CONTEC CMS50F Wrist-worn Pulse Oximeter User Manual
- June 3, 2024
- Contec
Table of Contents
CMSSOF User Manual
Pulse Oximeter
Instructions to User
Dear users, thank you very much for purchasing the Pulse Oximeter (hereinafter referred to as device). It is a medical device, which can be used repeatedly. The Manual describes, in accordance with the Pulse Oximeter’s features and requirements, main structure, functions, specifications, correct methods for transportation, installation, usage, operation, repair, maintenance, and storage, etc. as well as the safety procedures to protect both the user and equipment. Refer to the respective chapters for details. Please read the User Manual carefully before using this product. The User Manual which describes the operating procedures should be followed strictly. Failure to follow the User Manual may cause measuring abnormality, equipment damage, and human injury. The manufacturer is NOT responsible for the safety, reliability, and performance issues and any monitoring abnormality, human injury, and equipment damage due to users’ negligence of the operating instructions. The manufacturer’s warranty service does not cover such faults. Owing to the forthcoming renovation, the specific products you received may not be totally in accordance with the description of this User Manual. We would sincerely regret for that. Our company has the final interpretation to this manual. The content of this manual is subject to change without prior notice. Warnings:
Remind that it may cause serious consequences to a tester, user, or
environment
Explosive hazard—DO NOT use the device in an environment with inflammable gas
such as an anesthetic.
DO NOT use the device while examining by MIII and CT, as the induced current
may cause a burn.
Do not take the information displayed on the device as the sole basis for
clinical diagnosis. The device is only used as an auxiliary means in
diagnosis. And it must be used in conjunction with a doctor’s advice, clinical
manifestations, and symptoms.
The maintenance to the device or replacement of the battery (non-detachable
lithium battery) can only be performed by qualified service personnel
specified by the manufacturer, dangers (such as over-temperature, fire, or
explosion) may occur when replacing the battery by the personnel not fully
trained. Users are not permitted to maintain or refit the device by themselves
or replacement of the battery.
Uncomfortable or painful feelings may appear if using the device ceaselessly,
especially for microcirculation disturbance users. It is not recommended that
the sensor is used on the same finger for more than 2 hours.
For some special users who need a more careful inspection on the test site,
please don’t place the device on the edema or tender tissue.
Please do not stare at the red and infrared light emitter (the infrared light
is invisible) after turning on the device, including the maintenance staff, as
it may be harmful to the eyes.
The device contains silicone, PVC, TPU, TPE, and ABS materials, whose
biocompatibility has been tested in accordance with the requirements in ISO
10993-1, and it has passed the recommended biocompatibility test. The person
who is allergic to silicone, PVC, TPU, TPE or ABS can not use this device.
The disposal of scrap devices, their accessories, and packaging should follow
the local laws and regulations, to avoid polluting the local environment. And
the packaging materials must be placed in the region where the children are
out of reaching.
The device can not be used with equipment not specified in the Manual. Only
the accessories appointed or recommended by the manufacturer can be used,
otherwise, it may cause injury to the tester and operator or damage to the
device
The disposal of scrap devices. its accessories and packaging should follow
local laws and regulations, to avoid polluting the local environment. And the
packaging materials must be placed in the region where the children are out of
reaching.
The device can not be used with equipment not specified in the Manual. Only
the accessories appointed or recommended by the manufacturer can be used,
otherwise it any cause injury to the tester and operator or damage to the
device
The SpO, probe accompanied is only suitable for use with the device. The
device can only use the Sprat probe described in the Manual, so the operator
has the responsibility to check the compatibility between the device and the
SO: probe before use. incompatible accessories may cause device performance
degradation. device damage or user injury.
Do not reprocess the accompanying SpO, probe.
Check the device before use to make sure that there is no visible damage that
may affect the user’s safety and device performance.
When there is obvious damage. please replace the damaged parts before use. `
When the message “Senor or “Sensor Fault- appears on the screen, it indicates
that the $eth probe is disconnected or a line fault occurs. Check the
connection of the SpO, and probe and whether there is damage to the probe. if
necessary, please replace the probe to avoid risks.
The probe fault will not result in a safety hard.
