Introduction

Thank you for purchasing the Euromex Delphi-X Observer metallurgical The Delphi-X Observer series has been designed with all kinds of industrial applications and great durability in mind. This resulted in a modern, robust and high level microscope for advanced use, equipped with the best optical and mechanical components. The 25 mm field of view of the eyepieces and the plan apochromatic objectives enable observations with perfect color rendering at high resolving powers. Specific attention to production methods resulted also in an excellent price/performance ratio. Please read this manual carefully before using this product to ensure correct and save usage

  • The contents of this manual are subject to change without notice
  • The appearance of the actual product can differ from the models described in this manual
  • Not all equipment mentioned in this manual has to be part of the set you have purchased
  • All optics are anti-fungus treated and anti-reflection coated for maximum light throughput

General safety instructions

Dangers associated with the operation

  • Improper use could result in injury, malfunction or damage to property. It must be ensured that the operator informs every user of existing hazards
  • Danger of electrocution. Disconnect the power to the entire lighting system before installing, adding or changing any component
  • Not to be used in corrosive or explosive environments
  • Avoid direct exposure of eyes to the collimated light beam or direct light from the light guides or fibres
  • To avoid a hazard to children, account for all parts and keep all packing materials in a safe place

Disinfection and decontamination:

  • Exterior casing and mechanical surfaces must be wiped with a clean cloth, dampened with a disinfectant
  • Soft plastic parts and rubber surfaces can be cleaned by gently wiping a clean cloth, dampened with a disinfectant. Discoloration can occur if alcohol is used
  • The front lens of eyepieces and objectives are sensitive to chemicals. We recommend not to use aggressive disinfectants but to use lens paper or a soft fibre-free tissue, damped in cleaning solution. Cotton swabs may also be used. We recommend you use personal eyepieces without eyes hades in order to minimize risk
  • Never immerse or dip the eyepiece or objective into a disinfectant liquid! This will damage the component
  • Never use abrasive compounds or cleaners that may damage and scratch optical coatings
  • Properly clean and disinfect all possible contaminated surfaces of the microscope or contaminated accessories before storing for future use. Disinfection procedures must be effective and appropriate
  • Leave the disinfectant on the surface for the required exposure time, as specified by the manufacturer. If the disinfectant evaporates before the full exposure time, reapply disinfectant on the surface
  • For disinfection against bacteria, use a 70% aqueous solution of isopropanol (isopropyl alcohol) and apply for at least 30 seconds. Against viruses, we recommend to refer to specific alcohol or non-alcohol based disinfection products for laboratories

Before returning a microscope for repair or maintenance through a Euromex dealer, an RMA (return authorization form) together with a decontamination statement must be filled in! This document – available from Euromex for any reseller- must be shipped together with the microscope at all times

Handle with care

  • This product is a high quality optical instrument Delicate handling is required
  • Avoid subjecting it to sudden shocks and impacts
  • Impacts, even small ones, can affect the precision of the instrument

Handling the halogen bulb

Note: Always disconnect the power cord from your microscope before handling the halogen bulb and power unit and allow the system to cool down approximately 35 minutes to avoid burns

  • Never touch the halogen bulb with your bare hands
  • Dirt or fingerprints will reduce the life span and can result in uneven illumination, lowering the optical performance
  • Use only original Euromex replacement halogen bulbs
  • The use of other products may cause malfunctions and will void warranty
  • During use of the microscope the power unit will get hot; never touch it while in operation and allow the system to cool down approximately 35 minutes to avoid burns

Dirt on the lenses

  • Dirt on or inside the optical components, such as eyepieces, lenses, etc., affects the image quality of your system negatively
  • Always try to prevent your microscope from getting dirty by using the dust cover, prevent leaving fingerprints on the lenses and clean the outer surface of the lens regularly
  • Cleaning optical components is a delicate matter. Please, read the cleaning instructions further on in this manual

