains how to operate your unit. For more detailed instructions
on changing the parameters of the control panel, see page 10.
Once plugged in, the display will show the product temperature inside the
refrigerator and the compressor will start to run (the indicator light on the
control panel will turn on at the same time).
NOTE: This product utilizes two sensors: A Product Sensor (display
sensor) and an Air Sensor (control sensor).
For maximum performance, the refrigerator controls use a separate air sensor.
This allows it to rapidly respond to product loading and door openings. You
may need to adjust the controller lower than the desired product temperature
as the air temperature needs to be lower to pull the product temperature down.
The temperature displayed is the product temperature recorded by the product
sensor in glycol inside a vial located in the refrigerator. As the
refrigerator cools to the setpoint, the number on the display will decrease
accordingly. The product vial will take longer to cool than the air in the
refrigerator.
If power is cut off, wait for at least 5 minutes before plugging the unit in
again to avoid damaging the compressor.
CONTROL PANEL
Adjusting the Operating Temperature
To change the setpoint, press the SET key and the set-point temperature will
be displayed. To change the setpoint, within 10 seconds, press the up or down
arrow key until the desired temperature is displayed. The new set point will
be memorized when you push the SET key again.
Glassdoor models come preset to 4°C and solid door models are factory set to
5°C. Adjust the temperature up or down to achieve your desired product
temperature the displayed temperature.
High/Low-Temperature Alarm
If the temperature in your refrigerator drops below the air low alarm set
point, an alarm will sound. The front panel display will flash LO and the
current temperature will be displayed alternately. Also, if the temperature
rises above the high air alarm setpoint an alarm will sound.
The front panel display will flash HI and the current temperature will be
displayed alternately. if the temperature rises above the high product alarm
setpoint an alarm will sound. The front panel display will flash H2 and the
current temperature will be displayed alternately. There could be a number of
reasons for this which should be investigated immediately.
The refrigerator should be monitored to check that it is returning to the
correct temperature. Set the alarms values to meet your requirements in the
parameter set-up and settings.
Power Failure Alarm
After plugging in the equipment, move the “Reset Switch” located at the front
of the unit to the down position. When electrical power to the equipment is
lost, an alarm will sound, and it will silence when the power is restored.
Note the alarm is battery-powered and has a limited time to alarm.
Min/Max Temperature Recording
- Once connected to the power supply, your unit records the maximum high and minimum low temperatures reached inside the cabinet. The display will show the last HI and LO temperatures recorded until the refrigerator is reset.
- To display the HI reading, press and release the HI/LO button. To display the LO reading, press and release the HI/LO button again.
To Reset the Min/Max Temperature Recording
Hold the HI/LO button for 6 seconds. To confirm the operation, the display
starts blinking and the buzzer sounds. After 5 seconds, the normal display
will be restored.
Changing the Temperature Display
This unit ships with the temperature being displayed in Celsius. If you prefer
the display in Fahrenheit, you will need to enter your password and adjust the
parameters. Press and hold the “Set” button until it flashes “Pas,” then enter
the password (factory default is “15”) using the up and down arrows. Once you
enter the password, press “Set” about 25 times until “CF” displays, then
press the down arrow until the control displays “ºF”, then press “Set.” Wait 6
seconds and the setting will be recorded, with your unit resuming normal
operation.
Interior Light
Your unit includes an LED light that will automatically illuminate when the
door is opened. If you would like to keep the light on when the door is
closed, push the switch to the “on” position.
Door Alarm
If the door of the refrigerator is left open for more than 3 minutes, an
audible alarm will sound. The alarm will be turned off automatically when the
door is closed.
Remote Alarm
The electronic controller is equipped with dry contacts for alarm forwarding
to an external remote alarm system (terminal block on the back of the
refrigerator).
NOTE: The remote contacts are normally “closed” and “open” upon alarm. To
change the output to normally “open” and “closed” on alarm see item “C12” of
the Table of Parameters and Factory Settings. Change the “C12” setting to
“00” in lieu of “01”.
Summary of Alarm Types
ALARM TYPE
|
DISPLAY FLASH
---|---
Air temperature high alarm| HI delay alarm
Air temperature low alarm| LO delay alarm
The air temperature sensor short-circuited| 1H in time alarm
The air temperature sensor open-circuited| 1L in time alarm
The product temperature sensor short-circuited| 2H in time alarm
The air temperature sensor open-circuited| 2L in time alarm
Product temperature high temp alarm| H2 delay alarm
Product temperature low temp alarm| L2 delay alarm
Electric supply off alarm| EEL flash and alarm in time
Door open delay alarm| Dr flash and delay alarm
Storage of articles
- At maximum loading level, the content should not block the refrigerator’s airflow or be loaded above the load line.
- Leave space around the contents to allow a smooth flow of cold air inside the cabinet. To prevent freezing, be sure the articles do not touch the back of the interior cabinet.
- Store articles away from the inner fan.
- Shelves can be adjusted to allow proper airflow around the products being stored.
DETAILED TEMPERATURE CONTROLLER GUIDE
Front Panel Operation
Set temperature (compressor’s top temperature) adjustment:
- Press the SET button, the set temperature is displayed, then press the UP or DOWN arrow to store and memorize. Press the SET button to exit the adjustment status and display the product temperature. If no more buttons are pressed within 6 seconds, the refrigerator product temperature will be displayed.
- (Set temperature adjustment range: Parameter E1~E2)
Manually start/stop defrosting:
Press HI/LO button, then press the SET button and hold for 6 seconds to enter
defrost status or to stop defrost.
Display the sensor temperature (C13):
Press the DOWN arrow, and the product temperature (or cabinet temperature)
will flash display. After 6 seconds, will resume normal display.
Refrigeration LED:
During refrigeration, the cooling snowflake LED is on. When the temperature is
constant, the LED is off.
Defrost LED:
During defrosting (if activated), the defrost LED is on. During the delay
after defrosting, the LED flashes.
Parameter set-up and settings (for all settings other than temperature
setpoint):
Press the SET button and hold for 6 seconds to enter the parameter setting
(flash and display PAS).
After entering the correct password (factory default is 15), press SET, and
the display will show E1, E2, …~do3 PAS in sequence. Press the UP or DOWN
arrow and the value of the parameter will be displayed and can be modified and
stored. If no other buttons are pressed within 6 seconds, it will exit, and
the new value will be stored.
NOTE: Only when in the inner parameter menu (display PAS) and the correct
password is entered can the parameter value be adjusted. If the wrong password
is entered, the parameter modification will be exited, although the set
temperature adjustment is still active. If you forget the password, you will
need to resume the factory default settings.
The following are examples of how to change common settings:
A. Enter Password – Press and hold the “Set” button for 6 seconds until
it flashes “Pas”, then enter the default password “15” using the up and down
arrows. When “15” is reached press the “Set” button.
B. Change display to “°F” in lieu of “°C” – After entering the password
press the set button about 25 times to reach the “CF” display, then press the
up or down arrow unit the control displays “°F” then press set. Wait 6 seconds
and it will be set and the unit will go back to normal operation.
C. High Air Alarm Settings – After entering the password press the set
button until you reach the “C1” display, then press the up or down arrow unit
the control displays your desired setting then press set. Wait 6 seconds and
it will be set and the unit will go back to normal operation.
D. Low Air Alarm Settings – After entering the password press the set
button until you reach the “C2” display, then press the up or down arrow unit
the control displays your desired setting then press set. Wait 6 seconds and
it will be set and the unit will go back to normal operation.
E. High Product Alarm Settings – After entering the password press the
set button until you reach the “C3H” display, then press the up or down arrow
unit the control displays your desired setting then press set. Wait 6 seconds
and it will be set and the unit will go back to normal operation.
F. Low Product Alarm Settings – After entering the password press the set
button until you reach the “C3L” display, then press the up or down arrow unit
the control displays your desired setting then press set. Wait 6 seconds and
it will be set and the unit will go back to normal operation.