Functional testers can not be used to assess the accuracy of the Sp0; probe
and Pulse Oximeter. Ir Sonic functional testers or patient simulators can be
used to verify whether the device works normally. for example. INDE.X•21.FE
Simulator (software version: 1.00). please refer to the Manual for the
detailed operation steps. dr Some functional testers or patient simulators can
measure the accuracy of the device-copied calibration curve. but they can not
be used to evaluate the device’s accuracy.
When using the device. please keep it away from equipment that can generate a
strong electric field or strong magnetic field. Using the device in an
inappropriate environment may cause interference to the surrounding radio
equipment or affect its working.
When storing the device, keep it away from children. pets and insects to avoid
affecting its performance.
Do not place the device in places exposed to direct sunlight, or high
temperatures. humidity, dust, cotton wool, or easy-to-splash water, to avoid
affecting its performance. 4″ The measured accuracy will be affected by the
interference of electrosurgical equipment.
When several products are used on the same people simultaneously, danger may
occur which is arisen from the overlap of leakage current.
CO poisoning will appear excessive estimation so it is not recommended to use
the device 6″ This device is not intended for treatment.
The intended operator of the device may be a user. e Avoid maintaining the
device during use 6″ Users should read the product manual carefully before use
and operate according to the requirements
0verview
Insert the finger when measuring, and the device will directly display the
Sp02 value measured, it has higher accuracy and repeatability.
1.1 Features
A. Easy to use.
B. Small in volume, light in weight, and convenient to carry.
C. Low power consumption.
1.2 Intended purpose
The Pulse Oximeter can be used in measuring the pulse oxygen saturation and
pulse rate through the finger. The product is suitable for being used in
families, hospitals, oxygen bars, community healthcare, and physical care in
sports (It can be used before or after doing sports, and it is not recommended
to use the device during the process of having sports) and etc.
1.3 Environment Requirements
Storage Environment
a) Temperature : -40°C—+60°C
b) Relative humidity : <95%
c) Atmospheric pressure : 500hPa∼1060hPa
Operating Environment
a) Temperature: +10 °C— + 40 °C
b) Relative Humidity : <75 %
c) Atmospheric pressure: 700hPa-1060hPa
1.4 Precautions
1.4.1 Attention
Point out conditions or practices that may cause damage to the device or other
properties.
Before using the device, make sure that it locates in a normal working state
and operating environment.
In order to get a more accurate measurement, it should be used in a quiet and
comfortable environment.
When the device is carried from a cold or hot environment to a warm or humid
environment, please do not use it immediately, waiting four hours at least is
recommended.
If the device is splashed or coagulated by water, please stop operating
DO NOT operate the device with sharp things.
High temperature, high pressure, gas sterilizing, or immersion disinfection
for the device is not permitted. Refer to User Manual in the relative chapter
(6.1) for cleaning and disinfection. Please turn off the device before
cleaning and disinfection.
The device is suitable for children and adults.
The device may not be suitable for all users, if you can’t get a satisfactory
result, please stop using it.
Data averaging and signal processing have a delay in the upgrade of Sp02 data
values. When the data update period is less than 30 seconds, the time for
obtaining dynamic average values will increase, which is arisen from signal
degradation, low perfusion, or other interference, it depends on the PR value.
The device has a 3-year service life, and the date of manufacture sees the
label.
The expected service life of the attached parts or accessories of the
equipment is two year.
if the shelf life is less than the expected service life, the shelf life of
the attached parts or accessories of the equipment is two year.
This device has the function of prompting, users can check on this function
according to chapter 5.3.1 as a reference.
The maximum temperature at the Sp02 probe-tissue interface should be less than
41 °C which is measured by the temperature tester.
During measuring, when abnormal conditions appear on the screen, please pull
out your finger and reinsert it to measure again.
If some unknown error appears during measuring, press “RESET” button to reset
it.
Do not contort or drag the wire of the device.
The plethysmographic waveform is not normalized, as a signal inadequacy
indicator, when it is not smooth and stable, the accuracy of the measured
value may degrade. When it tends to be smooth and stable, the measured value
read is optimal and the waveform at this time is also the most standard.
The device can not be used during charging. g If necessary, please visit our
official website to get the information about Sp02 probe that can be used with
this device.
If the device or component is intended for single use, then the repeated use
of these parts will pose risks to the parameters and technical parameters of
the equipment known to the manufacturer.
If necessary, our company can provide some information (such as circuit
diagrams, component lists, illustrations, etc.), so that the qualified
technical personnel of the user can repair the device components designated by
our company.
The measured results will be influenced by the external coloring agent (such
as nail polish, coloring agent, or color skin care products, etc.), so don’t
use them on the test site.
As to the fingers which are too cold or too thin or whose fingernail is too
long, it may affect the measured results, so please insert the thicker finger
such as the thumb or middle finger deeply enough into the probe when
measuring.
The finger should be placed correctly (see Attached figure 4), as improper
installation or improper contact position for the sensor will influence the
measurement.
The light between the photoelectric receiving tube and the light-emitting tube
of the device must pass through the subject’s arteriole. Make sure the optical
path is free from any optical obstacles like rubberized fabric, to avoid
inaccurate results.
Excessive ambient light may affect the measured results, such as surgical
light (especially xenon light sources), bilirubin lamp, fluorescent lamp,
infrared heaters and direct sunlight, etc. In order to prevent interference
from ambient light, make sure to place the sensor properly and cover the
sensor with opaque material.
Frequent movement (active or passive) of the subject or severe activity can
affect the measurement accuracy.
The Sp02 probe should not be placed on a limb with the blood pressure cuff,
arterial ductus or intraluminal tube.
The measured value may be inaccurate during defibrillation and in a short
period after defibrillation, as it has not defibrillation function. The device
has been calibrated before leaving factory.
The device is calibrated to display functional oxygen saturation.
The equipment connected with the Oximeter interface should comply with the
requirements of IEC 60601-1.
Please select the medical power adapter to charge it, when connecting the
special adapter with the socket, make sure there is no shelter near the socket
and it is easy to plug and unplug, otherwise, the power will not be cut off in
time when necessary, causing damage.
1.4.2 Clinical restriction
A . As the measure is taken on the basis of arteriole pulse, the substantial
pulsating blood flow of the subject is required. For a subject with a weak
pulse due to shock, low ambient/body temperature, major bleeding, or use of
the vascular contracting drug, the Sp02 waveform (PLETH) will decrease. In
this case, the measurement will be more sensitive to interference.
B. The measurement will be influenced by intravascular staining agents (such
as indocyanine green or methylene blue), skin pigmentation.
C. The measured value may be normal seemingly for the tester who has anemia or
dysfunctional hemoglobin(such as carboxyhemoglobin (COHb), methemoglobin
(MetHb), and sulfhaemoglobin (Sul-lb)), but the tester may appear hypoxia, it
is recommended to perform further assessment according to the clinical
situations and symptoms.
D. Pulse oxygcn only has a reference meaning for anemia and toxic hypoxia, as
some severe anemia users still show better pulse oxygen measured valued.
E. Contraindication:
a. The person who is allergic to silicone, PVC, TPU TPE or ABS can not use
this device. b. The damaged skin tissue can’t be measured. c. During
cardiopulmonary resuscitation. d. When the user is hypovolemic. e. For
assessing the adequacy of ventilatory support. f. For detecting worsening lung
function in users on a high concentration of oxygen.
1.5 Clinical indications
The Pulse Oximeter can be used in measuring the pulse oxygen saturation and
pulse rate through the finger.
Principle
Principle of the Oximeter is as follows: An experience formula of data process is established taking use of Lambert-Beer Law according to Spectrum Absorption Characteristics of Reductive Hemoglobin (Hb) and Oxyhemoglobin (Hb02) in glow & near-infrared zones. The operation principle of the device is: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning & Recording Technology so that two beams of different wavelengths of light can be focused onto a human nail tip through a perspective clamp fmger-type sensor. Then measured signal can be obtained by a photosensitive element, information acquired through which will be shown on screen throw,* treatment in electronic circuits and microprocessor.
Functions
A. Sp02 value display
B. PR value and bar graph display
C. Pulse waveform display
D. Low-battery indication: low-battery indication appears when the battery
voltage is too low to work
E. Automatic standby function
F. display mode can be changed
G. Adjustable screen brightness
H. PR sound indication
I. Voice prompt for over-limit, sensor off/finger-out and low battery.
J. Memory function
K. The data can be uploaded to the terminal equipment by wired mode.
L. The data can be uploaded to the terminal equipment (Bluetooth wired
equipment) by wireless mode.
M. With clock function N. Charging function
Installation
4.1 Appearance
USB
interface: connect with USB cable or Sp02 probe Button: power on, pause sound
prompt, display clock, enter menu, menu operation. RESET hole: reset button
inside of it, to reset the device.
4.2 Interface introduction
4.3 Sp02 probe installation
Open the USB plug of the device, then insert the Sp02 probe interface into the
USB port of the device.
4.4 Connection of USB cable
Open the USB plug of the device, insert the micro end of the USB cable into
the Sp02 probe interface, and the other end into the computer or power
adapter.
4.5 Structure and accessories
A. Structure: main unit, SpO, probe, USB cable, power adapter (optional), and
Bluetooth adapter (optional).
B. Accessories: one Sp02 probe, one USB cable, one power adapter (optional),
one User Manual, and Bluetooth adapter (optional). Please check the device and
accessories according to the list to avoid the device can not work normally.
C.Software description Software name: CMS5OF embedded software Software
specification: no Release version: 2.0 Naming rule for version: V <Major
enhancive software upgrade>.
0perating
5.1 Measurement
A. Insert the finger into the probe as shown in Figure 4.
(The appearance of the actual probe may be different from the one shown as
Figure 4, please refer to the actual probe.) B.Press the button to turn on the
device, it displays the measurement interface. C.Wait a few seconds, and the
device directly shows measurement results on the screen.
Note: when inserting the finger, the light emitting from the sensor must
be directly irradiated to the side of the fingernail. Note: during measuring,
do not shake the finger and keep quiet, not move.
5.2Measurement interface
A. In the sound prompt “ON” state, when the sound prompt occurs, short press
the button to pause the sound prompt, and it will resume automatically after
about 60s. The sound prompt includes an over-limit prompt, low-battery prompt,
probe off or finger out prompt. B.If you want to turn off the sound prompt
permanently, please set it in the menu. If there is no sound prompt, short
pressing the button will directly enter the clock interface, and press again
to exit.
5.3 Menu operation
Under the measurement interface, long press the button to enter the main menu
interface as shown in Figure 5, sound, record, clock, system and Bluetooth,
etc. can be set, methods are as followings:
5.3.1 Sound prompt setting
Under the main menu, short press the button to select “Sound”, then long press
the button to enter its setting interface shown in Figure 6. Short press the
button to select the option to be adjusted, then long press the button to
change the value. “Direction”: direction, “up”: increase the value, “down”:
decrease the value “Sp02 HI”: upper limit prompt for Sp02 over-limit “Sp02
LO”: lower limit prompt for Sp02 over-limit “PR HI”: upper limit prompt for PR
over-limit “PR LO”: lower limit prompt for PR over-limit “Prompt Sound”:
prompt for over-limit, “off’: close, “on”: open. “Pulse Sound”: PR sound,
“off”: close, “on”: open. The lower limit can not exceed the upper limit, and
the upper limit can not be lower than the lower limit when adjusting the
values. Sp02 range: 0 % — 100 %, PR range: 0 — 254 bpm The values displayed in
Figure 6 arc the initial values of the ovcr-limit prompt. After setting, short
press the button to select “Exit”, then long press the button to exit sound
setting interface, and return to Main Menu interface.
5.3.2 Data storage
Under the main menu, short press the button to select “Record”, then long
press the button to enter the Record Menu interface as shown in Figure 7. It
indicates that the device is storing when the red dot “R.” in the measurement
interface flickers.
Short press the button to select the option to be adjusted, then long press
the button to change the value. “Mode”: record mode selection, including
“Auto” and “Manual” modes. Under “Manual” mode, select to turn on / off memory
by “Record”. Auto record: start recording after stable data appear, pull out
the finger to finish recording a group of data (99 groups of data at most),
the total duration does not exceed 72 hours. Manual record: store up to
24-hour data. When the memory is full, it will display “Memory is full!”, then
it will enter standby mode after several seconds. When exiting the standby
mode, it will display “Memory is HP” to give the user a prompt, press the
button again to enter the measurement interface.