Environment, storage and use

  • This product is a precision instrument and it should be used in a proper environment for optimal use
  • Install your product indoors on a stable, vibration free and level surface in order to prevent this instrument to fall thereby harming the operator
  • Do not place the product in direct sunlight
  • The ambient temperature should be between 5 to +40°C and humidity should be within 80% and 50%
  • Although the system is anti-mold treated, installing this product in a hot, humid location may still result in the formation of mold or condensation on lenses, impairing performance or causing malfunctions
  • Never turn the right and left focus knobs in opposite directions at the same time or turn the coarse focus knob past its farthest point as this will damage this product
  • Never use undue force when turning the knobs
  • Make sure that the microscope system can dissipate its heat (fire hazard)
  • Keep the microscope away from walls and obstructions for at least approximately 15 cm
  • Never turn the microscope on when the dust cover is in place or when items are placed on the microscope
  • Keep flammable fluids, fabrick, etc. well out of the way

Disconnect power
Always disconnect your microscope from power before doing any maintenance, cleaning, assembling or replacing bulbs to prevent electric shocks

Prevent contact with water and other fluids
Never allow water or other fluids to come in contact with your microscope, this can cause short circuiting your device, causing malfunction and damage to your system

Moving and assembling

  • This microscope is a relatively heavy system, consider this when moving and installing the system
  • Always lift the microscope by holding the main body and base of the microscope
  • Never lift or move the microscope by its focusing knobs, stage or head
  • When needed, move the microscope with two persons instead of one

Configuration, construction and controls

This chapter describes the main parts and functions of the Delphi-X Observer

1A Eyepieces 1H Nosepiece
1B Photoport 11 Objectives
1C Optical lightpath selector 1J Stage
10 Trinocular head 1K Long distance condenser
1E Metallographicattachment ll X-Y stagemovement
1F Reflected light stopbutton 1M Coarse and fine focusing knob
1G Optional slider (DIC,POL,etc) 1N Body
2A ND filter slots 21 Longdistance condenser
2B Slot for polarization slider 2J Light intensity controlknob
2C Slot for analyser slider 2K Upperand lower lightingswitch
2D Rotating turret 2L Condenser height control
2E Turret position indicator 2M Coarse and finefocusingknob
2F Slider fastening screw 2N Adjustable rackstop
2G Condenser diaphragm adjusting ring 20 Filter switchlevers
2H iCare sensor  
3A Optical pathselector 3J Earthingconnection screw
3B Viewing diaphragm 3K Lamphouse unit
3C Aperture diaphragm 3L USBinterface (future feature,not in use now)
3D Condenser fastening screw 3M Universal tool
3E Condenser heightcontrol 3N Lamphouse unit
3F iCare button 30 Fuse compartment
3G Field aperture control(fortransmitted lightmode) 3P Power socket
3H Lamphouse fastening screw 3Q Power switch
31 Light source plug  

Step 1 I Attaching the focus cassette

  • Attach the focus cassette according to the path shown in figure 1
  • The dovetail slot needs to be aligned with the slot of focusing cassette
  • Slide it down until it reaches the locking pin
  • Then use the hexagon wrench (Allen) tool to tighten the screw shown as I (in figure 2)

Step 2 I Attaching nosepiece (figure 3)

  • Slide the nosepiece into the slot
  • Fix into place with screw

Inserting the objectives
Insert the objectives ranging from low to high magnification (figure 4)

Placing the condenser

  • Use the condenser height control knob to lower the condenser holder to the lowest position
  • Unscrew the two parts of the condenser
  • Insert the top part of the condenser into the holder as shown in figure 5 and secure it by fixing the screw indicated in figure 6. (Centering the condenser is described later in this manual)
  • Then screw the lower part into the top part

Attaching the mechanical X/Y stage

  • Turn the coarse focus knob until the elevating section is brought to the lowest position
  • Attach the mechanical object stage according figure 7 by aligning the stage above the ring of the focus cassette
  • Fix the mechanical stage into place with screw at the back of the stage
  • Place the plass plate (figure 8)

Placing the reflected light attachment and lamp house
The metallographical attachment must be inserted between the body and the head of the microscope in accordance with the serial number in Fig. 5 and transferred to the appropriate position, then tighten the screw with hexagon wrench to fasten

Placing the microscope head

  • Place the head by loosening the screw (a)
  • Mount the head in its position inside the microscope arm
  • Secure it by tightening the screw again

Placing and mounting the eyepieces

  • First remove the dustcover of eyepiece tubes
  • Insert the eyepieces into the eyepiece tubes

Attaching the halogen lamp chamber

  • Slide the lampunit (Halogen) into position at back of microscope base
  • Use the wrench screw tool to secure bolt (VI)
  • Connect the plug

Placing the C-mountor photo port onto the microscope head

  • loosen the screw (a)
  • Place either the (-mount or photo port and tighten the screw

Connecting the power cord
The Delphi-X Observer microscopes support a wide range of operating voltages: 100 to 240 V. Please use a grounded power connection

  • Make sure the power switch is off before connecting
  • Insert the connector of the power cord into the Delphi-X Observer power socket and make sure it connects well
  • Insert the other connector into the mains socket and make sure it connects well
  • Put the power switch to ON

Operation

Placing the sample

  • lower the stage to make room for the sample
  • Bring the 4x objective (or lowest objective in your configuration) into the optical path by rotating the nosepiece until! the right objective clicks into position
  • Gently place the sample into position and make sure it will not damage the objective
  • Use the Xand Y axis control knobs of the mechanical stage to move to area of interest of the sample into the optical path

Switching between light sources
Next to the intensity controller there is a button for switching between transmitted and reflected illumination

Note :transmitted light is only available on some models

  • When the button is pushed in, the light is set to reflected mode
  • When the button is pushed out, the light is set to transmitted mode (standard)

Getting the specimen in focu

  • Use the coarse control knobs to adjust the focus quickly and roughly
  • Get the specimen into sight through the eyepieces
  • Then use the fine focus control knob to adjust the focus in detail

Adjusting the coarse focus tension
Next to the right side coarse focus know there is a ring for adjusting the coarse focus tensions This can be used to make the coarse control move lighter or heavier, according to user’s preference

Setting the focus lock
Next to the left side coarse focus knob there is a ring setting the focus lock. The focus lock can be used to limit the maximum position of the stage at a certain height in order to prevent objectives to get damaged, slides from breaking, or to set the stage at a reference height

  • Move the stage to the desired height then fix the ring tight to lock the mechanical stage’s maximum height
  • The stage still can be lowered but the highest position is now limited to the set position
  • Release the ring to undo the focus lock

Switching the fine focus knobs
The fine focus knobs can be switched from left to right to meet the user’s preference

  • Pull the knobs with moderate force to release the magnet holding the knobs onto the stand
  • Attach the magnets onto the holder and let it grab the knobs again to mount them onto the holder

Adjust interpupillary distance
The Delphi-X Observer has an interpupillary distance range of 47 to 78 mm. The correct interpupillary distance is reached when one round image is seen in the field of view This distance can be set by either pulling the tubes towards each other or pulling them from each other. This distance is different for each observer and this should be set individually. When more users are working with the microscope it is recommended to remember your interpupillary distance for a quick setup during new microscopy sessions

Adjustments of dioptre of a compound microscope
In order to compensate for human eye differences, distortion, thickness differences in cover glasses and tune for the best parfocality between objectives, one can use the dioptre to do so:

  • Take a good prepared slide for your reference

  • Set (both) the dioptre adjustments of the eyepieces to •o•

  • Select the 10x objective, look for a interesting area on the specimen and focus on this area

  • Select the 40x objective and focus on the specimen
    Warning: don’t touch the coarse and fine adjustment anymore

  • Select the 10x objective again

  • With your dominant eye open (close your other eye), rotate the dioptre adjustment from untill the selected area get as sharp as possible as with the 40x objective

  • If during this operation the image becomes unsharp, take your eyes from
    the eyepieces and turn the dioptre adjustment, without looking Into the eyepieces, a few divisions back from

  • Look into the eyepieces again and turn the dioptre adjustment from ‘+’ to ‘-‘ untill the selected area on your specimen gets the optimal sharpness

  • Repeat for your non-dominant eye, and with the second dioptre

Verification:

  • take your eyes from the eyepieces and look for 2 seconds to a far point in the room in order to “reset” your eyes
  • Look again into the eyepieces. If the adjustment is not good, repeat the operation till you reach the same sharpness for the 10x and 40x objective without touching the coarse and micrometric adjustments

The correct eye point
The eye point is the distance from the eyepiece to the user’s pupil. To obtain the correct eye point, move the eyes towards the eyepieces until a sharp image is reached at a full field of view

Select balance between eyepiece- and camera light throughput
The Delphi-X Observer gives users the option to select out of three output types, giving large flexibility when using cameras. The push/pull rod on the side of the microscope head can be set to three positions

  1. POSITION 1 I The optical light path is sent to the eyepieces only. Ideal for when no camera is used
  2. POSITION 21 The optical light path is sent to the eyepieces for 20% only. Ideal standard setting for when a camera is used
  3. POSITION 3 I The optical light path is sent to the camera only. Ideal for when camera is used at low light imaging These positions are Indicated on the head for the user’s convenience

|

Push rod in completely

| .

100/0

---|---|---

| Pullrod towards the middle| 20/80
 | Pullrod out completely| 0/100

Centering the condenser

  • Move the condenser to the top position (1)
  • Focus on a specimen using the smallest objective (f.e. 4x or 1 Ox objective)
  • Close the field diaphragm (2)
  • Use screws (figure 26) to move the field diaphragm into to view center
  • Open the field diaphragm (2) carefully to the outside of the field of view to ensure the field diaphragm is in the center and so the condenser has been centered properly

Using the aperture diaphragm
The aperture diaphragm (figure 27/3) should be used to adjust the numerical aperture, not to adjust image brightness. When the aperture diaphragm is opened to the 70 ~ 80% of objective aperture the ideal position is reached
The simplest way to do this is to use the markings on the condenser

Using halogen with LBD, ND 6 and ND 25 filters
(figure 28: Halogen version with LBD and two ND filters) The halogen version has three filter options:

  1. LBD is a filter for increasing color temperature
  2. ND25 is a filter with 25% light transmittance
  3. ND 6 is a filter with 6% light transmittance

iCare sensor
The unique iCare sensor is developed to avoid unnecessary loss of energy. The illumination of the microscope automatically switches off shortly after the user steps away from his or her position

  • Pushing the iCare button (1) will re-active the light
  • The iCare function is turned on by default
  • To turn off the iCare function push the iCare button for 4 seconds
  • The function will be deactivated and the bright LED (2) will dim to indicate the function has be turned off
  • Repeating this step will turn the function back on

Replacing the fuse
The fuse is placed in a drawer

  • To open it push the drawer aside with a screwdriver
  • Take out the drawer and replace the fuse gently

Reflected illumination unit

Use of the optional color filter

  • Insert the filter slider for the observation purposes into the filter slot {1, Fig. 31). Make sure you insert it from the left
  • The first “click” means that the empty position is in place. The second 0click” indicates that the filter is in the light path

Color temperature| Convert lighting to daylight for general observation and color photography
---|---
Greenfilter| Increase contrast for black-and-white photography
Yellowfilter| Increase contrast when semi-conductor wafer isobserved
Frosted sheet| Difusses light (but reduces intensity)
ND6| Reduce light intensity (transmittance:6%)
ND25| Reduce light intensity (transmittance:25%)

