SEALEY ADB02.V2 Adblue Digital Flow Meter User Manual
- May 15, 2024
- SEALEY
Table of Contents
ADB02.V2 Adblue Digital Flow Meter
User Manual ADBLUE® DIGITAL FLOW METER
MODEL NO: ADB02.V2
ADB02.V2 Adblue Digital Flow Meter
Thank you for purchasing a Sealey product. Manufactured to a high standard,
this product will, if used according to these instructions, and properly
maintained, give you years of trouble free performance.
IMPORTANT: PLEASE READ THESE INSTRUCTIONS CAREFULLY. NOTE THE SAFE OPERATIONAL
REQUIREMENTS, WARNINGS & CAUTIONS. USE THE PRODUCT CORRECTLY AND WITH CARE FOR
THE PURPOSE FOR WHICH IT IS INTENDED. FAILURE TO DO SO MAY CAUSE DAMAGE AND/OR
PERSONAL INJURY AND WILL INVALIDATE THE WARRANTY. KEEP THESE INSTRUCTIONS SAFE
FOR FUTURE USE.
Refer to instruction manual
SAFETY
‰ WARNING! Ensure health & safety, local authority, and general workshop
practice regulations are strictly adhered to when using this equipment.
Familiarise yourself with product application and limitations, as well as the
specifi c potential hazards peculiar to this product.
Maintain the meter in good condition (use an authorised service agent).
Replace or repair damaged parts. Use genuine parts only. Non authorised parts
will invalidate the warranty.
Use only to meter water/urea solutions.
Ensure safety eye protection and protective clothing are worn when using this
product.
Keep the work area clean, uncluttered and ensure there is adequate lighting.
Maintain correct balance and footing. Ensure the fl oor is not slippery and
wear non slip shoes.
Keep children and unauthorised persons away from the working area.
After use, drain any fl uids from the equipment before storage.
Dispose of waste liquids in accordance with local authority regulations.
DO NOT exceed the maximum pressure of 10bar.
DO NOT use the equipment near open fl ames.
DO NOT smoke whilst using this equipment.
DO NOT use for corrosive fl uids.
DO NOT dismantle, tamper with or adapt the equipment for any purpose
other than for which it is designed.
DO NOT use the unit if it has been dropped or mishandled, check the unit
to ensure there is no damage.
DO NOT use taper connections, use parallel connectors only.
DO NOT use compressed air on the turbine, the excessive rotation will
damage the unit. Keep the meter clean and store in a safe dry, childproof
location.
‰ WARNING! DO NOT allow uncontrolled discharge of fl uids thus
polluting the environment. All liquids must be disposed of according to local
authority regulations.
INTRODUCTION
Robust rubber housing with integral digital electronic metering. Easy-to-read large, 4-digit, 28mm high LCD display calibrated in litres.
Featuresflowrateindication.1”x1”BSPMale/Femalefittings.SuitableforusewithAdBlue®.Poweredby2xAAAbatteries(supplied).
IP55 Rated protection.
SPECIFICATION
Model No: ……………………………………………. ADB02.V2
Flow Rate: ……………………………………………. 7-120L/min
Maximum Working Pressure: ………………………… 145psi
INSTALLATION
4.1. Connect the ADB02 into the system, ensuring that the arrow on the side of the meter is pointing in the direction of the flow.
NOTE: Use 1” BSP threaded connectors only, DO NOT use taper threaded connectors.
NOTE: To improve the life of the turbine, a strainer should be fitted upstream of the meter.
4.2. Thefaciaofthemetercanberotatedtomatchtheorientationofthemeter(fig.2).Remove the four screws from the front of the unit and carefully separate from the rubber protection surround, rotate the facia to the desired position taking care to not trap the wires against the reedswitch.Pushgentlyintotherubbersurroundandrefitthescrews.
OPERATION
5.1. LCD DISPLAY (fig.2A)
The“LCD”oftheMETERfeaturestwonumericalregistersandvarious indications displayed to the user only when the applicable function so requires. Key:
-
Partial register (5 figures with moving comma FROM 0.1 to 99999) indicating the volume dispensed since the reset button was last pressed.
-
Indication of battery charge.
-
Indication of calibration mode.
-
Totals register (6 figures with moving comma FROM 0.1 to 99999), that can indicate two types of Total:
a) General Total that cannot be reset (TOTAL).
b) Resettable total (Reset TOTAL). -
Indication of total multiplication factor (x10 / x100 ).
-
Indication of type of total, (TOTAL / Reset TOTAL).
-
Indication of unit of measurement of Totals: L=Litres, Gal=Gallons
-
Indication of Flow Rate mode.
-
Indication of unit of measurement of Partial: Qts=Quarts, Pts=Pints, L=Litres, Gal=Gallons.
5.2. USER BUTTONS
5.2.1. TheADB02(fig.1)features two buttons(RESETandCAL)whichindividuallyperformtwomainfunctionsand,together,other secondary functions. The main functions performed are:
RESET key – resetting the partial register and resettable total (reset total).
CAL key – entering instrument calibration mode.
5.2.2. Used together, the two keys permit entering configuration mode, useful for changing the units of measurements and calibration factor.
5.3. STANDBYMODE
Whenthemediaisnotflowingthroughthemeter,themetershowsonlythewordTOTALonthedisplay.ThismodeiscalledSTANDBY and majority of adjustments are carried out in this mode.
5.4. MEASUREMENT UNITS CONFIGURATION
The user can select the main measurement unit, Quarts (Qts), Pints (Pts), Litres (L), Gallons (Gal); according to the following predefined combinations:
REF. NO| UNIT OF MEASUREMENT RESETTABLE BATCH TOTAL| UNIT OF MEASUREMENT
BATCH TOTALIZER
---|---|---
1| Litres (L)| Litres (L)
2| Gallon (Gal)| Gallon (Gal)
3| Quarts (Qts)| Gallon (Gal)
4| Pints (Pts)| Gallon (Gal)
5.5. SETTING THE UNIT OF MEASUREMENT
Wait for the meter to go to STANDBY mode (fig.3). Press the CAL and RESET keys
together. Hold it until the word “UNIT” appears on the screen together with
the current unit (fig.4). Press RESET key to scroll among the four
combinations of units of measurement as shown (fig.5). Press CAL key for more
than 2 seconds to store the new settings. The METER will return to the Stand
by Mode.
NOTE: No new calibration is required after changing the Unit of
Measurement.
5.6. NORMAL DISPENSING MODE
While the media is flowing through the meter, Batch Total and Reset Total are
displayed at the same time (fig.6). A few seconds after dispensing has ended,
the display switches from Reset TOTAL to TOTAL: the word RESET above the word
TOTAL disappears and the Reset TOTAL is replaced by TOTAL (fig.7). NOTE: This situation, where only “TOTAL” is displayed, is
called STANDBY mode. It remains stable until the user operates the meter
again.
5.7. RESETTING THE BATCH TOTAL
While in standby (i.e when the display shows TOTAL), press the RESET button
(fig.8). During reset, the display screen shows all the lit-up digits (fig.9).
After resetting, the display shows zero value on Resettable Batch TOTAL
(fig.10).After a few moments, the Reset TOTAL is replaced by TOTAL (fig.
11).
5.8. RESETTING THE RESET TOTAL
The Reset TOTAL can be reset by pressing the RESET key at length while the
display screen shows Reset TOTAL. The steps to be taken are:
5.8.1. Wait until the display shows TOTAL only (Standby mode) (fig. 12).
5.8.2. Press the RESET key. The display screen again shows all the segments of
the display followed by all the switched-off segments (fig.13).
5.8.3. While the display page showing the Reset TOTAL is displayed, press and
hold the Reset key again till the Resettable TOTAL turns to zero (fig.14).
5.8.4. Finally the page with the new Reset TOTAL is displayed (fig. 15).
5.9. CALIBRATION
In standby mode, press and hold the CAL key to see the current calibration
factor.
- Factory K Factor: Factory-set default factor.
It is equal to 1 (indicated as 1,000).
¢ User K Factor: Customized calibration factor, meaning modified by calibration.
5.9.1. CALIBRATION PROCEDURES
By pressing the CAL key while the meter is in Standby, the display shows the current calibration factor used. Two cases can occur:
Case 1: “FACT”. If no calibration has ever been performed, or the factory setting has been restored after previous calibrations, the following display page will appear (fig. 16).
NOTE: The word “FACT” is an abbreviation for “factory” shows that the factory calibration factor is being used.Case 2: “USER”. If, on the other hand, calibrations have been made by the user, the display page will appear showing the currently used calibration factor (in our example 0,998) (fig.17).
NOTE: The word “USER? indicates that a calibration factor, set by the user is being used.
. To confirm the choice of calibration factor, quickly press CAL while “USER” or “FACT” are displayed.
5.9.2. IN-FIELD CALIBRATION SEQUENCE
Wait until the METER comes in Standby (Display shows TOTAL) (fig.18). Press CAL key for more than 2 seconds. The METER enters calibration mode and shows “CAL”. The words “FACT” or “USER?” indicate which factor (factory or user) is currently in use(fig.19, fig.20). Press and hold RESET key. The METER shows “FIELD” and the Batch Total at zero. The meter is ready to perform in-field calibration (fig.21).
5.9.2.1. PROCESS FOR IN-FIELD CALIBRATION
5.9.2.2. Dispensing into sample container: without pressing any key, start dispensing into the sample container (fig.22).
5.9.2.3. Continue dispensing until the level of the fluid in the sample container has reached the graduated area (fig.23).
5.9.2.4. Press RESET key once. The METER detects that the calibration dispensing is finished. An arrow (up/down) appears which indicates the direction in which the value can be changed via steps 5.9.2.5 and 5.9.2.6. To calibrate the METER, the value indicated by the Batch total (example 9.800) must be forced to the Container value 9,860 marked on the graduated sample container (fig.24).
5.9.2.5. Press RESET key once. The arrow changes direction. The operation can be repeated to alternate the direction of the arrow (fig.25).5.9.2.6. Press “CAL” key to change the value in the direction indicated by the arrow. The reading changes either by one unit for every short press of CAL key or continually if the CAL key is kept pressed (fig.25).
5.9.2.7. Press and hold RESET key for more than 2 seconds. The METER is informed that the calibration procedure is finished. The meter calculates the new USER K FACTOR factor for a few seconds (fig.26).
5.9.2.8. The new USER K FACTOR is shown for a few seconds, after which the restart cycle is repeated to finally achieve standby condition (fig.27).
5.9.2.9. The METER stores the new calibration factor and is ready to begin dispensing (fig.28).
5.9.3. DIRECT CALIBRATION SEQUENCE
5.9.3.1. If normal METER operation shows a mean percentage error E,
(obtainable on the basis of several performed dispensing operations), this can
be corrected by applying a correction to the current calibration factor as
shown below :
5.9.3.2. Wait until the METER comes in Standby (Display shows TOTAL) (fig.29).
5.9.3.3. Press and hold CAL key. The METER enters calibration mode and shows “CAL”. The words “Fact” or “USER” indicate which factor (factoryoruser)is currently in use(fig.30).
5.9.3.4. Press and hold RESET key. The METER shows “CAL”, “Field” and the
Batch Total at zero. The meter is ready to perform in- field calibration
(fig.31).
5.9.3.5. Press and hold RESET key. “Direct” appears together with the Current
calibration factor. An arrow appears (upwards or downwards) defining the
direction (increases or decreases) of the reading (fig.32).
5.9.3.6. By pressing RESET key the user can change the direction of the arrow.
5.9.3.7. By pressing CAL key, the Meter value changes in the direction
indicated by the arrow, either one unit for every short press of CAL key or
continually if the CAL key is kept pressed. The speed increases or decreases
by keeping the key pressed.
5.9.3.8. Press RESET key for more than 2 seconds. The METER detects that the
desired reading has been set and the calibration procedure is finished
(fig.33).
5.9.3.9. At the end of the calculation, the new USER K FACTOR is shown for a
few seconds (fig.34).
5.9.3.10. The restart cycle is repeated to finally achieve standby mode
(fig.35).
5.9.4. MAINTENANCE
The Meter has been designed to require a minimum amount of maintenance. The
only maintenance jobs required are:
Battery change: Necessary when the batteries have run down.
Cleaning the turbine assembly: Due to the presence of solid particles
following bad filtering.
5.9.4.1. CHANGING THE BATTERY
The METER features two low-battery alarm levels
5.9.4.2. When the battery charge falls below the first level on the LCD, the
fixed battery symbol appears. In this condition, the METER continues to
operate correctly, but the fixed icon warns the user that it is time to change
the batteries (fig.36).
5.9.4.3. If meter operation continues without changing the batteries, the
second battery alarm level will be reached which will prevent any operation.
In this condition the battery icon starts to flash and is the only one to
remain visible on the LC (fig.37).![SEALEY ADB02 V2 Adblue Digital Flow Meter
- BATTERY](https://manuals.plus/wp-content/uploads/2024/03/SEALEY-ADB02-V2 -Adblue-Digital-Flow-Meter-BATTERY.png)
5.9.4.4. BATTERY REPLACEMENT PROCEDURE:
5.9.4.5. Press RESET to update all the totals
5.9.4.6. Remove the four screws from base and separate the battery cap.
5.9.4.7. Remove the old batteries.
5.9.4.8. Place the new batteries in the same position as the old ones, making
sure the positive pole is positioned as indicated.
5.9.4.9. Re-attach the battery cap and the four screws.
5.9.4.10. The METER will switch on automatically and normal operation can be
resumed.
NOTE: The old calibration will stay same as before.
5.9.5. CLEANING OF THE TURBINE ASSEMBLY
After removing the meter from pipes, any residual elements can be removed from
the turbine by simply washing it with water.
‰ WARNING Always make sure the liquid has been drained from the meter and
the line pressure is released before cleaning. Never use compressed air for
cleaning as it may damage the turbine assembly.
TROUBLESHOOTING
Problem | Possible Cause | Remedy |
---|---|---|
No LCD display | Bad battery connection | Check /Clean Battery terminals |
Batteries expired | Replace batteries | |
Incorrect measurements | Wrong K Factor | Refer to Section 5.9 Calibration |
Flow rate below meter limits | Increase flow rate | |
Reduced or zero flow rate | Turbine blocked | Clean turbine |
Meter not recording flow | Faulty PCB | Contact your Sealey dealer |
Faulty turbine | Contact your Sealey dealer |
Parts support is available for this product. Please log on to
www.sealey.co.uk, email
sales@sealey.co.uk or telephone 01284 757500
WEEE REGULATIONS
Dispose of this product at the end of its working life in compliance with the
EU Directive on Waste Electrical and Electronic Equipment (WEEE). When the
product is no longer required, it must be disposed of in an environmentally
protective way. Contact your local solid waste authority for recycling
information.
BATTERY REMOVAL
Under the Waste Batteries and Accumulators Regulations 2009, Jack Sealey Ltd
are required to inform potential purchasers of products containing batteries
(as defined within these regulations), that they are registered with Valpak’s
registered compliance scheme. Jack Sealey Ltd.’s Batteries Producer
Registration Number (BPRN) is BPRN00705.
ENVIRONMENT PROTECTION
Recycle unwanted materials instead of disposing of them as waste. All tools,
accessories and packaging should be sorted, taken to a recycling centre and
disposed of in a manner which is compatible with the environment. When the
product becomes completely unserviceable and requires disposal, drain any
fluids (if applicable) into approved containers and dispose of the product and
fluids according to local regulations.
Note: It is our policy to continually improve products and as such we
reserve the right to alter data, specifications and component parts without
prior notice.
Important: No Liability is accepted for incorrect use of this product.
Warranty: Guarantee is 12 months from purchase date, proof of which is
required for any claim.
Sealey Group, Kempson Way, Suffolk Business Park,
Bury St Edmunds, Suffolk. IP32 7AR
01284 757500
sales@sealey.co.uk
www.sealey.co.uk
ADB02.V2 Issue 2 (H,F) 08/08/23
Original Language Version© Jack Sealey Limited
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>