SKC AirChek 52 Sample Pump Instruction Manual

June 5, 2024
SKC

AirChek® 52 Sample Pump
Cat. No. 224-52 Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
724-941-9701
skcinc.com

Figure 1. AirChek 52 Sample Pump

Introduction

Description
The AirChek 52 Sample Pump (Figure 1) is designed for rugged industrial use at flows from 5 to 3000 ml/min. Ideal for on-worker applications, use the AirChek 52 for short-term or full-shift sampling with sorbent tubes, impingers, cyclones, PPIs, or filter cassettes.
Checking Pump/Kit Contents
Use the table below to verify that you received all items associated with the Cat. No. ordered. If you are missing items, contact SKC at 800-752-8472 (the U.S. only) or 724-941-9701.

If You Ordered Cat. No.|

Your Package Should Contain

---|---
224-52| AirChek 52 Sample Pump with NiMH battery pack and screwdriver set
224-52-S| AirChek 52 Sample Pump with NiMH battery pack and screwdriver set, single PowerFlex charger with cable, Tygon tubing (3 feet, 1/4-inch ID), and collar clip with a cable tie
224-52K| AirChek 52 Sample Pump with NiMH battery pack and screwdriver set, single PowerFlex charger with cable, filter cassette holder, Tygon tubing (3 feet, 1/4-inch ID), and soft-sided nylon carry case

Required Equipment
1/4-inch ID tubing
PowerFlex® Charger
**** Low flow accessories if sampling at 5 to 500 ml/min. See Accessories.

Getting Started

Charge the Battery Pack
For optimum charge, ensure that the pump is not running.

  1. Activate the charger (follow PowerFlex Operating Instructions 40069).

  2. Insert the charging cable connector into the charging port on the charger. See right.

  3. Insert the charging plug end of the cable into the jack on the back of the pump’s battery pack (Figure 1).

  4. Charge the battery completely (the LED at the port connected to the pump is a steady green) before using the pump.

  5. Disconnect the charging cable from the back of the pump’s battery pack.

Notes and Cautions

  • To comply with intrinsic safety regulations, do not charge or operate the pump from the charger in hazardous locations.
  • Using a non-approved charger void any warranty.
  • Using a repaired or rebuilt battery pack voids any warranty and the UL Listing for intrinsic safety.
  • Using any device other than the approved battery pack to power the pump voids the UL Listing for intrinsic safety and any warranty.
  • Ensure proper orientation of the charging cable before plugging it into the charging jack. Improper orientation/contact will short-circuit the battery and voids any warranty.
  • Short-circuiting the battery pack will render it immediately inoperative.
  • Failure to follow warnings and cautions voids any warranty
  • The battery pack may be kept on the SKC-approved charger for an indefinite time.

For more information on SKC pump battery packs, visitwww.skcinc.com/knowledgecenter.

Use the Pump with AC Power (Power option in non-hazardous locations)
Do not use the pump with the Battery Eliminator in hazardous locations. UL Listing for intrinsic safety is not in effect during pump operation with the Battery Eliminator.
The Battery Eliminator is an accessory that converts alternating current (AC) to direct current (DC) from which the pump can be operated for extended runs. The Battery Eliminator should be used in non-hazardous locations only. See Accessories.

  1. Remove the battery pack from the pump. See Replace the Battery Pack.
  2. The Battery Eliminator is comprised of two pieces, a wall cube, and a power adapter. Plug the wall cube into a standard wall outlet and insert its plug end into the power adapter.
  3. Fit the power adapter on the pump in place of the battery pack.

Operation

High Flow Applications (1000 to 3000 ml/min)
Set/Calibrate Flow Rate

  • Allow the pump to equilibrate after moving it from one temperature extreme to another.
  • Charge battery completely before calibration and sampling.
  1. Ensure that the battery is fully charged and that the pump has run for five minutes before calibrating. Leave the pump running.
  2. Prepare the calibrator. See calibrator instructions.
  3. Set up a calibration train: Using flexible tubing, connect the calibrator outlet (suction port) to the representative sample medium inlet. Using ¼-inch Tygon tubing, connect the sample medium outlet to the pump inlet. See right.
  4. Using a screwdriver, set the pump flow rate by turning the flow adjustment screw on top of the pump (Figure 1), clockwise to increase or counterclockwise to decrease flow, until the calibrator indicates the method-specified flow rate. Take a minimum of three readings and record the average flow rate, as per OSHA/NIOSH instructions.
  5. Press the on/off button to turn off the pump. Disconnect the calibrator and flexible tubing from the representative sample medium and proceed to Set Up/Sample.

Set-Up/Sample

  • Allow the pump to equilibrate after moving it from one temperature extreme to another.
  • Protect sample pump from the weather when sampling outdoors.
  • Powering the pump with any device other than the approved battery pack voids the UL Listing for intrinsic safety.
  • Do not use the pump with the Battery Eliminator in hazardous locations. UL Listing for intrinsic safety is not in effect during pump operation with Battery Eliminator.
  • Use of any device other than the approved battery pack or Battery Eliminator to power the pump voids any warranty.
  • Charge battery completely before calibration and sampling.
    1. Replace the representative sample medium with a new unexposed medium for sample collection.
    2. Place the sample medium were appropriate for sampling. a. For personal sampling, clip the sample collection medium to the worker in the breathing zone and the pump to the worker’s belt using the belt clip.

Clip sample medium to worker and pump to the belt.

When using an impinger, an in-line trap (SKC Cat. No. 225-22 or 225-22-01) is required between the pump and impinger to prevent fumes from accidentally being drawn into the sampler. Failure to use an appropriate in-line trap during impinger sampling voids any warranty.
b. For sampling with an impinger, mount the single or dual impinger and trap (SKC Cat. No. 225-20-01 or 225-20-02) to the face of the sampler using the accessory mounting screws (Figure 1) or place them in a holster at the worker’s waist.

Impinger holder on the pump with impinger and trap

3. Press the on/off button to clear the elapsed time displayed on the LCD and to start sampling. Record the start time and other pertinent information. The LCD will automatically display elapsed pump run time.
See below.
Possible Displays During Sampling
Flow or Battery Fault Shutdown – If the pump is unable to compensate due to excessive back pressure or a low battery condition exists, it will shut down and timing functions will pause. The LCD will display either a battery- shaped icon or a flow fault icon ( ) depending on the cause of the shutdown. Upon flow fault, the pump will attempt to restart up to five times. To restart from flow fault, correct the blockage and press the on/off button twice. The elapsed run time display will reset to 0 when the pump is restarted. If the battery icon is displayed, recharge the battery before sampling.
Displayed Elapsed Run Time – Elapsed run time is displayed continuously on the LCD. For elapsed times after 999 minutes, the display still shows elapsed time but alternates between displaying the first two digits of the elapsed time and the last three digits. For example, a pump that has run for 1,440 minutes would display first “01” and then “440.” The display maximum is 99 999, which is 99,999 minutes.
4. At the end of the sampling period, press the on/off button and record stop time and other pertinent information.
High Flow Applications (Cont)
5. Cap the sample and send it with blanks and pertinent sampling information to a laboratory for analysis.
6. Verify the flow.
a. Turn on the pump and reinstate the calibration train and sample medium.
b. Take three readings and record the average value as the post-sample flow rate. Do not adjust the pump flow rate during this step.
c. Compare the pre and post-sample flow rates. Note in sampling documentation if the values differ by more than ± 5%.

Low Flow Applications (5 to 500 ml/min)

  • Single-tube sampling requires the All-in-One Low Flow Holder; see the operating instructions for the All-in-One for details on its operation.
  • Multiple-tube sampling requires a Constant Pressure Controller (CPC) and a Dual, Tri, or Quad Adjustable Low Flow Tube Holder accessory; see the operating instructions for the CPC and Adjustable Low Flow Tube Holder for details on their operation.

Using the All-in-One Low Flow Holder (Figure 2)
Set/Calibrate Flow Rate for Single Tube
Calibration train with All-in-One Low Flow Holder connected to AirChek 52

  • Allow the pump to equilibrate after moving it from one temperature extreme to another. · Charge pump battery completely before calibration and sampling.
  • Calibrate/verify pump flow rate before and after each sampling operation using the tube holder and pump to be used for sampling.
  • Set the flow rate through the pump first and then calibrate the flow rate through the representative sorbent tube.
  • Two small inlet holes on the bottom of the built-in CPC of the All-in-One can become blocked. Periodically inspect and, if needed, clean with a small pick and blow particles away with a puff of air.

Pump Flow Rate

  1. Ensure that the battery is fully charged and that the pump has run for 5 minutes before calibrating. Leave the pump running.
  2. Prepare the calibrator (see calibrator instructions). Using flexible tubing, connect the calibrator outlet (suction port) to the pump inlet.
  3. Set the pump flow rate to 1.5 L/min. Using a small screwdriver, turn the flow adjustment screw on top of the pump (Figure 1), clockwise to increase flow or counterclockwise to decrease flow until the calibrator indicates 1.5 L/min.
  4. Remove tubing from the pump inlet.

Flow Rate Through Sorbent Tube

  1. Attach the Tygon tubing of the All-in-One to the pump inlet. Note: If the pump flow faults shortly after the holder is attached, check that the needle valve is open by using a small flat-head screwdriver to turn the flow adjust screw on the holder slightly counterclockwise. If the flow fault continues, check that the two small holes on the bottom of the built-in CPC are not blocked. If needed, clean them with a small pick and blow any particles away with a puff of air.

  2. Break tips off the representative sorbent tube and insert it into the rubber sleeve on the All-in-One (arrow on the tube points toward the holder). Note: Ensure that the sorbent tube fits snugly in the rubber sleeve prior to sampling to avoid air leakage. Two sleeves, each with a different inner diameter, are supplied with the All-in-One.

  3. Using flexible tubing, connect the calibrator outlet to the inlet of the representative sorbent tube in the holder. See calibration train above right.
    In the next step, do not shut off flow completely with a flow adjust screw or use an oversize screwdriver to adjust flow — valve or thread seat damage may result. Turn the screw on a holder to adjust flow.

  4. Do not adjust the flow on the pump. Adjust the flow rate through the representative sorbent tube using a small flat-head screwdriver to turn the flow adjust screw on the All-in-One (counterclockwise to increase, clockwise to decrease) until the calibrator displays the method-specified flow rate. Take a minimum of three readings and record the average flow rate, as per OSHA/NIOSH instructions.

  5. When calibration is complete, turn off the pump by pressing the on/off button. Disconnect the calibrator and tubing from the representative sorbent tube inlet.

Set-Up/Sample with Single Tube

  • Allow the pump to equilibrate after moving it from one temperature extreme to another.
  • Protect sample pump from the weather when sampling outdoors.
  • Use of any device other than the approved battery pack to power the pump voids the UL Listing for intrinsic safety and any warranty.
  • Charge pump completely before calibration and sampling.
  • Calibrate/verify pump flow rate before and after each sampling operation using the tube holder and pump to be used for sampling.
    Clip tube holder to worker and pump to belt.
  1. Replace the representative sorbent tube used for calibrating flow with a new unexposed sorbent tube for sample collection. Note: Ensure the sorbent tube fits snugly in the rubber sleeve of the All-in-One before sampling to avoid any air leakage. Two sleeves, each with a different inner diameter (ID), are supplied. Clip tube holder to worker and pump to belt.

  2.  Place the tube cover over the tube (Figure 2) and thread it into place on the All-in-One low flow holder.

  3.  Place the sorbent tube where appropriate for sampling. For personal sampling, clip the sample medium to the worker in the breathing zone and the pump to the worker’s belt.

  4. Press the on/off button to clear the elapsed time displayed on the LCD and to start sampling. Record the start time and other pertinent information. The LCD will automatically display elapsed run time. See below.
    Possible Displays During Sampling
    Flow or Battery Fault Shutdown – If the pump is unable to compensate due to excessive back pressure or a low battery condition exists, it will shut down and timing functions will pause. The LCD will display either a battery- shaped icon
    or a flow fault icon ( ) depending on the cause of the shutdown. Upon flow fault, the pump will attempt to restart up to five times. To restart from flow fault, correct the blockage and press the on/off button twice. The elapsed run time display will reset to 0 when the pump is restarted. If the battery icon is displayed, recharge the battery before sampling.
    Displayed Elapsed Run Time – Elapsed run time is displayed continuously on the LCD. For elapsed times after 999 minutes, the display still shows elapsed time but alternates between displaying the first two digits of the elapsed time and the last three digits. For example, a pump that has run for 1,440 minutes would display first “01” and then “440.” The display maximum is 99 999, which is 99,999 minutes.