To change any other parameters in the “Factory Settings” enter the Password and then press the “Set” button until the desired Parameter is displayed, then press the up or down arrow until the desired setting is displayed then press the “Set” key again. Wait 6 seconds and it will be set and the unit will go back to normal operation
Recording the highest and lowest temperature:
Once turned on and after the C5 delay, the unit will start to record the
highest and lowest temperature; the record will be refreshed at any time. The
values will be kept in memory even after the power is turned off. Press the
HI/LO button to see the highest temperature recorded. Press it again to see
the lowest temperature. Holding the HI/LO button for 6 seconds will cause the
display to flash and a buzzer to sound; the previous highest and lowest
temperature readings will be cleared, and a new record will begin. (Highest
and lowest temperature recording instructions: When parameter F4=0, the unit
can record the highest and lowest temperature at any time. When F4=1,2,3,
during defrost and delay locking 20 minutes, the highest and lowest
temperatures are not recorded. During air sensor failure, highest and lowest
temperatures are not recorded.) When set C13=00, the temperature in the liquid
bottle is recorded; when set C13=01, the temperature of the air in the cabinet
is recorded. The product temperature corresponds to the temperature of the
liquid in the bottle.
Factory default resumption
Press and hold the DOWN arrow; at the same time, press the UP arrow and hold
for 6 seconds. The display will flash and show 888. At this time, all
parameters will resume factory default values. After 6 seconds, the normal
operational mode will return.
To check or change the password:
Enter the inner parameter menu (PAS display) and enter the correct password.
After entering the inner parameter PAS, press the UP or DOWN arrow to display
and change the password, then press HI/LO to confirm and store the new
password.
Table of Parameters and Factory Settings
Parameter| Function| Set Range| Default| Parameter|
Function| Set Range| Default
---|---|---|---|---|---|---|---
Set| | | 40’C. For glass door models 5.0’C For solid door models| C3H|
Product Temp. High Temp. Alarm| -19.9-20.0°C / –
3.8-68.04F| 8.0’°C /
46.4t
PAS| Password| 00-99| 15| C3L| Product Temp. Low Temp. Alarm| -19.9-20.0″C / –
3.8-68.0-F| 2.0°C /
3.13T
El| Lowest set
point limit| -19.9°C / 3.8°F –
set temp.| 2.0•C / 35.6’F| C4| Hysteresis Alarm| 0.1-20.0″C /
1.8-36.0’F| 0.1°C F
0.18Y
E2| Higher set
point limit| 20.0°C / 68.0°F
-set temp.| 10.0°O / 50°P| C5| Start-up Temp. Alarm delay| 00-99 min| 30 min
E3| Temp.
Hysteresis| 0.1-20.0°C /
0.2-36.0°F| 2.0•C / 3.6- F| C6| Temp. Alarm
Delay| 00-99 min| 3 min
0.00E+00| Turning On
delay time| 00 -10 min.| 3 min.| C7| Power Off
relay Alarm| 00=do not alarm / 01=
alarm| 1
0.00E+00| Comp. start
delay time| 00 -10 min.| 3 min.| C8| Alarm relay close after. muffle alarm
relay switch| 01-0p811101=close| 8
E5| Offset on Air
Temperature| -10.0-10.04C/ –
18.0-18.0″F| 1.0°C / 1.8°F
For 1 a 3 cu. Ft.
nd models 2.0°C / 3.6°F For 6. EL 12 and 15 al. Ft. models| C9| Restart time
after buzzer
mute| 00=do not start / 01 -30
min=restart time| 10 min
E6| Offset on
Product Temp.| -10.13-10.0t / –
18.0-18.0°F| 0.0T/ 32°F| C10| Comp
Force stop time| 01-99 min| 99 min
FO| Defrost Type| 00=defrost by
turning of comp.| 0| C11| Comp
Force
Running
Time| 00=comp. Stop / 01-
99min=starting time| 0 min
Fl| Max defrost
duration| 01-60 min.| 20 min.| C12| Alarm
Output Type| 00=contact actuation
when alarm /01=contact disconnect when the alarm| 1
F2| Display
during
defrost| 00-24 hr.| 0| C13| Normal
State temp.
display type| 00=liquid bottle temp /
01=cablnet air temp.| 8
F4| Display
during
defrost| 00= closed room
temp. display
normally / 01=last
value before defrost
/ 02= fixed display /03= display defrost| F i| CF| Temperature
Unit| °C=Ceislus /
°F=Fahreehed| °C
Cl| High Temp
Air Alarm| C2-20.0°C 168.0°F| 9.0″C / 48.2-F| del| | 00=do not alarm /
01=99 min=delayed
alarm| 02 min
C2| Low Temp
Au Alarm| -19.9•C / 3.8t-C1| 1.0- C / 33.8’F| do2| Cop. status mist ien
door open| l:: stop / 01=original
status| 1
| | | | do3| Light status when door open| 00=start / 01=original
status| 1
FUNCTION DETAILS
Temperature Control: After turning on the unit and reaching the delay time
(parameter E40), the compressor starts operating when the cabinet temperature
is higher than (set temperature + hysteresis) and will be off when the
cabinet temperature is lower than the setpoint temperature.
To protect the compressor, it cannot be re-started unless the time after the
compressor stops is longer than the delay time (Parameter E41).
Defrost Control: After running for a defrost interval set time (parameter F2),
the unit will enter defrost and the compressor stops. When the defrost
duration time (parameter F1) ends, it will exit defrost status. After a
defrost period ends, there will be 2 minutes of dripping time before the unit
can enter refrigeration status.
-
When defrosting interval F2 is set to “00”, the automatic defrost by turning off the compressor will be canceled.
-
When setting parameter F4 is set to “0”, the temperature will be displayed normally during defrost.
-
When setting parameter F4 is set to “1”, the cabinet temperature is locked during defrosting, and the last value before defrost is displayed. When defrosting ends, the normal display will resume after a temperature display 20 minutes delay (or cabinet temperature lower than the set temperature).
The defrost LED flashes during the delay. -
When setting parameter F4 is set to “2”, the set temperature will be displayed during defrosting.
-
When defrosting ends, the normal display will be resumed after 20 minutes delay (or cabinet temperature lower than the set temperature). The defrost LED flashes during the delay.
-
When setting parameter F4 is set to “3”, dEF will be displayed during defrosting. When defrosting ends, the normal display will be resumed after 20 minutes of delay display dEF (or cabinet temperature lower than the set temperature). The defrost LED flashes during the delay.
Alarm Control: After turning on for the first time, the unit will need to
reach the C5 delay time setting, then the high/low-temperature alarm function
can be triggered (C1, C2, C3). After passing C5 delay, when the cabinet
temperature is abnormal (for example higher than the temperature alarm C1 or
lower than temperature alarm C2) and duration more than alarm delay time C6,
the unit will enter alarm status, alarm start. When a high-temperature alarm
alternate displays H1 and cabinet air temperature, the compressor starts to
run. When a low-temperature alarm, will alternate display L0 and cabinet
temperature, the compressor stops. When cabinet temperature is higher than the
(low-temperature alarm value C2 + alarm hysteresis C4), the low-temperature
alarm ends. When cabinet temperature is lower than (high-temperature alarm
value C1 –alarm hysteresis C4), the high-temperature alarm ends.
When the product sensor temperature is greater than or equal to the product
temperature high-temperature alarm value C3 and duration more than alarm delay
time C6, the unit enters alarm status and starts the alarm, alternate display
H2, and cabinet temperature. When the product sensor temperature is less than
or equal to the(high-temperature alarm value C3 – alarm hysteresis C4), it
will end the high-temperature alarm.
When the electric supply is off, flash and display EEL and alarm. When setting
do1 is set at “00”, the unit will not alarm when the door is open. When
setting do1 is set at “0”, once the unit reaches the delay time, it will flash
and display “dr alarm”, press the random button to cancel.
If setting C7 is set to “0”, then the unit’s relay does not alarm when power
is off.
If setting C8 is set to “0”, then after the alarm is canceled, the alarm relay
does not close. If C9 is set to “0”, then after the alarm cancel button is
pushed, the buzzer does not restart.
If setting other numbers, then after reaching delay time, the buzzer sounds
one more time. (Under the condition of alarm not terminating) Abnormal work
mode: When the air sensor is short-circuited or high temperature over the
limit (> 66°C/151°F), “1H” is displayed; when the air sensor is open-circuited
or low temperature over the limit (<40°C/°F) “1L” is displayed. The
compressorwill enter the force running mode, according to C10, and C11 set
parameters running in sequence.
When the product sensor is short-circuited, open-circuited or over-limit
(>66°C/151°F), alternate display 2H and air temperature, the product
temperature sensor open-circuited or low temperature over the limit
(<-40°C/°F), alternate display 2L and air temperature.
Instrument running alarm indication list
Alarm type | Compressor running | Display flash |
---|---|---|
Refrigerator high temp. alarm | Compressor on | H1 delay alarm |
Refrigerator low temp. alarm | Compressor off | LO delay alarm |
Air temp. sensor short-circuited | Press C10, C11 to run the | |
comp. | 1H in time alarm | |
Air temp. sensor open-circuited | Press C10, and C11 to run the comp. | 1L in |
time alarm
Product temp. sensor short-circuited| | 2H in time alarm
Product temp. sensor open-circuited| | 2L in time alarm
Product temperature high temp.| Limit alarm| H2 delay alarm
Product temperature low temp.| Limit alarm| L2 delay alarm
Electric supply off alarm| | EEL flash and alarm in
time
Door open delay alarm| | Dr flash and delay alarm
When the door is open, the light switch can turn the light on and off.
Circuit Diagram
When the door is closed, normal close contact When the door is open, contact is disconnected
Notes for installation:
- Sensor leads must be kept separately from main voltage wires in order to avoid high-frequency noise-induced. Separate the power supply of the loads from the power supply of the controller.
- When installing, the sensor must be placed with the head upward and the wire downward.
- The temperature controller must not be installed in an area where water drips.
CLEANING AND MAINTENANCE
Before cleaning, it is essential that you unplug the unit from the power line and transfer the contents where they can be stored and monitored at the correct temperatures.
- Wash the inner compartment with warm water and neutral detergent. DO NOT allow the control panel, cables or plug to get wet. NEVER use corrosive detergents, wire brushes, or abrasive scourers to clean your refrigerator. NEVER use metal or sharp implements to remove debris.
- Dry all surfaces thoroughly.
- To ensure trouble-free operation, the condenser should be cleaned every three months where appropriate using a vacuum hose. The condenser is located at the back of the cabinet. In exceptionally dusty locations, the condenser should be cleaned more often.
When transporting the unit, try to keep it vertical with the angle of inclination between the sides of the refrigerator and the horizontal plane not less than 50. Otherwise, the compressor may be affected, disrupting the normal operation of the refrigerator.
TROUBLESHOOTING
You can solve many common problems easily, saving you the cost of a possible service call. Try the suggestions below to see if you can solve the problem before calling the servicer.
Problems | Possible causes | Remedy |
---|---|---|
The unit does not operate | Bad connection of plug or burnt-out fuse | Unplug |
the power cord and reconnect the plug to
power supply or replace it with a new fuse
Abnormal operation of the compressor or there is a
buzzing sound| The power voltage is out of the rated range| Disconnect the
power supply immediately and reconnect with it after
normal voltage. It is necessary to have a
voltage stabilizer in case of a poor power supply.
The compressor operates for a long time and no frost on the surface of the
evaporator| The refrigeration system is at fault (leakage or blocked)| Call
for service
There is frost or ice on the walls of the inner cabinet and the internal
temperature is too low, the compressor never stops running.| The thermostat
does not work| Call for service
The temperature setting of the thermostat is too low.| Adjust to a higher
temperature
The internal temperature is too high, and the compressor never stops
operating.| Bad heat dissipation and ventilation of condenser| Improve
ventilation
Too many warm items were put in at one time| Remove some goods so air can
circulate
The door is being opened too frequently during the initial cool down| Permit
the unit to cool d
down adequately, prevent
product access during this
Too noisy| The unit is not level| Adjust the adjustable legs
The fastener of the unit is loose| Tighten the loose fastener
A pipe near the compressor is touching| Carefully separate the touching pipes
The side of unit is hot| The condenser in the sidewall gives out heat as part
of normal operation| Nothing to worry about.
