Fisher 88861023 Microplate Shaker Instruction Manual

June 14, 2024
Fisher

Microplate Shaker
Operations Manual

88861023 Microplate Shaker

Fisher 88861023 Microplate Shaker V1 12/21/17

This manual covers the model shown below

NA Model EU Model Voltage Description
88861023 N/A 100~240V Microplate Shaker US plug
88861024 15504070 100~240V Microplate Shaker Intl plug

Before using this product, read this entire operation manual carefully. Users should follow all of the operational guidelines contained in this manual and take all necessary safety precautions while using this product. Failure to follow these guidelines could result in potentially irreparable bodily harm and/or property damage.
Caution all internal adjustments and maintenance must be performed by qualified service personnel.
Material in this manual is for information purposes only. Fisher Scientific is committed to a continuing program of product development and improvement, and reserves the right to change information, such as specifications, appearance, and dimensions, described in this document without notice. Fisher Scientific makes no representations or warranties with respect to this manual. In no event shall Fisher Scientific be held liable for any damages, direct or incidental, arising out of or related to the use of this manual.
No part of this manual may be reproduced or transmitted in any form or by any means, including photocopying, recording, or using information storage and retrieval systems, for any purpose other than the purchaser’s own use, without the express written permission of the manufacturer.
Any other product names and services identified in this manual are trademarks or registered trademarks of their respective owners. No such use, or the use of any trade name, is intended to convey endorsement or other affiliation with Fisher Scientific.

Section 1 Inspection and Installation

  1. Inspect package and contents upon receipt of the instruments. If the package is severely damaged or if there are any missing pieces, please contact the manufacturer immediately.
  2. Unpack the instrument, ensure all parts of the instrument and accessories are not missing or damaged. Make sure to take out all the components before discarding the packing. If there are any missing or damaged pieces, please contact the manufacturer immediately.
  3. Place the instrument on a level and firm surface to avoid vibration and noise.

1.1 Packing List
Table -1 Packing List

Description NA Cat. No. NA/EU Cat. No. Figure
Microplate Shaker 88861023 US plug 88861024 /15504070 Intl plug
’96-well microplate holder 6 6
General Power Adaptor 1 1
US Plug 1 N/A
CN Plug N/A 1
EU Plug N/A 1
UK Plug N/A 1
‘M3x6 Screw 12 12
Screw Driver 1 1

*Has been installed on the instrument
1.2 Connections

Fisher 88861023 Microplate Shaker - Connections

1.3 Structure Diagram

Fisher 88861023 Microplate Shaker - Diagram

2.4 Capacity and Speed

Load Type Liquid Capacity Max Speed
96-well Microplate 1/2-well deep 1000rpm
96-well Microplate 2/3-well deep 800rpm
Fixed Tube Rack 40×Φ14 500rpm
Fixed Tube Rack 40×Φ16 450rpm
Adjustable Tube Rack 40×Φ14 500rpm
Adjustable Tube Rack 40×Φ16 400rpm

Note: These max. speeds are just for reference. We did the tests on the marble surface. Its friction coefficient is 0.32.
Warning: The rotation speed is inversely proportional to the load. When the instrument is running, it is recommended to adjust the rotation speed from low to high step by step and run the instrument at an appropriate speed to avoid spillage of liquids.

Section 2 Overview

2.1 Specifications
Category Number

Description NA Cat. No. NA/EU Cat. No.
Microplate Shaker 88861023 US plug 88861024 /15504070 Intl plug

Rotation Speed

Range 150 to 1000rpm
Orbit Diameter 2.5mm
Speed Accuracy ±1% of set speed up to 299rpm ±2% 300 to 1000rpm
Display LED
Display Accuracy 1 rpm

Note: Maximum speed may vary with heavy or unbalanced loads.
Load

Maximum Load 1 kg (Clamps included)
Maximum Capacity (Centered on tray) 96-well Microplate-4 standard 6 with

optional accessory

Power Supply

Requirement AC100-240V, 50Hz/60Hz, 72VA

Time

Timing Range 0min to 99h59min

Size

Overall Dimensions 415×293×137mm(16.3×11.5×5.4inch)
Tray Dimensions 275×235mm(10.8×9.3inch)
Packaging Dimensions 536×501×346mm(21.1×19.7×13.6inch)

Weight

Net Weight 11.5Kg(25.3lb)
Gross Weight 14Kg(30.8lb)

Others

Certificate ROHS, WEEE, cCSAus, CE Mark

2.2 Environmental Conditions
Application Environmental Conditions: indoor use

Altitude ≤2,000 m
Temperature 5℃ to 40℃
Humidity 20% to 85%

Storage Environmental Conditions

Altitude ≤2,000 m
Temperature 0℃ to 60℃
Humidity 20% to 90%, non-condensing

2.3 Safety Instructions
Please read the entire instruction manual before operating the Microplate Shaker.
WARNING DO NOT use the Shaker in a hazardous atmosphere or with hazardous materials for which the unit was not designed. Also, the user should be aware that the protection provided by the equipment may be impaired if accessories used are no provided or recommended by the manufacturer, or are used in a manner not specified by the manufacturer.
CAUTION ! To avoid electrical shock, completely cut off power to the unit by disconnecting the power cord from the unit or unplug from the wall outlet. Disconnect unit from the power supply prior to maintenance and service. Any spills should be removed promptly. Bio hazard spills should be cleaned using approved liquid promptly. Solvent spills are a fire hazard.
Stop the unit immediately, and DO NOT operate until clean up is complete and vapors have dissipated. Microplate
DO NOT immerse the unit for cleaning.
DO NOT operate the unit if it shows signs of electrical or mechanical damage.
Placement of Loads
Please place loads symmetrically during operation.
Symmetrical placement

Section 3 Operation

This chapter covers the control panel and its operation.
3.1 Control Panel

Fisher 88861023 Microplate Shaker - Control Panel

The front panel of the Microplate Shaker contains all the controls needed to operate the unit.

  1. Start/Stop button: Start or stop the instrument.
  2. RUN indicator light: The light is on when the instrument is running and off when the instrument is in standby.
  3. TIME display window: The window shows cumulative time (in continuous mode) or remaining time (in timer mode). The range of time displayed is 0 to 99 hours and 59 minutes. The accuracy is 1 minute.
  4. SPEED display window: The window shows set speed (when the instrument is in standby) or current speed (when the instrument is running).
  5. Set Time Buttons: UP/DOWN Arrow buttons are used to increase/decrease the set time of the instrument.
  6. Set Speed Buttons: UP/DOWN Arrow buttons are used to increase/ decrease the set speed of the instrument.

3.2 Installation

  1. Connect all the components according to the figures shown on page 4 of this manual. Use grounded power outlet.
  2. Press down the power switch on the back right side of the instrument and put it to the “|” state and then the instrument is in standby.

3.3 Settings
Time Settings

  1. Continuous mode
    Press the arrow button below the TIME display window. When the number shown on the display window starts flashing, press “ ” arrow button to decrease the time to 00:00 and then release the button. The time setting is finished after the number shown on the display window has flashed twice.

  2. Timer mode
    Press the arrow button below the TIME display window. When the number shown on the display window starts flashing, press arrow button to increase or decrease the time value. Release the button when the time shown on the display window reaches the set value. The time setting is finished after the number shown on the display window has flashed twice.

Speed Settings
Press the arrow button below the SPEED display window. When the number shown on the display window starts flashing, press arrow button to increase or decrease the speed value. Release the button when the speed shown on the display window reaches the set value. The speed setting is finished after the number shown on the display window has flashed twice
Note: press the arrow button for a longer time to accelerate the setting.
Run and Stop

  1. Continuous Mode
    Press button and the instrument will start running with the “ ” specified settings and the RUN indicator light will be on. The TIME display window will show the cumulative time and the SPEED display window will show the current peed. Press “ ” button again and the instrument will slow down until it stops. The instrument will then be in standby and the two display windows will show the set values.

  2. Timer Mode
    Press button and the instrument will The TIME display window will show the remaining time and the SPEED display window will show the current speed. Press button again and the instrument will slow down until it stops. The instrument will then be in standby and the two display windows will show the set values. “ ” start running with the specified settings and the RUN indicator light will be on.

Note:
To ensure shaking operation smooth and steady, it may take 1minute for the microprocessor control system to accelerate the tray to the set speed.
Finish Operation
After the operation is finished, press the power switch at the back right side the instrument and put it into the “O” state Unplug the instrument and store the instrument according to the storage guide.
Alarm Instructions

  1. If the actual speed of the instrument exceeds or is lower than the set speed to reach alarm range, alarm will be triggered.
  2. After the instrument alarms, LED RUN indicator light will go off, LED SPEED window will show “ERR1”, and the machine will stop running, press any key to put the instrument in standby.
  3. After the timer goes off, the buzzer will alarm and the instrument will stop running automatically.

Power Recovery
If the power supply is cut off suddenly while the instrument is in operation, the unit will automatically run at the previously set parameter upon power restoration.
The display window will flash. Press any button to stop flashing.
3.4 Installation of Accessories

  1. Loosen the screws of the tray with a cross screwdriver emplace the standard tray with platform large (figure 2).
  2. Place the test tube rack holder onto the tray and fasten the 4 sunk screws.
  3. Vertically clip the test tube rack into the holder and make sure there is no gap between the rack and the holder (figure 3)
  4. Insert the test tube (figure 4).

Note: When using single test tube, it is recommended not to set the speed too high because of load imbalance.

Fisher 88861023 Microplate Shaker - Fig1

Section 4 Safety Tips and Maintenance

Safety Tips

  1. Use independent power supply.
  2. Check if the local power supply voltage is suitable for use.
  3. Do not drag the power supply cable when unplugging.
  4. Do not use non-specified power cable or damage cable.
  5. Service should only be performed by a qualified professional.
  6. The power supply must be unplugged under the following situations:
    a. When the unit is moved
    b. When the electrical cabinet or the moving component is opened
    c. When the equipment is malfunctioning
    d. When the equipment is not in use

Maintenance
a. This instrument uses brushless DC motor. It is maintenance free and has a long service time, high quality, and low noise level.
b. Surface can be cleaned with a mild detergent and water.
Clean Spill
If accidental spillage of liquids caused by mishandling or contained breakage occurs on the surface of the instrument, please shut down the instrument and clean up the liquid immediately.
If the liquid has already spilled into the unit, cut off the power supply first and immediately clean up the liquid at the surface of the instrument. Place the instrument in a ventilated and dry environment for 24 hours before reuse. If the instrument is not  functioning after drying for 24 hours, please contact the manufacturer.
Warning: Disassembling/Assembling without a qualified professional’s guidance may cause malfunctioning of the instrument.

Section 5 Troubleshooting

Please refer to the following table to troubleshoot if any malfunction occurs. If the problem still exists, contact your local sales representative.

Error Cause Solution
Cannot start machine, LED display window off Power disconnected Connect the

power
Power switch off| Turn power switch on
Power adaptor failure| Replace power adaptor
No shaking of the tray| Over-weighted or unbalanced load| Adjust the weight and position of load, decrease rotation speed
Electrical malfunction| Contact Fisher Scientific
Mechanical malfunction| Contact Fisher Scientific
Loud noise| Microplate loose| Adjust position of the microplate holder
Tray loose| Fasten screws
Other| Keep record for maintenance

Section 6 Optional Accessories/Spare Parts

Description| NA Cat. No.| EU Cat. No.| Dimensions| Max. Qty| Figure
---|---|---|---|---|---
Platform Large| N/A| 15855911| 298×274mm| 1|
96-well microplate holder| N/A| 15865911| 54×19×19mm (7pcs/pack)| 7|
6 well platform assembly| 88861173| 15875911| 298×274×21.5mm| 1|
Fixed Tube Rack 40×Φ14| 88861148| 15594110| 262×112×140mm| 2|
Fixed Tube Rack 40×Φ16| 88861149| 15504120| 262×112×140mm| 2
Fixed Tube Rack 40×Φ18| N/A| 15783798| 262×112×140mm| 2
Fixed Tube Rack 27×Φ22| N/A| 15793798| 262×112×140mm| 2
Fixed Tube Rack 21×Φ30| N/A| 15703808| 262×112×140mm| 2
Adjustable Tube Rack| 88861150| 15514120| 262×112×171mm| 2|
Adjustable Tube Rack| 88861151| 15524120| 262×112×171mm| 2
Adjustable Tube Rack| N/A| 15713808| 262×112×171mm| 2
Adjustable Tube Rack| N/A| 15723808| 262×112×171mm| 2
Adjustable Tube Rack| N/A| 15733808| 262×112×171mm| 2
General Power Adaptor| 88861154| 15554120| AC 100~240V, 50/60HZ| 1|
Power Cable US Plug| 88861155| N/A| 125VAC, 10A, 1.8m| 1|
Power Cable CN Plug| 88861156| N/A| 250VAC, 10A, 1.8m| 1|
Power Cable EU Plug| 88861157| 15564120| 250VAC, 16A, 1.8m| 1|
Power Cable UK Plug| 88861158| 15574120| 250VAC, 13A, 1.8m| 1|

*Only used on platform large ( 15855911 ).

Section 7 Warranty

When used in laboratory conditions and according to these operation instructions and maintenance, this product is warranted for 24 months against defective materials or workmanship. The 24 month warranty period begins from the delivery date of this product.
For product quality or performance issues, contact Fisher Scientific Customer Service.

North America

United States
1-800-766-7000
fishersci.com| Canada
1-800-234-7437
fishersci.ca
Europe
Austria:
+43(0)800-20 88 40
at.fishersci.com| Netherlands:
+31(0)20 4887 70 00
nl.fishersci.com
Belgium:
+32(0)56 260 260
be.fishersci.com| Norway:
+47 22 95 59 59
fishersci.no
Denmark:
+45 70 27 99 20
fishersci.dk| Portugal:
+351 21 425 33 50
pt.fishersci.com
Germany:
+49(0)180 5258221
de.fishersci.com| Spain:
+34 002 239 303
es.fishersci.com
Ireland:
+959(0)1 885 5854
ie.fishersci.com| Sweden:
+46 31- 68 94 30
fishersci.se
Italy:
+39 02 950 59 478
it.fishersci.com| Switzerland:
+41(0)56 618 41 11
ch.fishersci.com
Finland:
+358(0)9 8027 6280
fishersci.fi| UK:
+44(1)1509 555 500
fisher.co.uk

Find out more at fishersci.com© 2017 Thermo
Fisher Scientific Inc. All rights reserved.
Trademarks used are owned as indicated at fishersci.com/trademarks.
2017 Fisher Scientific. All rights reserved.