  5. At the end of the sampling period, press the on/off button and record stop time and other pertinent information.

  6. Cap the sample tube and send it with blanks and pertinent sampling information to a laboratory for analysis.

  7. Verify the flow.
    a. Turn on the pump and reinstate the calibration train and sample media.
    b. Take three readings and record the average value as the post-sample flow rate. Do not adjust the pump flow rate during this step.
    c. Compare the pre and post-sample flow rates. Note in sampling documentation if the values differ by more than ± 5%.

Low Flow Applications (Cont)
Using a CPC and Dual, Tri, or Quad Adjustable Low Flow Tube Holder (Figure 3)

Figure 3. Quad Adjustable Low Flow Tube Holder

Set/Calibrate Flow Rate for Multiple Tubes

  • Requires Constant Pressure Controller (CPC) and Adjustable Low Flow Tube Holder (see Accessories). The low flow tube holder used with CPC allows up to four tube samples to be taken simultaneously, each at different flow rates if desired.
  • Allow the pump to equilibrate after moving it from one temperature extreme to another.
  • Charge pump battery completely before calibration and sampling.
  • Calibrate/verify pump flow rate before and after each sampling operation using the tube holder and pump to be used for sampling.
  • Set the flow rate through the pump first and then calibrate the flow rate through each representative sorbent tube.

Pump Flow Rate
Follow these important steps before proceeding.

  1. Ensure that the battery is fully charged and that the pump has run for 5 minutes before calibrating. Leave the pump running.

  2. Prepare a calibrator (see calibrator instructions). Using flexible tubing, connect the calibrator outlet (suction port) to the pump inlet.

  3. Calculate the sum of all-tube flow rates. The maximum flow rate through any one tube is 500 ml/min.
    a. If the sum is 1000 ml/min, set the pump flow rate to 1.5 L/min. b. If the sum is > 1000 ml/min, multiply that number by 0.15 and total the two numbers. Set the pump flow rate for the resulting new sum. (Example: Sampling with three sorbent tubes, each with a flow rate of 500 ml/min.
    The sum of the tube flow rates is calculated as 3 x 500 = 1500. Determining a 15% higher flow rate is calculated as 1500 x 0.15 = 225. Calculating the final pump flow setting would be 1500 + 225 = 1725 ml/min.) * Backpressure across some sample tubes can be higher than average. In these instances, the maximum flow rate of 500 ml/min per tube may not be achieved.

  4. Using a small screwdriver, turn the flow adjustment screw on top of the pump (Figure 1) (clockwise to increase flow or counterclockwise to decrease flow) until the calibrator indicates the required flow rate.

  5. Remove tubing from the pump inlet and calibrator outlet.

Flow Rate Through Sorbent Tubes

Figure 4. Calibration Train with Quad Adjustable Low Flow Tube Holder

  1. Attach the tubing on the CPC outlet (side of the CPC without a label) to the pump inlet. Attach the Adjustable Low Flow Holder to the CPC inlet (marked “To Sample”). See Figure 4.

  2. Break tips off the representative sorbent tubes and insert the tubes into the rubber sleeves on the holder (arrow on each tube pointing toward the pump). Place unopened tubes in any unused ports to “seal” them.

  3. Label all representative tubes and ports.

  4. Using flexible tubing, connect the exposed end of the first active tube to be calibrated to the calibrator outlet (suction port).

  5. Using a small screwdriver, loosen and turn the brass flow adjust screw (see above right) directly beneath the port holding the first active tube to be calibrated (clockwise to increase, counterclockwise to decrease) until the calibrator indicates the method-specified flow rate. Do not adjust the flow rate on the pump.
    Note: For tri and quad models, first rotate each anti-tamper cover to expose the flow adjust screws, then adjust the appropriate one until the calibrator indicates the desired flow (Figures 3 and 5).
    Figure 5. Cut-away of Tri/Quad Adjustable Low Flow Tube Holder

  6. Repeat Steps 4 and 5 for each active representative tube.
    Note: Changing the flow on one tube will not affect the flow rate through the remaining tubes.

  7. Once the flow is calibrated for each active tube, it is recommended practice to re-check the flow rate through representative tubes before removing them. Any adjustment should be minimal.

Set-Up/Sample with Multiple Tubes

Requires Constant Pressure Controller (CPC) and Adjustable Low Flow Tube Holder (see Accessories). The low flow tube holder used with CPC allows up to four tube samples to be taken simultaneously, each at different flow rates if desired.

  • Allow the pump to equilibrate after moving it from one temperature extreme to another.
  • Protect sample pump from the weather when sampling outdoors.
  • Use of any device or battery pack other than P85004 or P84002 to power the pump voids the UL Listing for intrinsic safety.
  • Charge pump completely before calibration and sampling.
  • Calibrate/verify pump flow rate before and after each sampling operation using the tube holder and pump to be used for sampling.
  1. Replace the representative sorbent tubes used for calibration with new unexposed sorbent tubes for sample collection.

  2. Place a protective tube cover over each tube and thread into place on the holder until secure.

  3. Place the adjustable holder with tubes that were appropriate for sampling. For personal sampling, clip the low flow tube holder to the worker in the breathing zone and the pump to the worker’s belt.

  4. Press the on/off button to clear the elapsed time displayed on the LCD and to start sampling. Record the start time and other pertinent information. Sample for the time specified in the method used. The LCD will automatically display elapsed run time.
    Possible Displays During Sampling
    Flow or Battery Fault Shutdown – If the pump is unable to compensate due to excessive back pressure or a low battery condition exists, it will shut down and timing functions will pause. The LCD will display either a battery- shaped icon or a flow fault icon ( ) depending on the cause of the shutdown. Upon flow fault, the pump will attempt to restart up to five times. To restart from flow fault, correct the blockage and press the on/off button twice. The elapsed run time display will reset to 0 when the pump is restarted. If the battery icon is displayed, recharge the battery before sampling.
    Displayed Elapsed Run Time – Elapsed run time is displayed continuously on the LCD. For elapsed times after 999 minutes, the display still shows elapsed time but alternates between displaying the first two digits of the elapsed time and the last three digits. For example, a pump that has run for 1,440 minutes would display first “01” and then “440.” The display maximum is 99 999, which is 99,999 minutes.

  5. At the end of the sampling period, press the on/off button and record stop time.

  6. Cap the sample tubes and send with blanks and pertinent sampling information to a laboratory for analysis.

  7. Verify the flow.
    a. Turn on the pump and reinstate the calibration train and sample media.
    b. Take three readings and record the average value as the post-sample flow rate. Do not adjust the pump flow rate during this step.
    c. Compare the pre and post-sample flow rates. Note in sampling documentation if the values differ by more than ± 5%.

Maintenance

Replace the Pump Inlet Filter
The AirChek 52 Sampler is fitted with a filter/trap just below the pump inlet. This prevents particles from being drawn into the pump mechanism. The filter should be replaced periodically as follows:

  1. Grip the inlet fitting (see right) and turn it counterclockwise to unthread it.

  2. Holding the pump upside down, use a small flat-head screwdriver or tweezers to remove the filter.

  3. Place a new filter in the inlet. Ensure that the rubber gasket inside the inlet fitting is still in place.

  4. Thread the inlet fitting back into place.

Replace the Battery Pack
Use of a repaired or rebuilt battery pack voids any warranty and the UL Listing for intrinsic safety.

  1. Loosen two security screws on the bottom of the pump case and allow them to fall to the out position.

  2. Pinch the battery pack release arms and pull the bottom half of the pump case away from the top half.

  3. Align arms on the top half of the case with slots on either side of the new battery pack housing. Ensure that the back of the battery pack with the charging jack is on the same side as the belt clip.

