Expert4house PH-W9909 Tuya Tester Instruction Manual
- June 1, 2024
- EXPERT4HOUSE
Table of Contents
Expert4house PH-W9909 Tuya Tester
SPECIFICATIONS
Range: pH 0.00-14.00pH EC 0-19000us/cm 0-199.0ms/cm TDS 0-19990ppm
0-199.0ppt SALT 0-199.9ppt 0%-20% S.G 0.990-1.400
Temp 0°C-50°C (32°F -122°F) Resolution: pH 0.01PH EC lus/cm (<1000us/cm)
10us/cm (> 1000us/cm) 0.1 ms/cm TDS 1ppm (<1000ppm) 10ppm (>1000ppm) 0.1ppt
SALT 1ppm (<1000ppm) 10ppm (>1000ppm) 0.1ppt 0.01% S.G 0.001 Temp 0.1°C (0.2°F)
- Accuracy: pH +0.03pH EC. TDS. SALT +2%F.S Temp #1.0°C
- Battery: DC6V
- Temperature Compensation: 0°C~50°C
- Operating Temperature: 0°C~50°C
- Dimensions: 130mmx77mmx31 mm
- Weight: 180g
- DC IN
- MODE/-
- TEMP/EC CAL/+
- WIFI
- pH CAL
- EC/TDS/SALT/G.S/TEMP
- TEMP-2
- pH
OPERATION
- Remove the protective cap. Connect pH electrode and temperature electrode.
- Clean the electrode with distilled water, and dry it well.
- Connect the power adaptor to turn on.
- Put the pH electrode, temperature electrode, and TDS electrode into the testing solution. Stir lightly and take a few seconds for a stable reading.
- Press the “ “key to choose the CAL/t °C/°F. Press” “for 5 seconds to calibrate the MODE TDS ( EC) value. Press the “ “key to choose different value display of “TDS/EC/ SALT/S.G. Press ” ” for 5 seconds to CAL calibrate the pH value.
- Turn off the meter after use. Clean the electrode in distilled water. Put on a protective cap.
PH CALIBRATION
- Pour standard buffer solution pH6.86, pH4.00 and pH9.18 (at 25°C) separately into three different clean beakers.
- For accurate calibration, fill one buffer solution into two beakers. One is for cleaning the electrode, and the other one is for calibration. By doing so, the pollution level can be reduced to the least.
- Connect to the power adaptor to turn it on.
- Immerse the pH electrode into standard buffer solution pH6.86, then stir gently until the value is stable. Long press the “ key for about 5 seconds. When the screen displays” “meter go into the pH6.8614.9 automatic correction mode (the meter can automatically identify standard buffer solution of pH4.01. pH9.18 and pH6.86). Display value corresponds to the standard buffer solution, the calibration is finished.
- Clean the pH electrode. Immerse the pH electrode into standard buffer solution pH4.01, then stir gently until the value is stable. Long press the ” “key for 5 seconds. When CAL the screen displays” meters go into the pH4.01 automatic correction mode. The display value corresponds to the standard buffer solution, the calibration is finished.
- Clean the pH electrode. Immerse the pH electrode into standard buffer solution pH9.18, then stir gently until the value is stable. Long press the” “key for 5 seconds. When CAL the screen displays” “meter goes into the pH9.18 automatic calibration mode. The display value corresponds to the standard buffer solution, the calibration is finished.
EC CALIBRATION
- Turn on the meter. Press the” “key to choose the EC-ms.
- Immerse the electrode into the distilled water. Activate it about five minutes.
- Immerse the electrode into EC calibration solution 12.88ms/cm (at 25°C), and stir it gently.
- When the reading is stable, long press the ” “key about 5 seconds. When CAL/+ the screen display” “or” “. Press the” ” or” “, until the display CAL/+ shows “12.80-12.90”
- Clean the electrode with distilled water and dry it with filter paper.
- Immerse the electrode into EC calibration solution 1.413ms/cm(at 25°C), and stir it gently. The display value is the same as the value of the calibration solution or approaching the value.
- Clean the electrode with distilled water and dry it with filter paper.
NOTICE
The electrode must be recalibrated: No calibration for a long time. Regular
and long-term continual use. High accuracy is required.
WARRANTY
All defects in materials and manufacture of this device are guaranteed for one year from the date of purchase. Within one year, please return the parts to the dealer or our office for free repair if the damage is not caused by the user’s negligence or wrong operation.
APP Multi-parameter Tester
Mobile phone connection APP
- Search “Tuya Smart” in the mobile APP store , find “and download.
- Click to enter after the download is tuyo complete” “Register an account, then log in to the homepage of Tuya , Click to add a device, long press on the meter key” “to search device automatically.
- The following icon appears” “Click to add, enter WiFi connection Mobile phone and device use the same WiFi. Click Next after entering the correct password.
- Page appears ” pet-w390s the icon appears after download “click Finish.Enter the program to view various parameters.
APP setting alarm
Enter Tuya APP on your mobile phone, click Smart” ” to create Smart Scene to enter” “click”Select pH value and imput a value at random,
click next to , Click “send a reminder notification reminder” to enter select the message reminder, the setting is complete. The current setting alarm is the pH value lower than 5.38. The APP will pop up an alarm message when the pH value is lower than 5.38