Melexis MLX90614 thermometer Evaluation Board User Manual
- June 5, 2024
- Melexis
Table of Contents
Microelectronic Integrated Systems
MLX90614 Single and Dual zone Infra-Red thermometer Evaluation Board
EVB90614
User Manual
Introduction
The EVB90614 is designed to support MLX90614 infrared thermometer modules.
The communication between the PC and the evaluation board is accomplished by
USB.
The Demonstration Kit contains the following items:
- FS USB demonstration board, pre-programmed with USB bootloader and demonstration firmware.
- A standard USB cable for use in communicating with the board.
- 1 pc. MLX90614AAA (single zone 5V)
- 1 pc. MLX90614BAA (single zone 3V)
- CD-ROM, containing the USB HID driver and EVB90614 software
The EVB90614 is designed to allow OEMs to configure the MLX 90614 IR
thermometer for virtually any application quickly. OEMs can quickly experiment
with temperature ranges, optics, etc. to find the best IR configuration to
meet the application needs. Once the best configuration is established, low-
volume OEMs can easily configure IR modules for their own use. For high-volume
OEMs, Melexis can supply specially configured modules, ready to install into
the customer’s application. Please contact IR Sales at Melexis for quotations.
The EVB90614 allows engineers to customize the MLX90614 for high-accuracy IR
measurements in an R&D or Laboratory setting.
Host computer requirements
To communicate with and program the EVB90614 evaluation board, the following hardware and software requirements must be met:
- PC-compatible system
- An available USB port
- CD-ROM drive (for use with the accompanying CD)
- Microsoft Windows 98, Second Edition (98SE), Windows 2000 Desktop or Windows XP
Note 1: EVB90614 is HID USB device and will require no drivers on PCs
that support HID class USB devices (except Microsoft Windows 98, Second
Edition (98SE)).
Check www.melexis.com for the most recent
release of the software.
Installing the software
As a USB device, the demonstration board requires very little effort to
install.
Most of the work is done by the operating system. The software is installed
simply by running the file “setup.exe”.
The software will be installed in a separate Melexis directory and can be
accessed from Start/Programs/Melexis/ Mlx90614 Configurator.
Note 2: If your version of Windows does not support the HID USB devices
you might need to install USB driver (check
www.melexis.com for such a driver).
**Getting started with the MLX90614 evaluation board
**
4.1 MLX90614 evaluation board overview
The top view of the evaluation board is shown in Fig.4.1.The main elements include:
-
Extension connector (70553-0003 from www.molex.com) allowing additional MLX90614 modules to be connected to SMBus.
Connector pin description:
pin1 – SCL/Vz (pin 1 is marked on the plastic body of the connector)
pin2 – PWM/SDA
pin3 – Vdd
pin4 – Vss -
ZIF socket for MLX90614
-
Button “next”
-
Button “start”
-
Button “reset”
-
RGB status LED
-
USB “B” Receptacles
The EVB90614 board receives its power supply only from the USB cable (Bus- Powered Device). An external power supply source is not needed.
4.2 Connecting the EVB90614 evaluation board
To connect the evaluation board:
- Unbox and unwrap the board, and set it on a non-conductive surface near the host system.
- Connect the USB cable (supplied in the kit) to an open USB port on the host system or a USB hub connected to the host system, and to the USB connector on the board. The LED will shine in BLUE.
- The EVB90614 evaluation board is an HID-compliant device so a special USB device driver is not needed (except when using Window 98 & 98SE).
- Check the board connection. This can be done from Start/Settings/Control Panel/ System/Hardware/Device Manager (Fig. 4.2).
Getting started with MLX90614 Configurator
5.1 MLX90614 Configurator overview
The MLX90614 Configurator is a Windows-based software application designed to
be used with the EVB90614 evaluation board for evaluating Melexis MLX90614
infrared thermometers. Using this software, the features and performance of
MLX90614 infrared thermometers can be evaluated. All basic configurations of
MLX90614 are also supported in a user-friendly format. Configuration is
written in the EEPROM of each MLX90614 and will be the POR default afterward.
This way the MLX90614 can be further used without EVB in target applications.
5.2 Working with MLX90614 Configurator
Insert an MLX90614 thermometer in the ZIF socket (see Fig. 4.1). Make sure
that the module benchmark coincides with the mark on the PCB.
