MTTS PT-CL-MAN-01 Colibri Phototherapy User Manual
- June 14, 2024
- MTTS
Table of Contents
Colibri
phototherapy
User Manual
Infant Phototherapy for Jaundice Treatment
Colibri
111720640961
Effective date: July 2023
Issue Number: 01 Version: 12 EN
Issue Date: October 2023
Company Information
MEDICAL TECHNOLOGY TRANSFER AND SERVICES Co., LTD
House No. 26, Alley 41, An Duong Vuong Street, Tay Ho District, Hanoi City,
Vietnam
SRN: VN-MF-000036669
Tel: +84 24 3766 6521
Fax: +84 24 3718 8050
Email: assistance@mtts-asia.com
www.mtts-asia.com
Logic s.r.l.
Via Antonio Pigafetta 1
34147 Trieste, Italy
SRN: IT-AR-000012301
Introduction
Intended purpose
The Colibri Phototherapy is intended to provide phototherapy to treat infant
jaundice when a phototherapy is required and prescribed by a physician.
Intended users
a) The device is to be operated by qualified personnel only.
b) Knowledge (minimum):
- Understands basic principles of phototherapy
- Understands the specialized care needed for pre-term and full-term neonates
c) Language Understanding: Can read and comprehend user manual
d) Experience (minimum): Qualified medical professional with experience
neonatal care
Intended patient population
- Age: Infants
- Weight: < 10kg
- Nationality: Any
- Patient State: patient is not user
Indications
Infant with high serum bilirubin or a rapidly rising bilirubin level.
Contraindications
Few contraindications to phototherapy are recognized. These include:
- Neonates with congenital erythropoietic porphyria
- Family history of porphyria
- Concurrent treatment with photosensitising drugs
Limitations
- The effective surface area is 50 cm x 30 cm (1500 cm²).
- The maximum height of the light source is 140 cm.
- The minimum height of the light source is 115 cm.
Clinical Benefit
The clinical benefit to the patient is the degradation of total serum
bilirubin for the treatment of infant jaundice.
Side effects
Side effects of phototherapy may include: the interference in the
maternal–infant relationship, imbalance of thermal environment and water loss,
skin lesions, bronze baby syndrome, Disorder of circadian rhythms, patent
ductus arteriosus, ocular effects, allergic diseases such as asthma, rhinitis,
and conjunctivitis.
NOTE: Any serious incident that has occurred in relation to the device
should be re- ported to MTTS and the competent authority of the Member State
in which the user and/or patient is established.
Device Description
Overview
Control Panel
1| On/Off Button| Press this button to turn Colibri Phototherapy display and
functions on and off.
---|---|---
2| Therapy Mode Button| Press this button to switch between standard and
intensive therapy mode.
3| Display| Shows therapy mode and treatment time.
Safety Information
Warnings
- Colibri Phototherapy should be used only by appropriately trained personnel and under the direction of qualified medical personnel familiar with currently known risks and benefits of infant phototherapy equipment use.
- Always cover the patient’s eyes with a shield or patch whenever the patient’s eyes can be exposed to Colibri Photothrapy light. Periodically, you may want to ensure that the patient’s eyes are protected and infection free.
- In the event of a power outage, eye coverings should be left on the patient whenever they are under Colibri Phototherapy to protect the patient’s eyes in case the power returns.
- An infant’s water balance may be influenced by phototherapy. Regular feeding times may need to occur more often in order to prevent dehydration.
- Patient’s bilirubin values should be measured regularly while undergoing phototherapy treatment.
- Regularly monitor the patient’s temperature and fluid status.
- Bilirubin photoisomers may be toxic.
- Patients adjacent to the Colibri Phototherapy may need to be protected with protective shields or eye protection.
- The operator may experience some effects during prolonged exposure to the area irradiated by the Colibri Phototherapy.*
- Turn the unit off when checking the baby’s condition and visualizing skin color; blue light can hinder clinical observations by masking skin colour changes, such as cyanosis.
*Sensitive individuals may experience headache, nausea or mild vertigo if he/she stays too long in the illuminated area. Using Colibri Phototherapy in a well-lighted area or wearing glasses with yellow lenses can alleviate potential effects. Guard Dog Bones glasses (p/n 413BB) are recommended and are available online at www.safetyglasses.com.
