SellEton SL-720 Weighing Kit Instruction Manual
- June 1, 2024
- selLEton
Table of Contents
SellEton SL-720 Weighing Kit
Specifications
Feature | Specification |
---|---|
Maximum Capacity | 10,000 lbs |
Overload Protection | 200% |
FAQ
Can the SL-720 kit be used with any platform?
Yes, the kit allows you to use any platform of your choice, as long as it can support the maximum weight capacity.
What is the maximum capacity of the SL-720 weighing kit?
The maximum capacity is 10,000 lbs.
Can the weighing capacity of the kit be increased?
Yes, by adding more load cells or purchasing higher capacity kits.
Are customization options available for the SL-720 kit?
Yes, the kit can be customized with accessories such as printers, scoreboards, software, custom ramps, stands, and more.
OVERVIEW
- Our SL-720 floor scale weighing kit has everything you need to build your own weighing scale. It enables you to use any platform of your choice (platform not included with the kit). This weighing kit is mostly used for floor scale, livestock scale applications or to turn any object, or platform into a scale with Maximum capacity of 10,000 lbs.
- You can increase the kit’s weighing capacity by adding more load cells or buying higher-capacity kits. The kit can also be customized to Accessories such as printers, scoreboards, software, custom ramps, stands & much more! All load cells come with 200% overload protection.
ACCESSORIES
- x 4 load cells with 10′ cable
- x 4 load spacers ( installed in between the load cell & the mounting block )
- x 4 load cell Mounting blocks ( typically welded to steel structures ) or bolted to wood.
- x 8 Hexagon bolts
- x 8 washers
- x 8 lock washers
- x 1 Junction Box ( Plastic or Stainless steel )
- x 1 Quick Connect Cable ( you will hard wire it to the Junction Box, Refer to the wiring diagram within this Manual.
- x 1 Digital Weight Indicator ( Either in mild steel or Stainless steel LCD or LED, depending on your purchase.
OVERALL INSTALLATION
FOR WOODEN PLATFORMS
-
Step 1: Make sure the wooden platform you’ve built – or are in the process of building – can withstand the maximum weight you’ll put on it.
-
Step 2: Turn the platform upside down.
-
Step 3: Drill 2 holes ( Same size & spacing between the holes as the mounting block. Use the mounting block to draw out the holes first & then drill. Ultimately they need to match up to the load cells, be sure to take time to do it correctly. This can impact readings if load cells are not positioned correctly.
-
NOTE: At all four corners of the platform (x 2 holes should be drilled on every corner). Each hole should be drilled as close to its respective edge as possible. You need to purchase longer bolts for thicker platforms. Measure the mounting block & load cell along with the thickness of the platform & purchase the correct length of bolts.
-
So that the load cell, mounting block & platform are all connected firmly.
-
Step 4: Use the 1/16″ load spacers to bolt the load cells onto the platform. The spacers should be placed between the load cell & the mounting block. This is necessary to give the load cells adequate room for accurate readings.
( refers to the picture in this manual) -
Step 5: Take out the junction box. See the below diagram to understand where each of the 4 load cell cables will need to go into the junction box.
IF YOU’RE USING A STEEL PLATFORM
You will need to weld each mounting block to the corner of the platform. Only
weld on the outside of the mounting block & ensure that all blocks are
positioned in the same way, with the feet of the load cell located in the
corner of the platform (not facing inside the platform).
Picture below demonstrates the correct installation format
WARNING: Do not weld the mounting blocks to the platform with the load cell attached to it, it will damage the load cells.
Please refer to the diagram below for the indicator cable & load cell wiring:
CAPACITY & ACCURACY CHOICES
- 1000 lb Capacity x .2 lb Accuracy
- 2500 lb Capacity x .5 lb Accuracy
- 5000 lb Capacity x 1 lb Accuracy
- 10,000 lb Capacity x 1 lb Accuracy
CALIBRATION PROCEDURE
-
Turn on the scale by holding ON/OFF for 2 seconds.
-
Press HOLD and PRINT together to access the setup menu.
-
If done correctly, the display should now show C01.
-
Press PRINT to access the C1 channel. The display should show [C I #].
-
Press ZERO to choose which unit you want to calibrate in (1 = kg, 2 = lb).
-
Press PRINT to set the value. The display will now show C02.
-
Press PRINT to access the C2 channel. The display should show [C2 #].
-
Press ZERO to change the setting to the decimal places desired (The C2 channel is used to adjust the decimal point on the scale. A value of 1 means there is one digit behind the decimal point.)
-
Press PRINT to set the value. The display will now show C03.
-
Press PRINT to access the C3 channel. The display should show [C3 #].
-
Press ZERO to cycle through the values until the desired graduation appears. (The C3 channel adjusts the divisions on the scale. A value of 1 selected and C2 set to 1, the scale will read in 0.1 lb increments.)
-
Press PRINT to set the value. The display will now show CO4.
-
Press PRINT to access the C4 channel. The display will show [######].
-
Enter in the maximum capacity you want to use for this scale by using UNIT and COUNT to move the cursor left and right, and TARE and ZERO to move the values down and up. (The C4 channel is used to enter in the max capacity of the scale; Make sure this doesn’t exceed the max capacity of the scale; Max capacity divided by the increment set in C02 and C03 above cannot exceed 5000.)
-
Press PRINT to set the value. The display will now show C05.
-
Press PRINT to access the CS channel. The display should show [C5 0].
-
The CS channel calibrates zero on the scale. Make sure the scale is empty.
-
Press ZERO a to change the value to 1.
-
Press PRINT The display will count down from 10-1 while the scale is calibrating to zero. When the display shows O the zero calibration is complete.
SL-7510 offers 2 calibration methods, Single Point which uses one weight to calibrate and Linear Calibration, which uses multiple (2-7) weights for a more accurate calibration.
To Calibrate using only 1 calibration weight (Single Point Calibration) -
Press PRINT to continue. The display will now show C06.
-
Press PRINT to access the CO6 channel. The display will show [[6 0].
-
The C6 channel is used to calibrate the scale with a known weight. Press ZERO A to set the value of C6 to [C6 1]. Press PRINT . The display will flash SPRN, and then show [######].
TROUBLESHOOTING
Error Codes
For Additional support please reach out to tech@selletonscales.com or Call 844-735-5386.
Selleton Scales is a leading manufacturer of scales and wrapping solutions for commercial and industrial applications. We help businesses integrate scales and measuring devices into their business operations, ensuring high levels of reliability and improved productivity. Our mission is simple — to provide businesses with extremely accurate scales at an affordable price!