Note: under manual mode, when “Record” is “ON”, the device will prompt to
clear the data stored last time. It will display “Recording” when there is no
operation under the record state for the 30s, then it will enter energy-saving
mode after several seconds, long press the button to exit this mode; short
press the button, it will display “Recording”.
Note: under data recording state, after the display screen turns off
automatically, in order to save power, pulse sound indication will turn off
automatically.
“Seg”: data segment. After setting, short press the button to select “Exit”,
long press the button to exit the record menu, and return to the main menu.
“Delete All”: delete all records (auto record mode is shown as Figure 7).
Note: Please upload data in time after recording, otherwise the data may
be covered when the storage space is full.
Note: the historical data will be deleted once switching the record mode.
Under the recording state, the record mode can not be switched; under manual
mode, the “Record” should be turned off before switching mode.
5.3.3 Cock setting
Under main menu, short-press the button to select “Clock”, long press the
button to enter its sub-menu as shown in Figure 8.
Short press the button to select the option to be adjusted, then long press
the button to change the value. The device adopts a 24-hour clock. “set time”:
set the time, “yes”: to set the time, “no”: not set the time “set year”: set
the year “set month”: set the month “set day”: set the day “set hour”: set the
hour “set minute”: set the minute After setting, short press the button to
select “Exit”, then long press the button to exit clock setting interface. If
the user changes the setting, the current set time will be displayed, then
press the button to return to the main menu; if the time is not set, it
directly returns to the main menu. 5.3.4 System setting and other options
introduction Under main menu, short press the button to select “System”, then
long press the button to enter the interface as shown in Figure 9. Short press
the button to select the option to be adjusted, then long press the button to
change the value. “Hard.Ver.”: hardware version “Soft.Ver.”: Software version
“ID”: user name “Demo”: set the Demo mode, “on”: turn on the Demo mode, “off”:
turn off the Demo mode. “Sound Volume”: set the sound volume, adjustable
range: 1 — 3 “Brightness”: set the screen brightness, adjustable range: 1 4
After setting, short press the button to select “Exit”, then long press the
button to exit the system menu and return to the main menu.
5.3.5 Bluetooth setting Under main menu, short-press the button to select
“Bluetooth”, then long-press the button to enter its selection interface as
shown in Figure 10 and Figure 11. When the Bluetooth is “ON’, if there is no
data to be transmitted, the Bluetooth will turn off automatically after a few
seconds. Note: Under the data transmitting state by Bluetooth, the
Bluetooth can not be turned off.
5.3.6 Shutdown
Under the main menu interface, short press the button to select “Power of 1 ’,
then long press the button to turn off the device.
5.3.7 E\il main menu
Under main menu, short press the button to select “Exit”, then long press the
button to exit the main menu and return to the measurement interface.
5.4 Data upload
5.4.1 Wired transmission
Connect the device to the computer by the USB cable, and upload the data after
connecting with the PC software properly, refer to “Software operating
instructions” for details.
Note: The PC software can be downloaded from our official website.
5.4.2 Bluetooth wired transmission
Turn for d ou the device Bluetooth and the PC software to upload data, refer
to “Software operating instruction’ details.
5.5 Charging
A power adapter can be selected to charge the device. It indicates that the
device is charging when the indicator is orange, and the charging is finished
when the indicator turns to green. |
5.6 Reset
Use a pointed and hard object (for example, a paper clip) to press the reset
button inside of the RESET hole, to reset the device.
6.1 Cleaning and Disinfecting
The device must be turned off before cleaning, and it should not be immersed
in liquid. Please take out the internal battery before cleaning, do not
immerse it in liquid.
Use 75% alcohol to wipe the device enclosure, and use liquid soap or
isopropanol to wipe the watchband for disinfection, nature dry or clean it
with a clean and soft cloth. Do not spray any liquid on the device directly,
and avoid liquid penetrating into the device.
6.2 IVIainlcnancc
A.Check the main unit and all accessories periodically to make sure that there
is no visible damage that may affect user’s safety and monitoring performance.
It is recommended that the device should be inspected weekly at least. When
there is obvious damage, stop using it.
B. Please clean and disinfect the device before/after using it according to
the User Manual (6.1).
C.Plcasc charge the battery in lime when a low battery appears.
D. Recharge the battery soon after over-discharge. The device should be
recharged every three months when it is not used for some time. It can extend
the battery life following this guidance.
E.The device does need not to be calibrated during maintenance.
6.3 Transportation and Storage
A. The packed device can be transported by ordinary conveyance or according to
io the transport contract.
During transportation, avoid strong shock, vibration, and splashing with rain
or snow, and it can not be material. transported mixed with toxic, harmful,
B. The packed device should be stored in room with no corrosive gases and good
ventilation. Temperature: -40°C~60°C; Humidity: <95
Turboshooting
Trouble | Possible Reason | Solution |
---|---|---|
The values can not be displayed normally or stably | 1. The finger is not |
properly positioned.
2. The finger is shaking or the user is moving.
3. The device is not used in the environment required by the manual.
4. The device works abnormally.| 1. Please insert the finger properly and
measure again.
2. Lcl the user keep calm
3. Please use the device in normal orminsnmaL
4. Please contact the after-sales.
The device can not be turned on| 1. Low battery or the battery is drained
away.
2. The device works abnormally.| 1-Please charge the battery.
2-Please, contact the after-sales.
The display disappears suddenly.| I . The device enters into energy-saving
mode.
2. Low battery.
3. The device works abnormally.| 1, Please contact the after-sales.
2. Please charge the battery
3. Please contact die after-sales.
The device can not be used for full-time after charge.| 1 The battery is not
charged fully.
2. The device works abnormally.| 1. Please charge the battery.
2. Please contact the after-sales.
The battery can not be fully charged even after 10 hours of charging
lime.| The battery works abnormally.| Please contact the after-sales.
The data can not be stored.| 1. The device is not operated according to lo
the manual.
2.’The device works abnormally.| 1 . Please operate the device according to
the manual.
2. Please contact the after-sales.
Key of Symbools
Symbols | Meaning | Symbols | Meaning |
---|---|---|---|
Caution, consult accompanying documents | bpm | Pulse rate (bpm) | |
Pulse oxygen saturation (%) | |||
Manufacturer | Fully charged | ||
SN | Serial Mimi | Use-by date | |
**** | Recycling garbage WEEE (2012/19/EU) | USB | |
--- | --- | --- | --- |
+ | Battery anode | Battery cathode | |
IP22 | It means this pulse altimeter is protected | ||
against harmful effects of dripping water when tilled at 15° . | RST | RESET |
hole
| Temperature limitation| | Humidity limitation
| Atmospheric pressure limitation| | This way up
fraga, a tale with we| | keep any from rear
| Low bona)| | Close the sound prompt
| Pause the sound prompt| | Open the sound prompt
Sensor Off| Menu/ Power button| | Close the PR sound
Recyclable| | Open the PR sound
Bluetooth icon (Bluetooth wired device)| Manufacture Dale|
The probe is disconnected.| Finger Out| The finger is not inserted.
R*| Recording| Sensor
Fault| Probe failure
P/N| Material code| LOT| Batch No.
****| 1. The finger clip falls off 4 Do finch mutat II
2. Probe error
3. Signal inadequacy indicator| | Alarm inhibit
Note: Your device may not contain all the following symbols.
sPo: Fee note I I|
---|---
Display range| O%-99%
Measured range| 0%-4.0(P/o
Accuracy see note 2]| 70%-100%: 2%; 0%-69A: unspecified.
Resolution| 1%
PR
Display range| 30bpm 250bpra
Measured range| 30bpm-250bpm
Accuracy see note 31| 12 bpm during the pulse rate range of 30 bpm 99 bpm and
*2% during the pulse rate range of 100 bpm – 250 bpm.
Resolution| I by
Accuracy under low perfusion [sec note 41| Low perfusion 0.4%: 40: <re.,
PR: ±2 bpm during the puke rate range of 30 bpm – 99 bpm and 2%a during the
pulse rate range of 100 bpm 250 bpm.