Sometimes a light sound of water flowing will be heard| The refrigerant flows
inside the pipe.| Nothing to worry about.
There may be
condensation on the glass door| High ambient temperature or humid conditions|
Dry with a cloth
The display shows “So” and the compressor does not run| The temperature sensor
is an open circuit| Call for service
The display shows “SC” and the compressor does not| The temperature sensor is
short-circuited.| Call for service
The display shows “HH” and the compressor does not run| The ambient
temperature is above 40°C/104°F| Stop using immediately
The display shows “LL” and the compressor does not run| The ambient
temperature is below 40°C/-40°F.| Stop using immediately
If you continue to experience issues, contact Accucold’s technical support department at 718-8933900.
NOTES
LIMITED WARRANTY
TWO-YEAR LIMITED WARRANTY
Within the 48 contiguous United States, for two years from the date of
purchase, when this unit is operated and maintained according to instructions
attached to or furnished with the product, the warrantor will pay for factory
specified parts and repair labor to correct defects in materials or
workmanship. Service must be provided by a designated service company. Outside
the 48 states, all parts are warranted for two years from manufacturing
defects. Plastic parts, shelves, and cabinets are warranted to be manufactured
to commercially acceptable standards and are not covered from damage during
handling or breakage.
FIVE-YEAR COMPRESSOR WARRANTY
- The compressor is covered for five years.
- The replacement does not include labor.
ITEMS WARRANTOR WILL NOT PAY FOR:
- Service calls to correct the installation of your equipment, to instruct you how to use your equipment, to replace or repair fuses or to correct wiring or plumbing.
- Service calls to repair or replace unit light bulbs or broken shelves. Consumable parts (such as filters) are excluded from warranty coverage.
- Damage resulting from accident, alteration, misuse, abuse, fire, flood, acts of God, improper installation, installation not in accordance with electrical or plumbing codes, or use of products not approved by the warrantor.
- Replacement parts or repair labor costs for units operated outside the United States.
- Repairs to parts or systems resulting from unauthorized modifications made to the unit.
- The removal and reinstallation of your unit if it is installed in an inaccessible location or is not installed in accordance with published installation instructions.
DISCLAIMER OF IMPLIED WARRANTIES; LIMITATION OF REMEDIES
THE CUSTOMER’S SOLE AND EXCLUSIVE REMEDY UNDER THIS LIMITED WARRANTY SHALL BE
PRODUCT REPAIR AS PROVIDED HEREIN. IMPLIED WARRANTIES, INCLUDING WARRANTIES
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED TO ONE
YEAR. WARRANTOR SHALL NOT BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.
SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR
CONSEQUENTIAL DAMAGES, OR LIMITATIONS ON THE DURATION OF IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS, SO THESE EXCLUSIONS OR LIMITATIONS MAY NOT APPLY
TO YOU. THIS WARRANTY GIVES YOU SPECIFIC LEGAL RIGHTS AND YOU MAY ALSO HAVE
OTHER RIGHTS, WHICH VARY FROM STATE TO STATE.
WARNING: This product can expose you to chemicals including Nickel (Metallic)
which is known to the State of California to cause cancer.
For more information go to
www.P65Warnings.ca.gov
Note Nickel is a component in all stainless steel and some other
metallic compositions
Accucold Division of Felix Storch, Inc.An ISO 9001:2015
registered company770 Garrison Avenue
Bronx, NY 10474
www.accucold.com
For parts and accessory ordering, troubleshooting, and helpful hints, visit:
www.accucold.com/support
9202019
References
- Medical & Scientific Cold Storage Solutions | Accucold® Medical Refrigerators
- Medical & Scientific Cold Storage Solutions | Accucold® Medical Refrigerators
- Service and Support | Accucold® Medical Refrigerators