IRTECH C16 C Series Pocket Thermal Imaging Camera User Manual
- June 5, 2024
- IRTECH
Table of Contents
C-series pocket thermal camera
User manual
! Warnings, cautions, and notes
Definition
! Warning Represents a dangerous situation or act that may result in
personal injury or death.
! Caution Represents a situation or action that could result in damage to the
camera or permanent loss of data.
! NOTE Indicates a useful hint for the user.
Important information – Please read before using the instrument
! Caution – As the camera uses a very sensitive thermal sensor, the lens
must not be pointed directly at a strong source of radiation (e.g. sun, direct
laser beam, or reflection) under any circumstances (on or off) as this will
cause permanent damage to the camera!
! Caution – The camera must be transported in its original box and must
not be shaken or bumped strongly during use or transport.
! Caution –It is recommended that the camera be stored in its original
case in a cool, dry, ventilated environment free from strong electromagnetic
fields.
! Caution – Avoid staining the surface of the lens with oil and various
chemicals and damaging the surface. Close the lens cap after use.
! Caution – To prevent the potential risk of data loss, always copy (back
up) data to your computer
! NOTE – The camera may require a 3-5 minute warm-up process before
accurate readings can be taken.
! NOTE – Every thermal imaging camera is factory calibrated for
temperature and it is recommended that this be done annually.
! Caution – Do not open the case or make modifications without
permission, repairs should only be carried out by authorized personnel
Introduction
Thank you for choosing the C16/C25 series pocket Thermal Imaging Camera from
Zhejiang Dali Technology Co., Ltd.
Main features
Measurement of maximum temperature, minimum temperature, and multiple movable
points
Multiple movable area temperature measurement
Arbitrary line temperature measurement 6 color codes to choose
Infrared and visible image fusion
Temperature setting
Focus free
3.5-inch color LCD touch screen
Adjustable LCD brightness
LED fill light
WiFi wireless network
Gravity sensor
Enhanced Mode
Multiple language options
Applications:
Preventive maintenance
-
Electric power industry: thermal state inspection of transmission lines and power equipment, fault diagnosis.
-
The electrical system Identifies circuit overloads in advance before they happen.
-
Mechanical system: reduce downtime and prevent failure.
Building science -
Roof: Detect water leakage quickly and efficiently.
-
Building structure: Infrared energy assessment survey for commercial buildings and residential buildings.
-
Moisture detection: finds the root of dampness and mildew.
-
Remediation: Evaluate remedies and make sure the area is completely dry.
Other applications -
Iron and steel industry: monitoring the process of steel-making and rolling, diagnosing the damage to the hot blast stove, detecting the temperature of sheet and embryo.
-
Fire protection: forest fire prevention and potential fire source search, special materials spontaneous combustion prevention and testing, electrical fire safety testing.
-
Medical: human body surface temperature detection and temperature field distribution analysis.
-
Petrochemical industry: oil pipeline status inspection, material interface inspection, heat leakage, and insulation structure, power equipment status inspection, etc.
1.1 Standard configuration
- Pocket thermal camera
- TYPE-C data cable
- Packaging box
- Certificat
Thermal camera introduction
2.1 Key function
[1] power key
Used for thermal imager on and off, standby, and wake up. Press this switch
for more than three seconds to start/close the thermal imager; Short press to
enter standby/wake up (any other key can also wake up).
! Note: After power off, it is recommended to wait at least ten seconds
before starting up again to ensure the safety of the thermal imager
2.2 Interface
[3] USB
For data transfer and charging
[4] Indicator
For charging indication
Basic operation
3.1 The battery cannot be replaced
3.2 Quick Start
3.2.1 Obtaining thermal images
- Long press the power key (more than 3 seconds) until the power light turns green. About 10 seconds later, the device is initialized and enters the working state.
- The thermal imager is free of focusing and can make the target image clear without focusing operation.
3.2.2 Temperature measurement
- Move the cross temperature measuring point on the screen to the target object and the “XX” displayed in the upper left corner of the screen is the temperature of the target.
- When the target temperature is greater or less than the upper or lower limit of the temperature measured by the thermal imager, the screen temperature will display > XXX℃ or < XXX℃ to prompt the user.
Menu for setting temperature measurement area properties Click to select the temperature measuring point, temperature measuring area, and temperature measuring line and drag the whole temperature measuring object. You can drag the angle of the temperature measurement area by holding down any point of the four corners of the area to change the size of the area. Hold down a section of the temperature line for an extension.
3.2.3 Image storage
To store the current image, you can choose the following methods.
-
Click the button on the screen, and then click to take photos for storage .
3.2.4 Image playback -
Click the playback button to pop up the image library and select the folder to play back the image
3.2.5 Image export
- The USB data cable can be used to connect the computer to operate the stored data in the thermal imager, including image export, deletion, formatting, and other operations
3.2.6 LCD settings
- You can adjust the screen brightness by exhaling the drop-down menu and sliding the brightness.