The software can be launched from Start/Programs/Melexis/ MLX90614
Configurator.
The Configurator automatically recognizes the module type, address, and
voltage supply (Fig. 5.1). The status LED is shining GREEN.
If there is no thermometer module inserted on the EVB or the module is not an MLX90614, the user will be notified (Fig. 5.2). Click the button OK. The main panel looks as shown in Fig. 5.3.
IF a module is put incorrectly the user will be notified (Fig. 5.4).
Note that the buttons Power off and Refresh are alternating.
The MLX90614 Configurator gives the opportunity to connect a few MLX90614
modules in a network (Fig.5.5). Note that if in a network both 3V
(MLX90614Bxx) and 5V (MLX90614Axx) modules are present, currently, only the 3V
modules will be available.
Note 3: When more than one module is connected on the bus an error
message (see Fig. 5.4) will be shown.
Note that this doesn’t mean mandatory that there is an incorrectly put module
but the user is encouraged carefully to check the modules before clicking the
button Yes.
Note 4: If in an SMBus network there is a module with address zero it
will be not reported. A module with SMBus address zero will be reported if it
is the only one connected to the board. Also, modules with the same SMBus
addresses connected to the network will cause conflict on the bus when
operation reading is done.
Note 5: It is strongly recommended to save the factory default EEPROM of
every new MLX90614 before doing anything else with it (See below for how to do
that).
Note 6: schematics for PCBs to attach multiple modules using connector 1
can be found on page 20.
Main panel description
| Turn off the module power supply.
---|---
| (Re-)start the system to refresh information about the modules.
| Opens the “Console Utility” window for the manual sending of commands.
See Appendix A for commands set.
| Opens “Configuration uploader” window for quick replication of MLX90614
infrared modules’ configuration.
| Opens the “Advanced” window from which the user can upgrade EVB firmware,
change the measuring unit of the temperature scale and relocate the log file.
| Opens current EVB90614 User Manual. It is recommended to check
www.melexis.com for updates and related
documents.
| Opens the “Configuration utility” window used for configuring MLX90614
infrared thermometers.
| Opens the “Measure utility” window for temperature measurements.
| Dual-zone module identification picture.
| Single zone module identification picture.
| Cyclop module identification picture.
| Module type and SMBus slave address identification fields.
| Show which EVB is active (more than one EVB can be connected to one PC).
5.3 Configuration of MLX90614 modules
From the main panel push the button Config. ”Configuration utility” window is
opened (Fig. 5.6).
From this window, the user can configure the MLX90614 module. A short description of every setting field follows below. For a detailed description of the settings please refer to the MLX90614 datasheet. These settings are stored in MLX90614 EEPROM and will be the power-up default configuration.
-
To the max, To min – defines the temperature ranges for object temperature in PWM mode.
-
Ta max, Ta min – defines temperature ranges for ambient temperature in PWM mode.
-
PWM/SMBus/Relay – chooses between SMBus, PWM, and Thermo Relay mode.
-
PWM mode – defines single or extended PWM mode
-
PWM struct – selects what data will be transmitted via the PWM.
-
PWM period – defines the PWM period. A list of available settings as a drop-down menu.
-
PWM period repetitions – defines how many times each PWM period will be repeated.
Any even number of repetitions from 0 to 62 is valid. -
PWM pin configuration – chooses between NMOS “open drain” and “push-pull” PWM output.
-
Sensor zone – defines the number of IR sensors. This number is factory programmed and is recommended not to be altered.
-
SMBus address – set SMBus slave address. All values between 0 and 127 are valid. All MLX90614 will accept communication with slave address 0. Therefore in the SMBus network, a slave address 0 will be useless.
-
F.I.R filter- drop-down list of settling time available via embedded in MLX90614 digital filter. Slower settling comes with less noise.
-
I.I.R filter [%] – drop-down list of available values for impulse limit. 100% passes all impulses, and 50% cuts impulse magnitude in half via embedded in MLX90614 IIR filter. Note that the settling time depends on both FIR and IIR filter settings.
Some options are available from the drop-down menu while others must be typed
by the user. For example to change the object temperature range double click
on the To min or To max field and type the new range.
To change between PWM and SMBus mode double click on PWM/SMBus/Relay field and
select from the drop-down menu the desired option (Fig.5.8).
Note that when the Thermo relay option has been selected the fields To max and
To min change in Hysteresis and Threshold respectively (Fig.5.7). Hysteresis
adds to the threshold on each side. Trip points will be Threshold+Hysteresis
for rising temperature and Threshold-Hysteresis for decreasing temperature.
Tick to view the full EEPROM memory in hexadecimal view (Fig.5.9).
Note that only highlighted cells are accessible by the user (see MLX90614
datasheet for more information and Note 7). All other EEPROM cells contain
factory calibration values and are protected against erasing and writing.
Note 7: Bits <3>, <13:11> and <15> of ConfigRegister1 (EPROM address 0x05)
can’t be changed from this window because contains a specific information.
They can be changed only using the command we (see 5.4 Manual sending
commands)
Note 8: Do not write SMBus address 0x00 in any module when you make a sensor
network because reading this module will cause all devices on the network to
reply. This will result in confusion on the SMBus. Click the button Read to
read the complete EEPROM memory. Starting the SW or clicking “Refresh” after
“Power Off” will automatically read the EEPROM, too.
Click the button Write to write changed EEPROM cells. Factory calibration
values will not be altered. Writing EEPROM cell is automatically preceded by
erasure.
Click the button Save file to save EEPROM in a bin file. Refer to Fig.5.10.
Click the button Load file to load a bin file. If the loaded bin file is to be
written into MLX90614
EEPROM a “Write” must be clicked afterward. Refer to Fig.5.11.
5.4 Manual sending commands
From the main panel push the button Console. ”Console utility” window is
opened (Fig.5.12).
Type a valid command in the Command field, push Execute (or button ‘Enter’ on the keyboard). The result will be reflected in the Information field. The supported commands are described in APPENDIX A. The console utility is provided for extended functionality. All read/configure/write as well as measure/log data operations can be done without this utility.
5.5 Demonstration of SMBus and PWM temperature measuring
From the main panel push the button Measure. The ”Measure utility” window is
opened (Fig.5.13).
Note 9: If the used module is a single zone only two thermometers will be present on the screen (Fig.5.14).
Push the button Start to start temperature measurement. With the MLX90614 configured in PWM mode, the PWM output depends on the ranges written in EEPROM. These ranges are automatically read and filled in the Min/Max boxes on this screen. An example is shown in Fig.5.15.
In the field “Measure period” the user can make the measuring faster or
slower.
Max and Min’s fields permit users to change graphic scales.
When the measurement is started and the Do log field is ticked data will be
acquired in a log.csv file. If the SMBus mode is selected the following RAM
registers will be logged in the log.csv file (See MLX90614 datasheet for more
information):
- Ambient sensor data – SMBus RAM address=0x03h
- IR sensor 1 data – SMBus RAM address= 0x04h
- IR sensor 2 data – SMBus RAM address= 0x05h
- Linearized ambient temperature Ta – SMBus RAM address =0x06h
- Linearized object temperature (IR1) TOBJ1 – SMBus RAM address= 0x07h
- Linearized object temperature (IR2) TOBJ2 – SMBus RAM address = 0x08h
If PWM mode is selected logged data will depend on the chosen PWM structure.
Every time a measurement is started the log file is continued, with the entire
EEPROM saved as a new heading row.
Note 10: In Regional and Language Options (available from
Start/Settings/Control panel) list separator needs to be “,” and decimal
symbol – “.” in order to get the CSV log file opened in an easy-to-use form in
EXCEL.
Note 11: Note that the CSV log file is the same for all measurements,
there is not a separate file for every module that is measured.
To change the temperature measuring unit (°C, °F, or K) and relocate the
log.csv file:
- Click the button Advanced from the main panel
- Click the button Options are available in the “Advanced” window.
- From “Software settings” window choose the desired settings (Fig. 5.16).
The Thermo Relay option can be simulated with MLX90614 Configurator. For
example, let’s choose the Thermo Relay option in the “Config” window and set
the following parameters:
Threshold=27ºC, Hysteresis=1ºC. Let open the “Measure utility” window (Fig.
5.17) and press the button Start. The Thermo Relay output should be in logical
“0” if the object temperature is below 26ºC how as shown in Fig. 5.17. When
the object temperature becomes above 28ºC the Thermo Relay output is switching
over in logical “1” (Fig. 5.18). Min and Max’s fields show the min and max
temperature threshold by which the Thermo Relay output will switch over in
logical “0” and “1” respectively. Note that the module is indeed in SMBus mode
by SMBus Request command and the shown below pictures shows how the real
Thermo Relay output should keep itself.
To use EVB90614 with modules with fixed PWM output (no SMBus access is
available) the following option can be used:
1. Click the button Advanced from the main panel
4. Click the button PWM-M available in the “Advanced” window.
The window “PWM detached measurement module” is opened (Fig. 5.19).
All settings must be entered manually by the user. In the Slowdown field, the user can write values between 0 and 100. A bigger value in this field will make the thermometer readings slower and easier.
5.6 Quick customization of MLX90614 thermometers
Push the button Uploader on the main panel. The “Configuration uploader”
window is opened (Fig. 5.20).
This window is very similar to the “Configuration utility” window but here the
user cannot change the EEPROM.
Click the button “Select EEPROM” to open the window “Open EEPROM file” (See
Fig.
5.11). Browse to select the desired EEPROM configuration file and click the
button “Open”.
The chosen EEPROM configuration is loaded (Fig. 5.21).
Tick to view the full EEPROM memory in hexadecimal view if desired.
The user has two choices to load an EEPROM configuration in an MLX90614 module:
- Upload
- Quick upload
When the Upload button has been pushed the software automatically checks the
module power supply before writing the EEPROM memory of the module. This
process takes some time. This option is useful when the user wants to
configure MLX90614 modules of different versions. It is recommended to use
this option as it provides additional safety (a 3V module would not be
subjected to overvoltage stress unintentionally).
When long series of MLX90614 modules need to be configured with the same
configuration the Quick upload option might be preferred. In this case, the
user must initially choose the module voltage supply manually in the field
below the button Quick upload (See Fig. 5.18 or Fig. 5.19) and push the button
Quick upload. In this case, the software does not make module power supply
verification so the time of module configuration is reduced. Tick the SMBus
increment field if you want the SMBus address field to be incremented
automatically.
Note that in both cases the users cannot enter the SMBus device address. And
indeed this is not necessary because the software automatically uses address
0x00 which is the universal address for all MLX90614 devices. The users must
realize also that when using the “Configuration uploader” window only one
MLX90614 device should be present or all devices in the network are changed.
If you work in a network and want to change the configuration of a specific
module in the network use the” Configuration utility” window (see 5.3
Configuration of MLX90614 modules).
5.7 Updating EVB90614 firmware
For possible future upgrades of the EVB FW, a bootloader is programmed into
the EVB MCU. This makes it possible for the user to load updates into the EVB
using the same
software.
This can be done in two ways:
Updating EVB90614 firmware – method 1
- Start MLX90614 Configurator
- Push the button Advanced from the main panel. “Advanced” window is opened (Fig.5.22).
- Push the button Upload HEX. A warning message appears (Fig. 5.23).
- Click Yes and in a new browser window to find the firmware file will appear, select the correct file and click the button Open (Fig. 5.24).
Wait until a message box appears (Fig. 5.25).
Close this window. The EVB90614 board is updated and ready for work.
If an incompatible file is loaded warning messages will appear.
Follow these messages to load a compatible file.
Updating EVB90614 firmware – method 2
-
Close MLX90614 Configurator
-
Press and hold the button to start on the board
-
Press and release the button reset on the board
-
Release the button and start
After these steps, the board is manually entered into bootloader mode. The LED is turned off. -
Start MLX90614 Configurator. The following message appears (Fig. 5.25).
-
Click OK and follow step 4 in method 1
5.8 Automatic emissivity compensation
This option allows automatic emissivity compensation in an MLX90614 when an
object with emissivity lower than 1 is measured. Most materials have an
emissivity close to 1 and a correction is not needed.
This routine can also be used to compensate for IR windows or IR lenses one
wants to place between the object and the sensor in order to protect the
sensor or make changes in the FOV. This can only be successful if one takes
care to keep the window or lens at the same temperature as the sensor. The
emissivity compensation will be active for all ambient and object temperatures
measurable by the thermometer but will have the highest accuracy at the object
and ambient temperatures used during the “calibration”…
To determine the emissivity it is necessary to know the object’s temperature
accurately. This can be done by either using a contact thermometer or by
painting part of the object black with a thin layer of high-temperature paint.
The real temperature can then be measured with the MLX90614 pointing at the
painted area and the emissivity can be determined with the MLX90614 pointing
at the unpainted area.
It is assumed that the MLX90614 sensor has the same temperature as the area
surrounding the object and sensor. The calculation of the emissivity
correction is highly sensitive to this temperature. The EVB can heat up the
sensor 1-2degrees. For this specific measurement, it is advised to mount the
sensor away from the EVB and make the SMBus interface using connector 1) on
the EVB (see figure 4.1).
The emissivity coefficient is calculated using the formula:
Where:
Tomas – the object temperature measured from MLX90614
Ta – the ambient temperature measured from MLX90614
To read – the real object temperature
The automatic emissivity correction option is available from Advanced
->Emissivity.
The “K-Emissivity correction” window appears (Fig. 5.27) inviting the user to
enter the real temperature Toreal of the object. Push the OK button to start
the measurement of Tomas and Ta and the calculation of Ke. The sensor must be
looking at the object at that moment.
After the measurement finishes the following window appears:
Choose one of the proposed options.
If the temperature difference between Ta and Toreal is lower than 5 °C a
warning appears (Fig. 5.29).
Troubleshooting
If the USB communication with EVB90614 fails or the MLX90614 Configurator hangs, close the program, push the button reset on the board (the LED must shine BLUE), and launch the program again.
Disclaimer
Devices sold by Melexis are covered by the warranty and patent indemnification
provisions appearing in its Term of Sale. Melexis makes no warranty, express,
statutory, implied, or by description regarding the information set forth
herein or regarding the freedom of the described devices from patent
infringement. Melexis reserves the right to change specifications and prices
at any time and without notice. Therefore, prior to designing this product
into a system, it is necessary to check with Melexis for current information.
This product is intended for use in normal commercial applications.
Applications requiring extended temperature range, unusual environmental
requirements, or high-reliability applications, such as military, medical
life-support, or life-sustaining equipment are specifically not recommended
without additional processing by Melexis for each application.
The information furnished by Melexis is believed to be correct and accurate.
However, Melexis shall not be liable to the recipient or any third party for
any damages, including but not limited to personal injury, property damage,
loss of profits, loss of use, interrupt of business or indirect, special
incidental or consequential damages, of any kind, in connection with or
arising out of the furnishing, performance or use of the technical data
herein. No obligation or liability to the recipient or any third party shall
arise or flow out of Melexis’ rendering of technical or other services.
© 2006 Melexis NV. All rights reserved.
For the latest version of this document, go to our website at www.melexis.com
Or for additional information contact Melexis Direct:
Europe, Africa, Asia:
Phone: +32 1367 0495
E-mail: sales_europe@melexis.com| America:
Phone: +1 603 223 2362
E-mail: sales_usa@melexis.com
---|---
ISO/TS 16949 and ISO14001 Certified
♦APPENDIX A – EVB90614 commands set
-
Read MLX90614 RAM address
Syntax: RR address
Operands: 0(0x0) ≤ address ≤31(0x1F) -
Read MLX90614 EEPROM address
Syntax: re address
Operands: 0(0x0) ≤ address ≤31(0x1F) -
Write MLX90614 EEPROM address
Syntax: we address value
Operands: 0(0x0) ≤ address ≤31(0x1F)
0(0x0) ≤ value ≤65535(0xFFFF) -
Restart MLX90614 (turn off-turn on module power supply)
Syntax: rt
Operands: no operands -
Send Request (switch to SMBus mode)
Syntax: RQ
Operands: no operands -
Capture PWM
Syntax: cp
Operands: no operands
♦ APPENDIX B – EVB90614 schematics
Note that the components marked “Optional” are not populated. They are assumed
to give the user the option to develop his own applications with the board.
Note: If the user decides to add their own firmware using EVB90614 he must use program memory above 0x4000 (see PIC18F4550 datasheet). Also, EEPROM memory addresses between 0x40 and 0x43 must be maintained at 0x00 or 0xFF otherwise the EVB90614 will not be recognized by the host when it is in bootloader mode. For the latest information about this opportunity check www.melexis.com.