Labocult H770 Cheap Incubator Instruction Manual

May 15, 2024
Labocult

Labocult H770 Cheap Incubator

Labocult-H770-Cheap-Incubator-PRODUCT,,..,.,.,.,

Product Information

Specifications

  • Applied in laboratories, industrial and mining, universities, colleges, research institutions, medicine & health
  • High-quality cold rolling steel electrostatic spraying exterior
  • Stainless Steel inner chamber with semicircular arcs for easy cleaning
  • Intelligent PID temperature controller with timing and alarm functions
  • Magnetic door seal for secure closure
  • Heat insulation for energy efficiency and environmental protection

Technical Parameters

  • Voltage:
  • Power consumption:
  • Typical consumption:
  • Temperature range:
  • Environmental temperature:
  • Humidity:
  • Accuracy:
  • Inner chamber size:
  • Exterior size:
  • Timing:

Safety Information

  • Install external grounding protection
  • Avoid using inflammable, explosive, noxious, or corrosive materials
  • Ensure horizontal installation
  • Maintenance by authorized specialists only
  • Read instructions carefully before operation

Operation Cautions

  1. Avoid modifying internal parameters without permission
  2. Avoid overloading trays and using corrosive materials for cleaning
  3. Cut off power when equipment stops and keep interior and exterior clean

Controller Operation

The TS-1000 temperature controller is user-friendly with super fuzzy PID control.

Panel Brief Introduction

  • Shape size : 109 mm x 125 mm
  • Temp. Range:
  • Time range :
  • Accuracy :
  • Environment Temperature :
  • Relative Humidity: < 85% RH

Panel Instructions

Indicators:

  • RUN /AT: Indicates running status or auto-tuning of P.I.D
  • OUT : Indicates heater status

SUMMARY

lt is applied in laboratories, in industrial and mining, universities and colleges, research institutions, medicine & health and other units for storage and culturing.

STRUCTURE FEATURES

  1. High-quality cold rolling steel electrostatic spraying exterior.
  2. Stainless Steel inner chamber, semicircular arcs at corners for easy cleaning.
  3. Intelligent PID temperature controller, with timing, alarm indicates, Temperature deviation trimming, self-tuning and etc. Cut off the power automatically, when over-temperature is limited, ensuring the safety of experiments and personnel.
  4. Equipped with a magnetic door seal to ensure the door stay closed.
  5. Heat insulation to prevent the loss of heat too quickly. Energy efficient, low carbon, and environmental protection.
  • Voltage: 220-240V, 50 Hz

  • Power consumption: 300 Watt max

  • Typical consumption: 8 Watt

  • Temperature range: Room temperature +5 °C – 70 °C8

  • Environmental temperature :0 °C – 40 °C

  • Humidity: ≤80% RH / non condensing

  • Accuracy: +0.5 °C

  • Inner chamber size :230×200×200 mm

  • Exterior size: 300x330x330 mm

  • Timing: 0-9999min

TEMPERATURE DISTRIBUTION

Labocult-H770-Cheap-Incubator-FIG1

Note : the measured value has a little difference because of different models

ENVIRONMENTAL CONDITIONS

  • Environment temperature: 5 ~40
  • Relative humidity: ≤80%RH
  • Pressure: 80-106Kpa

SAFETY INFORMATION

  1. In order to ensure the safety of equipment and experiment, please install external grounding protection and supply power according to the requirements of equipment specifications.
  2. Don not use with inflammable and explosive materials, noxious goods and strong corrosive articles by this equipment.
  3. Ensure the horizontal installation.
  4. Service and maintenance only by authorized specialists.
  5. Don’t make compulsory startup, must eliminate the alarm reminder.
  6. Read this instruction carefully before operating this equipment.

OPERATION CAUTIONS

  1. For the initial startup, don’t modify the internal parameters of the program controller except for the permission in the instructions.
  2. The workroom adopts a vertical ventilation cycle. Each tray can not be placed too much, the total test load should not be bigger than 1/2 of the tray.
  3. The environment temperature must 5 °C lower than the setting temperature, then it can work in normally.
  4. Don’t use acid, alkali, and other corrosive articles to scrub the internal surface and external surface. The neutral washing detergent could be used for regular cleaning and then wiped with dry cloth.
  5. When the equipment stops, cut off the power and keep the interior and exterior dry and clean.

CONTROLLER OPERATION

The series of TS-1000 temperature controllers adapt to laboratory and analysis instruments and operate very easily. Temperature controllers adapt to “super fuzzy PID control” much better than traditional PID control.

PANEL BRIEF INTRODUCTION

  1. Shape size: 109 mm x 125 mm

  2. Temp. Range: TS – 10xx, 0 °C -100°C

    • Time range: 0 -9999 Minutes
  3. Accuracy: < 0,5 ° C

  4. Environment: Mainboard 220V -, ± 10 % AC

    • Temperature: 0 ° -50 °C
    • Relative Humidity: < 85% RH

PANEL INSTRUCTIONS

INDICATOR DEFINITION

  1. RUN/AT is bright when the controller is running, when the runtime is over, this indicator turns oft. When the controller enters the auto-tuning of P.I.D this indicator is flashing.
  2. OUT is on when the heater turns on rise this indicator is oft.
  3. ALM when the over-temperature alarm occurs, this indicator turns on.

USAGE

  1. When the controller is switched on, display windows show “lnP”” and the value of the temperature range for 3 seconds, then it starts running.

  2. Temperature and time settings.
    12  Press the “Set” button, and the controller runs into the temperature setting state. Re-press the”Set” button, the controller runs into the time setting state. In the setting state, you can use the “ ” buttons to get the required settings. Press the “set” button again, it returns from the setting state and the settings are saved automatically.
    lf the time is set as “0”, the controller will run continuously, the display window of “SV” will display the set point temperature. lf the time set value is not equal “O”, timers start time when the measuring temperature reaches the set point temperature, the display window of “SV” will display the runtime.
    When the runtime is over, the “sV” window will display “End”, and the buzzer will sound for 30 s, Press the button “ ” for 3 s to restart.

    • (1) The normal display
    • (2) The temperature setting state
    • (3) The time-setting state
    • (4) Timing Display
  3. When the temperature alarm, the buzzer will sound,” ALM” lights. lf a change in temperature setting and over-temperature alarm,” ALM” lights up, but no song buzzer.

  4. When the buzzer sounds, it can be muted by pressing any button.

  5. “◄” button: In the setting state, it can shift the set value by pressing the button.

  6. “:’

    1. button: In the setting state, it can reduce the set value by pressing the button. lf press and hold the button, the set value will reduce continuously. The timing state, lang press the button for 3 seconds and can make the program stop.
  7. “ .” button: In the setting status, it can increase the set value by pressing the button. lf press and hold the button, the set value will increase continuously.

  8. In the setting state, the controller will return to run status if without any key press in one minute.

  9. lf the display window shows “—-“, it indicates the fault ot temperature.

Labocult-H770-Cheap-Incubator-FIG5 Labocult-H770-Cheap-Incubator-FIG6

SYSTEM AUTO TUNING {AT)

lt the temperature control result is not perfect, auto-tuning can be performed. Auto-tuning may cause the temperature to increase more. The user should consider this factor before auto-tuning. When auto-tuning is completed, it is not possible to return to the manufacturer’s original setting. Autotuning can be performed only when the door ot the product is closed. Autotuning must be performed

when the device has been switched on just before. The room temperature should be about 25 C. lt the unit has already reached a higher temperature or the room temperature is too low or too high, all these tractors will attract the result ot the autotuning. The result ot the temperature control will then not be optimal.
Now you can start setting the autotuning

In the non-set state, press the ,,” button for 6 seconds, the „AT” indicator will flash, it will turn off automatically when the autotuning process is completed. The auto-tuning process takes about half an hour. During the auto- tuning, better PID parameters are automatically set, so that the temperature control result is optimized. The new PID parameter is automatically stored in the system and the unit automatically returns to normal operation mode. During the autotuning process, if the user wants to stop this process, he must press the ,,◄” key for 6 seconds to interrupt the autotuning process. When the autotuning is finished, it is not possible to delete the new PID parameter.

lf an overtemperature alarm occurs during the autotuning process, the „ALM” indicator will not light up and no beep will sound. The heater alarm relay will still be turned oft automatically.
During the auto-tuning process, the „SET” key has no function. lf a time is set before auto-tuning starts, the display of „SV” does not show the time but the temperature setting value.

INTERNAL PARAMETERS SETTINGS

Note : All the internal parameters has been adjusted when the factory testing. Forbidden to modify them except Sensor Correction parameter.

Press the “Set” button for 3 seconds, controller will display the password prompt “Lc”. Adjust the password to the required value, then press the “Set” button again, it will run into the internal parameter setting state. if press the “Set” button for another 3 seconds, it will return to the running state.

PARAMETER LIST-1:

IMPORTANT! Please leave the adjustment to a trained service person

IMPORTANT! Please leave the adjustment to a trained service person.

Labocult-H770-Cheap-Incubator-FIG7

PARAMETER LIST-2:

IMPORTANT! Please leave the adjustment to a trained service person Labocult-H770-Cheap-Incubator-FIG8

ENGLISH NAME AND PARAMETER INDICATING THE SYMBOL TABLE

Labocult-H770-Cheap-Incubator-FIG9

WIRING

Labocult-H770-Cheap-Incubator-FIG10

PROBLEM ANALYSIS

Labocult-H770-Cheap-Incubator-FIG11

  • Electromagnetic Directive 2014/30/EU
  • Low Voltage Directive 2014/35/EU

servoprax GmbH

Am Marienbusch 9, 46485 Wesel, Germany

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>