COREMORROW E53.C Series Piezo Controller
DECLARATION
This user manual is a integrated user manual of the E53.C series piezoelectric controller. Please read this user manual carefully before using this controller. Follow the instructions in the manual during use. If there is any problem, please contact us for technical support. If you do not follow this manual or disassemble and modify the product yourself, the company will not be liable for any consequences arising therefrom. Please read the following to avoid personal injury and to prevent damage to this product or any other product connected to it. In order to avoid possible hazards, this product can only be used within the specified range.
Notice
Do not touch any exposed ends of the product and its accessories. There is
high voltage inside. Do not open the case without permission. Do not connect
or disconnect input, output, or sensor cables with power on. Please keep
surface of E53.C clean and dry, don’t operate in humid or static environment.
After use, output voltage should be cleared to zero before turning off the
controller switch, such as switching the servo state to the open-loop state.
Danger
The piezoelectric power amplifier described in this manual is a high-voltage
device capable of outputting high currents, which can cause serious or even
fatal damage if not used properly. It is strongly recommended that you do not
touch any parts that connect to the high voltage output. Special Note: If you
connect it with other products in addition to our company, please follow the
general accident prevention procedures. Operating the high-voltage
amplification requires training professional operators.
Warning
If the voltage exceeds the PZT’s tolerable range, it will cause permanent
damage to the PZT. Before adding voltage to the PZT poles, it must be ensured
that the positive and negative poles of the PZT are connected correctly and
the operating voltage is within the allowable range of this PZT.
Cautious
E53.C housing should be installed on a horizontal surface in an area with a
3CM air flow area to prevent internal convection in the vertical direction.
Insufficient airflow can cause equipment to overheat or premature instrument
damage.
Security
Introduction
Please keep surface of E53.C clean and dry. Do not operate in the humid or
static environment. E53.C is used to drive capacitive loads (such as piezo
actuators). E53.C should not be used in user manuals of other products of the
same name. Pay special attention that E53.C cannot be used to drive resistive
or inductive loads. E53.C could be used for static and dynamic operating
applications.
Safety Instructions
E53.C is based on the national safety standard. Improper use may cause
personal injury or damage to the piezo controller. The operator is responsible
for the correct installation and operation of the piezo controller. Please
read the user manual in detail. Please eliminate any faults and potential
safety hazards caused by the faults. If the protective ground wire is not
connected or connected incorrectly, there will be a possibility of leakage. If
you touch the E53.C piezo controller, it may cause serious or even fatal
injuries. If the piezo controller housing is opened without permission,
touching the live parts may cause electric shock, resulting in serious or even
fatal injury or damage to the piezo controller. Only authorized professional
technicians with corresponding qualifications could open the piezo controller.
When opening E53.C series controller, please disconnect the power plug. Please
do not touch any internal parts when operating under bare conditions.
Notes
The contents in the user manual are all standard descriptions, and the
customized parameters are not explained in detail in this manual.
The latest user manual is available for download on CoreMorrow website. When
operating theE53.C, the user manual should be placed near the system for easy
reference in time. If the user manual is missing or damaged, please contact
CoreMorrow customer service department. Please timely add all the information
given in the manufacturer’s user manual, such as supplements or technical
descriptions. If your user manual is incomplete, it will miss a lot of
important information, cause serious or fatal injuries, and cause property
damage. Please read and understand the content in the user manual before
installing and operating the E53.C. Only professionals who are authorized to
meet the technical requirements could install, operate, maintain and clean the
E53.C.
Introduction
Features
- 1 channel small size
- 24V20~30V1.5A
- 36W Peak current
- 1A Ave current 60mA
- Unload bandwidth 10KHz
- Output short circuit protection
Applicaitons
Driving piezo actuators Driving piezo objective scanners
Order information
E53.C—-Open loop Analog/software input control
Accept customized according to requirements
12bit gain/-20 120V output voltagedefault
15bit gain/-20 150V output voltage
Driving Principle
Appearance
Power Calculation
- Average outputSine wave operation mode Pa Upp · Us · f· Cpiezo
- Pa=Average output[W]
- Cpiezo=Piezo actuator capacitance[F]
- Upp=Peak and peak drive voltage [V]
- f=Operating frequency of the sine wave[Hz]
- Us=Drive voltage[V]Vs+-Vs-
Panel Introduction
No. | Function | Description |
---|---|---|
① | Power indicator | Green, lights when power’s on |
② | USB port | MicroUSB port |
③ | RS-232/422 | See interface pin definition |
④ | Analog input | Analog voltage input interface |
⑤ | Piezo connector | Driving piezo actuaotr |
⑥ | Limit | Over-current indicator |
No. | Function | Description |
---|
1
|
Signal source control switch
| M:Signal source selection by communication
instruction
D:Digital circuit as signal source
A:Using an external analog signal as a signal
source
Notes and suggestions
- E53.C cannot be used to drive inductive loads. If the inductive loads are driven, the product may be damaged.
- If there is no need, please do not twist the potentiometer easily.
Contact us
Harbin Core Tomorrow Science & Technology Co., Ltd. Tel: +86-451-86268790 Email : info@coremorrow.com Website : www.coremorrow.com Address: Building I2, No.191 Xuefu Road, Nangang District, Harbin, HLJ, China
CoreMorrow Official and CTO WeChat are below:
Documents / Resources
|
COREMORROW E53.C Series Piezo
Controller
[pdf] User Manual
E53.C Series, Piezo Controller, E53.C Series Piezo Controller, Controller
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>