illumina Local Run Manager CF 139-Variant 2.0 Analysis Module Software User Guide
- June 3, 2024
- illumina
Table of Contents
illumina Local Run Manager CF 139-Variant 2.0 Analysis Module Software
This document and its contents are proprietary to Illumina, Inc. and its
affiliates (“Illumina”), and are intended solely for the contractual use of
its customer in connection with the use of the product(s) described herein and
for no other purpose. This document and its contents shall not be used or
distributed for any other purpose and/or otherwise communicated, disclosed, or
reproduced in any way whatsoever without the prior written consent of
Illumina. Illumina does not convey any license under its patent, trademark,
copyright, or common-law rights nor similar rights of any third parties by
this document.
The instructions in this document must be strictly and explicitly followed by
qualified and properly trained personnel in order to ensure the proper and
safe use of the product(s) described herein. All of the contents of this
document must be fully read and understood prior to using such product(s).
FAILURE TO COMPLETELY READ AND EXPLICITLY FOLLOW ALL OF THE INSTRUCTIONS
CONTAINED HEREIN MAY RESULT IN DAMAGE TO THE PRODUCT(S), INJURY TO PERSONS,
INCLUDING TO USERS OR OTHERS, AND DAMAGE TO OTHER PROPERTY, AND WILL VOID ANY
WARRANTY APPLICABLE TO THE PRODUCT(S).
ILLUMINA DOES NOT ASSUME ANY LIABILITY ARISING OUT OF THE IMPROPER USE OF THE
PRODUCT(S) DESCRIBED HEREIN (INCLUDING PARTS THEREOF OR SOFTWARE).
© 2021 Illumina, Inc. All rights reserved.
All trademarks are the property of Illumina, Inc. or their respective owners.
For specific trademark information, see
www.illumina.com/company/legal.html.
Local Run Manager CF 139-Variant 2.0 Analysis Module Workflow Guide
Overview
The Local Run Manager CF 139-Variant 2.0 analysis module is for use with
TruSight Cystic Fibrosis 139-Variant Assay. The assay detects 139 clinically
relevant cystic fibrosis disease-causing mutations and variants of the cystic
fibrosis transmembrane conductance regulator (CFTR) gene in genomic DNA
isolated from human peripheral whole blood specimens. The analysis module
performs secondary analysis and report generation from sequencing runs that
use TruSight Cystic Fibrosis. See the package insert TruSight Cystic Fibrosis
Kit (Document # 1000000097720).
The analysis module evaluates short regions of amplified DNA, or amplicons,
for variants. Focused sequencing of amplicons enables high coverage of
particular regions across a large number of samples.
About This Guide
This guide provides instructions for setting up run parameters for sequencing
and analysis for the CF 139-Variant 2.0 analysis module. Use of the software
requires basic knowledge of the current Windows operating system and web
browser-based user interface. For information about the Local Run Manager
dashboard and system settings, see the Local Run Manager Software Reference
Guide for MiSeqDx
(document # 1000000011880).
Enter Run Information
Set Parameters
-
Log in to Local Run Manager.
-
Select Create Run, and then select CF 139-Variant 2.0.
-
Enter a run name that identifies the run from sequencing through analysis.
Use alphanumeric characters, spaces, underscores, or dashes (40 characters or less). -
[Optional] Enter a run description.
Use alphanumeric characters, spaces, underscores, or dashes (150 characters or less). -
Enter the lot number and expiration date for the library prep kit.
Specify Samples for the Run
Specify samples for the run using one of the following options.- Enter samples manually—Use the blank table on the Create Run screen. Suggested sample wells are highlighted.
- Import samples—Navigate to an external file in a comma-separated values (*.csv) format. A template is available for download on the Create Run screen.
Enter Samples Manually
-
Enter a unique sample name in the Sample Name field.
Use alphanumeric characters, dashes, or underscores (40 characters or less). -
Right-click and select positive and negative control samples.
In order to save a run, it must have at least one positive and one negative control. -
[Optional] Enter a sample description in the sample Description tab.
Use alphanumeric characters, dashes, or underscores (50 characters or less). -
[Optional] Select an Index 1 adapter from the Index 1 (i7) drop-down list.
This step is optional because the i7 and i5 index combinations autopopulate with a default layout. -
[Optional] Select an Index 2 adapter from the Index 2 (i5) drop-down list.
This step is optional because the i7 and i5 index combinations autopopulate with a default layout. -
Select the Print icon to display the plate layout.
-
Select Print to print the plate layout as a reference for preparing libraries.
-
[Optional] Select Export to export sample information file.
-
Select Save Run.
If less than 24 samples have been entered, the Insufficient Sample window will display. Select Proceed to continue, or select Cancel to edit the samples.
CAUTION
Proceeding with less than 24 samples is not validated by Illumina. See
TruSight Cystic Fibrosis Package Insert (Document # 1000000097720) for more
information.
Import Samples
Sample information can be imported from two types of files:
- A sample information file that was previously exported from the CF 139-Variant 2.0 module using the Export feature.
- A template file, which can be generated by selecting Template on the Create Run screen. The template file contains the correct column headings for import, with placeholder information in each column. Use an external editor to customize the template file:
- Add sample information for each sample in the run.
- Once all sample information has been added, delete any remaining placeholder information in unused cells.
- Save template file.
To import sample information:
- Select Import Samples, and then browse to the file and select it.
- Select the Print icon to display the plate layout.
- Select Print to print the plate layout as a reference for preparing libraries.
- [Optional] Select Export to export sample information to an external file.
- Select Save Run.
If less than 24 samples have been entered, the Insufficient Sample window will display. Select Proceed to continue, or select Cancel to edit the samples.
CAUTION
Proceeding with less than 24 samples is not validated by Illumina. See
TruSight Cystic Fibrosis Package Insert (Document # 1000000097720) for more
information.
Edit a Run
For instructions on editing the information in your run before sequencing, see
Local Run Manager
Software Reference Guide for MiSeqDx (document # 1000000011880).
Analysis Methods
For a description of the analysis methods for TruSight Cystic Fibrosis, see
the package insert TruSight Cystic Fibrosis Kit (Document # 1000000097720).
View Run and Results
-
From the Local Run Manager main screen, click the run name.
-
From the Run Overview tab, review the sequencing run metrics.
-
[Optional] Select the Copy to Clipboard icon to copy the output run folder path.
-
Select the Sequencing Information tab to review run parameters and consumables information.
-
Select the Samples and Results tab to view the analysis results.
The tab provides a sample list that summarizes the call rate, performance, and information on controls.
If analysis was repeated, expand the Select Analysis drop-down list and select the appropriate analysis. -
[Optional] Double-click the Comment field to enter a comment about a sample in the list. Click Save Changes.
-
Select samples in the list to display detailed variant information. Sample variant information displays below the sample list.
NOTE Each time changes are saved, an analysis results file will be generated in the Alignment folder. This file includes a time stamp appended to the file name, with the format YYMMDD_HHMMSS.
For more information on the Run Overview and Sequencing Information tabs, and
how to requeue analysis, see the Local Run Manager Software Reference Guide
for MiSeqDx (document #
1000000011880).
Lot Tracking Data File
Once analysis is complete, choose one of the following options to view the
LotTracking.txt file.
-
From the Sample and Results tab, click Export Lot Tracking Data.
-
Open the file in the Alignment folder.
The path to the Alignment folder is displayed on the Samples and Results tab in the Analysis Folder field. -
[Optional] Select the Copy to Clipboard icon to copy the analysis folder path.
When analysis is complete, the lot tracking data file is written to the Alignment folder for the run. For example: MiSeqAnalysis\<Run Folder Name>\Alignment_N\YYMMDD_HHMMSS. N is a sequential number that increases if the analysis was requeued. YYMMDD_HHMMSS is the timestamp of the run.
Results Report
After secondary analysis is complete, analysis results are summarized on the
Samples and Results tab. An analysis results (output) file for the assay also
summarizes results in a one tab-delimited text file called
TruSightCF139VariantAssay.txt.
Results in the analysis results file include the information found on the
Sample and Results tab. Choose one of the following options to view the
analysis results file TruSightCF139VariantAssay.txt:
- From the Sample and Results tab, click Export Data.
- Open the file in the Alignment folder.
The path to the Alignment folder is displayed in the Analysis Folder field of the Samples and Results tab.
NOTE
Select the Copy to Clipboard icon to copy the analysis folder path.
The analysis files for each sequencing run are stored on instrument, with
files saved to
MiSeqAnalysis\
Name>\Alignment_N\YYMMDD_HHMMSS. N is a sequential number that increments if
the analysis was requeued. YYMMDD_HHMMSS is the timestamp of the run.
Analysis Result File Information
The analysis results file TruSightCF139VariantAssay.txt includes three
sections: the file header, sample information, and sample variants
information. The file header is listed first, and includes general information
about the run. Note that there is only one file header per run. In contrast,
sample information and sample variants information are included for every
sample used in the analysis. These sections are paired together for each
sample.
The following table provides a description of each row in the file header,
which includes information about the run:
Table 1 File Header (Run Information)
- Test Describes the test that was performed. Run ID Run ID that was generated by the MiSeq Operating Software (MOS) at the beginning of the sequencing run.
- Run Date Date (YYMMDD) that the sequencing run was started in MOS.
- Analysis Version Software version that was used for analysis.
The following table provides a description of each row in the sample information sections: Table 2 Sample Information
- Sample ID The sample name provided when the run was created, combined with the analysis ID assigned by the Local Run Manager software. If you requeue a run, the sample name part of the Sample ID stays the same, while the analysis ID changes.
- This field is in the *.txt file only.
Sample Name The sample name provided when the run was created.
The following table provides a description of each column in the sample variants sections: Table 3 Sample Variants Information
Revision History
- Document Date Description of Change
- Document # 1000000100945 v01 August 2021 Updated EU Authorized Representative address.
- Document # 1000000100945 v00 December 2019 Initial release.
Technical Assistance
For technical assistance, contact Illumina Technical Support.
- Website: www.illumina.com
- Email: techsupport@illumina.com
Illumina Customer Support Telephone Numbers
Region | Toll Free | Regional |
---|---|---|
North America | +1.800.809.4566 | |
Australia | +1.800.775.688 | |
Austria | +43 800006249 | +43 19286540 |
Belgium | +32 80077160 | +32 34002973 |
China | 400.066.5835 | |
Denmark | +45 80820183 | +45 89871156 |
Finland | +358 800918363 | +358 974790110 |
France | +33 805102193 | +33 170770446 |
Germany | +49 8001014940 | +49 8938035677 |
Hong Kong, China | 800960230 | |
Ireland | +353 1800936608 | +353 016950506 |
Italy | +39 800985513 | +39 236003759 |
Japan | 0800.111.5011 | |
Netherlands | +31 8000222493 | +31 207132960 |
New Zealand | 0800.451.650 | |
Norway | +47 800 16836 | +47 21939693 |
Singapore | +1.800.579.2745 | |
South Korea | +82 80 234 5300 | |
Spain | +34 911899417 | +34 800300143 |
Sweden | +46 850619671 | +46 200883979 |
Switzerland | +41 565800000 | +41 800200442 |
Taiwan, China | 00806651752 | |
United Kingdom | +44 8000126019 | +44 2073057197 |
Other countries | +44.1799.534000 |
Safety data sheets (SDSs)—Available on the Illumina website at support.illumina.com/sds.html. Product documentation—Available for download from support.illumina.com.
References
- Support Center
- Support Center
- Safety Data Sheets (SDS)
- Illumina | Sequencing and array solutions to fuel genomic discoveries
- Illumina | Sequencing and array solutions to fuel genomic discoveries
- Legal | Legal Information for this website