- Varying the ambient conditions (ambient temperature and/or different radiation sources) may negatively affect the patient. Refer to your hospital phototherapy policy and procedure regarding appropriate ambient conditions.
- Due to photo effects, drugs and infusion liquids should not be stored in the radiation area.
- Do not use the in the presence of potentially combustible gases including oxygen, nitrous oxide or other anaesthetics.
- Using reflective foils to increase the efficacy of phototherapy may cause hazardous increase in patient’s body temperature.
- When the Colibri Phototherapy is used in combination with thermotherapy devices (incubators, radiant warmers, or heated mattresses), The light may impact the heat supply in these devices and the patient’s body temperature. The use of baby controlled mode of these devices is recommended unless manual mode specifically prescribed.
Caution
- Always lock the caster wheels of the Colibri Phototherapy before starting the treatment.
- Do not position the device in a location that impairs its ventilation.
- Do not cover the device with blankets during operation.
- Do not modify this equipment without authorization of the manufacturer.
- DO NOT disassemble any part of the Colibri Phototherapy. None of the parts of the device are designated to be repaired or replaced by service personnel. In case of any problems, contact the nearest MTTS representative.
- DO NOT use non-standard power cords with Colibri Phototherapy. Using nonstandard power cords can result in damaging the device.
- Do not use Colibri Phototherapy if it appears that any part of the device is not functioning properly or if any parts appear damaged. Contact MTTS.
- In case of any malfunction DO NOT attempt to service this machine. Colibri Photothrapy should only be serviced by a qualified MTTS technician.
- LED lifetime warning: The blue LEDs in Colibri Phototherapy are rated to provide effective treatment for 60,000 hours.
If the Colibri Phototherapy has been used for over 60,000 hours, after turning ON the display will show the “total device hours” for 3 seconds and ‘Error 4’ message after that. The error message will continue to appear every second during treatment.
NOTE: The estimated 60,000 hour life span is equal to a full 7 years of nonstop use. After this period of time the LED phototherapy lights will continue to function, however treatment will be significantly less effective and Colibri Phototherapy should be replaced.
Setting up Phototherapy
Step 1 – Remove components from the box
Box contains:
- Stand
- Base
- Canopy set
Step 2 – Stand assembly preparation
1. Remove protection label.
2-3. Unscrew slowly height adjustment knob.
While unscrewing the knob secure inner pole with hand in order to prevent
uncontrolled releasing.
4. Remove height adjustment knob.
Step 3 – Base assembly
1. Insert the stand into the base.
2. Screw in 3 bolts into the thread hole.
3-4. Connect and secure DC connector.
5. Tighten the bolts using wrench (size 13).
Step 4 – Stand assembly
- Insert stand support ring.
- Screw in height adjustment knob.
Step 5 – Canopy set assembly
1. Pull out plastic bracket with DC connector from inside of the stand.
2. Connect it with the canopy set wire.
3-4. Insert connected wire back into the stand and cover it with the plastic
bracket.
5. Attach canopy set to the stand and secure it with screws.
2 people may be required to perform operations 2-4.
Step 6 – Plug in and power up
The connector plugs into the back of the Colibri Photothrapy base and then
directly into the power outlet in the wall.
Colibri Phototherapy should only be used with the power outlets that have a
protective earth (ground) pin.
Using Phototherapy
Starting and stoping the therapy
- Plug in Colibri Phototherapy to the power outlet. Turn on the device by pressing the “On/Off Button”.
The display shows the “total device hours”. This will be displayed for 3 seconds.
- The blue phototherapy lights will now start to shine in STANDARD MODE.
The display will start counting the “treatment time” in STANDARD MODE.
This example shows 1 minute of STANDARD therapy.
- The phototherapy treatment will continue until the doctor decides therapy is no longer necessary and the device can be switched off by pressing “On/Off Button”.
This example shows 72 hours 30 minutes of STANDARD therapy.
This example shows the OFF message.
