FKM6016B Digital Clamp Meter Multimeter 4000 Counts Amp Voltage Tester User Manual

May 15, 2024
FKM

FKM6016B Digital Clamp Meter Multimeter 4000 Counts Amp Voltage Tester

Overall, Safety Standards, Safety Symbols, Notices

Overall
This digital AC clamp meter is a stable and reliable instrument with a micro smart IC and dual integral AD transformer as the core, coming with a full range overload protection circuit. It can be used to test AC, AC voltage, resistor, diode, circuit continuity.

Safety Standards
This instrument has been designed and manufactured with safety standards of IEC61010, IEC61010-2-032, dual insulation CAT III 600V, and pollution class 2.

Safety Symbols

  • Warning symbols, cautiously operate.
  • High voltage danger symbol.
  • Allow to be used nearby by conductors that are not dangerous to life.
  • Dual insulation (Class I| safety equipment)
  • Ground.

Notice

  • Read the manual carefully, especially notice the “ ” contents with” ” symbols. Please follow the instructions.
  • If the test pen needs to be replaced, replace a new one with the same model number or the same specifications

Accessories

  • Check the instrument and test pen before use. If exposure of test leads, broken cover, or abnormal display is found, do not use it.
  • In the test, do not touch the terminal not in use.
    • When testing a DC voltage higher than 60V or an AC voltage higher than 30V, do not reach your fingers beyond the safety barrier.
  • When the measured range is unknown, put the test range at the maximum.
  • Do not test voltage higher than the maximum range.
  • The test lead should be away from the tested circuit before the switch range selector.
    • Before testing the diode in a live circuit, cut off the power and discharge all capacitors.
  • Do not expose the instrument in strong light, high temperature, or damp environments.
  • Do not touch naked circuit lines, connectors, or measured circuits.

Accessories

  1. Instruction manual
    • 1
  2. Test lead
    • 1
  3. thermocouple
    • 1
  4. Package.
    • 1
  5. 1.5V SIZE AAA battery
    • 2

Description

  1. Jaw Assembly : sampling AC.
  2. Safety Barrier : prevent touching live conductors in the test.
  3. Rotatory Range Selector : select measuring functions and ranges.
  4. Data Hold : press “H” button, and the last reading keeps on displaying the “H” symbol. Press the “H” button again to resume normal test mode.
  5. Backlight : press and hold 2 seconds to turn on the backlight, and press and hold 2 seconds again to turn off.
  6. LCD : Maximum display 4000, reading height 12mm.
  7. Input Terminal : red test lead positive input terminal in testing voltage, diode, transistor, and continuity.
  8. COM terminal : black test lead negative input terminal except for AC measurement.
  9. Flashlight : Press this button to turn on the flashlight. Press again to turn it off.
  10. SEL : in test mode, select between diode, resistor, and continuity.
  11. Warning Light : in the continuity test, the red light turns on with the NCV warning.
  12. Lever for Jaw Opening Closing : press to open and release to close the jaw.
  13. Clamp Body Light
  14. NCV Sensor Probe : when a strong AC signal is detected, the red light turns on and a buzzer sounds.

Panel description

FKM6016B-Digital-Clamp-Meter-Multimeter-4000-Counts-Amp-Voltage-Tester-
FIG-3

Instructions

AC and DC Voltage Measurement Insert the readout test line into the “ ” jack and black test lead into the “COM” jack.

DC Voltage Measurement
Turn the rotatory range selector to , and connect the test lead to the voltage to be measured. Read the measurement value and polarity of the red test lead from the display.

AC Voltage Measurement
Turn the rotatory range selector to , and connect the test lead to the voltage to be measured. Read measurement value from the display.

Caution

  • If the measured voltage range is not known in advance, turn the rotatory range selector to the maximum. Then decrease gradually to get a satisfactory resolution.
  • Beware of electric safety in measuring high voltage.

AC measurement

  1. Turn the rotatory range selector to AC;
  2. Clamp the jaw around the conductor to be measured. Do not more than one conductor at the same time.
  3. Read measurement value from the display.

Notice

  • If the measured current range is not known in advance, turn the rotatory range selector to the maximum. Then decrease gradually to get a satisfactory resolution.

Resistor Measurement

  1. Insert the red test lead into the jack and the black test lead into the “COM” jack.
  2. Turn the rotatory range, selector, to” “, press the “SEL” button to switch to Ω, and place the test probe tips into contact with the sample to be measured
  3. Read measurement value from the display.

Notice:

  • If the measured sample has higher resistance beyond the maximum range, “OL” will be displayed. Please change to use a meter with a higher measurement range.
  • When measuring resistance, make sure the power is off and all capacitors are fully discharged.
  • When measuring resistance over 1M OHM, it might take a few seconds to stabilize the reading.

