euromex IS.1153-PLi iScope Polarisation Microscope User Manual

June 5, 2024
euromex

iScope
polarisation (iscope)

euromex IS 1153 PLi iScope Polarisation Microscope supplementary
user manual

 Introduction

This manual is meant as a supplement to the standard scope manual and only describes the functions and use of the polarization elements of your iscopeeuromex IS 1153 PLi iScope Polarisation Microscope -
fig

Basic controls

Stage Rotation When the stage rotation clamping knob (1) is loosened, the stage can be rotated 360° horizontally
2.1 Using the Bertrand Lens
By revolving the Bertrand lens dial (3), the Bertrand lens can be selected. At the ID position, the lens is removed from the light path. At the “B” position, the lens is engaged
2.2 Focusing the Bertrand Lens
During endoscopic observation, to focus the conoscopic image, turn the Bertrand lens focusing ring (4) slightly until a clear interference image is obtained in the eyepiece

euromex IS 1153 PLi iScope Polarisation Microscope -
fig1Adjustments before observation

3.1 Centering the objectives
All 360° centering objectives of the Euromex IS-series material microscopes are pre-centered in our factory. However, during transportation or after a long period of inactivity, the centering of these objectives may have been shifted
Please follow the steps below in order to re-center the objectives of a polarization microscope

  1. Remove one eyepiece from the microscope head
  2. Insert the widefield eyepiece with crosshairs into the tube of the eyepiece you just removed
  3. Position a microscope slide with a crosshairs reticule under the stage clamps
  4. Check if the 4x objective is positioned in the optical path
  5. Position the round stage with the vernier on its ‘0″ positioneuromex IS 1153 PLi iScope Polarisation Microscope - fig2
  6. Position the middle of the crosshairs of the microscope slide on top of the cross-hairs of the eyepiece (a)
  7. Rotate the stage by 180°. A displacement might be observed (b)
  8. Move by hand, the middle of the crosshairs of the microscope slide, approximately halfway to the eyepiece crosshairs (c)
  9. Rotate the stage back to its °O’ position
  10. The 4x objective is equipped with two adjustable screws inside the revolving nosepiece for centering the objective. Use the centering screws to move the center of the eyepiece crosshairs towards the center of the crosshairs of the microscope slide
  11. Repeat steps 7 to 10 until the objective is centered
  12. Repeat for the other objectives using the same procedure

If the centering cannot be done correctly:

  1. Check if the 4x objective is set in the central position of the centering screw’s correction range. This means that when the centration is changed by using the screws in the revolver, the cross of the slide should be able to move in all directions in an equal amount
  2. Repeat this for the other objectives
  3. If the centering is still unsuccessful, please check if the mechanical stage is correctly centered. The stage is fixed by four screws at the bottom of the stage. Untighten the screws so the stage can be moved and align the stage around the lens of the condenser visually.
    Note that the condenser should first be aligned in the correct manner, the procedure is described in the iScope user manual

Orthoscopic observation

Orthoscopic observation is available for 4x to 100x objectives

  1. Revolve the Bertrand lens dial (1) to ‘0″ position in order to remove the Bertrand lens from the optical path
  2. Swing out the top lens of the condenser
  3. For the reflected illumination system, the polarizer is fixed and the analyzer can be rotated 360 degrees. Rotate the analyzer (2) until complete extinction is obtained
  4. For transmitted illumination, the polarizer can be rotated 360 degrees, turn it until complete extinction is obtained
  5. Place the specimen for orthoscopic observation
  6. Insert test plates for further observation, test, and studyeuromex IS 1153 PLi iScope Polarisation Microscope - fig3

5.0 Conoscopic observation

Use 20x to 100x objectives

  1. Engage the polarizer and analyzer for extinction position
  2. Swing the condenser top lens into the light path
  3. Revolve the Bertrand lens dial (1) to the ‘B’ position, to engage the lens in the light path
  4. Open the aperture iris diaphragm (4) to its largest size
  5. Revolve the focusing dial (S) of the Bertrand lens to focus on the conoscopic imageeuromex IS 1153 PLi iScope Polarisation Microscope - fig4

Note: If the periphery of the conoscopic image is dark, move the condenser vertically to find the position where the periphery is brightest
Euromex Microscopen by • Papenkamp 20 • 6836 BD Arnhem • The Netherlands T +31 (0) 26 323 22 11 • info@euromex.comwww.euromex.com

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals