NEUATION iFUGE M24PR Micro Centrifuge User Manual
- June 5, 2024
- NEUATION
Table of Contents
NEUATION iFUGE M24PR Micro Centrifuge
INTRODUCTION
This manual provides important safety information for this centrifuge. It should be kept near the centrifuge for quick & easy reference. This centrifuge is equipped with maintenance free brush less DC motor drive with microprocessor-controlled that can speed up to 15,000 rpm and a safety feature of imbalance detection. This centrifuge is equipped with refrigeration which can be maintained as low as up to -20°C up to 40°C.
INTENDED USE
This centrifuge is designed for usage in Research labs & institutes to
separate the biological samples such as cell suspensions, macromolecules such
as DNA, RNA & proteins and other biochemical suspended solutions.
NOTE: Before using the instrument, please read this user manual carefully. This user manual is intended to assist with the operation and care of the unit only and not its repair. For repair please contact the supplier.
FEATURES
- Maximum volume of 44 x 1.5/2 ml*
- Temperature ranges from -20°C to 40°C
- Temperature of 4 •c maintained even at high speed
- Imbalance detector with auto cutoff for safe operation
- Brushless DC motor for maintenance-free performance
- Large LCD display with speed and time indicator
- Speed setting from 500-15000 rpm
- One press changeover from rpm to RCF functionality
- Digital Countdown Timer
- Reduction adaptors for 0.1/0.2 & 0.5ml tubes
- Emergency lid release during power failure
- Automatic internal diagnosis & error display
- Silent operation
(*Available with 4 optional rotors)
STANDARD ACCESSORIES
- T – Allen wrench.
- Adaptors for 0.4/0.5 ml tubes.
- Adaptors for 0.1/0.2 ml tubes.
- Product user manual and warranty card.
- Power cord: 220-240V, SOHz.
NOTE: Rotors has to be ordered separately. (optional accessory)
TECHNICAL SPECIFICATIONS
Motor Type | Brushless DC Motor |
---|
Rotor Capacity
| 24 x l. 5/ 2ml tube, 44 x l. 5/ 2ml tube,
8 x 4 x 0.1/0.2ml PCR tube/strips, 8 x 5ml tube
Speed Setting| Variable 500 – 15000 rpm
22388 x g (Max RCF)*
Speed Accuracy| ± 100 RPM
Temperature Accuracy| ± 1 ·c
Temperature Range| -2o· c to 40°c.
Run Time| 30s to 999min & 59s
Acceleration Time| < 72 second s
Deceleration Time| < 74 seconds
Noise Level| <60dB
Ambient Temperature| 5 ·c to 40 ·c
Permissible Relative Moisture| 20to 80%
Size (W X D X H)| 325 x 720 x 305 mm
Weight| 43 kgs
Input Power| 220-240V, 50Hz
Power Consumption| 650W
SAFETY PRECAUTIONS
- Never use the centrifuge in any manner not specified in this manual.
- Using equipment in any manner not specified in this manual or by the manufacturer will void the warranty.
- Never move the centrifuge while the rotor is spinning.
- The rotor and the rotor lid must always be securely fastened. If the centrifuge makes unusual noise during operation, the rotor or rotor lid fit needs to be checked. Switch OFF the device immediately by pressing STOP, check rotor fit & fasten it well.
- The rotors must be loaded symmetrically. Each tube should be counterbalanced by another tube of the same weight.
- Do not use the centrifuge or rotor that have not been correctly installed or damaged.
- Repairs must only be performed by authorized service technician.
- Using incorrect rotors & wrong spare parts will void the warranty.
- Centrifuge may be used for the specified applications only. It must not be It operated in a hazardous or flammable environment and must not be used ffi to centrifuge explosive or highly reactive substances.
- If liquids are spilled on the rotor or rotor chamber, the centrifuge must be cleaned carefully and properly before being used again.
- Prior to centrifugation, the tubes should be visually inspected for material damage. Damaged tubes may not be centrifuged. This is because broken tubes can, in addition to sample loss, create imbalance which can result in further damage to the centrifuge and accessories.
- The maximum capacity of tubes must not be exceed its corresponding rotor capacity. Do not use liquid with density higher than 1.2g/ml for full load operation.
- Do not lean on the equipment. It may damage the equipment or the harm the operator.
- When moving the centrifuge from a cold room to a normal room, run the centrifuge for 30 minutes beforehand in the cold room to avoid condensation. Alternately, allow it to warm up in the lab for at least 3 hours before use, but do not plug in the centrifuge in order to prevent possible damage by condensation.
- Be sure to close the tubes lid tightly prior to centrifugation. Open tubes lid can be torn off during centrifugation and can damage the rotor lid or centrifuge.
- Rotors and rotor lids are high-graded components which are subject to extreme mechanical strain. Even slight scratches and tears can lead to serious internal material damage. Ensure to check the rotor for any signs of damage before use. Rotor & rotor lid showing visible signs of corrosion or mechanical damage should not be used, contact your local supplier or mfg for replacement rotor & parts.
- Do not fill tubes while they are in the rotor. Liquid spillage may harm the device.
- In the event of contamination caused by aggressive agents, the rotor must be cleaned immediately using a natural cleaning liquid (like water). This is particularly important for the bores of the tubes. If any damage is seen, contact the service technician.
- Before using cleaning or decontamination methods, other than those stipulate by the manufacturer, contact the manufacturer to ensure that the intended method will not damage the centrifuge.
- The power cord given with centrifuge unit is designed to use for that particular centrifuge. Do not use any other power cord, using any other power cord /adaptor may damage the centrifuge and will void the warranty.
INSTALLATION
Open the box, then remove the packaging and gently take the centrifuge out of the box. Before using this centrifuge open the centrifuge & remove all packaging from the rotor chamber & ensure rotor is firmly tightened. The user manual and accessories should be kept near the centrifuge. Please keep all packaging in safe storage for at least 2 years for warranty purpose.
MOUNTING
Place the centrifuge on a flat and leveled surface, ensure that the four feet of this centrifuge stand on the surface firmly. Avoid installing on a slippery surface or surface prone to vibration.
- Ideal ambient temperature is 20°c ± S°C, avoid placing the centrifuge in direct sunlight.
- Keep clearance of at least 30-35 cm on both sides and at least 20 cm behind it to guarantee cooling efficiency.
- Keep away from heat or water to avoid sample temperature issues or centrifuge failures
- Do not place the centrifuge in any area where operating the unit maybe difficult.
CONNECTING POWER CORD
STANDARD PARTS LISTING
USER INTERFACE & DISPLAY
Item | Name | Function |
---|
1
|
RPM Button
| Single press, speed display will beblink to set. To select the RPM value to
be set for centrifugation
2| RCF Button| Long press RPM/RCF will toggle between RPM to
RCF. To select the RCF value to be set for centrifugation
3
|
TEMP. Button
| Single press button to set the temperature from -20°C to40°C.
4| PROG| Press button to enter in Program Mode and to select any Program out
of the 99 available programs.
5| LINE| Press button to select line of any specific program out
of 4 available lines in each program.
---|---|---
6
| ****
RPM/RCF
| Press button to select SPEED value in rpm mode. And press+/- key to set desired run RPM. Press and Hold to
switch to RCF mode. Press+/- key to set desired RCF.
7
| TEMP| Press button to set the temperature value. And press
+/- key to set desired temperature.
8
| TIME| Press button to selectT IM E mode. And press+/- key to
set desired run TIME.
9| START| Press button to START to start the centrifuge.
10
| ****
STOP/OPEN
| Single press Stop/Open button, will stop motor. Second press Stop/Open button, will open the lid once the motor stop. Also used for lid open if centrifuge motor is
not running.
11| +/- Keys| Use to set speed, Temperature and time values. Press “+” to in cre aseand ” -” to decrease values.
12
| QUICK COOL| To start a temperature cont ro l run at optimized (S000)rpm to achieve maximum cooling
13
| QUICK SPIN| To set short spin on a set speed. Press short spin continues
press for Quick spin operation.
14| ROTOR| To select among multiple rotor listed below.
ROTOR INSTALLATION
ROTORS AND ACCESSORIES
The table below shows rotors compatible with the centrifuge and the Max. RCF
with different tubes and adaptors.
NOTE: Always use recommended accessories for best results and product safety. Optional rotors & additional accessories must be ordered separately.
ROTOR REMOVAL AND REPLACEMENT PROCESS
The rotor comes pre-installed with the centrifuge. If you want to remove or
replace the rotor, follow the instructions below.
REMOVING THE ROTOR
- Do not remove or loosen the rotor lid before attempting to remove the rotor.
- Using the T -Allen Key, loosen the rotor nut by turning it counterclockwise. Do not try to pull the rotor, the rotor will come up automatically.
- Once the rotor nut is loosen completely, pull up the rotor vertically.
REPLACING THE ROTOR
- To replace or install the rotor, take the rotor and load vertically onto the shaft.
- Place the rotor nut in the center hole of the rotor onto the motor shaft.
- Put T-Allen Key into the rotor nut and turn clockwise to tighten and counter clockwise to loosen the rotor.
- After properly fastening the rotor, place the rotor lid on the rotor lid nut by hand and rotate the rotor lid nut clockwise.
NOTE:
- Check the rotor is firmly tightened before running the next program.
- Do not remove or loosen the rotor lid before removing the rotor.
BALANCING THE ROTOR
- Always balance the rotor before centrifugation. Above are examples of properly balanced rotors.
- The samples in the tubes should be of equal volume.
- If the tubes are not loaded correctly – vibration or imbalance can occur which can cause serious damage to the centrifuge.
- If the tubes are not loaded symmetrically then the imbalance detector will cut off the running centrifuge for device & user safety. This will stop the centrifuge and Err 55 will be seen indicating tubes are not loaded symmetrically. To resume operation, load tubes symmetrically & restart the centrifuge.
- Incorrect method of loading tubes in centrifuge rotor:
OPERATING THE CENTRIFUGE
STARTING THE CENTRIFUGE
After connecting the power cord, switch ON the power at the rear side of the
centrifuge. Make sure to check the rotor fit before use. Centrifuge will not
operate with an open lid.
NOTE: Maintain a gap af 3 seconds between turning aft and turning on the
centrifuge. DO NOT turn off and turn on the centrifuge instantly.
REMOTE OPERATION & PROGRAMMING
SWITCH ON THE CENTRIFUGE
After connecting the power cord. Switch ON the main power supply & then Switch
on the power switch located on the right side of the instrument. Make sure to
check the rotor fitment before use. Centrifuge will not operate with open lid.
NOTE: Maintain a gap of 3 seconds between switch OFF and switch ON again. DO NOT switch OFF and ON again instantly.
SPEED SETTING
After closing the centrifuge lid, press “RPM/ RCF” to select speed setting in
RPM/ RCF mode. Press “+” to increase the speed value and press “-” to decrease
the speed value. The minimum and maximum speed (RPM) of the centrifuge is 500
rpm and 15000 rpm respectively. You can quickly change the increment/
decrement value by rotating setting knob quickly. Value is accepted when left
idle indicated by multiple blinks of the set value.
The speed can also be changed while the centrifuge is under operation. Press
the RPM / RCF button & use setting knob to change speed as instructed above.
Changing the speed between the ongoing centrifugation will run the centrifuge
at updated speed for the remaining time indicated by the timer.
TIMER SETTING
Press “TIME” button to set the run time. After selecting press “+” to increase
the timer value and press”-” to decrease the timer value. Value is accepted
when left idle indicated by multiple blinks of the set value. The centrifuge
timer can be set from 30 s to 999mins and 59s. The timer in the centrifuge is
countdown timer and time in the display will be in “Min/Sec”. The same will
showed on the display.
- To set the timer in minutes, pressing the time button “ONCE”. For example: if time is005:00 (5 Mins 0Sec), then next time will be 006:00 (6 Mins0 Sec).
- To set the timer in seconds, pressing the time button “TWICE”. For example: if time is005:00 (5 Mins 0 Sec), then next time will be 005:01 (5 Mins 1 Sec).
- Pressing time button more than 2 times will start the process from point 1 again.
- The input will be accepted if we leave the setting knob idle for 3 seconds. The value be saved automatically after five blinks.
TEMPERATURE SETTING
Press temperature button for setting the temperature from -20 °C up to 40 °C.
The parameters can be set when the Temperature value blinks by pressing “+”
key for increasing the temperature & “-” to reduce the temperature. The values
will automatically saves after screen stops blinking.
REMOTE OPERATION & PROGRAMMING
QUICK COOL SETTING
PRESS Quick Cool Key, indicated by (cubic rotation) on top right (Indication
will be in dotted rotation in place of rotor digit indication). During quick
cool operation rpm:FIX 5000 & TEMP:-On set value. Press START button for
centrifuge to start at FIX RPM & set TEMP. After reaching set TEMP, motor
stops thus RPM becomes zero but compressor continues to keep the chamber
temperature at the set value. The quick cool works for 1 hour only. (for
indication of refrigeration running process, TEMP display will blink) If
timers stops and TEMP does not reach to set value, TIME will switch-over from
beginning. “RPM/RCF, Timer, Rotor, Line, Prog”, “Quick Spin” button not work
during “Quick Cool” operation.
SHORT SPIN OPERATION
Short Spin Centrifugation is the feature for short/pulse/quick run. It will
run as long as the button is pressed. Set rotational speed prior to short spin
as required. During short spin the timer will be in incremental mode. After
releasing short spin button the time in the display will show duration of
short spin.
ROTOR SELECTION
Different rotors has different radius which means that RCF for that rotor
would also be different. Proper selection of rotor is mandatory to achieve
proper RCF. Press “Rotor” button and then rotate setting knob to select the
rotor from available 4 rotors. Press “Rotor” button . See below chart to
identify rotor no. from its volume:
Rotors
Rotor No.| Name| rpm| RCF at max rpm
1| 24 slot x 2ml rotor| 15000| 21420 g
2| 4 x 8 PCR rotor| 15000| 15397 g
3| 8 slot x 5 ml roto r| 15000| 18461g
4| 44 slot x 2ml rotor| 15000| 22388 g
–
seconds of the adjustment. Then so the max value of speed and acc/dcc to be changed as per the rotor selection. (Max. Speed (RPM/RCF) and Ace/Dec vaule for all the rotors have been added in the software) Now set the desired speed (RPM/RCF), time & temp and press START button to store and run the selection.
START AND STOP OPERATION
Press “START BUTTON” to start operation and press “STOP/OPEN BUTTON” to stop
the ongoing operation. When the centrifuge is running the symbol “0” will
rotate.
Pressing the “STOP/OPEN BUTTON” will stop the operation. Once operation is
stopped, press Button again to open the centrifuge lid. If the time gets over,
centrifuge will stop automatically. When the centrifuge is not running the
symbol ” 0 ” will be idle. To open the lid in non-operating stage, press the
“STOP/OPEN
BUTTON”.
After completing run, before any another operation, it is mandatory to open
the lid once before starting new operation. Centrifuge will not do 2nd
operation if lid is not opened and closed for at least one time after
completion of 1’1 operation.
NOTE: It is mandatory to open the lid once after completion of operation for 2nd operation. Centrifuge will not start if lid is not opened and closed after completion of a operation.
PROGRAM MODE
Long press “PROG” button to enter into program mode. Short press select
program and rotate setting knob (clockwise & anticlockwise) to select desire
program out of 99 available programs. User can select any program from 1 to 99
programs and can have user specific setting parameter for all programs. Each
program has 4 lines indicated by Ll, L2, L3, L4. Press “LINE” button to enter
in line selection mode and rotate setting knob to select line out of 4 lines.
Every line of the program can have different values for all setting
parameters. User can set and save different speed, different time, different
temperature for any line of program. Only Rotor will not remain save for any
program. Rotor selection is not Program specific. Rotor selection is universal
and it should be set before setting any other parameters.
On first time usage, all programs will have zero (0) values and once the
“PROG” button is pressed below display will appear indicating selection of
program mode.
As per above image, the circle O left of “P0l Ll” is blank indicating no values in any line of specific program. Once any line of Program in filled or set, below change of display will appear.
The circle @) left to “P0l L3” will have a tick mark indicating that any one line of the specific program is filled. If any line leading to selected line is set, then the arrow below the line number will appear.
If no leading line is filled or last line L4 is selected then the arrow below
Line will not appear. It indicates there is no further operation left as
leading lines of the specific program are empty.
program and selected rotor remains active for all programs until new rotor
number is not selected. Select the rotor, Press “PROG” button to enter in
Program mode, select specific program using setting knob, Press “LINE” button
to select specific line and set different parameter for that particular line
of program. Values for any parameter gets save after 3 blinks. Once Program is
set, press “START” button to start program.
After completion of any operation, it is mandatory to open the lid at least
once for another operation. Open the lid, close it again and press START for
another operation.
NOTE:
- Rotor is not program-specific. Selected rotor will be active for any new operation.
- It is mandatory to open the lid once after completion of operation for 2nd operation. Centrifuge will not start if lid is not opened and closed after ter completion of an ongoing operation.
IMBALANCE DETECTION
The centrifuge is equipped with an imba’lance detection safety feature. When
the rotor is not loaded symmetrically, the imbalance detector gets activated
and will cut.
REMOTE OPERATION & PROGRAMMING
off the centrifugation. The error “Err 55” will be shown on the display. First
correct the imbalance load then switch OFF & switch it ON again. The values
will be same as set before imbalance. The imbalance detection feature cannot
be deactivated, as it is factory fitted.
OPENING CENTRIFUGE LID IN POWER FAILURE
Disconnect the centrifuge from the main power supply. Wait until the rotor has
come to a standstill (this may take longer time). Once the rotor has stopped,
Pull the Emergency Lid release lever by hand or by a small Allen Key located
on the left side of the machine as show in image below. This will open the
centrifuge lid.
NOTE: This method of opening the lid should be used only in case of emergency or power failure.
MAINTENANCE AND CLEANING
- The rotor and the outside of the centrifuge should be cleaned regularly with a mild wet (with water) cloth.
- Ensure that while cleaning the unit is not plugging in.
- Wear protective glove & safety glass while operating & cleaning the device.
- The brushless motor in the centrifuge requires no routine maintenance. Any required service should be performed by authorized, qualified personnel only. Repairs performed by unauthorized personnel may void the warranty.
- Always keep the centrifuge housing, rotor chamber and rotor clean. All parts should be wiped down periodically with a soft cloth. For more thorough cleaning, use a neutral cleaning agent (Ph between 6 and 8) and clean with a soft cloth. Exclusive amounts of liquid should be avoided.
NOTE: Liquid should not come into contact with the motor.
- After cleaning, ensure that all parts are dry before re-use.
- Regularly cleaning of the rotor is important.
- If the rotor chamber needs cleaning, clean with cloth or sponge moistened with a neutral detergent solution.
- Do not place the rotor into the cleaning solution.
- If corrosive, toxic or pathogenic bacteria are accidentally spilled in the rotor or rotor chamber the centrifuge must be decontaminated throughly.
TROUBLESHOOTING
This centrifuge has a self- diagnostic function. If a problem occurs, an error/warning code will be displayed on the display screen and the operator can determine the malfunction with the warning code below.
ERROR | PROBLEM | SOLUTION |
---|
No display
| No main power connection.| Power check & prope r plug-in of main cable at
both ends.
Power failure.| Check the mains fuse.
Improper connection.| Connect adaptor properly.
| Lid not closed correctly.| Close lid correctly and restat
unit.
Error with lid closing and opening mechanism.|
Contact service.
Error 1
| Latch jammed or any
limit switch of latch got damaged.
|
Contact service representat ive.
Error 41
|
Selected temp value not reachabl e for selected set speed.
| Set temp value will be taken automatically that it can be archived after 1 hour, Error indication will be displayed
every 15 second. restart unit.
Error 42
|
Temperature is out of control.
| Error indication will be displayed every 15 second.
If it persist in every operation consistently, contact service
representative. restart unit.
Error 43
| Over temperature inside chamber.| Turn off the centrifuge and wait until temperature goes
down. restart unit.
Error 44 | Temperature sensor failure. | Contact service representative. |
---|
Error 52
| ****
Rotor stuck or incorrect operating voltage.
| Turn off the centifuge, check rotor fitment or apply correct 230 VAC ± l0VAC operating
voltage. restart unit
Error 55| Rotor not loaded
symmetrically.
| Load rotor symmetr ically &
restart centrifuge. restart unit.
Centrifuge lid cannot be opened.
| Rotor is still spinning.| Wait for the rotor to come to a stop.
Power Failure.| Emergency lid relea se after rotor stop.
Centrifuge shakes during acceleration & exceptional running noise.
| Rotor not loaded
symmetrically.
| Load rotor symmetrically &
restart operation.
Either a brok en tube, damage to the rotor or motor is cause for run
noise.
| ****
Replace broken tube. For damaged rotor/motor contact service representative.
Rotor damaged.| Remove & change rotor.
Display Error.| Loose connection of
display.
| Contact service representative.
Power tripping.| Cable is not fit properly.| Remove cable and connect
properly.
System gets hang.| Electronic error.| Switch OFF centrifuge and then switch it ON again. If the error still shows, contact service representative.
IMPORTANT NOTE
- Maintain a 3 second gap between restarting the centrifuge. Instant ON-OFF can lead to a reset, erasing last run memory.
- If the motor gets hot due to which there will be a fluctuation in speed value then allow the centrifuge to cool for at/east 30 minutes. Do not do any operation for 30 minutes.
- Do not use liquids with density higher than 1.2g/ml for full load operation.
WARR ANTY STATEMENT
This product is warranted to be free from defects in material and workmanship
for a period of two (2) years from date of purchase. Your product will be duly
repaired upon prompt notification in compliance with the following conditions:
This warranty is valid only if the product is used for its intended purpose
and within the guidelines specified in this instruction manual. This warranty
does not cover damage caused by accident, neglect, misuse, improper service,
natural forces or other causes not arising from defects in original material
or workmanship. This warranty does not cover any incidental or consequential
damages, commercial loss or any other damages from the use of this product.
The warranty is invalidated by any non-factory modification, which will
immediately terminate all liabilities on us for the products or damages caused
by its use. The customer shall be responsible for the product or use of
products as well as any supervision required for safety. If requested the
products must be returned to the distributor in well packed and insured manner
and all shipping charges must be paid.
NOTE: Some states da not allow limitation on the length of implied
waranties or the exclusion or limitation of incidental ar consequential
damages. This warranty gives you specific legal rights. This warranty is given
expressly in lieu of all other warranties, expressed or implied.
The purchaser agrees that there is no warranty of merchantability or of
fitness for any intended purpose and there are no other remedies or
warranties, implied, which extend beyond the description on the face of the
agreement. This warranty is only applicable to the original purchaser.
Products received without proper authorization will not be processed for
warranty or service. All items returned for service should be sent with
postage prepaid in the original packaging or other suitable packaging, padded
to avoid damage. We will not be responsible for damage incurred by improper
packaging.
NOTE: This warranty is valid only if the waranty is registered with the supplier within 30 days fram the date of purchase.
Fo r you ref erenc e , make a not e of serial number , dat e o f purchas e an d supplier here.
Serial No.: Purchase Date :
Supplier:
PRODUCT DISPOSAL
In case the product is to be disposed of, the relevant legal regulations are
to be observed.
Information on the disposal of electrical and electronic devices in the
European Community.
The disposal of electrical devices is regulated within the European Community by national regulations based on EU Directive 2012/19/EU on waste electrical and electronic equipment (WEEE). According to these regulations, any devices supplied after 13.06.05 in the business to business sphere, to which this product is assigned, may no longer be disposed off in municipal or domestic waste. They are marked with the following symbol to indicate this.
As disposal regulations within the EU may vary from country to country, please contact your supplier if necessary.
TR TRANSPORTATION &STORAGE
This devices is heavy & weighs approximately 43kgs. Care must be taken while lifting up. Always take few people help to lift this instrument. The refrigerated centrifuge must be kept always in upright position while storing/transporting from place to place.
- Use only original packaging during transportation
- For longer distance take transportation aid like hard trucks
- Avoid knocking, harsh shaking or jolting the device
- Always retain the packaging material & transportation protections for longer storage or transportation
- The transportation conditions for the instrument is -25° C to 60° C with a Relative Humidity up to 80% & max pressure of 106kPa for both general & air transportation
- For storage the instrument is recommended to be stored in original package. the instrument is -S°C to 45°C with a Relative Humidity up to 70% & max pressure of 106kPa
Neuation Technologies Pvt. Ltd.
Plot No. 15, GIDC Electronic Park SEZ Kolavada Road, Gandhinagar – 382026,
Gujarat, India
Website: www.neuation.com