Light interference| under normal and ambient light conditions, the Spia?
deviation -I%
Pulse Intensity| Continuous bar-graph display, the higher display indicates a
stronger pulse.
Upper and lower limit of measured values
SpO :| 0% 100%
PR| 0 bpm – 254 bpm
Optical sensor !see note 51
Red light| Wavelength: about 660 tun, optical output power :< 6.65 mW
Infrared light| Wavelength: about 905 nm. optical output power < 6.75 mW
Memory| Up to 99 groups of data under the auto mode, total duration does not
exceed 72 hours. Up to 24-hour data under manual mode.
Safety class| Internally powered equipment, type BF applied part
International Protection| I P22
Working voltage| DC 3.6 V -4.2 V
Working current| < 100 mA
Power supply
A rechargeable lithium battery (3.7V) (The red wire on the battery denotes
anode, the black wire on the bancry denotes
| cathode.)
---|---
Battery life| Charge and discharge:no less than 500 times
Adapter specification| Output voltage: DC 5V Output current: 1000 mA
Dimensions and Weight
Dimension| 61 mm(L) , :56 mat(W) x 24 mm(H)
Weight| About 60 g (including a lithium battery)
Note 1: the claims of SpC>2 accuracies shall be supported by clinical
study measurements taken over the full range. By artificial inducing, get the
stable oxygen level to the range of 70 % to 100 % Sp0:, compare the SpO:
values collected by the secondary standard pulse oximeter equipment and the
tested equipment al the win time, to form paired data, which ore used for (he
accuracy analysis.(li is applicable for the probes equipped. There are 12
healthy volunteers (male: 6. female: 6; age: 18-50; skin color: black: 2,
light: 8, while: 2) data in the clinical report. Note 2: because pulse
oximeter equipment measurements are statistically distributed, only about hvo-
thirds of pulse oximeter equipment measurements can be expected to fall within
±Amis of the value measured by a CO-OXIMETER.
Note 3: Patient simulator has been used to verify the pulse rate
accuracy, it is stated as the root-mean-square difference between the PR
measurement value and the value set by simulator. Note 4: percentage
modulation of infrared signal as the indication of pulsating signal strength,
patieut simulator has been used to verify ils accuracy under conditions of low
perfusion. SpO? and PR values are different due to low signal conditions,
compare Ihcm with the known SpO: and PR values of input signal. Note 5:
optical sensors as the light-emitting components, will affect other medical
devices applied the wavelength range. The information may be useful for the
clinicians who carry out the optical treatment. For example, photodynamic
therapy operated by clinician.
Specification| |
---|---|---
State| Prompt condition delay| Prompt signal generation delay
Low voltage prompt| Is| 20ms
SpO: prompt| 330ms| 20ms
Pulse rate prompt| 330ms| 20ms
Probe error prompt| 16ms| _ 20ms
Tabic 1:
Guidance and manufacturer s declaration -electromagnetic emission
The Pulse Oximeter is intended for use in the electromagnetic environment
specified below. The purchaser or the user of the device should assure that it
is used in such an environment.
Emission test| Compliance
RF emissions C1SPR 11| Group
Table 2:
Guidance and manufacturer’s declaration-electromagnetic immunity|
---|---
The Pulse Oximeter is intended for use in the electromagnetic environment
specified below. The purchaser or The user of the Pulse Oximclcr should assure
that it is used in such an environment.|
Immunity test| IEC60601 test level| Compliance level
Electrostatic discharge (ESD)
IEC 61000-4-2| ±8kV contact
土 15 kV air| ±8kV contact
土 15kV air
Power frequency (50 / 60Hz)
magnetic field
IEC 61000-4-8| 30 A/m| 30A/m
Table 3:
Guidance and manufacturer’s declaration — electromagnetic immunity
The Pulse Oximeter is intended for use in the electromagnetic environment
specified below. The customer the user of the Pulse Oximeter should assure
that it is used in such an environment.
