PeakTech 5115 Digital Thermometer User Manual
- June 13, 2024
- PeakTech
Table of Contents
5110/5115
Digital-Thermometer
Operation Manual
Safety Precautions
This product complies with the requirements of the following European
Community Directives: 2014/30/EU (Electromagnetic Compatibility) as amended by
2014/32/EU (CE-Marking).
To ensure safe operation of the equipment and eliminate the danger of serious
injury due to short-circuits (arcing), the following safety precautions must
be observed.
- Do not operate the meter before the cabinet has been closed and screwed safely.
- Check test leads and probes for faulty insulation or bare wires before connection to the equipment.
- Comply with the warning labels and other info on the equipment.
- Keep the equipment dry.
- CAUTION! Repeated sharp flexing can break the thermocouple leads. To prolong lead life, avoid sharp bends in the leads, especially near the connector.
- Do not subject the equipment to direct sunlight or extreme temperatures, humidity or dampness.
- Do not subject the equipment to shocks or strong vibrations.
- Do not operate the equipment near strong magnetic fields (motors, transformers etc.).
- Keep hot soldering irons or guns away from the equipment.
- To avoid electrical shock, do not use this instrument when voltages at the measurement surface exceed 24V AC or DC.
- To avoid damage or burns, do not make temperature measurement in microwaves ovens.
- Allow the equipment to stabilize at room temperature before taking up measurement (important for exact measurements).
- Replace the battery as soon as the battery indicator “BAT” appears. With a low battery, the meter might produce false reading that can lead to personal injury.
- Fetch out the battery when the meter will not be used for long period.
- Periodically wipe the cabinet with a damp cloth and mid detergent. Do not use abrasives or solvents.
- The meter is suitable for indoor use only
- Do not store the meter in a place of explosive, inflammable substances.
- Do not place the equipment face-down on any table or work bench to prevent damaging the controls at the front
- Do not modify the equipment in any way
- Opening the equipment and service- and repair work must only be performed by qualified service personnel
- Measuring instruments don’t belong to children hands.
Introduction
These compact and handy digital thermometers are best suited for temperature
measurements with Type K temperature sensors and have a very wide measuring
range of – 50 °C to 1300 °C.
The measured value can be toggled between ° C and °F at the press of a button,
and also in Kelvin with the PeakTech 5110 model.
The PeakTech 5115, on the other hand, has two input channels which can be
measured separately or can be used by mathematical function (T1-T2) to measure
a temperature difference.
The housing is made of sturdy plastic and secured with a rubber protective
cover against damage.
Specifications
Display| 3 ½-digit liquid crystal display (LCD) with maximum reading of 1999
and backlight
---|---
Overload- Display| LCD-Displays shows “1”
Temperature scale| °C, °F or K (only P 5110) user-selectable
Measurement range| -50° C….1300° C; -58° F….2000° F 223K … 2000 K (P 5110)
-50° C….1300° C; -58° F….2000° F (P 5115)
Resolution| 0,1°C / 1° C; 0.1°F / 1° F; 1 K (P 5110)
0,1°C / 1° C; 0.1°F / 1° F (P 5115)
Temperature Coefficient| 0.1 times the applicable accuracy specification per
°C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F)
Input protection| 24 V AC or 60 V DC maximum input voltage on any combination
of input pins
Reading rate| 2.5 times per second
Input Connector| Accepts standard miniature thermocouple connectors (flat
blades spaced 7,9 mm, center to center).
Operating Temperature| 0 … +50°C (32°F … 122°F) < 80 %
Storage Temperature| -20°C … +60°C (-4°F … 140°F) < 70 %
Battery| Standard 9 V battery (NEDA 1604, IEC 6F22, 006 P)
Battery Life| 200 hours typical with carbon zinc battery
Dimensions| 162 (H) x 76 (W) x 39 (D) mm
Weight| 210 g
Accessories| Holster, Battery, Type-K thermocouple (P 5115 2 St.)
3.1. Technische Daten
PeakTech® 5110:
Range | Accuracy |
---|---|
-50°C … 0°C | ± 2°C |
0°C … 1000°C
1000°C … 1300°C| ± 0,5% rdg. + 1°C
± 0,8% rdg. + 1°C
-58°F … +32°F| ± 4°F
32°F … 2000°F| ± 0,5% rdg. + 2°F
223K … 273K| ± 5 K
273K … 2000 K| ± 1,0% rdg.. + 2K
PeakTech® 5115:
Range | Accuracy |
---|---|
-50°C … 0°C | ± 2°C |
0°C … 1000°C
1000°C … 1300°C| ± 0,5% rdg. + 1°C
± 0,8% rdg. + 1°C
-58°F … +32°F| ± 4°F
32°F … 2000°F| ± 0,5% rdg. + 2°F
Accuracy is specified for operating temperatures of 18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple errors.
K-Type Probe | Accuracy |
---|---|
TF-50 (-50°C…250°C) | ± 0,5% rdg. |
TF-55 (-20°C…300°C) | ± 2,5% rdg. |
Operating Instructions
4.1. Selecting the temperature scale
Readings are displayed in either degrees Celsius (°C) or degrees Fahrenheit
(°F). When the thermometer is turned on, it is set to the temperature scale
that was in use when the thermometer was last turned off. To change the
temperature scale, press one of the following keys °C or °F (additional P 5110
is supplied with: K).
4.2. Selecting the display resolution
The thermometer allows two choices of resolution:
high resolution: 0.1°C or 0.1°F
low resolution: 1°C / 1°F (1 K: only P 5110)
To select the alternate display resolution, press the corresponding “0.1°/1°”
key (P 5115) / (P 5110: 0,1° or 1° key).
4.3. MAX Mode
Pressing the Max key to enter the MAX mode. The thermometer then records and
updates the maximum absolute value and the MAX annunciator appears on the
display. Pressing the MAX key again to exit the MAX recording mode.
For breading the auto. updating of the saved value (new maxvalue shouldn’t be
saved), press button “HOLD”. For returning to auto. updating during a
measuring of a new Max-value press “HOLD”-button again.
4.4 HOLD Mode
Pressing the HOLD key to enter the Data Hold mode, the “HOLD” annunciator is
displayed. When HOLD mode is selected, the thermometer held the present
readings and stops all further measurements.
Pressing the HOLD key again cancels HOLD mode, causing the thermometer to
resume taking measurements.
4.5. Backlight Mode
Pressing the Backlight key to turn on the LCD-backlighting Function. Pressing
the Backlight key again to turn off the LCDbacklighting function
4.6. Tc (Temperature Compensator) (only P 5110)
Checking Mode.
Pressing and hold the “Tc” key to enter the temperature compensator-checking
mode. The thermometer will display the temperature inside.
4.7. Preparation
- Ensure battery is in place, if LCD-display shows the battery symbol, the battery should be replaced.
- Make sure all function keys are set properly. (HOLD) switch is on an “OFF” position, make sure no “DATAHOLD” displayed on LCD.
- Inspect whether temperature probe has been placed securely in the temperature outlet, (the “+” outlet of temperature probe should be connected to the “+” of the temperature outlet, the “-” outlet of the temperature probe should be connected to the “-” of temperature outlet).
4.8. Measurement
Temperature measuring:
- Turn on the thermometer
- Connect the outlet of test probe “K-Type” to the outlet of temperature outlet.
- Measure environment temperature with temperature probe, the reading will appear on LCD.
4.9. Adjustment for T1-T2 measurement (only P 5115)
The choice of the input connectors with the correspond indicated buttons T1,
T2 or T1-T2 at the unit. To measure the differential temperature connect the
thermocouples to the inputs and press “T1-T2”. The indicated temperature value
is the difference between T1 and T2.
Remark:
-
For an accurate reading of T2, with liquid temperature probe (see Fig. 7)
-
When the superficial temperature is taking, the top of temperature probe should have a close contact with the surface to be measured (see Fig. 8).
4.10. Battery Replacement
Power is supplied by a 9 V “transistor” battery (NEDA 1604, IEC 6F22). The battery sign appears on the LCD when replacement is needed.
To replace the battery, remove the screw from the back of the meter and lift off the battery cover. Remove the battery from its contacts.
Caution! Batteries, which are used up dispose duly. Used up batteries are hazardous and must be given in the for this being supposed collective container.
4.11. Statutory Notification about the Battery Regulations
The delivery of many devices includes batteries, which for example serve to operate the remote control. There also could be batteries or accumulators built into the device itself. In connection with the sale of these batteries or accumulators, we are obliged under the Battery Regulations to notify our customers of the following:
Please dispose of old batteries at a council collection point or return them to a local shop at no cost. The disposal in domestic refuse is strictly forbidden according to the Battery Regulations. You can return used batteries obtained from us at no charge at the address on the last side in this manual or by posting with sufficient stamps.
Contaminated batteries shall be marked with a symbol consisting of a crossed- out refuse bin and the chemical symbol (Cd, Hg or Pb) of the heavy metal which is responsible for the classification as pollutant:- “Cd” means cadmium.
- “Hg” means mercury.
- “Pb” stands for lead.
All rights, also for translation, reprinting and copy of this manual or parts
are reserved.
Reproduction of all kinds (photocopy, microfilm or other) only by written
permission of the publisher.
This manual considers the latest technical knowing. Technical alterations
reserved.
We herewith confirm, that the units are calibrated by the factory according to
the specifications as per the technical specifications.
We recommend to calibrate the unit again, after 1 year.
© PeakTech® 11/2019/MP/HR/EHR
PeakTech Prüf- und Messtechnik GmbH
– Gerstenstieg 4 – DE-22926 Ahrensburg /Germany
+49 (0) 4102 97398-80 +49 (0) 4102 97398-99
info@peaktech.de
www.peaktech.de
Read User Manual Online (PDF format)