3.2.7 Menu Operation Description
Touch screen operation:
- Click the icon at the bottom right of the touch screen and click each icon to perform corresponding operations. Enter the setting menu, click the menu item to set, and enter the sub-menu of the next level, the last level menu for function switch, parameter setting, option selection, and other operations; Click the arrow before the interface name to return to the previous menu.
- Click the touch screen drop-down shortcut menu to switch on and off WiFi and LED lights. Click the clear screen icon to clear the main interface UI, lock the screen and adjust the brightness
Operation guide
4.1 Operation Interface Description
4.1.1 User interface of working stat
The user interface of the working state of the thermal imager
[1] Temperature unit: A unit of temperature
[2] Upper limit temperature: upper limit temperature value of the color label.
[3] Temperature color code: the color bar of the current image display
[4] Lower limit temperature: lower limit temperature value of the color label.
[5] Temperature measurement object: you can choose point temperature
measurement, line temperature measurement, and region temperature measurement.
[6] Menu buttons: the menu includes Settings, mode selection, color code
selection, temperature measurement object selection, etc
[7] Storage button: it can take photos and store them.
[8] Playback button: You can view stored photos.
4.1.2 Drop-down menu
Drop-down menu interface
The drop-down menu interface: Click the corresponding icon to turn on/off
WIFI, LED fill light, clear screen, and lock screen. Adjust screen brightness
by sliding.
4.1.3 Main menu
Main menu interface
Main menu interface: The ICONS of the main menu on the right are image playback, photo saving, and menu button in sequence
4.2 Temperature Measurement Mode
Add or delete a temperature measurement object, including a temperature
measurement point, temperature measurement area, and temperature measurement
line.
Tap the touch screen. Click the temperature measurement mode icon on the touch
screen, and the following interface appears
Temperature measurement mode interface
Click the ICONS of temperature measurement objects such as temperature measurement point, temperature measurement area, and temperature measurement line on the touch screen to add temperature measurement objects such as temperature measurement point, temperature measurement area, and temperature measurement line on the screen. Click the delete icon on the far right to clear all the added temperature objects
4.3 Color code-switching
This function is used to select the color code of the image display. There are
six color codes to choose from.
Click the touch screen to pop up the main menu. Click the color label on the
touch screen to pop up the following interface
Color code-switching interface
According to the display requirements of the target scene, directly click the corresponding color bar on the touch screen to switch to the selected color bar
4.4 Image adjustment
This function is used for manual/automatic mode selection of image adjustment.
Click on the touch screen to pop up the main menu. Click on the touch screen
to pop up the ICONS of automatic /manual mode as shown below. Click the mode
icon to select, and switch to the selected mode.
To adjust the mode manually, click to select the upper and lower temperature
values of the color label and slide your finger on the right side of the color
label to adjust the upper and lower temperature values of the color label
separately or simultaneously. Click the selected temperature value again to
deselect it
Image adjustment interface
4.5 Image Storage
This feature is used to save images in JPG format.
Tap the touch screen and tap the photo icon . Click on the touch screen to
save the current image directly
Photo interface
4.6 Image playback
This function is used to browse and playback the saved image file.
Click the playback icon on the touch screen to pop up the picture library
interface, as shown below. Click the corresponding folder to open the folder,
and click the icon to delete the folder.
Photo gallery interface
4.7 System Settings
This function is used to set various parameters of the thermal imager,
including temperature measurement parameter setting, temperature measurement
function setting, network setting, storage setting, alarm setting, device
setting, etc.
Click the touch screen to pop up the main menu. Click the setting icon ****on
the touch screen to pop up the system setting menu interface, as shown below.
System Setting interface
4.7.1 Temperature Measurement Parameters Settings
In the temperature measurement parameter setting menu, there are temperature
measurement gear settings, specific emissivity settings, ambient temperature
settings, background temperature settings, humidity settings, temperature
measurement distance settings, and temperature correction settings.
Temperature measurement parameter setting menu
Temperature measurement gear setting : The temperature gear can be
switched.
Radiation coefficient setting: You can select the emissivity value of the
target through the touch screen.
Ambient temperature setting: You can select the ambient temperature by
touching the screen.
Background temperature setting: You can select the desired background
temperature value by touching the screen.
Humidity setting: You can select the relative humidity of the environment
through the touch screen.
Distance setting : You can select the distance of the target by touching
the screen.
Temperature correction: You can select the required correction
temperature value by touching the screen.
4.7.2 Temperature measurement settings
In the temperature measurement setting menu, there are automatic zero
adjustment intervals, automatic enhancement mode, and high and low-temperature
point settings for temperature measurement lines,s, and temperature
measurement areas. Temperature measurement setting menu
Auto-zero interval: You can select the time interval to be zero-adjusted
by touching the screen.
Auto Enhance Mode: On/Off can be selected via touch screen.
High and low-temperature points of the temperature measurement line: You
can set the display conditions of the high and low-temperature points.
High and low-temperature points in the temperature measurement area: You
can set the display conditions of the high and low-temperature points.
4.7.3 Network Settings
In the network settings menu, WIFI settings can be performed.
Network settings menu
WIFI setting : You can turn on/off WIFI through the touch screen. The default password of WIFI is 12345678, which can be connected through the mobile APP and realize remote monitoring and other functions
4.7.4 Storage settings
In the storage settings menu, there are image storage format settings, memory
card capacity display, and format settings.
Storage settings menu
Image storage settings: You can set a separate JPG format, or you can set
an associated visible light format.
Memory Card Capacity: Displays the available memory capacity.
Format: You can format the memory card.
4.7.5 Alarm Setting
In the alarm setting menu, there are temperature alarm modes, high and low- temperature alarm settings, alarm temperature settings, alarm linkage interval settings, and alarm buzzer settings.
Alarm Settings Menu
Temperature alarm mode: On/off can be selected by touch screen.
Alarm type: High /low-temperature alarm can be selected by touch screen.
Alarm temperature: The alarm temperature threshold can be selected
through the touch screen.
Alarm linkage interval: The alarm linkage interval can be selected by
touching the screen.
Alarm buzzer: The alarm buzzer on/off can be selected by touch screen.
4.7.6 Device Settings
In the device settings menu, there are language and time settings, brightness
settings, shutdown time settings, sleep time settings, screen rotation lock
settings, color code bar display settings, update and upgrade settings, and
restore factory settings.
Image settings menu
Language and time settings: you can choose the time system, you can
choose the date system, time settings, distance units, and temperature units.
Brightness setting: You can slide the screen to adjust the brightness.
Shutdown time setting: you can choose the shutdown time of 5 minutes, 30
minutes, 60 minutes, and never.
Sleep time setting: you can choose the sleep time of 5 minutes, 30
minutes, and never.
Screen rotation lock settings: Screen rotation lock can be selected to be
on/off.
Color code bar display settings : The color code bar display can be
selected on/off.
Update/upgrade settings: FPGA upgrade and ARM upgrade can be selected.
Restore factory settings: You can choose to confirm or cancel, the device
will restart after confirmation.
4.7.7 System Information
System information includes device model, device number, software version,
logical version, and soft-core version.
RFQ
If you encounter problems with the camera, use the table below for troubleshooting. If the problem cannot be solved, please disconnect the power and contact the maintenance department of our company.
fault phenomenon | Reasons and solutions |
---|---|
The thermal imager failed to start | • The battery runs out of power. |
→Charge the device.
The thermal imager automatically shuts down| • The battery runs out of power.
→Charge the device.
• The device is set to be powered off.
→Select “never” here.
The battery is running down too fast| • The ambient temperature is too low.
• The charging cable is damaged, and the device failed to be charged.
→Replace with a new data cable.
Thermal imager screen brightness dimmed| • Check the brightness setting of
the thermal imager screen.
- Adjust the camera screen brightness setting.
The thermal image turns black and white| • Black and white labels are selected.
→Select the normal color code.
Appendix A
The emissivity of common materials (for reference only)
Material name | Surface condition | Temperature(℃) | Emissivity(ε) |
---|---|---|---|
aluminum | non-oxidizing | 100 | 0.20 |
oxidation | 100 | 0.55 | |
brass | polished brown | 20 | 0.40 |
matt | 38 | 0.22 | |
oxidation | 100 | 0.61 | |
copper | severely oxidized | 20 | 0.78 |
iron | oxidation | 100 | 0.74 |
rust | 25 | 0.65 | |
cast iron | oxidized | 200 | 0.64 |
not oxidized | 100 | 0.21 | |
wrought iron | hairy | 25 | 0.94 |
polished | 38 | 0.28 | |
nickel | oxidized | 200 | 0.37 |
stainless steel | oxidation | 60 | 0.85 |
steel | oxidation at 800°C | 200 | 0.79 |
general brick | surface | 20 | 0.93 |
concrete | surface | 20 | 0.92 |
glass | polished flat | 20 | 0.94 |
paint | white | 100 | 0.92 |
natural black | 100 | 0.97 | |
carbon | smoke black | 25 | 0.95 |
candle smoke | 20 | 0.95 | |
graphite rough surface | 20 | 0.98 | |
paint | 16 colors average | 100 | 0.94 |
paper | white | 20 | 0.93 |
Sandy soil | surface | 20 | 0.90 |
wood | planed | 20 | 0.90 |
water | distilled water | 20 | 0.96 |
skin | humanity | 32 | 0.98 |
ceramics | thin | 21 | 0.90 |
thick | 21 | 0.93 |
FCC Warning
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions:
- This device may not cause harmful interference, and
- this device must accept any interference received, including interference that may cause undesired operation.
Any changes or modifications not expressly approved by the party responsible
for compliance could void the user’s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
The SAR limit adopted by the USA is 1.6 watts/kilogram (W/kg) averaged over
one gram of tissue. The highest SAR value reported to the Federal
Communications Commission (FCC) when it is tested for use in the body is 0.432
W/kg.
The device complies with the RF specifications when the device is used near
your eye or at a distance of 5 mm from your body. Ensure that the device
accessories such as a device case and a device holster are not composed of
metal components.