RX3000 Station Key Components
1. Log in to HOBOlink.
Log in to an existing account at www.hobolink.com
or create a new one. You’ll receive an email to activate the new account.
2. Register the station.
In HOBOlink, click Devices, RX Devices, and then the Register a Device link.
Give the station a name and enter the serial number and device key from the
label inside the station door.
Install optional modules or user-supplied SIM.
Ensure the station is powered off (unplug the charging device and then
disconnect the battery). Insert optional RXMOD-x analog, relay, water level
sensor, or RXW Manager modules into the Module 1 or 2 slots by plugging the
connector into the receptacle on the station. Screw the module in place with
the provided screw. If there is just one relay module, install it on the
right. If you installed an RXW Manager
module, plug the cable from the RXW Manager mote into the jack on the module,
making sure the cable is inserted through the upper cable access opening on
the bottom of the station case. Do not reconnect the power on the station
until the mote is plugged in as shown. WARNING: If you
inadvertently install modules while the power is on, you must disconnect and
then reconnect the battery and charging device to guarantee proper operation.
Note: If you are using your own SIM card, follow the instructions at
https://www.onsetcomp.com/support/manuals/installing-simrx3000-station
before continuing.
4. Plug in the battery and charging device.
Connect the battery port connector followed by the appropriate charging port
connector. Once the battery cable is plugged in, “Initializing System” will
flash on the LCD. A checkmark appears next to “System” after the system
initialization is complete.
Tip: Refer to the RX3000 manual for instructions on wire routing and
environmental sealing of wires.
Check or configure device communications.
For RX3003 or RX3004 cellular models:
The station will automatically connect to HOBOlink after powering up the
station in the previous step. Continue to step 6. For RX3001 Ethernet models:
a. Plug in the Ethernet cable.
Note: The RX3001 Ethernet station is automatically configured to use
DHCP. If your network uses static IP addresses, contact your Network
Administrator for the correct addresses and follow the Ethernet configuration
steps in the RX3000 manual at
www.onsetcomp.com/manual/rx3000.
b. Press the Connect button (under the cloud icon) on the station to connect
to HOBOlink. Wait for a checkmark to appear next to Last Connection as shown
in the example in this step.
For RX3002 Wi-Fi models:
a. Connect the station to the computer with the USB cable.
b. Select Manage RX Station from the Device menu. (Select Domain networks and
click Allow Access if you see a Windows® Firewall warning.)
Note: To use the RX Manager, use HOBOware version 3.7.23 or earlier. It
is not available in the latest version, 3.7.24.
c. In the RX Station Manager, click the Actions button and select Network
Access.
d. Enter the Security information for your Wi-Fi network. Type the Network
Name, select the Security Type, and type the Security Key. Consult your
Network Administrator or wireless router documentation for help with
determining your network security type.
Note: The station uses DHCP by default. If your network uses static IP
addresses, contact your Network Administrator for the correct addresses to
enter in the Networking fields.
e. Click Save in the RX Station Manager. Click Done and then close the RX
Station Manager.
f. Disconnect the USB cable.
g. Press the Connect button (under the cloud icon) on the station to connect
to HOBOlink. Wait for a checkmark to appear next to Last Connection as shown
in the example in this step.
Plug in and search for any smart sensors.
Important: If this RX3000 station is a replacement for an existing U30
station, it is imperative that you power down the U30 station you are
replacing (disconnect the battery and AC adapter or solar panel) before you
remove the smart sensors and connect them to the RX3000. Do not repower the
U30 station or allow it to connect to HOBOlink again unless at least one
different smart sensor is connected to the U30 station first.
a. Plug smart sensors into the ports at the bottom of the case. Tip: Refer to the RX3000 manual for instructions on wire
routing and environmental sealing of wires.
b. Press the Select button to view the smart sensors on the LCD and then press
the Search button (the magnifying glass icon should be visible as in the
example below). The station will search for all connected smart sensors and
show the number of channels after a few seconds. Note that some smart sensors
have multiple channels (such as the temperature/RH sensors). 7. Add any wireless sensor motes.
Important: Keep the mote(s) near the RX3000 station while completing
these steps.
a. Press select to switch to the module where the RXW Manager is installed
(Module 1 or Module 2) and then press the Search button to wait for motes to
join the network. b. Install the rechargeable batteries
in the mote and press the button on the mote for 3 seconds.
c. Watch the mote LCD during the process of joining the network.
|
---|---
This signal strength icon blinks while searching for a network.| Once a
network is found, the icon will stop flashing and the bars will cycle from
left to right.
|
This network connection “x” icon blinks while the mote completes the
registration process, which may take up to five minutes.| Once the mote has
finished joining the network, the “x” icon is removed and the channel count on
the station LCD increases by the number of measurement channels for the mote
plus the battery.
Repeat steps to add other motes. Press the Search button on the station when
finished adding motes.
8. Connect water level sensors, analog sensors, or relay devices.
Note: You may wait to connect these until you are at the deployment site.
Power down the station (unplug any charging device and then disconnect the
battery). Connect any sensors or devices to the optional modules. For a water
level sensor, feed the cable through the top cable access opening. Slide the
connector at the end of the cable into the receptacle on the module. Insert
the cable jack on the other end into the water level sensor and screw on the
locking nut. See RX3000 station manual for details. Power the station back up
after connecting sensors/devices.
9. Connect to HOBOlink.
Use the Select button to return to the main LCD screen that shows all sensors
and modules. Press the Connect button (the cloud icon should be visible as
shown in the example). This is necessary for HOBOlink to identify newly added
sensors. 10. Configure the station in HOBOlink.
Go to Devices > RX Devices. Click the icon next to your station. Make changes
and click Save as you configure:
- System settings and connection interval.
- Smart sensor logging interval and sampling interval.
- Settings for each sensor (e.g., labels, graph, and scaling).
- Optional modules and associated channels.
Any configuration changes will be transferred to the station the next time it
connects with HOBOlink.
11. Start logging.
Press the Start button on the station to start logging. For details on using
HOBOlink including checking the latest conditions, creating dashboards, or
setting up a map,
go to www.hobolink.com and click Help.
For additional information: See the HOBO RX3000 Remote Monitoring Station Manual for specifications and detailed operation, including deployment guidelines, mounting instructions, and steps for installing the weatherproof rubber cable channels. Go to www.onsetcomp.com/manual/rx3000 or scan the code at the left.
http://www.onsetcomp.com/manual/rx3000
© 2015–2022 Onset Computer Corporation. All rights reserved. Onset, HOBO, and
HOBOlink are trademarks or registered trademarks of Onset Computer
Corporation. All other trademarks are the property of their respective
companies.
This product has been manufactured by Onset Computer Corporation and in
compliance with Onset’s ISO 9001:2015 Quality Management System.
1-800-LOGGERS (564-4377)
508-759-9500
www.onsetcomp.com/support/contact
18254-K MAN-QSG-RX3000
References
- HOBOlink
- Onset HOBO and InTemp Data Loggers
- HOBO RX3000 Remote Monitoring Station (RX3001-00-01, RX3002-00-01, RX3003-00-01, RX3004-00-01) User's Manual - 18255-RX3000-Manual | Onset Data Loggers
- Technical Support | Onset's HOBO and InTemp Data Loggers
- HOBO RX3000 Remote Monitoring Station User's Manual | Onset's HOBO and InTemp Data Loggers
- Product Manuals | Onset HOBO Data Logger
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>