Installation and setting
- Remove the brightfield condenser by unwinding the screw marked with the arrow. Set up the dark field condenser and tighten the screw
- Turn the condenser to the highest position and place a drop of immersion oil on the lens
- Lower the condenser slightly, place the slide. In this way, the immersion oil should not yet touch the object rinse
- Gently turn the condenser up again. When the oil touches the slide, it is clearly visible through the slide
- Focus the image with the least magnification (probably with 4x lens)
- Center the condenser with the screws of the condenser
- The condenser is now ready for use
- To use the lOOx lens and/or the special lOOx lens with built-in iris aperture, you must also put immersion oil on the prepared slided
- Focus the lens. On the lens 86.521 (A) and A IS. 7200-i (B) you can adjust the background by changing the opening of the inner lens aperture. The aperture can be changed by turning the ring
Euromex Microscopen b
Papenkamp 20
6836 BD Arnhem
The Netherlands
T +31 (0) 26 323 22 11
info@euromex.com
www.euromex.com