euromex iScope Face to Face Microscope User Manual

October 30, 2023
euromex

euromex iScope Face to Face Microscope

euromex-iScope-Face-to-Face-Microscope-PRODUCT

Installation and setting

  1. Remove the brightfield condenser by unwinding the screw marked with the arrow. Set up the dark field condenser and tighten the screw
  2. Turn the condenser to the highest position and place a drop of immersion oil on the lens
  3. Lower the condenser slightly, place the slide. In this way, the immersion oil should not yet touch the object rinse
  4. Gently turn the condenser up again. When the oil touches the slide, it is clearly visible through the slide
  5. Focus the image with the least magnification (probably with 4x lens)
  6. Center the condenser with the screws of the condenser
  7. The condenser is now ready for use
  8. To use the lOOx lens and/or the special lOOx lens with built-in iris aperture, you must also put immersion oil on the prepared slided
  9. Focus the lens. On the lens 86.521 (A) and A IS. 7200-i (B) you can adjust the background by changing the opening of the inner lens aperture. The aperture can be changed by turning the ring

Euromex Microscopen b
Papenkamp 20
6836 BD Arnhem
The Netherlands
T +31 (0) 26 323 22 11
info@euromex.com
www.euromex.com
All information may be changed without prior notice v.224211euromex-iScope-
Face-to-Face-Microscope-FIG- \(9\)

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals