H2S sensor calibration kit
For H2S and SULF sensors
Manual
Warranty and liability
1.1 Notice to Purchaser
This product is for research use only. Not for use in human diagnostic or
therapeutic procedures.
1.2 Warning
Microsensors have very pointed tips and must be handled with care to avoid
personal injury and only by trained personnel. Unisense A/S recommends users
to attend instruction courses to ensure proper use of the products.
1.3 Warranty and Liability
The H2S Calibration Kit is guaranteed to give the concentration indicated on
the package label until expiry as indicated on the package label. The warranty
does not include replacement necessitated by accident, neglect, misuse,
unauthorized repair, or modification of the product. In no event will Unisense
A/S be liable for any direct, indirect, consequential or incidental damages,
including lost profits, or for any claim by any third party, arising out of
the use, the results of use, or the inability to use this product.
Support, ordering, and contact information
If you wish to order additional products or if you encounter any problems and
need scientific or technical assistance, please do not hesitate to contact our
sales and support team. We will respond to your inquiry within one working
day.E-mail: sales@unisense.com
Unisons A/S
Langdyssen 5
DK-8200 Aarhus N, Denmark
Tel: +45 8944 9500
Fax: +45 8944 9549
Further documentation and support are available at our website:
www.unisense.com.
Content of the calibration kit
Item | Number |
---|---|
Exetainer with ZnS suspension – H2S stock solution | 10 |
Exetainer with HCl (pH = 2.1) and glass beads – Mixing vial | 10 |
Calibration cap with O-ring and 3 cm Viton tubing | 1 |
10 ml syringe | 1 |
1 ml syringe | 2 |
80 x 2.1 mm needle (green) | 1 |
30 x 0.6 mm needle (blue) | 1 |
50 x 1.2 mm needle (red) | 1 |
Figure 1: Calibration kit contents: A: Calibration kit box with Exetainers, B: 80 x 2.1 mm needle (green), C: 1 ml syringes, D: 10 ml syringe, E 50 x 1.2 mm needle (red), F: 30 x 0.6 mm needle (blue), G: Calibration Cap with tubing, H: O-ring.
Principle of calibration
Unisense Hydrogen Sulfide sensors (H2S and SUF) respond linearly to H2S
concentrations within their linear range (see specifications for your sensor
at https://www.unisense.com/H2S). Therefore, a twopoint calibration is
sufficient. One calibration point is the signal for zero H2S, which can be
water equilibrated with atmospheric air, and the other calibration point is
the signal for one known H2S concentration.
In this calibration kit, the sulfide is shipped in the form of zinc sulfide
precipitate (ZnS) which is insoluble in water (Solubility product = ca. 2 10
). During the calibration procedure, ZnS is injected into a dilute Hall (pH =
2.1). This causes the ZnS to dissolve and H2S is formed quantitatively.
H2S will react slowly with O2 if this is present. To minimize this, the ZnS
suspension and the dilute Hall are anoxic when dispensed into the Exetainers.
However, during storage and handling a small amount of O2 will enter the
solution. Therefore, the calibration must be performed relatively rapid, as
described below, to keep this reaction at a negligible level.
Correction for salinity
During calibration, a relationship between sensor signal and concentrations is
established. However, the sensor responds to the partial pressure of H2S which
at a given concentration varies with salinity. Therefore, a correction for
salinity must be applied.
Henrys law describes how partial pressure depends on the concentration and
solubility of a gas: (Equation 1)
The calibration solution, prepared as described in section 6.1 below, will have the H2S concentration stated on the package label. The solubility of H2S will decrease with increasing salinity. Therefore, at a given concentration, the partial pressure will increase with increasing salinity. As the sensor responds to partial pressure, the sensor signal will increase as well when salinity increases at a constant H2S concentration. The conversion of sensor signal to concentration must, therefore, take the relationship between salinity and solubility into account.
The salinity in the calibration solution (0.36‰) corresponds to freshwater. If
measuring at other salinities, a correction must be made. This correction is
done by assuming a H2S concentration different from that given on the package
label. This Virtual concentration is the one that, at the measuring salinity,
gives the same partial pressure of H2S as that in the calibration solution. If
the measuring salinity is higher than that in the calibration solution, the
Virtual concentration will be lower than that given on the package label. The
reason is that when salinity is higher than in the calibration solution, a
lower H2S concentration is needed to give the same partial pressure as that in
the calibration solution (see Henrys law above).
The Virtual concentration is calculated as (see Appendix 1):
(Equation 2)
(Equation 3)
where Conc. (Virtual) is the Virtual concentration, Conc. (kit) is the H2S
concentration given on the calibration box label, Sol. (Virtual) is the
solubility of H2S at the measurement salinity and temperature, Sol. (Kit) is
the solubility of H2S at the salinity in the calibration solution and the
measurement temperature, and Corr. factor is the correction factor. Note that
calibration and measurements must be carried out at the same temperature.
The Virtual concentration, calculated from Equation 3, is the concentration
that must be entered as the Known value (µmold/L) in the Sensor Trace
software.
Example:
Measure at 30‰ and 20°C.
Calibrate at 20°C.
Concentration of H2S in the calibration solution (Conc. (Kit)) = 104.3 µM
Correction factor (Corr. factor, Table 1) = 0.871
Virtual H2S concentration (Conc. (Virtual)) = 104.7 µM ⨉ 0.871 = 91.2 µM
In this example a concentration of 91.2 µM will give the same sensor signal as the concentration of 104.3 µM in the calibration solution due to the difference in solubility of H2S at the two salinities. Therefore, 91.2 µM is the value to be entered as the Known value in the SensorTrace software.
Calibration procedure
For calibrating the H2S sensor, a low and a high calibration point are needed. Note that MicroRespiration sensors and sensors with flow cells will not fit the calibration cap and must be calibrated as described in section 6.4 and 6.5 This calibration kit is not for calibrating High Range H2S sensors that work in the mM range.
6.1 Preparation of the H2S calibration solution
- Mount the 50 x 1.2 mm needle (red) on one 1 ml syringe, mount the 30 x 0.6 mm needle (blue) on the other 1 ml syringe, and mount the 80 x 2.1 mm needle (green) on the 10 ml syringe.
- Adjust the temperature of the Exetainers with ZnS and HCl to the wanted calibration emperature (see note A, section ).
- Shake the Exetainer with the ZnS precipitate vigorously for 30 seconds (see note B, section ).
- Open the Exetainer with the ZnS precipitate.
- Aspirate ca. 0.3 ml with the 1 ml syringe with the red needle, turn the needle upwards, tap it gently to get the bubbles to the top and eject these.
- Empty the syringe and aspirate 1 ml ZnS suspension and eject this back into the Exetainer with the needle immersed. Repeat this three times before filling the syringe.
- Adjust the volume in the syringe to exactly 1.0 ml.
- Insert the blue needle, mounted on the second 1 ml syringe, through the septum of the Exetainer with HCl. Leave the tip of the needle right below the septum.
- Insert the red needle, mounted on the syringe with ZnS, fully into the HCl Exetainer and inject the ZnS. The excess liquid is pushed into the empty syringe (See note C, section ).
- Remove first the ZnS syringe, then the syringe with the excess liquid.
- Shake the Exetainer vigorously for 10 seconds.
- Leave the Exetainer for 15 minutes for the formation of H2S from ZnS to complete.
Figure 2: Sulfide sensor with the calibration cap mounted. Calibration solution is injected with the 10 ml syringe.
6.2 Preparation of the hydrogen sulfide sensor
IMPORTANT:
- The pre-polarization period of the H2S sensor must have been completed before doing the calibration. See the H2S sensor manual for details: https://www.unisense.com/manuals/
- The temperature of the low and high calibration solutions must be the same.
- Perform the calibration at the same temperature as the measurements if possible. The UniAmp series of amplifiers has a built-in temperature compensation within –3°C of the calibration temperature.
- If doing the low calibration point with the calibration cap, make sure to do this before the high calibration point to avoid carry over from the H2S standard.
6.3 Calibrating most hydrogen sulfide sensors
(All sensors except those in Flow Cells and for the Microrespiration System –
see 6.4 and 6.5)
6.3.1 Obtaining the low calibration point
6.3.1.1 Using the Unisense Calibration Chamber
- Place the sensor with the protection tube in sulfide free water (see the H2S microsensor manual (https://www.unisense.com/manuals/))
- Allow the sensor to respond and stabilize and record the calibration value in SensorTrace (see the SensorTrace manual for details: https://www.unisense.com/manuals/)
6.3.1.2 Using the Calibration Cap
- Mount the calibration cap on the protection tube with the H2S sensor (Figure 2). Make sure that the O-ring is in place at the bottom of the calibration cap creating a seal between this and the protection tube.
- Fill the 10 ml syringe with sulfide free water.
- Inject this water into the calibration cap until the sensor tip is immersed at least 2-3 cm.
- Allow the sensor to respond and stabilize. Then record the calibration value in Sensor Trace (see the Sensor Trace manual for details: https://www.unisense.com/manuals/)
6.3.2 Obtaining the high calibration point
- Mount the calibration cap, if not already mounted, as described above in 6.3.1.2.
- Open the Exetainer with the H2S calibration solution prepared as above in 6.1.
- Aspirate ca. 10 ml of the H2S calibration solution with the syringe and needle.
- Keep the syringe vertical and avoid mixing of the calibration solution with the air bubble inside.
- Remove the needle and attach the 10 ml syringe to the calibration cap tubing.
- Inject the calibration solution slowly until the sensor tip is immersed at least 2-3 cm.
- Allow the sensor to respond and stabilize. Then record the calibration value in Sensor Trace (see the Sensor Trace manual for details: https://www.unisense.com/manuals/) (See note D, section ). If the response is much slower than expected, the ZnS may not have been converted fully into H2S (See note E, section )
- Remove the H2S calibration solution with the syringe.
- Rinse the calibration cap and protection tube carefully, removing all the H2S calibration solution using the 10 ml syringe.
6.4 Calibrating hydrogen sulfide sensors for the Microrespiration system
Sensors of the Microrespiration type (SULF-MR or H2S-MR) cannot be calibrated
using the calibration cap. Instead, it is recommended to follow the procedure
outlined below. Pay attention to the general information in section 6.1 and
6.2, the Notes and recommendations in section , and follow the procedure below
Figure 3: H2S sensor in the Micro respiration guide.
6.4.1 Obtaining the high calibration point
- Prepare the H2S calibration solution as described in section 6.1.
- Aspirate 5 ml of the H2S calibration solution with the syringe and needle. Do this slowly to avoid bubble formation.
- Dispense the H2S calibration solution into a Micro Respiration chamber. Place the needle at the bottom of the chamber, filling from below, to avoid bubbles and splashing.
- Mount the lid in the Micro Respiration chamber making sure that no air bubbles are trapped.
- Place the Micro Respiration chamber in the stirrer rack.
- Place the H2S sensor in the stirrer rack with its plastic tip in the opening of the lid.
- Insert the sensor into the chamber.
- Allow the sensor to respond and stabilize and record the calibration value in Sensor Trace (see the Sensor Trace manual for details: https://www.unisense.com/manuals/).
- Retract the sensor tip and remove the sensor from the stirrer rack.
6.4.2 Obtaining the low calibration point
6.4.2.1 Using the Unisense Cal300 Calibration Chamber
-
Place the sensor in the Cal300 Calibration Chamber containing H2S free water (see the H2S microsensor manual (https://www.unisense.com/manuals/).
• The H2S sensor must be mounted in the blue Micro respiration guide and the tip must be retracted.
• Temperature of the water must be the same as where the measurements are done. -
Allow the sensor to respond and stabilize and record the calibration value in Sensor Trace (see the Sensor Trace manual for details: https://www.unisense.com/manuals/)
6.4.2.2 Using a Micro respiration Chamber
- Prepare a volume of H2S free water at the same temperature as the H2S calibration solution used in 6.4.1.
- Transfer this water to a Micro Respiration chamber and mount the lid.
- Place the Micro Respiration chamber in the stirrer rack
- Place the H2S sensor in the stirrer rack with its plastic tip in the opening of the lid.
- Insert the sensor into the chamber.
- Allow the sensor to respond and stabilize and record the calibration value in Sensor Trace (see the Sensor Trace manual for details: https://www.unisense.com/manuals/) (See note D, section ). If the response is much slower than expected, the ZnS may not have been converted fully into H2S (See note E, section )
- Retract the sensor tip and remove the sensor from the stirrer rack.
6.5 Calibrating hydrogen sulfide sensors in flow cells
Sensors with flow cells cannot be calibrated using the Calibration Cap.
Instead, it is recommended to follow the procedure outlined below. Pay
attention to the general information in section 6.1 and 6.2, the general
calibration procedure in section 6.3, the Notes and recommendations in section
, and follow the procedure below.
6.5.1 Create a calibration setup
In order to calibrate a sensor in a flow cell, the sensor tip must be exposed
to the calibration liquid prepared in the Exetainer (described in section
6.1). The optimal way to do this depends on the actual setup, however, it is
recommended to make a setup that allows calibration of the sensor without
removing it from the flow cell and from the setup. Generally, this could be a
Luer connector connected to the flow cell, directly or via tubing, that allows
injection of the calibration liquid into the flow cell. A three-way valve on
either side of the flow cell will allow easy injection of the calibration
liquid with the sensor and flow cell in place.
Connection of the syringe with calibration liquid to the flow cells:
- PEEK flow cells: The syringe may be attached directly to the flow cell via a Luer adapter that is mounted directly in the flow cell (Figure 4, left)
- Glass and Swagelok stainless steel flow cells: The syringe may be attached via rubber tubing. The syringe may be connected directly to the rubber tubing or via a barbed Luer adaptor (Figure 4, right)
Figure 4: Left: Luer adaptor for direct mounting in the flow cell (e.g., IDEX P-624). Right: Barbed Luer adaptor for tube connection
6.6 Notes and recommendations
A. Perform the calibration at the same temperature as the measurements if
possible. The UniAmp series of amplifiers has a built in temperature
compensation within –3°C of the calibration temperature.
B. It is important to shake the ZnS suspension very well and obtain a
homogenous distribution of the precipitate. If not, the aliquot of suspension
transferred to the HCl Exetainer will not contain the correct amount of ZnS
and the final H2S concentration will be wrong.
C. Injection of the H2S containing water is done with the needle inserted
fully while the blue needle on the empty syringe is inserted just below the
septum. Thereby the injected H2S containing water will not be lost.
D. The calibration point should be saved within ca. 30-60 seconds after
injecting the H2S calibration solution. The period between injection and
saving the calibration point should be long enough to obtain almost full
response which depends on the sensor response time. Furthermore, it should
not be longer than needed because the H2S calibration solution will be
contaminated with O2 during handling, and consumption of H2S due to reaction
with O2 will occur. This reaction is slow but can be seen as a slow decrease
in signal over a few minutes.
E. If the sensor signal keeps rising for several minutes, the formation of H2S
from ZnS may not have been completed. Leave the sensor in the H2S until the
signal is stable. For the next calibration, extend the time for the ZnS to be
converted to H2S (Section 6.1).
Calibrating at other concentrations
It is possible to obtain a lower concentration than that obtained using the
standard procedure in section. This may be done either injecting less than 1.0
ml or by diluting the calibration solution obtained using the standard
procedure in section.
Dilution of the solution made in section 6.1.
- Prepare two Mixing Exetainers.
- Transfer 1.0 ml the ZnS suspension into the first Mixing Exetainer as described above in section 6.1.
- After shaking the Exetainer vigorously to obtain a homogenous distribution of H2S, open the Exetainer and aspirate a volume with a syringe.
- Inject a known volume into the second Exetainer with the blue needle on a syringe inserted just below the septum to collect the excess liquid (this volume of this syringe must be sufficient to accept all the excess liquid).
- Shake the Exetainer vigorously.
Now the sensor may be calibrated, as described above for the different types of sensors, using the calibration solution in the second Exetainer.
The concentration of H2S in the second Exetainer may be calculated as:
Final conc. (µM) = Inj. vol. (ml)/Exetainer vol. (ml) x Certified conc. (µM)
where Final conc. (µM) is the concentration obtained in the second Exetainer,
Inj. vole (ml) is the volume injected transferred from the first to the second
the Exetainer, Exetainer vol. (µM) is the volume of the second Exetainer and
Certified conc. (µM) is the H2S concentration obtained when following the
standard procedure in section. The Certified conc. and the Exetainer volume
are shown on the label on the calibration kit box.
Please note that when injecting less than the full volume of a syringe, the accuracy of the injection will become lower. Therefore, always use a syringe with a full volume that is close the amount to be injected. E.g., if 3 ml is injected with a 10 ml syringe the accuracy is low.
Specifications
- Volume of ZnS suspension in Exetainer 1 : 12.5 ml
- Volume of HCl (pH = 2.1) in Exetainer 1 : 12.5 ml
- Lifetime of the calibration kit: See label on the calibration box
- Concentration of H2S in calibration solution 2 : See label on the calibration box
1The zinc sulfide suspension and the HCl solution are made anoxic with Ar
bubbling before these liquids are dispensed into the Exetainers.
2The actual concentration of H2S in the final calibration solution is
determined from measurements with a H2S sensor calibrated with a certified
calibration gas. The number of this certificate is specified on the
calibration kit box. The H2S concentration in the final calibration solution,
following the procedure in section 6.1, is specified in the label on the
calibration kit box.
Appendix 1: Calculation of the correction factor for salinity
The relationship between partial pressure, concentration and solubility for a
gas is given by Henrys law: Part. press. = Conc./Sol.
In the calibration solution:
During measurement at a different salinity: The aim is to calculate which concentration at the measuring salinity that will give the same partial pressure as in the calibration solution. The partial pressure at the measurement salinity and that at the calibration salinity are thus set to be equal:
Example
Concentration calculated based on the calibration = 100 µM
Conditions at measuring: Temperature = 20°C, Salinity = 32‰ => Correction
factor = 0.863 (Table 1)
Virtual concentration at 20°C, 32‰ = 100 µM x 0.863 = 86.3 µM
Table 1. Correction factors for calculating virtual concentration
Calculated from: Morse, J. W., F. J.
Miller, J. C. Cornwell, and D. Rickard. 1987. The chemistry of the hydrogen
sulfide and iron sulfide systems in natural waters.
Earth-Science Rev. 24: 1–42. https://doi.org/10.1016/0012-8252(87)90046-8
References
- Microsensors, instruments, and meters for microscale measurements
- Microsensors, instruments, and meters for microscale measurements
- Redirecting
- H2S Microsensor for hydrogen sulfide research - Unisense
- Manuals - Unisense