TEMPO FIP100 Fiber Inspection Probe User Guide
- June 17, 2024
- Tempo
Table of Contents
TEMPO FIP100 Fiber Inspection Probe
Specifications
General Characteristics:
- Operating Temperature: 0 to 50°C
- Operating Humidity: 90% at 30°C, non-condensing
- Storage Temperature: -40 to +70°C
- Storage Humidity: Up to 95% at 35°C, non-condensing
Product Usage Instructions
1. General Information
1.1 General
The Fiber Inspection Probe FIP100 is designed to be connected to a mobile device. It is used for inspecting fiber connections.
1.2 Conditions of Application
The Probe should be operated in temperatures ranging from 0 to 50°C with a humidity level of 90% at 30°C, non-condensing. Storage temperature should be between -40 to +70°C with humidity not exceeding 95% at 35°C.
2. Performance Specifications
Performance specifications details are not provided in the extracted text.
3. Safety Information
Safety information details are not provided in the extracted text.
4. FIP100 Overview
4.1 Controls and Interfaces
Ensure the optical input is protected with the dust cover when the device is not in use. Make sure there is no optical power in the measuring fiber before connecting the device.
4.2 Multifunctional Button Operation
The multifunctional button on the FIP100 serves various purposes:
- Powers FIP100 ON / OFF.
- In Live Image mode, allows for auto-focus, image capture, analysis, and save results.
- In Captured Image mode, press to return to Live Image mode.
4.3 Multifunctional LED Indicator States
Nr. | Action | State |
---|---|---|
1 | Power ON | Firmware loading |
FAQs
- Q: What adapter types are available for the FIP100?
- A: The available adapter types include SCUPC, LCUPC, 2.5MMUPC, 1.25MMUPC, SCAPC, 2.5MMAPC, 1.25MMAPC, and LCAPC.
“`
Fiber Inspection Probe FIP100
USER GUIDE-2-
GENERAL INFORMATION
1.1. General The Fiber Inspection Probe FIP100 (Probe) connected to a mobile
device, is designed to:
– visually characterize the end surface cleanliness of fiber optic connectors,
– visually display the position of defects on the ferrule surface, –
automatically identify ferrule defects, – evaluation of the ferrule to meet
the requirements of existing standards or user criteria. The probe is a
portable meter with a self-powered, plug-in wireless Wi-Fi network to the
mobile device (iOS or Android). Probe allows to store measured data (the image
of the optical fiber end face in a digital form) and the analysis results in
the internal memory of the mobile device, and transmit stored data report
using a mobile communication device application.
Features of this product are protected by the Patents: US9217688B2;
GB201016991D0.
Each FIP100 includes:
FIP100 main unit
1 pc
2.5mm Universal UPC Tip (2.5MMUPC)
1 pc
Retractable Lanyard
1 pc
Power Supply with NA, UK and EU adapters
1 pc
Quick Reference Card
1 pc
Soft carrying case
1 pc
Available adapter types:
SCUPC BH LCUPC BH 2.5MMUPC 1.25MMUPC SCAPC BH 2.5MMAPC 1.25MMAPC LCAPC BH
SC/UPC Bulkhead Tip LC/UPC Bulkhead Tip 2.5mm Universal UPC Tip 1.25mm Universal UPC Tip SC/APC Bulkhead Tip 2.5mm Universal APC Tip 1.25mm Universal APC Tip LC/APC Bulkhead Tip
1.2. Conditions of application 0 to 50oC; humidity 90% at 30oC, non- condensing. 1.3. Storage Temperature -40 to +70oC; humidity non-condensing up to 95% at temperature 35oC.
-3-
PERFORMANCE SPECIFICATIONS
2.1. Field of View – Live mode and Captured: 612 x 460 m 2.2. Manual Detection
Capability (minimum) – 0.25 m 2.3. Auto Analysis Resolution <1.0 m 2.4. Stored
Image Size – 2592 x 1944 (5M) pixels 2.5. Maximum Allowable Live Fiber Power
Level – +20 dBm (Image cannot be viewed if fiber is live) 2.6. Wi-Fi
Characteristics – IEEE 802.11bng 2.7. Button Functionality – Power On/Off (>3
secs); Capture/Analysis/Auto-save/Live 2.8. Main LED Functionality – Blue =
Power On, Green = Pass, Red = Fail, Cyan = No Fiber 2.9. Live Image Frames per
Second – 15 2.10. Magnification – Variable from 80X to 700X, in Live and
Capture modes 2.11. Applications Compatibility – Android 4.0.3, iOS 8.1 2.12.
Image Capture with Pass/Fail Analysis – IEC 61300-3-35 (2015), user-set
criteria 2.13. Image File Format – JPEG, GIF 2.14. Image & Pass/Fail Results
Storage – Yes 2.15. File Storage Capacity – Remote Device limited 2.16. Result
Manager – Storage, rename, delete, transfer 2.17. Reporting – Built-in
fillable PDF reporter 2.18. Supported Languages – English, French, German,
Japanese, Korean, Russian, Spanish
GENERAL CHARACTERISTICS
2.19. Battery type – Li-Ion, non-field replaceable 2.20. Maximum Charger
Current Draw – 1.2A, battery charge current + device consumption current 2.21.
Operating Time (typical) – 60 hours; 8 hours continuous 2.22. Recharge Time –
4 hours 2.23. Low-Battery Warning – Viewed on smart device 2.24. Charging LED
Status
Red = Charging, Green = Fully Charged, Blinking Red/Green = Battery Fault.
Charging status viewed on smart device. 2.25. Power Save Features (Controlled
by iOS & Android App) Probe Auto-Off disabled, 1, 5, 30 minutes; Probe Wi-Fi
Not Connected 5 minutes 2.26. AC Charger – 100-240VAC, 50/60Hz, 5VDC, 2A
2.27. Weight / Dimensions – 150 g / diam.40 x 226 mm
-4-
SAFETY INFORMATION
It is important to always keep the optical sensor window clean and free of
debris. Use lint free wipes and an optical cleaning fluid such as 99%
isopropyl alcohol to clean the sensor window. Keep all surfaces of the Probe
clean and free of dust and water whenever possible. Always store the Probe in
the provided carry case to avoid exposure to contaminants. Prior to use,
carefully check cleanliness of sensor surface. Do not use nonstandard optical
connectors and plugs with badly polished butt-ends. They can damage the sensor
surface. Scratched or contaminated optical connectors can impact the
performance of the device.
ATTENTION! ! Never look directly into the optical outputs of fiber optic
network equipment, test equipment, patch cords and jumpers. Laser radiation is
harmful to eyes. ! Never inspect connectors with live fiber power level more
+20 dBm to avoid a damage device. ! Use only the specified AC adapter. Use of
another type of AC adapter can damage the instrument and create the danger of
fire and electrical shock. To avoid the danger of fire and electrical shock:
Never use a voltage that is different from that for which the AC adapter is
rated. Do not plug the unit into a power outlet that is shared by other
devices. Never modify the power cord or excessively bend, twist, or pull it.
Do not allow the power cord to become damaged. Do not place heavy objects on
the power cord or expose it to heat. Never touch the AC adapter while your
hands are wet. Should the power cord become seriously damaged (internal
wiring exposed or shorted), contact the manufacturer to request servicing.
-5-
OVERVIEW
FIP100 OVERVIEW
4.1. Controls and interfaces
1 Multifunctional button Press and release the button to turn the FIP100
probe On. Press and hold button for over 3 seconds to turn the probe OFF. 2
LED status indicator will display Flashing Blue light 3 Optical input.
ATTENTION! It is necessary to protect optical input with dust cover (4) if the
device is not used.
4 Dust cover 5 Optical adapter 6 Retractable Lanyard
ATTENTION! Be sure that there is no optical power in measuring fiber before
connecting the device to this fiber.
) General appearance
b) Marking and recharger’s compartment cover 7 5 VDC input jack 8
Recharger’s compartment cover 9 Charging indicator 10 Label with serial
number
-6-
4.2. Multifunctional Button Operation -1 (Picture a)
– Powers FIP100 ON / OFF. – In the Live Image mode, press the Multifunctional
button to perform the following: – Auto Focus image (if the Auto Focus option
is enabled) – Capture the displayed image and enter the Captured Image mode –
Analyze image (if the Pass/Fail option is enabled) – Save results and in the
memory of WiFi connected Smart Device (if Auto Save is enabled) – In the
Captured Image mode, press the Multifunctional button to return to the Live
Image mode.
4.3. Multifunctional LED Indicator States
Nr. Action
State
1 Power ON
Firmware loading
2 Press button 3-5 times
Upgrade mode, loading Firmware
3 Firmware flashing
Firmware Flashing. After flashing FIP100 enter to Powered ON
4 Powered ON
Ready to connect, send Wi-Fi SSID beacon
5 Smart device Wi-Fi connected Wi-Fi connected, Apps not active
6 Smart device Wi-Fi disconnected
Wi-Fi disconnected, Wi-Fi SSID beacon
7 App starts
Live mode
8 App closed/standby
Wi-Fi connected, Apps not active
9 Image capture and Analysis Pass
10 Image capture and Analysis Fail
11 Image capture and Analysis No Fiber
12 Image capture without Analysis, Menu
Not Analyzed, Menu
LED color/Behavior Blue blinks 3 times: 1s ON / 1s OFF Blue blinks: 0.5s ON /
0.5s OFF Blue blinks 0.2s ON / 0.2s OFF
Blue blinks: 2s ON / 1s OFF White lit steadily Blue blinks: 2 s ON / 1s OFF
Blue lit steadily White lit steadily Green lit steadily Red lit steadily Cyan
lit steadily Purple lit steadily
4.4. Battery Charging and Operation
To charge battery of the FIP100 probe: 1. Plug the included AC Charger into AC
outlet. 2. Connect charger plug into 5 VDC jack on the FIP100 probe. 3.
