SprintRay SRE-1001462 IBD Clear Transparant Instruction Manual
- May 15, 2024
- SprintRay
Table of Contents
SprintRay SRE-1001462 IBD Clear Transparant
Indication
SprintRay EU IDB is a resin for the generative production of dental indirect bonding trays based on image projection systems.
Contraindication
SprintRay EU IDB is contraindicated for every application that is not part of the indication (see above).
Requirements
Software:
- RayWare (CAD)
Hardware:
- Pro Desktop 3D (3D Printing)
- Pro Wash / Dry (Cleaning)
- ProCure (Post Curing)
SprintRay, Inc., 3577 North Figueroa St., Los Angeles, CA 90065
Material
SprintRay EU IDB consists of functional (meth)acrylic resins.
Geometric presetting
According to the software design specifications.
Material parameters
Radiation penetration depth controllable by exposure time.
- 50 μm
- 100 µm
Manufacturing process (fi g. 1-10)
-
Prepare data (RayWare).
-
Choose process parameters (build style, etc.).
-
Transfer prepared data to 3D printing device.
-
Prepare 3D printing – shake bottle.
-
Fill resin tank of 3D printing device.
-
Build the parts.
-
Clean parts (ProWash / Dry):
- blow surpluss resin out of cavities with pressured air before washing
- wash 5 min. / 5 min.
-
Dry parts (ProWash / Dry):
- 3 min.
- additional: dry with pressured air as required
-
Post curing (ProCure): 5 min. / 60°C.
-
Finish parts.
Finishing processes
No applicable.
Notice
Please follow the instructions provided by the software manufacturer in terms of parameter settings and design recommendations. Please follow the instructions provided by the hardware manufacturer in terms of parameter settings/printing & post curing recommendations. To avoid detrimental effects on material quality do not expose the liquid material to irradiation under any circumstances. Deviations from the described manufacturing processes or storing conditions may lead to different mechanical and optical properties of the material. Ensure personal protective gear during processing.
Caution: Polymerized resins are chemically resistant – avoid stains on clothing! Avoid any contact with skin and eyes! In case of accidental contact, rinse with adequate running water, consult a doctor if necessary. The lot number and the best before date are indicated on each packaging. In case of claims please always indicate the lot number of the product. Do not use the product after expiry of the best before date. Dispose of contents/container according to official regulations.
Hazard statements
- Causes skin irritation. May cause an allergic skin reaction.
- Causes serious eye irritation. May cause respiratory irritation.
Physical properties
SprintRay EU IDB
-
Elongation at break (ISO 37)/ca. 50 %**
-
Tensile strength
MPa (ISO 37): ca. 6.2** -
Shore hardness D
(ISO 48-4): ca. 40** -
Viscosity
(DIN 53019-1): ca. 2.5**
Ordering information
SprintRay EU IDB
1 kg:
Λ ≤ 405 nm
clear-transparent/
REF: SRE-1001462
- These data come from measurements of a representative sample that was determined as part of our quality assurance.
** According to internal design and requirements specifications.
Symbols
-
This side up.
-
Temperature limit
-
CE mark
-
Warning
-
Do not use if package is damaged.
-
Consult instructions for use
-
Keep away from sunlight.
Rx only
QTY: 1EA
-
Medical Device
-
Use by date
-
Catalogue number
-
Lot number
-
Date of manufacture
Customer Support
Manufacturer:
pro3dure medical GmbH
Am Burgberg 13 · 58642 Iserlohn, Germany
Phone: +49 (0)2374 920050-0 · Fax +49 (0)2374 920050-50
info@pro3dure.com |
www.pro3dure.com | Made in Germany
Distributor:
SprintRay Europe GmbH Brunnenweg 11 · 64331 Weiterstadt · Germany
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>