Immunity test| IEC 60601 test level| Compliance level
Radiated RF
I EC61000-4-3
…| 10 V/m 80 MHz’. 2.7 GHz| 10 V/m80 MHz- 2.7 GHz
,
NOTE I Al 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These
guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcasts and TV broadcasts cannot be predicted theoretically with
accuracy. To assess the electromagnetic em environment due to fixed RF
transmitters. all electromagnetic site surveys should be considered. If the
measured field strength in the location in which the Pulse Oximeter is used
exceeds the applicable RF compliance level above, the Pulse Oximeter should be
observed to verify normal operation. If abnormal performance is observed.
additional measures may be necessary, such as reorienting or relocating the
Pulse Oximeter. h Over the frequency range 150 KHz to 80 MHz. field strengths
should be less than 3V/m.
Table 4:
Guidance and manufacturer’s declaration – electromagnetic Immunity
The [Code SO is intended for use in the electromagnetic environment specified
below. The customer or the user of the Pulse Oximeter should assure that it is
used in such an environment
| Test Freqte ncy
(MHz)| Band
a)
(MHz)| Service a)| Modulation b)| Mock,’ alien b) (W)| Distance
(m)| IMMUNI TY
TEST
LEVEL
(Wm)
| 385| 3390| 80
TETRA 400| Pulse modulation
18 Hz| K8| 0,3| 27
Radiated| 450| 430
-470| GMRS 460, FRS 460| FMc)
±5 kHz deviation
I kHz sine| 2| 0.3| 28
RF| 710| 704
-787| LTE Band
13,17
780| Pulse modulation b)
Hz| 2| 0.3| 9
IEC610| 745
00-4-3 (Test| 780
sped flea tions for| 810| | GSM
800/900, TETRA
800.
iDEN 820. CDMA 850, LIE Band 5| Pulse modulation b)
18 Hz| C-,1| 0.3| 28
870
ENCLO SURE PORT
I MMUN ITYto
RF
wireless common actions equipment in)| 930| 800
960 –
1720| | GSM 1800; CDMA
1900;
GSM 1900; DECT;
LTE Band I.
3,4,25;
UMTS| Pulse modulation b)
217 Hz| 2
28| It .3|
1845|
1970| 1700
-1990
2450| 2400
-2570| Bluetooth, WLAN,
802.11
big/n,
RFID 2450,
LTE Band 7| Pulse modulation b)
217 Hz| 2| 3| 28
5240| | WLAN
802.11
atn| Pulse modulation b)| 217 Hz 0.2| 0.3| 9
5500| 5100
-5800
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 at
The I m test distance is pcnnidcd by IEC 61000-4-3.
a)For some services, only the uplink frequencies are included.
b)The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)As an alternative to FM modulation. 50% pulse modulation at 18 Hz may be
used because it does not represent actual modulation. it would be the worst
case.
The MANUFACTURER should consider reducing the minimum RISK MANAGEMENT, and
using higher IMMUNITY TEST reduced SWIIIIIInn separation distance. Minimum
separation distances LEVELS shall be calculated using the following equation:
Where P is the maximum power in W, d is the minimum IMMUNITY TEST LEVEL in
Vim.| separation
separation| LEVELS distance, based on
that are appropriate for the for higher IMMUNITY TEST distance in ra, and E is
the
Wiaming
Don’t near active HFSURGICAL EQUIPMENT and the KF shielded room of an ME
SYSTEM for magnetic resonance imaging, where the intensity of EM DISTURBANCES
is high. Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such use
is necessary, this equipment and the other equipment should be observed to
verify that they are operating normally.
Use of accessories, transducers, and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.”
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the device including cables specified by Che manufacturer.
Otherwise, degradation of the performance of this equipment could result.
Active medical devices are subject to special EMC precautions and they must be
installed and used in accordance with these guidelines.
A Note:
When the device is disturbed, (the data measured may fluctuate, please measure
repeatedly or in another environment to ensure its accuracy. Bluetooth
Specification Working frequency: 2402 MHz 2480 MHz Modulation mode: GFSK
Transmitting power:-6 dBm, +4 dBm Receiving sensitivity: -93 dBm
FCC Caution
§ 15.19 Labeling requirements.
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
§ 15.21 Information to user.
Any Changes or modifications not expressly approved by the party responsible
for compliance could void the user’s authority to operate the equipment.
§ 15.105 Information to the user.
Note: This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment to an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
The device has been evaluated to meet the general RF exposure requirements. The device can be used in portable/mobile exposure conditions without restriction
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>