Reflective optical path selection
Rotate the disc (4) in the metallographic attachment according to the required observation path

  • BF1: reflected brightfield observation (transmittance: 6%)
  • BF2: reflected brightfield observation
  • DF: reflected darkfield observation

“BF2” is the standard brightfield position. In case of high i lght intensity “BFI” can be used as it contains a neutral density filter to lower the light intensity

Diaphragm against mid-field view

  1. Put the reflected light stop on the “e” position
  2. Rotate the darkfield/brightfield quick selector (disc) to the BF position
  3. Put the reflected light stop on the “O” position and let the light through
  4. Rotate the nosepiece to place the 10x objective into the light path. The sample is then placed on the stage and the image is roughly focused
  5. Pull the lever (3, fig. 32) until the aperture diaphragm is at its smallest diameter
  6. Use a hexagonal wrench to insert the two field diaphragm centering holes (2, fig. 32).Adjust the image of the aperture to the center of the field of view (fig. 33 shows the adjustment process)
  7. Push the rod (3, fig. 32) to open the aperture until the image of the aperture and the perimeter of the field of view are inscribed. If the image is not correctly aligned, re-align until it is centered
  8. Open the diaphragm so that the image is exactly connected to the field of view

Field stop
The field stop adjusts the lighting area to produce high-contrast images. According to the objective in use, adjust the rod of the field of view (no. 3, fig. 32) of the reflecting illuminator until the aperture image is outside the field of view, to screen out unnecessary light

Note
The diaphragm knob must be pushed forward to fully open the diaphragm

Use of the aperture diaphragm

  1. Move the reflected light stop to the position thus blocking the light path
  2. Rotate the darkfield/brightfield quick selector (disc) to the BF position
  3. Move the reflected light stop to the position to ad mid the light
  4. Turn the nosepiece to bring the 10x objective into the light path, then place the sample on the stage to roughly focus the image
  5. Move the reflected light stop to the “e’ position and close the light path
  6. Rotate the darkfield/brightfield quick selector (disc) to the BF position
  7. Move the reflected light stop to the ·o· position and admit the light
  8. Turn the nosepiece to bring the 10x objective into the light path, then place the sample on the stage to roughly focus the image
  9. Adjust the aperture diaphragm untill it is at its minimum diameter
  10. Adjust the field diaphragm untill it is at its minimum diameter. Meanwhile, the aperture diaphragm image can be seen as well
  11. Insert the hex wrench into two aperture diaphragm alignment holes (3), the position can now be adjusted to the center position (see 35)

Use of the polarizer and analyzer

  1. Remove the dustcover in the polarizer slot then place the polarizer slider with the text facing the operator into the polarizer slot (1). The slider has two positions, one contains the polarizer the other contains a hole for when no polarization is wanted
  2. Remove the dust caps that close off the slot for the analyser then place the analyzer into the slot (text upwards) (2)
  3. Turn the wheel of the analyzer to change the orientation of the polarization filter

Cleaning optics

How to keep the optics clean?
Dust and dirt particles have a negative affect on the image quality. Keeping the optical system of your microscope clean is essential for the best image quality and overall lifetime of your microscope. Dust and dirt on optical elements such as lenses, prisms and filters that are left unattended can become difficult – or even impossible – to remove and may cause mold

  • Place your objective or eyepiece on a secure location
  • Objectives can be screwed into the cover of an objective case
  • Eyepieces can be placed in the microscope box
  • Condensers and collector lenses can remain in place in the microscope
  • To prevent scratches on coatings and optical glass try to remove dirt and dust that sticks to the optical surface first with an air-blower or with pressurized dry air (oil-free and under moderate pressure version only)
  • Use an absorbent lens paper or cotton swap. Dampen a swap or towel with a small amount of lens cleaning fluid or cleaning mixture (either pure iso-propanol or a mixture of 7 parts ether and 3 parts alcohol)
  • Clean the lens by using the tip of the cotton swap or the lens paper. Use enough lens paper so that solvents do not dissolve oils from your hands which can make their way through the paper on to the coated surface
  • When cleaning a large lens surface, wipe with little pressure from the centre towards the periphery in a circular motion. Do not use ziq-zaq motion
  • Discard each lens paper or cotton swap after a single use • Wait until the cleaning fluid is evaporated, or speed up this process by using pressurized dry air
  • Check if the surface is clean by using a magnifying glass
  • Place the cleaned item back on the microscope

Please note that cleaning of the optical surfaces indicated in this instruction only applies to external surfaces of objectives, eyepieces, filters and condensers. Internal surfaces must always be done by yourEuromex microscopes distributor

Troubleshooting

Proper use and maintenance ensure best performance of your Delphi-X Observer. If problems occur this chapter explains how to resolve most issues. Please make sure this chapter is read and checked before contacting your Euromex distributor for service. If a problem is not described in this list or the suggested solution does not bring the result needed, please contact your Euromex distributor

|

Check if the powercableisconnected well, try other powercable

---|---
There i s n o li g h t fromthe l a m p| No power
The bulb is not inserted| Remove the bulband place it back
The bulb isdefective| Remove the bulb and placeitback
The transmitted-reflected illumination switch is in the wrong position| Change position of the switch
The fuse hasblown| Replace the fuse
There is no powerfrom the mains connection| Replace it
The light selector is in the wrong position| Press thelightselector
Shutterisclosed| Open theshutter
The l a mp burns out suddenly| Poor bulbquality| Use the specified lamp to replace it. If the problemis not solved,contact your reseller
Th e bulb flickers o r t h e brightn ess i s

ve rti g in o u s

| The bulb is at the end of it’s life span| Replacethebulb
---|---|---
The bulb is not inserted entirely into the holder| Removethebulband replaceit
iCare se n so r does

notturn off t h e illumination

| There are otherthings in front of microscope (within 1 meter)| Clear all object in a 1 meterradius
The iCarefunction isturned off

..

| Press and hold the iCare button for3 seconds to

activatethe function

llJ’r.fflhrn|  | Bmm
Th e e d ge of the field of v i ew i s dark o r t h e bri g h t n ess is not uniform| The nosepiece isnot in the located position

(objective and light path are notcoaxial)

| Locate the nosepiece properly where it cl i cks
The image of the lampis not centered| Center the l amps
Thelens(objective, condenser,eyepiece or collector) is dirty| Clean it thoroughly
Find dust and sta in

in t h e field of v i ew

| First rotate the eyepieces, if the dust moves:| Clean theeyepieces
Next move the stagewith slide,if the dust

moves:

| Clean the sl i de orreplacetheslide
Next move the condenser up and down, if the dust moves (using 4 or 10xobjective):| Clean the top of the condenser
Next change objective, if the dirt isno longer visible:| Clean thebottom lens of the objective
If problem remains:| Clean the collector lens
Im age quality is not o p t im a l ( r eso lu t i o n o r contrast)| There are stainson the lens(including condenser, objective, eyepiece and collector)| Clean the opticalelements
Thesizeof the aperture diaphragm istoo large| Close the diaphragm
The sizeoftheaperture diaphragm is too small| Open the diaphragm
The position of the condenser is too low| Adjust the position
Periphery of the im age i s d a rk / unclear (unevenly illumin ate d )| Diaphragm(s) closed too far| Open the diaphragm(s)
Lamp unit is not placed correctly| Take out lamp unit andre-install
Incorrect position of the light path switching lever| Set to right position
The nosepieceis not in the right position| Turn the nosepiece u nti l it ‘clicks• into posit i on
Slider(s) is (are) not in correct position| Place inright position
One side of t h e image i s dark| Thecondenser is not centered correctly| Center the condenser
The condenser is placed inclined in it’s holder| Install the condenser againand center it
The nosepieceis not in the right position| Turn thecondenser until it 1 cl i cks into position
Diaphragm(s) is (are) not centered| Center diaphragm(s)
Slider(s) is (are) not in correct position| Place inright position
One part of t h e im age is n ot in focus.Part of the im age becomes out offocus w hil e moving specimen| The condenser is placed inclined in it’s holder| Install the condenser aga i n and center it
The stageistilted| Re-install thestagemak i ng sure it is levelled
The specimen slideis not placed flaton the stage| Replace the slide on thestage
The nosepieceis not in the right position

Thespecimen slideis not prepared well

| Turnthenosepiece until * _it clicks”_ into posit i on
Try aspecimen of known quality and confirm
Im a g e can n ot b e focused w hil e stage i s in hi g h es t**

position

| Focus lock system is secured at the wrong

position

| Release the focuslock , focus and lock again
Thestage is notinstalled correctly| Re-install thestagemaking sure it is levelled
Th e image through the eyepieces is s h ow n asa double im age or half moons appear| lnterpupillary distance hasnot been set correctly| Perform interpupillaryadjustment
Dioptre adjustment has not been done correctly| Performdiopteradjustment
E yes ar e g e tting tir e d| lnterpupillarydistance hasnot been set correctly| Perform interpupillaryadjustment
---|---|---
Dioptre adjustment has not been donecorrectly| Perform diop t eradjus t ment
Brightness is not correct| Adjus t brightn e ssusing intensity control knobor filters
The im a ge i s too dark| Too low intensity set on intensity controller| Increase theintensity by rotating t hecon t roller
The size of the aperture diaphragm is too small| Adjus t again
The position of the condenser is too low| Adjust the posit i on
Poor bulb quality| Use thespecified lamp
Diaphragm(s)closed too far| Open thediaphragms
Light path selector set in wrong position| Sele c t t he 100:0or 20:BOposi t ion
The bulb isat the end of it’s lifespan| Replace thebulb
The Kohler incident light is not in the centre| Adjus t thebolt o f Kohlerincident ligh t
Th e imag e i s too bright| Too high intensity set on intensity controller| Decrease thein t ensity by ro t ating thecontroller
The size of the aperture diaphragm is too large| Adjus t again
The position of the condenser is too high| Adjus t the position
The image appear s blu e- i s h , y e llo w -i s h or or a ng e- i s h| Too low or too highintensity set on intensity controller (Halogen illumination only)| Increase or decrease the in t ensity by rotating the

controller , andusethe NO filters

The bulb isat the end ofit’s lifespan| Replacethebulb
Acolor filter is inserted| Remove filter
The imag e cannot

b e focu s s e d wh e n u s ing high m a gnific a tion

obj e cti ve s

| Focus lock system is secured at the wrong

p o siti o n

| Release the focuslock , focus andlockagain
The stage is not in the highest position| Adjust heightof stage
Thesampleistoo small| Change sample
The objectiv e touch es the s p e cim e n wh e n th e m a gnificati o n is

b e ing changed

| Focus lock system is secured at the wrong position| Release the focuslock , focus andlockagain
Larg e focu s devi a tion whil e changing

objectiv es

| An objective is placed incorrectly, not screwed in allthe way| Make sure t he rightobjec t ive isused and screwitall thewayintotherevolving nosep i ece
Diopter adjustment has not been donecorrectly| Perfo r m d i op t eradjus t ment
Th e slid e does not mov e, o rmov e s too h ea vily| The tension of the X/Y controls of the stageare set too tight| Adjusttension t o proper setting

Documents / Resources

| euromex DX.2153-PLi Delphi-X Metallurgical Observer Clinical Microscope [pdf] User Manual
DX.2153-PLi, Delphi-X Metallurgical Observer Clinical Microscope
---|---

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals