accucold TMD-01 Thermometer Alarm User Manual
- June 3, 2024
- accucold
Table of Contents
accucold TMD-01 Thermometer Alarm
Specifications
- Large LCD that simultaneously displays minimum, maximum, and current temperatures
- User-defined high and low-temperature limits
- Visual and audible indicators for temperatures outside the set range
- A min/max feature to monitor and store the highest and lowest temperature readings (until the memory is cleared)
- A temperature sensor enclosed in a glycol-filled bottle protects it from rapid temperature changes when the refrigerator/freezer door is opened
- A low battery alert function (battery symbol flashes)
- Alarm notification for probe malfunction or disconnection
- Measuring temperature range: Probe 1(default) -30 ~ 70 oC (or -22 ~ 158 oF)
- Measuring temperature range: Probe 2 (optional) -50 ~ 45 oC (or -58 ~ 113 oF)
- Accuracy (probes 1 & 2): ± 0.3 oC (-10 ~ 10 oC or 14 ~ 50 oF), in other ranges ± 0.5 oC
- The sampling rate of 20 seconds
- Powered by two durable 1.5V AAA lithium batteries
- Select table temperature scale (°C/°F)
- Large display with illumination
Package Contents
- TMD-01 Thermometer/Alarm
- Temperature sensor (NTC) in glycol-filled bottle
- Two AAA Lithium batteries (1.5V)
- Instructions manual
- ISO/IEC 17025:2017 NIST certificate (calibrated units only)
Set-Up
NOTE: Before use, remove and discard the clear plastic protective sheet from the LCD Display.
- Battery Installation: Remove the battery compartment cover located on the rear of the unit, and remove the insulation strip from the battery compartment. The unit will beep and all segments of the LCD screen will be activated.
Functions/Controls
Programming the Alarm
Set the high and low alarm temperature limits above and below the refrigerator/freezer operating temperature to prevent false alarms. Refrigerators are designed to run between 2-8 °C (36-46 °F) while freezers run below -15 °C (5 °F).
Eliminating the Alarm Indicators
When an alarm condition occurs LO-ALARM/HI-ALARM icons will appear on the display along with a beeping sound to alert the user. Press any key ONCE to stop the beeping sound. The beeping sound lasts for 60 seconds, but the high and low indicators will remain on display even when the unit gets back in range. Pressing any key will silence the beeping sound, but it will be reactivated every 30 minutes if the alarm event still exists.
Press the keys simultaneously to clear the LO-ALARM/HI-ALARM indicators. The alarm will not be clear if the alarm event still exists.
Using the Alarm
- Placement: Once the thermometer has been programmed, place the temperature sensor (in a glycol bottle) in the location to be monitored, such as inside the refrigerator or freezer. Attach the thermometer to the door of the unit using its magnet.
- If the temperature measurement rises above or falls below the set temperature range, the alarm will sound. To silence the alarm, press any key ONCE.
- The alarm/thermometer’s maximum and minimum readings reflect the highest and lowest temperatures since the battery was inserted or since the RESET button was last pressed.
Troubleshooting
Issue | Possible Solution |
---|---|
The display shows NP. | _The temperature sensor is not installed |
correctly_
The display screen is not working.| Make sure the batteries are installed
correctly. The batteries may need to be replaced.
The “Low Battery” indicator is flashing.| The batteries may need to be
replaced.
Care & Maintenance
- Do not disassemble the product, as product damage may result.
- Store the product where it will not be exposed to direct sunlight, dust or high humidity.
- Do not wash or expose the product to water or other liquids.
- Clean the product by wiping it with a soft, dry cloth.
- Never use volatile or abrasive liquids or cleaners to clean the product.
- Do not drop the product or subject it to sudden shock or impact.
Customer Support
- For technical support, please call 800-932-4267 (U.S. and Canada) or email info@summitappliance.com
- For more information, please call ACCUCOLD at 718-893-3900 (U.S. and Canada). See us on the web at www.accucold.com
- For re-calibration and certification questions, please email us at calibration@summitappliance.com
Limited Warranty
SUMMIT products have a limited warranty period of 1 year against defects in materials and workmanship from the date of purchase. Accessory items and sensors have a limited warranty of 3 months. Repair services have a limited warranty period of 3 months against defects in materials and workmanship. SUMMIT shall, at its option either repair or replace hardware products that prove to be defective, if a notice to that effect is received within the warranty period. SUMMIT makes no other warranties or representations of any kind whatsoever, expressed or implied, except that of title, and all implied warranties including any warranty of merchantability and fitness for a particular purpose are hereby disclaimed.
WARNING
This product can expose you to chemicals including Nickel (Metallic) which is
known to the State of California to cause cancer. For more information go to
www.P65Warnings.ca.gov
Note
Nickel is a component in all stainless steel and some other metallic
compositions.
Accucold, Div. of Felix Storch, Inc. An ISO 9001:2015 registered company 770 Garrison Ave. Bronx, NY 10474 Tel: 718-893-3900 • Fax: 844-478-8799
References
- Medical & Scientific Cold Storage Solutions | Accucold® Medical Refrigerators
- Medical & Scientific Cold Storage Solutions | Accucold® Medical Refrigerators
- P65Warnings.ca.gov
- p65warnings.ca.gov/