Diode and Continuity Test
Insert the readout test line into the ” ” jack and the black test lead into the “COM” jack. The polarity of the red test lead is “+”

  1. Turn the rotatory range selector to” “, and press the “SEL” button to switch to . Bring the red test lead in contact with the positive electrode and the black test lead in contact with the negative electrode. Read the forward voltage drop value from the display.
  2. Turn the rotatory range selector to” “, and press the “SEL” button to switch to . Place the test leads in contact with two points of the measured circuit. If the resistance is
    Instructions, Auto Off
    less than 50 OHM, the beeper emits continuous sound.

Capacitance measurement

  1. Insert the red pen and black pen into the “INPUT” and “COM “terminals.
  2. Set the function range switch to “ ” range, press to”SEL” switch to “ “, and connect the marker to two points of the measured capacitor.
  3. Read the measurement results from the monitor.

Note

If the measured capacitance exceeds the maximum value of the selected range, the display will appear as “OL”. Upon checking the in-line capacitance, the power supply to the line under test must be cut off and all capacitors fully discharged.

Temperature measurement

  1. Set the functional range switch to the TEMP range position. The LCDs are the current ambient temperature.
  2. When the thermocouple is needed to measure the temperature, the red plug of the K-type thermoelectric couple can be inserted”INPUT” jack, the black plug into the “COM “jack, and use thermocouple probe to contact the measured object or area for measurement.
  3. Display the reading on the LCD.

Auto Off

The unit powers off automatically when staying idle for over 15 minutes to save power.

Accuracy Specification

  • Accuracy: + (a%+cts)
  • Warranty: 1 year
  • Ambient Temperature: 18°C~28°C, Ambient humidity: Less than 75%
  • Temperature co-efficient : 0.1 xaccracy/1°C

Notice:
In measuring AC, place the measured conductor at the center o the jaw, otherwise there might be an offset of 1.5%

DC Voltage

FKM6016B-Digital-Clamp-Meter-Multimeter-4000-Counts-Amp-Voltage-Tester-
FIG-17

  • Input Resistance : 10MΩ
  • Max. Input Voltage : 600V DC or 600V rms AC

AC Voltage

FKM6016B-Digital-Clamp-Meter-Multimeter-4000-Counts-Amp-Voltage-Tester-
FIG-18

  • Input Resistance : 10MΩ
  • Frequency Range : 40Hz~400Hz
  • Max. Input Voltage : 600V DC or 600V rms AC

AC Current

FKM6016B-Digital-Clamp-Meter-Multimeter-4000-Counts-Amp-Voltage-Tester-
FIG-19

  • Frequency Range : 50Hz~60Hz
  • Max. Input current : 120% of full scale, less than 60 seconds.

Resistance

FKM6016B-Digital-Clamp-Meter-Multimeter-4000-Counts-Amp-Voltage-Tester-
FIG-20

Overload: 600V DC or 600V rms AC

Capacitance

FKM6016B-Digital-Clamp-Meter-Multimeter-4000-Counts-Amp-Voltage-Tester-
FIG-21

Overload: 600V DC or 600V rms AC

Temperature

FKM6016B-Digital-Clamp-Meter-Multimeter-4000-Counts-Amp-Voltage-Tester-
FIG-22

Accuracy does not include the error of the thermocouple probe.

Overload: 600V DC or 600V rms AC

Diode

FKM6016B-Digital-Clamp-Meter-Multimeter-4000-Counts-Amp-Voltage-Tester-
FIG-23

Overload: 600V DC or 600V rms AC

Technical Specifications

Continuity

FKM6016B-Digital-Clamp-Meter-Multimeter-4000-Counts-Amp-Voltage-Tester-
FIG-24

  • Overload : 600V DC or 600V rms AC
  • General : Max. The voltage between input and ground is CATII 600V DC and 600V AC
  • Display : LCD, max reading 4000
  • Principle : Dual Integral A/D transformer, auto range
  • Measuring frequency : 3 times/second
  • Unit Display : Display function and unit symbols
  • Electrode : negative input shows “—”
  • Overload Display : “OL”
  • Data Hold Display : ” H”
  • Low Power Display : ” “
  • Power supply : DC1.5V x2 Size AAA
  • Dimension: 185mm x 71mm x35mm
  • Max. Jaw Opening Size : 26mm
  • Temperature Environment for Use : 5C-35C
  • Temperature for Storing : -10C-50C

Maintenance and Battery Replacement

Maintenance

  1. Before opening the back cover, take away the test leads from the measured circuit.
  2. Use a wet cloth and a little detergent to clean the instruments. Do not use chemical or grinding solvents.
  3. Stop using if any abnormal conditions occur.
  4. Calibration or maintenance can only be performed by professionals.

Battery Replacement

Warning
To avoid electric strikes, take away test leads from the measured circuit before opening the battery cover. Only replace batteries with the same type and electric specifications.

When the low power icon” ” appears, batteries need to be replaced immediately.

  1. Take away test leads from the measured circuit. Turn the rotatory range selector to OFF. Take away let leads from input jacks.
  2. Use a screwdriver to unscrew the batter cover.
  3. Take out used batteries and replace them with a new 1.5V Size AAA battery.
  4. Put on the cover and fasten the screw tightly.

Get 24/7 Assistance at

Use a camera or QR scanner to scan

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals