UNISENSE Redox Electrode Calibration Kit Instruction Manual
- June 5, 2024
- UNISENSE
Table of Contents
- UNISENSE Redox Electrode Calibration Kit
- Warranty and liability
- Support, ordering, and contact information
- Safety
- Content of the calibration kit
- Principle of calibration
- Calibration procedure
- Specifications
- The redox potential of the calibration solution at different temperatures
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
UNISENSE Redox Electrode Calibration Kit
Warranty and liability
Notice to Purchaser
This product is for research use only. Not for use in human diagnostic or
therapeutic procedures.
Warning
Microsensors have very pointed tips and must be handled with care to avoid
personal injury and only by trained personnel. Unisense A/S recommends users
to attend instruction courses to ensure proper use of the products.
Warranty and Liability
The Redox Calibration Kit is guaranteed to give the redox potential given in
section 8 until expiry as indicated on the package label. The warranty does
not include replacement necessitated by accident, neglect, misuse,
unauthorized repair, or modification of the product. In no event will Unisense
A/S be
liable for any direct, indirect, consequential or incidental damages,
including lost profits, or for any claim by any third party, arising out of
the use, the results of use, or the inability to use this product.
Support, ordering, and contact information
If you wish to order additional products or if you encounter any problems and
need scientific or technical assistance, please do not hesitate to contact our
sales and support team. We will respond to your inquiry within one working
day.
E-mail: sales@unisense.com
Unisense A/S
Tueager 1
DK-8200 Aarhus N, Denmark
Tel: +45 8944 9500
Fax: +45 8944 9549
Further documentation and support are available at our website:
https://www.unisense.com/.
Safety
The calibration solution contains sodium hydroxide.
Warning
- Hazard Statements
- H319 – Causes serious eye irritation
- H302 – Harmful if swallowed
- H315 – Causes skin irritation
- H360FD – May damage fertility. May damage the unborn child See the full Materials Safety Data Sheet at: https://www.unisense.com/calibration_kits/
- NOTE: The calibration solution is iodine-based and strongly colouring. Wipe off any spills immediately!
Content of the calibration kit
Item | Number |
---|---|
Exetainer with Redox buffer solution | 10 |
Exetainer with low Redox test solution | 10 |
Calibration cap with O-ring and 3 cm Viton tubing | 2 |
Y-connector with 3 cm Viton tubing | 1 |
10 ml syringe | 1 |
80 x 2.1 mm needle (green) | 1 |
10 cm transparent tube for a reference electrode | 1 |
Principle of calibration
The redox potential must always be reported relative to the Standard Hydrogen
Electrode (SHE).
However, during measurement it is not possible to use the SHE and, another
type of reference electrode must be used. This can be any reference electrode
(Ag/AgCl, Calomel, etc.) which all have a different half-cell potential than
the SHE. The measured redox potential must be corrected for this
difference.
The calibration solution in this calibration kit has a well-defined redox
potential relative to the SHE (see table in paragraph 8) and this can be used
to determine the offset between the measured redox potential and the redox
potential relative to the SHE.
The offset is calculated as:
- Offset (mV) = PotSHE (mV) – PotCAL (mV)
where PotCAL is the measured potential in the calibration solution using the
redox electrode and the Ag/AgCl electrodes and PotSHE is the defined redox
potential for the calibration solution relative to the SHE found in section 8.
Once the offset is determined, redox potentials measured relative to a
Ag/AgCl reference electrode can be recalculated relative to the SHE as:
- PotTRUE (mV) = PotMEAS (mV) + Offset (mV)
where PotTRUE is the measured potential relative to the SHE and PotMEAS is the redox potential measured in the sample using the redox electrode and the Ag/AgCl reference electrode.
Calibration procedure
For calibration of redox electrodes, it is only necessary to determine the offset relative to the standard hydrogen electrode and to verify that the redox electrode responds to a change in redox potential (see section 5).
NOTE: For calibration in the Unisense software, the 1-point calibration box must be checked (Figure 2).
Calibrating a redox microelectrode with the external reference electrode
Estimating the offset
-
Connect the two calibration caps with the Y-connector.
-
Mount one calibration cap on the 10 cm transparent tube.
-
Mount the other calibration cap on the redox electrode protection tube.
-
Place the redox electrode and 10 cm transparent tube upright.
-
Place the reference electrode in the 10 cm transparent tube.
-
Mount the needle (green) on the 10 ml syringe.
-
Open the Exetainer with the Redox buffer and aspirate ca. 10 ml.
-
Remove the needle and attach the syringe to the black tubing on the Y-connector.
-
Inject the calibration solution into the calibration caps until the tips of both electrodes are immersed at least 1 cm.
-
Allow the redox electrode to respond and stabilize. Then record the calibration value in SensorTrace (see the SensorTrace manual for details: https://www.unisense.com/manuals/).
The offset is now determined and SensorTrace will automatically apply this to the measurements.
NOTE: The 1 point calibration box must be checked (Figure 2). -
Remove the redox buffer from the calibration caps.
-
Rinse the 10 ml syringe and the calibration caps thoroughly with water.
Verifying sensor response
- Open the Exetainer with the Redox test solution and aspirate 10 ml.
- Remove the needle and attach the syringe to the black tubing on the Y-connector.
- Inject the Redox test solution into the calibration caps until the tips of both electrodes are immersed at least 1 cm.
- Allow the redox electrode to respond. This response is much slower and more variable than for the Redox buffer.
- The signal should now be at least 200 mV below the signal for the Redox buffer. If this is the case the redox electrode is working fine.
Calibrating a redox microelectrode with a built-in the reference electrode
Estimating the offset
-
Mount a calibration cap on the redox electrode protection tube (see note A below).
-
Place the redox electrode upright.
-
Mount the needle (green) on the 10 ml syringe.
-
Open the Exetainer with the Redox buffer and aspirate ca. 10 ml.
-
Remove the needle and attach the syringe to the black tubing on the calibration cap.
-
Inject the calibration solution into the calibration cap until the tip of the electrode is immersed at least 1 cm.
-
Allow the redox electrode to respond and stabilize. Then record the calibration value in SensorTrace (see the SensorTrace manual for details: https://www.unisense.com/manuals/).
The offset is now determined and SensorTrace will automatically apply this to the measurements.
NOTE: The 1-point calibration box must be checked (Figure 2). -
Remove the redox buffer from the calibration cap.
-
Rinse the 10 ml syringe and the calibration cap thoroughly with water.
Verifying sensor response
- Open the Exetainer with the Redox test solution and aspirate 10 ml.
- Remove the needle and attach the syringe to the black tubing on the calibration cap.
- Inject the Redox test solution into the calibration cap until the tip of the electrode is immersed at least 1 cm.
- Allow the redox electrode to respond. This response is much slower and more variable than for the Redox buffer.
- The signal should now be at least 200 mV below the signal for the Redox buffer. If this is the case the redox electrode is working fine.
NOTES:
A. The protection tube of redox microelectrodes with built-in reference is
filled with 2 M KCl which keeps the reference electrode hydrated. This liquid
may be saved in a container and used for filling the protection tube after
use.
Calibrating a redox microelectrode for the Microrespiration System
Estimating the offset
-
Fill the Redox buffer into a Microrespiration chamber using the syringe and needle. Make sure that the chamber including the capillary opening in the lid is filled (see note A below).
-
Place this chamber in the Microrespiration stirrer rack.
-
The stirrer rack with chamber must be immersed in a water bath and the reference electrode must be dipped into the water bath (see note B below).
-
Place the redox microelectrode in the stirrer rack with the plastic tip in the opening of the Microrespiration chamber lid.
-
Insert the redox microelectrode into the redox buffer.
-
Allow the redox microelectrode to respond and stabilize. Then record the calibration value in SensorTrace (see the SensorTrace manual for details: https://www.unisense.com/manuals/).
The offset is now determined and SensorTrace will automatically apply this to the measurements.
NOTE: The 1-point calibration box must be checked (Figure 2). -
Pull the redox microelectrode back into the blue shaft and remove it from the rack.
-
lean the redox microelectrode by pushing it out of the blue shaft and dipping it into a beaker with water
Verifying sensor response
- Fill the low Redox test solution into a Microrespiration chamber using the syringe and needle. Make sure that the chamber including the capillary opening in the lid is filled.
- Repeat points 2 – 5 in “6.3.1 Estimating the offset” above.
- Allow the redox microelectrode to respond and stabilize.
- The signal should now be at least 200 mV below the signal for the Redox buffer. If this is the case the redox electrode is working fine.
- Pull the redox microelectrode back into the blue shaft and remove it from the rack.
- Clean the redox microelectrode by pushing it out of the blue shaft and dipping it into a beaker with water.
NOTES:
A. Microrespiration chambers from 400 μl to 4 ml may be used. Fill the chamber
and mount the lid making sure that the capillary opening in the lid is filled
with redox buffer.
B. There must be a continuous liquid connection between the redox electrode
and the reference electrode. This is ensured by completely filling the
Microrespiration chamber, including the capillary in the lid, with the redox
buffer. The water in the water bath is then in contact with the redox buffer
in the capillary opening of the lid, ensuring this continuous connection.
Specifications
- The volume of Redox buffer in Exetainer: 12.5 ml
- The volume of Redox test solution in Exetainer: 12.5 ml
- A lifetime of the calibration kit: See label on the calibration box
- The redox potential of the Redox buffer: See table in section 8 below.
The redox potential of the calibration solution at different temperatures
Redox potential of the Redox buffer solution relative to the Standard Hydrogen Electrode as a function of temperature
Temp (°C)| Eh (mV)| | Temp (°C)| Eh (mV)
0| 438.0| | 26| 418.9
1| 437.4| | 27| 418.0
2| 436.8| | 28| 417.2
3| 436.1| | 29| 416.3
4| 435.5| | 30| 415.4
5| 434.8| | 31| 414.5
6| 434.2| | 32| 413.6
7| 433.5| | 33| 412.7
8| 432.8| | 34| 411.8
9| 432.1| | 35| 410.9
10| 431.4| | 36| 409.9
11| 430.7| | 37| 409.0
12| 430.0| | 38| 408.0
13| 429.2| | 39| 407.1
14| 428.5| | 40| 406.1
15| 427.7| | 41| 405.1
16| 427.0| | 42| 404.1
17| 426.2| | 43| 403.1
18| 425.4| | 44| 402.1
19| 424.7| | 45| 401.1
20| 423.9| | 46| 400.1
21| 423.1| | 47| 399.0
22| 422.2| | 48| 398.0
23| 421.4| | 49| 396.9
24| 420.6| | 50| 395.9
25| 419.8| | 51| 394.8
References
- Microsensors, instruments, and meters for microscale measurements
- Accurate, fast, and easy calibration of your microsensors
- Manuals - Unisense
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>