  4. Slide battery pack up until arms click into place.

  5. Push each security screw into the battery pack housing and tighten.

Once installed, completely charge the new battery pack before operating the pump (see Charge the Battery Pack). It may be necessary to charge it a few times before achieving maximum battery capacity.
For information on SKC pump battery packs, visit www.skcinc.com/knowledgecenter.
Pump Service
Pumps under warranty should be sent to SKC Inc. for service. See Limited Warranty and Return Policy.

Accessories/Replacement Parts

Accessories
Medium Flow chek-mate Calibrator with CalChek, 0.50 to 5 L/min, includes a 9-volt alkaline battery

with NIST standard traceable calibration certificate
with UK standard traceable calibration certificate
with ISO standard traceable calibration certificate| | 375-0550N
375-0550
375-0550S
---|---|---

Low Flow Sampling
Low Flow (5 to 500 ml/min) Kit includes an All-in-One low flow holder and Type-A tube cover
Adjustable Low Flow Tube Holders
Use with CPC listed below. Require separate tube covers listed below

Dual Holder
Tri Holder
Quad Holder| | 224-26-02
224-26-03
224-26-04
224-26-CPC
---|---|---

Constant Pressure Controller (CPC)
Protective Tube Covers for Sorbent Tubes
Use with adjustable low-flow tube holders listed above.

Type A – 6-mm OD x 70-mm L, included in Low Flow Kit above
Type B – 8-mm OD x 110-mm L
Type C – 10-mm OD x 150-mm L
Type D – 10-mm OD x 220-mm L| | 224-29A
224-29B
224-29C
224-29D
---|---|---

Battery Maintenance
PowerFlex Charging System for SKC Personal Pumps

5-Station, 100-240 V
Single, 100-240 V
PowerFlex Pump Cable,
for AirChek 52| 223-1000
223-2000
223-1004
223-300
---|---

Battery Eliminator, * connects the pump to line power for extended sampling, 115 V
Pump Operation

Screwdriver Set included with pump
Protective Nylon Pouch with belt and shoulder strap
Black
High-profile Red
Noise-reducing Black| ****| 224-11
224-88
224-96A
224-96C
---|---|---

  • Not UL Listed for intrinsic safety

Replacement Parts

Use only SKC-approved parts to ensure reliable performance. Failure to do so voids the UL Listing for intrinsic safety and any warranty.
Use of a repaired or rebuilt battery pack voids any warranty and the UL Listing for intrinsic safety.
Failure to follow warnings cautions, and notes voids any warranty.

Internal Stack

No.
1A/1C
2
2A
3
4
5A/5B
6
7
8
9| Description
Stack screws, pk/12
Inlet/Hose Connect includes filter and gasket
Replacement Inlet Filter, pk/10
Pulsation dampener
Stack plate
Valve plates (top and bottom)
Diaphragm/Yoke assembly
Pump body
Motor/Eccentric
Pump base plate
Full Stack without Motor| Cat. No.
P51891
P20106
P40370
P2010802
N/A
P213201
P2129B
P22417G
P518803
P20102
P20102A|
---|---|---|---

Parts not indicated in the illustration:

N/A
N/A| PC Board
Screw and O-ring Set (complete for one pump)| P79592
P22402
---|---|---

External

No.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15| Description
Battery Pack, NiMH
Battery Pack Release Arms
Battery Pack Security Screws
Compensation Pot A
Compensation Pot B
Power Plug for Battery Pack, pk/5
Charging Jack (on the rear of a battery pack)
Impinger Holder Mounting Screws
Inlet (includes filter and gasket)
Control Cover
Flow Adjustment Screw
Power Switch (on/off)
Belt Clip (not shown)
LCD
External Case| Cat. No.
P78011AMH
N/A
N/A
N/A
N/A
P20107
N/A
N/A
P20106
P20105
N/A
N/A
P20104
P72391
P2010003|
---|---|---|---

SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to skcinc.com/warranty.

Appendix

Performance Profile

Flow Range| 1000 to 3000 ml/min 5 to 500 ml/min requires low flow accessories. See Accessories.
---|---
Flow Control| Holds constant flow to ± 5% of set-point after calibration
Compensation Range
(back pressure capability)| 1000 ml/min up to 25 inches water back pressure 2000 ml/min up to 25 inches water back pressure 2500 ml/min up to 20 inches water back pressure 3000 ml/min up to 15 inches water back pressure
Typical Back Pressure of Sampling Media (inches water)| Compare the information in this table to the pump compensation range to determine appropriate applications.| Flow Rate (L/min)| 1.0| 1.5| 2.0| 2.5
---|---|---|---|---
Filter/Pore Size ( μ m)| | | |
25-mm MCE/0.8| 6| 9| 12| 15
25-mm MCE/0.45| 14| 22| 28| 35
37-mm MCE/0.8| 2| 3| 4| 5
37-mm PVC/5.0| 1| 1| 2| 2
Run Time| Nickel-metal hydride (NiMH) battery: 12 hrs minimum at 2000 ml/min and up to 25 inches water back pressure; depends on media used. See Table 1.
Battery Eliminator: The pump  provides extended runs.
Power Supply| Rechargeable 4.8-V NiMH battery pack, 3.3-Ah capacity
A Battery Eliminator is available (see Accessories); use voids the UL Listing for intrinsic safety.
Battery Charging Time ( varies with battery capacity and level of discharge)| 6 to 8.5 hrs with PowerFlex charger
Temperature| Operating: 32 to 104 F (0 to 40 C)
Storage: -4 to 113 F (-20 to 45 C)
Charging: 50 to 113 F (10 to 45 C)
Humidity| 0 to 95% non-condensing
Flow Fault| If the pump is unable to compensate for longer than 15 seconds due to excessive back pressure, the pump enters flow fault. During flow fault, the pump stops, the fault icon appears on the display, and elapsed time remains on the display. Auto-restart is attempted up to 5 times. Note: Elapsed time information will be lost on restart.
Low Battery Fault| LCD displays a low battery icon and the pump shuts down. LCD remains on.
Time Display| LCD shows elapsed pump run time in minutes up to 99,999.
Note: Older pump models display elapsed time up to three digits (999) and then roll back to zero.
LCD Indicator| LCD always remains on. Displayed elapsed run time information resets when the pump is started.
Noise Level| 62.5 dBA – pump without case 55 dBA – pump housed in the noise-reducing case (optional accessory Cat. No. 224-96C, see Accessories)* Measured 3 ft (1 m) from pump operating at 2 L/min with a 37-mm, 0.8-μm MCE filter cassette
RFI/EMI Shielding| 27 to 1000 MHz
Intrinsic Safety| UL Listed for Intrinsic Safety: Class I, Division 1 and 2, Groups A, B, C, D; Class II, Division 1 and 2, Groups E, F, G; and, Class III, Temperature Code T3C. ATEX-approved models are available. Contact SKC.
Dimensions| 5 x 3 x 1.75 in (12.7 x 7.6 x 4.4 cm)
Weight| 20 oz (567 grams)
Multiple-tube Sampling| Requires additional accessories to sample with up to four sorbent tubes simultaneously at flows from 5 to 500 ml/min, dependent on backpressure. See Accessories.
Tubing| Requires 1/4-inch ID tubing
Certifications| • Intrinsic safety: UL Class I, Groups A, B, C, and D; Class II, Groups E, F, and G; Class III hazardous locations when used with SKC battery pack model P78011AMH (SKC Cat. No. P78011AMH)
• CE marked
• ATEX-approved models available

Appendix (Cont)
Typical Run Times
Table 1 contains the typical run times achieved when using a fully charged NiMH battery pack. Data is sorted by type of sample media. All run times are listed in hours. Results are obtained using a new pump and new fully charged battery. Pump performance may vary. Increases in back pressure during sampling due to the buildup of samples on the filter can decrease battery life.

Table 1. Pump Run Time in Hours with NiMH Battery
Mixed Cellulose (MCE) filter, 0.8-m pore size

**Flow Rate (L/min)** Filter Diameter
37 mm 25 mm
2.0 40
2.5 38
3.0 34

Polyvinyl Chloride (PVC) filter, 5.0-m pore size

**Flow Rate (L/min)** Filter Diameter
37 mm 25 mm
2.0 56
2.5 48
3.0 42

skcinc.com
Form 37714 Rev 211108

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>