Charging indicator displays the battery charging status as follows:
LED color/Behavior Off Red Green Red/Green flashing
Battery Charging States
AC is not connected
Battery is charging
Battery is fully charged
Charging error: · Verify that correct 5VDC 2A charger is used · Allow the
FIP100 probe to cool before charging
4. FIP100 probe charges while operating.
-7-
5. FIP100 APP OVERVIEW
5.1. Downloading FIP100 App Download FIP100 App from Google Play
or Apple Store
5.2. Connecting the FIP100 to a Smart Device On FIP100 probe:
– Press and release the / button to power On; – LED status indicator will
display Flashing Blue light indicating that FIP100 is in “WiFi Beacon Mode”
ready to connect to your smart device.
On your Smart Device:
– Navigate to Settings; – Tap Wi-Fi. Make sure Wi-Fi is turned on; – Locate
your FIP100 device (as example screen below) and tap on it; – Enter password;
– Tap Connect; – Note the `Connected’ status; – Your FIP100 probe is now
connected to your smart device.
On FIP100 probe:
– Note that LED status indicator changes to White indicating that FIP100 is
connected, but FIP100 App is not active;
– Launch the FIP100 App on your smart device; – Note that LED status indicator
changes to Blue indicating that FIP100 is connected and FIP100 App
is active in Live Mode.
-8-
6. FIP100 APP USER INTERFACE OVERVIEW
6.1. Mobile Device Screen Format FIP100 App supports vertical and horizontal
screen formats. Switching from vertical format to horizontal is performed by
turning the mobile device vertically/horizontally. Note: `Screen Rotation’
feature needs to be enabled in the mobile device Settings menu.
6.2. Supported Touch Screen Gestures
The FIP100 App display is controlled from the touch screen of the mobile device using a supported set of gestures.
Touch Gesture
Description
Tap Briefly touch surface with fingertip
Function Select
Screens All screens
Double tap
Enter edit mode
Rapidly touch surface twice
with fingertip
Captured Image,
Drag Move fingertip over surface without losing contact
Flick/Swipe Quickly brush surface with fingertip
Pan, scroll Switching modes
Live Image, Captured Image Captured Image
Pinch Touch surface with two fingers and bring them closer together
Spread Touch surface with two fingers and move them apart
Zoom out Zoom in
Live Image, Captured Image Live Image, Captured Image
-9-
LIVE IMAGE MODE
7.1. Live Image Screen Features FIP100 App starts in the Live Image mode,
which enables the user to view live images of a fiber end-face and perform
visual inspection of the displayed fiber end-face image on a mobile device
1. Screen title: Indicates that FIP100 operates are in the Live image mode.
2. Live video streaming status: This row indicates frame per second data and
kB/s transfer rate, focus position and signature for auto focusing. 3. Battery
status field: indicates percentage of battery charge
– percentage of charge, – green color battery charge >20%, – red color low
battery <20% warning, – yellow color battery charging. 4. AF (auto-focus)
control: Tap to initiate auto-focus. 5. Focus control slider: Drag Focus
slider or tap ( + ) / ( ) to increase/decrease focal length . 6. Zoom
controls: · Tap ( + ) / ( ) for increasing/decreasing the scale of the live
fiber-end image . · Tap A to reset the nominal Zoom 7. Live image of the fiber
that is being inspected. · The fiber image can be scaled and positioned on the
screen by pinching/spreading/dragging the displayed fiber image. · Tapping on
the fiber image will turn ON/Off the Focus and Zoom control buttons. 8. Menu
on-screen button: Tapping Menu displays the Main Menu that allows the user to
set preferences, manage saved test results and perform other non-test
functions. 9. Capture on-screen button: Tap to initiate Image capture and
enter the Capture screen. 10. Back on-screen button: Tap to return to the
previous screen.
-10-
MAIN MENU
8.1. Main Menu Overview
The Main Menu is accessed from Live Image mode by tapping the Menu on-screen
button. Main Menu contains various Settings that are used to select user
preferences, perform general settings, manage saved test results, and perform
other non-test functions.
The Main Menu screen contains the following controls:
–
/ (On/Off) -Tapping this control enables/disables the selected Setting.
– Right arrow – Tapping this control will either display a pop-up menu or a sub-screen allowing to configure the selected Setting.
– Back on-screen button – Tap this control to return to the previous screen – Live Image mode. The Main Menu allows configuring Settings as follows
Auto Focus 1. Tap the / control to enable / disable auto focus. · If set to On, then pressing the Capture button will automatically focus before capturing an image. · When set to Off, the fiber image will be captured at the current focus position.
Auto Save
2. Tap on the Right arrow to display a pop-up menu · Choose from the
available Auto Save options: All, Pass only, None. · When Auto Save is
enabled, then tapping Capture button from Live Image mode will cause FIP100 to
perform the following: Auto Focus (if enabled) Capture image Perform
Pass/Fail Analysis (if enabled) Save image and pass/fail results to the
configured Job/Cable folder
1
2
-11-
Pass / Fall 3. Tap the
/ control to enable/disable Pass/Fail Analysis.
3. Tap on the Right arrow to display a sub-screen that allows configuring Pass/Fail Criteria
Beeper 4. Tap the / control to enable/disable a sound signal that indicates
Pass/Fail Analysis results. When
enabled, FIP100 will produce High tone beep for Passed results and Low tone
beep for Failed or No Fiber results
Auto-Power Off
5. Tap on the Right arrow to display a pop-up menu – Choose from the
available Auto-Power options:1 min, 5 min, 30 min, Never.
Language 6. Select the desired Language option from the list of available
options.
-12-
Report Settings
7. Configure reports by adding various data: Customer, Company, Location,
Operator, and Comments. These report settings will be applied to all saved
result.
Result Manager
8. Tap on the Right arrow results.
to display the Result Manager, which allows the user to manage stored
Last Image
9. Tap on the Right arrow to recall the most recently viewed fiber end-face image.
Import 10. Allows the user to import saved fiber images from the Photo Gallery.
Device Info 11. Tap on the Right arrow
to view Application, WiFi SSID and Device information.
-13-
CAPTURED IMAGE MODE
9.1. Captured Image Views
Once a fiber end-face image has been captured and analyzed (if enabled),
inspection test results will be displayed in the Captured Image screen.
Tapping on the displayed fiber end-face image allows the user to switch the
display modes sequentially, as follows:
1 – Image of the inspected fiber end-face with Pass/Fail Analysis overlay
(default display mode) 2 – Pass/Fail Analysis results only 3 – Image of a
fiber end-face only
1
2
3
Captured Image screen features three Display Views Identified by View Tabs: 1
– Fiber Image View identified by Fiber Image Tab
2 – Pass/Fail Results Table View identified by Pass/Fail Results Table Tab 3 –
Image Information View identified by Image Information Tab Switching between
Display Views is done by tapping on the View Tab icon or by swiping the screen
to the left or right. The active Tab icon is displayed on a highlighted in
Orange color background
1
2
3
-14-
9.2. Fiber Image View 1 – Screen title: Indicates date and time of the test –
[Captured @ hh:mm:ss] 2 – Pass/Fail Analysis indication: Shown only if
Pass/Fail Analysis option is enabled in the Main Menu. Depending on performed
analysis, this field will display:
Pass for Passing fiber, Fail for Failing fiber, No fiber No fiber if No Fiber
detected during Pass/Fail Analysis 3 – Display Tabs icons with the Image tab
selected (the active tab icon will is displayed on a highlighted in Orange
color background A. Switching between views is done by swiping the screen to
the left or right or by directly taping on the tab icon.
Fiber End-face Image; Pass/Fail Results Table; Image Information. 4 –
Zoom controls: · Tap ( + ) / ( ) for increasing/decreasing the scale of the
live fiber-end image . · Tap B to reset the nominal Zoom 5 – End-face image:
displays end-face image with Pass/Fail overlay (if enabled) – Passing
scratches/defects highlighted in GREEN C – Failing scratches/defects
highlighted in RED D – Analyzed zones represented by BLUE circles E 6 –
Analyze on-screen button: Tap to perform analysis of defects / scratches. 7 –
Save on-screen button: Only displayed is Auto Save is disabled in Main Menu.
Tap to manually save captured image and Pass/Fail results (if performed). 8 –
Back on-screen button: Tap to return to the Live Image mode.
-15-
9.3. Pass / Fall Results Table View 1 – Screen title: Indicates date and time
of the test – [Captured @ hh:mm:ss] 2 – Pass/Fail Analysis indication: Shown
only if Pass/Fail Analysis option is enabled in the Main Menu. Depending on
performed analysis, this field will display:
Pass for Passing fiber, Fail for Failing fiber, No fiber No fiber if No Fiber
detected during Pass/Fail Analysis 3 – Display Tabs icons with the Pass/Fail
display tab selected (the active tab icon will is displayed on a highlighted
in Orange color background A. Switching between views is done by swiping the
screen to the left or right or by directly taping on the tab icon.
Fiber End-face Image; Pass/Fail Results Table; Image Information. 4 –
Analysis Rule applied to determine Pass/Fail 5 – Analysis Zones (if enabled):
A – core, B – cladding, C – adhesive, D – contact area. 6 – Scratch analysis
results for each zone: – Reports the number of detected scratches exceeding
limit for each region. – Highlights failed rules in RED. 7 – Defect analysis
results for each zone: · Reports the number of detected defects exceeding
limit for each region. · Highlights failed rules in RED. 8 – Analyze on-screen
button: Tap to perform analysis of defects/scratches. 9 – Save on-screen
button: Only displayed is Auto Save is disabled in Main Menu. Tap to manually
save captured image and Pass/Fail results (if performed). 10 – Back on-screen
button: Tap to return to the Live Image mode.
-16-
9.4. Image Information View
1 – Screen title: Indicates date and time of the test – [Captured @ hh:mm:ss]
2 – Pass/Fail Analysis indication: Shown only if Pass/Fail Analysis option is
enabled in the Main Menu. Depending on performed analysis, this field will
display:
Pass for Passing fiber, Fail for Failing fiber, No fiber No fiber if No Fiber
detected during Pass/Fail Analysis
3 – Display Tabs icons with the Pass/Fail display tab selected (the active tab
icon will is displayed on a highlighted in Orange color background A.
Switching between views is done by swiping the screen to the left or right or
by directly taping on the tab icon.
Fiber End-face Image; Pass/Fail Results Table; Image Information.
4 – Results ID field: Parameters displayed in this field Job, Scope End, Far
End, Cable, Fiber identify the directory (Job/Cable/Fiber) into which the
inspected fiber end-face image and Pass/Fail Analysis results will be saved.
Tapping on icon of any parameter in this field will display the Result Manager
screen, which allows the user to view/change/edit details as needed Tapping on
Swap Ends B will swap test ends.
5 – Time & Date field: indicates Time & Date of the displayed captured image.
6 – Rule field: indicates the applied Pass/Fail Analysis rule. Tapping on Pass/Fail Rules screen, which allows the user to view/change/edit Rules.
icon displays the
7 – Report Settings: Tap on icon to customize reports by adding
Customer/Company/Location/Operator information. Customized report information
will be applied to the currently captured image only. For next captured image,
will be used report information defined in the Menu > Report settings
8 – Analyze on-screen button: Tap to perform analysis of defects/scratches 9 –
Save on-screen button: Only displayed is Auto Save is disabled in Main Menu.
Tap to manually save captured image and Pass/Fail results (if performed). 10 –
Back on-screen button: Tap to return to the Live Image mode.
-17-
10. PASS/FAIL CRITERIA 10.1. Pass/Fail Screen Features Configuring Pass/Fail
criteria is done in the Pass/Fail screen, which is accessed from Live Image
mode > Main Menu screen. A From the Live Image screen, tap Menu B In the Main
Menu, tap on the Right arrow next to the Pass/Fail option to display a sub-
screen that allows the user to enable, select and configure Rules for
Pass/Fail Analysis.
1 – On/Off / control – Tap the to enable/disable Pass/Fail Analysis. · If
enabled, then tapping on the Capture button in the Live Image screen will
automatically start analysis of the fiber end-face image according to the
selected Rule. · If disabled, then no analysis will be performed on the
Capture button press. 2 – IEC Rules: a set of Rules per IEC 61300-3-35
standard. These Rules are not editable. · Tap the desired IEC Rule (shown as
Folder image) to select. · The selected Rule will be marked with Red check
mark (as in screen example – C · Double tap the desired IEC Rule to view. 3 –
User Rules: a set of Rules created by users. These Rules are editable. · Tap
the desired Rule (shown as Folder image) to select. · The selected Rule will
be marked with Red check mark. · Double tap the desired Rule to view and/or
edit criteria. 4 – Tap Back button to return to the previous screen.
-18-
10.2. IEC Pass/Fail Analysis
IEC 61300-3-35 defines connector inspection Pass/Fail criteria. Pass/Fail
criteria depends on: · Fiber type (SMF or MMF) · Connector end-face Zones
A: Core B: Cladding C: Adhesive (between cladding and ferrule) D:
Physical contact area · Type of end face flaw: Scratches Defects
(contamination, particles, chips) · Size of flaw
10.3. Example IEC Rule
Shown below is the display when the IEC rule for single-mode PC connector with
return loss 45 dB is selected. The table identifies the number of scratches or
defects of a certain size, which may be allowed in each of the core, cladding,
and contact regions.
Zone A: Core 0-25 microns
B: Cladding 25-115 microns
D: Contact 135-250 microns
Scratch >0 µm: 0 No scratches >0 µm allowed >3 µm: 0 No scratches >3 µm
allowed
Blank No limitations on
scratches
Defect
0 µm: 0 No defects >0 µm
allowed 2-5 µm: 5 Up to 5 defects 2-5 µm diameter allowed >5 µm: 0 No defects 5 µm allowed >10 µm;0 No defects >10 µm allowed
-19-
10.4. Configuration Pass/Fail Criteria Enable/Disable Pass/Fail Analysis ·
Access the Main Menu screen. · Tap the / control A to enable/disable Pass/Fail
Analysis. Select Rule · Tap on the Right arrow to display a list of available
Rules (preset IEC and user-set rules). · Tap the desired Rule (shown as Folder
image) to select. · Note that selected Rule will be marked with Red check mark
B. View Rule · To view Rule details double tap on the desired Rule folder. ·
Settings for Zone, Scratches and Defects will be displayed C. · Tap on desired
field to view settings. Zone settings D Displays enabled for analyses zones
(A, B, C, D) and inner and outer diameters of zones. Scratches settings E
Displays configured rules for scratches for appropriate zone, scratches size
and numbers. Defects settings F Displays configured rules for defects for
appropriate zone, defects size and numbers. · When done viewing, tap Back
button to return to the previous screen.
-20-
Edit User Rule: · To edit User Rule double tap on the desired User Rule A or
similar Standard Rule B (any available rule). · Settings for Zone, Scratches
and Defects will be shown. · Tap on desired field to view and edit settings.
Rule name C: allows to change Rule name. Standard rule name or selected
User Rule name will be replaced by new User Rule name. Zone settings D:
allows to turn On/Off zones A, B, C, D for analyses and change inner and outer
diameters of zones. Scratches settings E: allows to set rules for scratches
for appropriate zone, scratches size and numbers. Defects settings F:
allows to set rules for defects for appropriate zone, defects size and
numbers.
-21-
· While in any of Settings screen, Zone/Scratches/Defects: Tap the On/Off –
/ control to enable/disable the desired Rule parameter. Once enabled, tap
on the desired field to edit the parameter value. Using on screen Editor,
change the selected value as needed. Tap Done G. · Tap Back H to return to
the previous screen. · Tap Save J to save edits.
-22-
RESULT MANAGER
11.1. Result Manager Screen
To access the Result Manager
· From the Live Image mode, tap Menu
to display the Main Menu.
· Tap on the Right arrow next to the Result Manager option. · Result Manager screen displays saved test results organized into Jobs, Cables, Fibers. Screen Features 1 – Jobs field: Displays a list of saved Jobs (folders). · Drag Jobs field horizontally to navigate in Job folders. · Field title A indicates the number of Jobs saved and the name/or number of Job(s) selected. 2 – Cables field: Displays a list of saved Cables (folders). · Drag Cables field horizontally to navigate in Cables folders. · Field title B indicates the number of Cables saved and the name/or number of Cable(s) selected. 3 – Fibers field: Displays a list of saved fiber end-face inspection results. · Drag Fiber field vertically to navigate in the Fibers. · Field title C indicates the number of Fibers saved and the name/or number of Fiber(s) selected. 4 – Add button: Double tap this button to create new Job or Cable. 5 – Group button: Tapping on this button enables multi-select functionality 6 – Share button: Tap on this button to display a pop-up menu that allows the user to Rename (Single Item Selection only), Send to, Send Report, or Delete operations. 7 – Back button: Tap this button to return to the previous screen.
-23-
11.2. Creating a New Job/Cable Folder 1 – In the Result Manager screen, double
tap Add in the Jobs field. 2 – In the Job Editor screen, note that new Job is
created with a default name format: Job YYYMMDD 3 – Edit the default Job name
as needed, then tap Accept A or Done B. 4 – When back in the Result Manager
screen, note that the new Job folder is created and selected (displayed in
Blue color). 5 – In the Cables field, double tap Add . 6 – In the Cable Editor
screen, you may either change numbers for Scope End/Far End/Cable or accept
default numbers, then tap Accept C or Done D. 7 – When back in the Result
Manager screen, note that the new Job folder is created and selected
(displayed in Blue color). Note: The selected Job/Cable folder will become
current and all following tests will be saved to this currently selected
Job/Cable folder.
-24-
11.3. Renaming a Job/Cable/Fiber To Rename Saved Test Results 1 – Navigate to
the desired item, Job/Cable/Fiber; tap to select. 2 – Tap on the Share button.
3 – Tap RENAME. 4 – Using the Editor displayed on screen, make changes as
needed. 5 – Once done editing, tap Accept A or Done B .to save changes.
-25-
11.4. Saving Test Results Saving Test Results for the First Time If test
results are saved for the first time and no Job/Cable folder previously
created in the Result Manager, then default settings for Job/Scope End/Far
End/Cable/Fiber will be used, where: 1 – Job Folder name = YYYYMMDD; 2 – Scope
End = 100 3 – Far End = 200; 4 – Cable number = 001; 5 – Fiber number = 1 The
default settings for Job/Scope End/Far End/Cable/Fiber may be changed as
needed. · When the captured fiber end image is displayed, tap on the
Information tab to display the Information Details screen. · Tap on the Right
arrow in the desired parameter line. · Double tap on the desired
Job/Cable/Fiber. · Make edits as needed using the on-screen Editor. · When
done, tap Accept A or Done B. · Tap Check to save changes. · Tap On Report
Settings if need to change Customer, Company, Location, Operator or Comments
for this result · Tap Save to save test results.
-26-
11.5. Saving Test Results to a New Job/Cable/Folder
To save test results to a new Job/Cable folder 1 – Capture an image. Once the
captured image is displayed, tap on the Information tab. 2 – In the
Information Details screen, tap on the Right Arrow in the Job line. 3 – Double
tap Add . 4 – Note that new Job name is displayed in the default format –
YYYYMMDD. 5 – Edit Job name as needed. Tap Accept or Done. 6 – The new Job
folder is created and selected (displayed in Blue color). 7 – Tap Check to
save changes. 8 – Note that newly created Job folder is now current. If
needed, you may change Scope End, Far End, Cable, and Fiber. Make sure to
tap Check to save changes. If needed, you may change Report Settings –
Customer, Company, Location, Operator or Comments for this result
9 – When done, tap Save to save test results in the newly created folder.
Note: this will make the newly created Job/Cable folder current and all
following test results will be saved to this Job/Cable folder.
1
2
3
4
5
6, 7
8, 9
-27-
11.6. Saving Test Results to an Existing Job/Cable/Folder
To save test results to an existing Job/Cable folder 1 – Capture an image.
Once the captured image is displayed, tap on the Information tab. 2 – In the
Information Details screen, look at the Job Folder name – this is the current
folder where the test result will be saved. 3 – To select a different
destination Job folder, tap on the Right Arrow . 4 – Note that the current Job
folder is displayed in Blue color. 5 – Locate the desired destination Job
folder and tap on it to make it current. 6 – Tap Check . 7 – When back in the
Information Details screen, note that the Job line is now displays the
selected folder as current. 8 – If needed, you may change Report Settings –
Customer, Company, Location, Operator or Comments for this result 9 – Tap Save
to save test results in the selected Job folder. Note: all following test
results will be saved to the selected Job/Cable folder.
1
2, 3
4
5, 6
7, 8, 9
-28-
11.7. Result Manager Functionality To Select a Single Item – Job/Cable/Fiber ·
Tap on the desired Job/Cable/Fiber icon to select it. If a Job (Cable) folder
contains Cables (Fibers) then tapping on that folder will expand it to show
contents. · Note that color of the selected item changes from Yellow (folder)
/ White (fiber) to Blue.
/ – Item is not selected / – Item is selected To View Saved Test Result ·
Navigate to the desired fiber. · Note that fiber icons indicate results as
follows:
– Passing fiber; – Failing fiber – Fiber is not analyzed · Double tap on the
desired Fiber to view saved fiber end-face image. To Select Multiple Items –
Jobs, Cables, Fibers · Tap on the Group button – A to enable multi-select
functionality. · Note that button color changes to Blue – B. · Tap on as many
items as needed to multi-select them. · Note that color of the selected items
changes to Blue – C. · Note that name of the selected group changes to reflect
the number of selected items – D.
-29-
To Delete Saved Test Results 1 – Select a single item or multiple items for
deletion. To select a single item, navigate to the desired item
(Job/Cable/Fiber), and then tap on it to select. To select multiple items,
tap on the Group button to enable multi-select functionality, and then tap on
as many items as needed to multi-select. 2 – Tap on the Share button. 3 – From
the displayed pop-up menu, select DELETE. 4 – When prompted, confirm deletion.
-30-
Sharing Captured Test Results Saved test results may be transferred to another
location using your mobile device applications that support transfer
functionality. Note: Prior to transferring, make sure your mobile device is
connected to the Internet. 1 – Select a single item or multiple items for
transfer. To select a single item, navigate to the desired item
(Job/Cable/Fiber), and then tap on it to select. To select multiple items,
tap on the Group button to enable multi-select functionality, and then tap on
as many items as needed to multi-select. 2 – Tap on the Share button. 3 – From
the displayed pop-up menu select:
– SEND TO to share captured test results, or – SEND REPORT to share .pdf
reports. 4 FIP100 will generate zipped file with Job-Cable-Fiber path. 5 –
From the displayed list, select the desired transferring option. Follow on-
screen instructions.
–
-31-
Shared Test Results and Reports Results or reports zip file contains
appropriate Job>Cables>Fibers folder structure for all selected
Job/Cables/Files. This allows easy to add new fibers result into existed
folders on PC/Database.
Job folder of shared results contain at least one Cable folder:
Cable folder of shared results contain: 1 at least one jpg file of captured
2592×1944 pixels image; 2 at least one gif file with analysis data of captured
2592×1944 pixels image
1
2
In addition to
_1N for not analyzed results;
_1P for passed results:
_1F for failed results.
-32-
Each Job folder of shared reports contain at least one Cable folder. A
summary.pdf file also included to Job folder, if multiply results pdf report
was generated:
The summary.pdf file contain this Job Summary Report with Pass/Fail status,
names and Data/Time for result files:
Cable folder of shared report contain at least one pdf report:
In addition to
-33-
Fiber .pdf report include: Global Pass/Fail verdict; General Information with
fillable fields. Fields marked blue contain Report Settings information of
saved result and may be changed later by user; Graphics images with and
without Analysis; Analyzes Results sheet; Device info with device model, SN,
FW version and App version information.
-34-
12. Operations with adapter tips Select Adapter Tip for connector to be inspected
Description SC/UPC Bulkhead Tip
TEMPO CAT NO Tip SCUPC BH
LC/UPC Bulkhead Tip
LCUPC BH
2.5mm Universal UPC Tip 2.5MMUPC 1.25mm Universal UPC Tip 1.25MMUPC
SC/APC Bulkhead Tip
SCAPC BH
2.5mm Universal APC Tip 2.5MMAPC
1.25mm Universal APC Tip 1.25MMAPC
LC/APC Bulkhead Tip
LCAPC BH
Connector
Installing Adapter Tips – While holding adapter tip in place, raise captive
nut engages with adapter tip threads – Turn captive nut counter-clockwise
(viewed from front/top) – Hand-tighten (do not over-tighten)
Removing Adapter Tips – Probe tips are held in place with a captive nut – Turn
captive nut clockwise (viewed from front/top) – Continue until captive nut is
free of adapter tip – Remove adapter tip by pulling it gently straight off the
probe base
Recommendation on use Universal APC patchcord tips – Insert ferule of
connector to be inspected to Universal patchcord tip
-35-
– Set focus position in the middle of scale – Slow rotate ferule of connector
in tip to get contrast image – Press AF to focus at ferule end-face
Recommendation on use APC bulkhead tips – Align keys on tip and bulkhead –
Insert tip into bulkhead – ferule of connector to be inspected to Universal
patchcord tip
13. Inspecting fibers with FIP100 solution
1. Get your FIP100 Probe ready. · Make sure battery is fully charged. · Make
sure that optical port is equipped with the appropriated adapter tip.
2. Connect FIP100 probe to the FIP100 App installed on your mobile device.
3. Insert the fiber under test into adapter tip.
4. On your FIP100 App you should see Live video of the fiber being inspected.
You may perform visual inspection of the displayed fiber end-face image.
5. From the Live Image Screen display Main Menu. Configure test options as
needed.
6. Tab Back button to return to the Live Image mode.
7. When back in the Live Image mode, initiate test by either pressing the
Multifunctional button on the FIP100 Probe or tapping the Capture on-screen
button on the Live Image screen.
8. According to test configurations set in the Main Menu (Auto Focus,
Pass/Fail, Auto Save). FIP100 will capture, inspect and save the fiber end-
face image and Pass/Fail Analysis.
WARRANTY
The device is guaranteed for two years from the date of original sale. The manufacturer will repair or replace the device free of charge if it fails because of a manufacturing defect. This warranty is valid if the device is used for its intended purposes only (as described in the Instruction manual) and is not damaged or misused.
Address:
TEMPO COMMUNICATIONS, INC
1390 ASPEN WAY VISTA CALIFORNIA 92081
United States
Fiber Inspection Probe FIP100 S/N ___
Date: __
-36-