Selecting therapy mode
Colibri Phototherapy can be operated in 2 modes:
– STANDARD: ~37 µW/cm2/nm
– INTENSIVE: ~53 µW/cm2/nm
- To select the mode press the “Therapy Mode Button” Colibri Phototherapy starts with STANDARD MODE once it is turned on. Pressing the “Therapy Mode Button” will switch it to INTENSIVE MODE.
STANDARD MODE is indicated by the letter “n” on the display (“n” for “normal”)
INTENSIVE MODE is indicated by the letter “b” on the display (“b” for “boost”)
The display shows the “total device hours of INTENSIVE MODE”. This will be displayed for 3 seconds.
- The blue phototherapy lights will now start to shine in INTENSIVE MODE.
The display will start counting the “treatment time” in INTENSIVE MODE.
This example shows 1 minute of INTENSIVE therapy.
- The phototherapy treatment will continue until the doctor decides therapy is no longer necessary and the device can be switched off by pressing the “On/Off Button”.
Pressing the “Therapy Mode Button” will switch back the device to STANDARD MODE and restart the treatment counter.
This example shows 13 hours 20 minutes of INTENSIVE therapy.
If there is a power interruption, when the power is restored the device will
return to the mode it was in before the interruption of power occured.
Height adjustment
Colibri Phototherapy can be moved up and down in order to select preferred
distance from the light source to the patient. The height of the device is
between 115 and 140 cm.
NOTE: Recommended distance from the light source to the patient is 30-40
cm.
NOTE: Exercise caution when adjusting the height of the Colibri
Phototherapy.
Battery backup (optional)
Battery Backup
The battery backup allows operating the device without AC power supply:
– 3 hours in STANDARD MODE
– 2 hours in INTENSIVE MODE
The backup activates automatically as soon as the AC power supply is off.
To check the battery charge level and obtain additional information, follow
the steps below:
Action | Button | Display | Information |
---|---|---|---|
Press | Device turned on in |
STANDARD MODE
1 min of treatment time [see p.121
Hold 3 seconds| | | Software version number
Press| | | Remaining battery capacity percentage
Press| | | Cell 1 voltage
Press| | | Cell 2 voltage
Press| | | Cell 3 voltage
Press| | | Cell 4 voltage
Press| | | Device in STANDARD MODE 2 min of treatment time
Battery backup (optional)
Low battery alarm
When the lowest cell voltage reaches 3V, the device informs about low
batteries.
“Lo bAt’ and audible alarm lasts 10 seconds after which the devices turns off
by itself.
Cleaning and disinfection
Cleaning and disinfection
-
If visibly soiled, clean the affected parts with soap and water using a soft cloth.
Then wipe with a wet cloth and dry with a soft cloth. For difficult areas use a standard soft-bristled brush. -
To loosen heavy, dried-on soil, you may first need to saturate the spot with water.
When cleaning the device disconnect the power cord.
Irradiance measurement
NOTE: To ensure that the device is functioning properly, regularly
monitor Colibri Phototherapy in the following way.
Measurement instructions:
To test the irradiance of Colibri Phototherapy, measure using a light meter
specifically designed to measure wavelengths for phototherapy.
A stabilization time of 2 hours of continuous operation is required for
accurate measuring of the irradiance.
When measuring the irradiance, prop up the light meter 30-40 cm from the top
of the canopies to achieve the irradiance seen in the pictures below. Measure
the irradiance of Colibri Phototherapy in each of the sections seen in the
pictures below.
The lights will decay gradually over time, reaching approximately 70% of their
original irradiance after the Colibri Phototherapy device has been used for
60,000 hours. After 60,000 hours of use, the light source will start to
degrade rapidly. It is recommended that the device be replaced after 60,000
hours of use. If you must keep using the device after 60,000 hours of use, you
should increase the frequency at which you measure the irradiance of the
device.
Specifications
Performance Specifications
Illumination source | High-power blue LEDs, 1-1.25W standard operating | |
---|---|---|
Peak wavelength | 440 – 460 nm | |
Lamp duration | 60,000 hours | |
Total spectral irradiance | Standard Mode | |
Intensive Mode | 37.0 µW/cm2/nm |
53.0 µW/cm2/nm
Effective surface area| | 50 cm x 30 cm
Relative local distribution| Standard Mode| 0.59 (IEC Compliant >0.4)
| Intensive Mode| 0.60 (IEC Compliant >0.4)
Time Totalizer| | Treatment time, total device hours, total device hours in
Intensive Mode
Measured with| | MTTS LM-800 at 40cm
Physical Specifications
Dimensions (LxWxD) | max height | 69.5 cm x 60.2 cm x 155 cm |
---|---|---|
Height of light source | maximum | 140 cm |
minimum | 115 cm | |
Total unit mass | 16 kg |
Electrical Specifications
Power characteristics
Off-the-shelf internal power supply| 60W, 100-240VAC, 47/63Hz
Part No. Sinpro MPU60A-105
International medical safety approvals
(ANSI/AAMI/EN 60601-1, UL/TUV)
Class I construction is standard (ground required)
100k hours MTBF
Energy Star Efficiency Level V compliant
RoHS 3 compliant
Over voltage / over current protection
Continuous service|
---|---|---
Power Cords| North America| (part 50-07302-01)
| C.E.| (part
| U.K.| 50-07305-01)
| | (part 50-09273-01)
Environmental Specifications
Operating| Ambient temperature +10°C to +35°C
Humidity: 0% to 90% RH non condensing
Atmospheric Pressure: 70-106kPa
---|---
Transport and storage| Ambient temperature 0°C to +50°C
Humidity: 0% to 90% RH non condensing
Atmospheric Pressure: 70-106kPa
Exclusions| None
Explanation of Symbols
| This statement is a warning. Not following this statement could result in
injury to the patient or operator, or damage to the device.
---|---
| CE Marking with Notified Body Number.
| Do not disassemble the device unless you are an MTTS trained technician or
have been instructed to by qualified personnel (excluding the removal of the
bassinet).
|
Consult instructions for use or consult the electronic insturctions for use
| This device contains electronic parts. Do not dispose of it with normal
waste. Dispose of it according to local guidelines for disposal of eletronics.
Dispose according to the WEEE directive in the European Union.
| Medical device
| Unique identification number
| Website where a user can obtain additional information on the medical
product
| Model number
| Catalogue number
| Authorized Representative in the European Community
| Device manufacturer.
| Date of manufacture.
| Serial number.
| An eye patch should always be placed on the patient before commencing
treatment.
| Keep out of direct sunlight.
| Liquid ingress protection – level 1 (dripping water).
| Country of manufacture
| Temperature limit
| Humidity limitation
| Atmospheric pressure limitation
Warranty Policy
General Terms
This MTTS Limited Warranty gives you, the customer, express limited warranty
rights from MTTS, the manufacturer for the duration specified on the Warranty
Card. Please refer to the MTTS Website for an extensive description of your
limited warranty entitlements. In addition, you may also have other legal
rights under applicable law or special written agreement with MTTS. MTTS MAkES
NO OTHER ExPRESS WARRANTY OR CONDITION WHETHER WRITTEN OR ORAL AND MTTS
ExPRESSLY DISCLAIMS ALL WARRANTIES AND CONDITIONS NOT STATED IN THIS LIMITED
WARRANTY. TO THE ExTENT ALLOWED BY LOCAL LAW OF jURISDICTIONS OUTSIDE VIETNAM,
MTTS DISCLAIMS ALL IMPLIED WARRANTIES OR CONDITIONS, INCLUDING ANY IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. FOR ALL
TRANSACTIONS OCCURRING IN VIETNAM ANY IMPLIED WARRANTY OR CONDITION OF
MERCHANTABILITY, SATISFACTORY QUALITY, OR FITNESS FOR A PARTICULAR PURPOSE IS
LIMITED TO THE DURATION OF THE ExPRESS WARRANTY SET FORTH ABOVE. SOME
COUNTRIES DO NOT ALLOW A LIMITATION ON HOW LONG AN IMPLIED WARRANTY LASTS OR
THE ExCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES FOR
CONSUMER PRODUCTS. IN SUCH COUNTRIES, SOME ExCLUSIONS OR LIMITATIONS OF THIS
LIMITED WARRANTY MAY NOT APPLY TO YOU. FOR CONSUMER TRANSACTION, THE LIMITED
WARRANTY TERMS CONTAINED IN THIS STATEMENT, ExCEPT TO THE ExTENT LAWFULLY
PERMITTED, DO NOT ExCLUDE, RESTRICT, OR MODIFY BUT ARE IN ADDITION TO THE
MANDATORY STATUTORY RIGHTS APPLICABLE TO THE SALE OF THIS PRODUCT TO YOU. This
Limited Warranty is applicable in all countries and may be enforced in any
country or region where MTTS or its authorized service providers offer
warranty service for the same product model number (subject to the terms and
conditions set forth in this Limited Warranty)
Under this Limited Warranty, products purchased in one country or region may
be transferred to another country or region where MTTS or its authorized
service providers offer warranty service for the same product model number.
Warranty terms, service availability, and service response times may vary from
country or region to country or region. Standard warranty service response
time is subject to change due to local parts availability. If so, your MTTS
authorized service provider can provide you with details. MTTS will not alter
form, fit, or function of this MTTS product to make it operate in a country
for which it was never intended to function for legal or regulatory reasons.
MTTS is not responsible for any tariffs or duties that may be incurred in
transferring the products. MTTS guarantees that the product that you have
purchased or leased from MTTS is free from defects in materials or workmanship
under normal use during Limited Warranty Period. The Limited Warranty Period
starts on the date of purchase or lease from MTTS, or from the date MTTS
completes installation. Your dated sales or delivery receipt, showing the date
of purchase of the product, is your proof of the purchase or lease date. You
may be required to provide proof or purchase or lease as a condition of
receiving warranty service. You are entitled to hardware warranty service
according to the terms and conditions of this document if a repair to your
MTTS product is required within the Limited Warranty Period. Unless otherwise
stated, and to the extent permitted by local law, new MTTS product may be
manufactured using new materials and used materials equivalent to new in
performance and reliability. MTTS may repair or replace MTTS products (a) with
new or previously used products or parts equivalent to new in performance and
reliability, or (b) with equivalent products to an original product that has
been discontinued. Replacement parts are warranted to be free from defects in
material or workmanship for ninety (90) days or, for the reminder of Limited
Warranty Period of the MTTS product they are replacing or in which they are
installed, whichever is longer.
MTTS will, at its sole discretion, repair or replace any components or product
that manifests a defect in materials or workmanship during the Limited
Warranty Period. All component parts removed under this Limited Warranty
become the property of MTTS. In the unlikely event that your MTTS product has
recurring failures, MTTS at its sole discretion, may elect to provide you with
(a) a replacement unit selected by MTTS that is the same or equivalent to your
MTTS product in performance or (b) to give you a refund of your purchase price
or lease payments (less interests) instead of a replacement. This is your
exclusive remedy for defective products.
Exclusions
MTTS DOES NOT WARRANT THAT THE OPERATION OF THIS PRODUCT WILL BE
UNINTERRUPTED OR ERRORFREE. MTTS IS NOT RESPONSIBLE FOR DAMAGE THAT OCCURS AS
RESULT OF YOUR FAILURE TO FOLLOW THE INSTRUCTIONS INTENDED FOR THE PRODUCT.
This Limited Warranty does not apply to expendable or consumable parts and
does not extend to any product from which the serial number has been removed
or that gas been damaged or rendered defective (a) as a result of accident,
misuse, abuse, contamination, improper or inadequate maintenance or
calibration (if required) or other external causes; (b) by operation outside
the usage parameters stated in the user documentation shipped with the
product; (c) by software, interfacing, parts or supplies not supplier by MTTS;
(d) improper site preparation or maintenance; (e) virus infection; (f) loss or
damage in transit ; or (g) by modification or service by anyone other than (i)
MTTS personnel, (ii) an MTTS authorized service provider, or (iii) your own
installation of end-user replaceable MTTS or MTTS approved parts if available
for your MTTS product in the servicing country or region.
Limitation of Liability
IF YOUR MTTS PRODUCT FAILS TO WORk AS WARRANTED ABOVE, THE MAxIMUM
LIABILITY OF MTTS UNDER THIS LIMITED WARRANTY IS ExPRESSLY LIMITED TO THE
LESSER OF THE PRICE YOU HAVE PAID FOR THE PRODUCT OR THE COSTS OF REPAIR OR
REPLACEMENT OF ANY HARDWARE COMPONENTS THAT MALFUNCTION IN CONDITIONS OF
NORMAL USE. ExCEPT AS INDICATED ABOVE, IN NO EVENT WILL MTTS BE LIABLE FOR ANY
DAMAGES CAUSED BY THE PRODUCT OR THE FAILURE OF THE PRODUCT TO PERFORM,
INCLUDING ANY LOST PROFITS OR SAVINGS OR SPECIAL, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES. MTTS IS NOT LIABLE FOR ANY CLAIM MADE BY A THIRD PARTY OR MADE BY YOU
FOR A THIRD PARTY. THIS LIMITATION OF LIABILITY APPLIES WHETHER DAMAGES ARE
SOUGHT, OR CLAIM MADE, UNDER THIS LIMITED WARRANTY OR AS A TORT CLAIM
(INCLUDING NEGLIGENCE AND STRICT PRODUCT LIABILITY), A CONTRACT CLAIM, OR ANY
OTHER CLAIM. THIS LIMITATION LIABILITY CANNOT BE WAIVED OR AMENDED BY ANY
PERSON. THIS LIMITATION OF LIABILITY WILL BE EFFECTIVE EVEN IF YOU HAVE
ADVISED MTTS OF THE POSSIBILITY OF ANY SUCH DAMAGES. THIS LIMITATION OF
LIABILITY, HOWEVER, WILL NOT APPLY TO CLAIMS FOR PERSONAL INjURY. THIS LIMITED
WARRANTY GIVES YOU SPECIFIC LEGAL RIGHTS. YOU MAY ALSO HAVE OTHER RIGHTS THAT
MAY VARY FROM COUNTRY TO COUNTRY. YOU ARE ADVISED TO CONSULT APPLICABLE
COUNTRY LAWS FOR A FULL DETERMINATION OF YOUR RIGHTS.
Limited Warranty Period
The Limited Warranty Period for this MTTS product is a specified, fixed
period commencing of the date of purchase and specified on the Warranty Card.
The date on your sales receipt is the date of purchase unless MTTS or your
reseller informs you otherwise in writing.
Customer Responsibilities
In order to avoid the risk of charges for issues not covered by your limited
warranty (issues that are not due to defects in materials and workmanship on
MTTS product), you will be asked to assist MTTS as follows: Verify
configurations, load most recent firmware, install software patches, run MTTS
diagnostics and utilities; Implement temporary procedures or workarounds
provided by MTTS while MTTS works on permanent solution; Cooperate with MTTS
in attempting to resolve the problem using online chat, email, or telephone.
This may involve performing routine diagnostic procedures, installing
additional software updates or patches; Perform additional tasks as defined
within each type of warranty service provided by MTT and any other actions
that MTTS may reasonably request in order to best perform the warranty support
CUSTOMER IS RESPONSIBLE FOR DELIVERING THE PRODUCT (AND ALL COSTS INVOLVED)
FROM HIS LOCATION TO THE MTTS AUTHORIzED SERVICE POINT.
Contacting MTTS
If your MTTS product fails during the Limited Warranty Period and the suggestions in the product documentation do not solve the problem, you can receive support by doing one of the following:
- Locate and contact your nearest MTTS service provider via MTTS website: http://www.mtts-asia.com/support/
- Call the Technical Support Center: +84 43 7666521
Before calling MTTS or an MTTS authorized service provider please have the following information available:
- Product serial number and model name
- Applicable error messages
- Detailed questions
MEDICAL TECHNOLOGY TRANSFER AND SERVICES Co., LTD
House No. 26, Alley 41, An Duong Vuong Street, Tay Ho District, Hanoi City,
Vietnam
SRN: VN-MF-000036669
Tel: +84 24 3766 6521
Fax: +84 24 3718 8050
Email: assistance@mtts-asia.com
www.mtts-asia.com
Logic s.r.l.
Via Antonio Pigafetta 1
34147 Trieste, Italy
SRN: IT-AR-000012301