GREENLEE CMI-1000 Industrial Clamp Meters Instruction Manual
- June 17, 2024
- GREENLEE
Table of Contents
INSTRUCTION MANUAL CM-1360 • CM-1560
CMH-1000 • CMI-1000
Clamp Meters
Read and understand all of the instructions and safety information in this manual before operating or servicing this tool.
Description
The Greenlee CM-1360, CM-1560, CMH-1000, and CMI-1000 Clamp Meters are hand-
held testing devices with the following measurement capabilities: AC and DC
voltage, AC current, AC current AmpTip™, frequency, non-contact electric field
(EF) detection, resistance, and capacitance. They also check diodes and verify
continuity. These meters are designed to be placed on or removed from
insulated or uninsulated conductors.
Additional functionality specific to each model includes:
Function | CMI-1000 | CM-1560 | CM-1560 |
---|---|---|---|
AC + DC Voltage | • | • | |
DC Current AmpTip™ | • | • | |
DC + AC Current AmpTip™ | • | • | |
DC Current | • | • | |
Flashlight | • | • | |
µDC Current | • | • | |
Temperature (K-style thermocouple) | • | • | |
Motor Rotation | • | • | |
Phase Rotation | • | • | |
Peak RMS Capture | • | • | • |
5 ms Crest | • | • | • |
Other capabilities include:
- Hold Mode to capture the present LCD value
- Record Minimum, Maximum, or Average Mode
- Zero Mode
- Intelligent Auto Power Off
- Automatic ranging
- Backlit LCD
Safety
Safety is essential in the use and maintenance of Greenlee tools and
equipment. This instruction manual and any markings on the tool provide
information for avoiding hazards and unsafe practices related to the use of
this tool. Observe all of the safety information provided.
Purpose of This Manual
This instruction manual is intended to familiarize all personnel with the safe
operation and maintenance procedures for the Greenlee CM-1360, CM-1560,
CMH-1000, and CMI-1000 Clamp Meters.
Keep this manual available to all personnel. Replacement manuals are available
upon request at no charge at www.greenlee.com.
**** Do not discard this product or throw away!
For recycling information, go to www.greenlee.com.
All specifications are nominal and may change as design improvements occur.
Greenlee Textron Inc. shall not be liable for damages resulting from
misapplication or misuse of its products.
® Registered: The color green for electrical test instruments is a registered
trademark of Textron Innovations Inc.
AmpTip is a trademark of BTC.
Important Safety Information
SAFETY ALERT SYMBOL
** This symbol is used to call your attention to hazards or unsafe
practices which could result in an injury or property damage. The signal word,
defined below, indicates the severity of the hazard. The message after the
signal word provides information for preventing or avoiding the hazard.
DANGER
Immediate hazards which, if not avoided, WILL result in severe injury or
death.
WARNING
Hazards which, if not avoided, COULD result in severe injury or death.
CAUTION
Hazards or unsafe practices which, if not avoided, MAY result in injury or
property damage.
WARNING
Read and understand this material before operating or servicing this
equipment.
Failure to understand how to safely operate this tool could result in an
accident causing serious injury or death.
WARNING
Electric shock hazard:
Contact with live circuits could result in severe injury or death.
WARNING
Electric shock and fire hazard:**
- Do not expose this unit to rain or moisture.
- Do not use the unit if it is wet or damaged.
- Clamp meter, test leads or any other clamp accessory, when used to make a measurement, create a System. The System is rated for CAT III 1000 V or CAT IV 600 V when using the test leads or accessories provided with the meter. The System CAT and voltage rating is limited by the lowest rated component in the System when using test leads or accessories not provided with the meter.
- Inspect the test leads or accessory before use. They must be clean and dry, and the insulation must be in good condition. Do not use the test lead if the contrasting inner layer of insulation is visible.
- Use this unit for the manufacturer’s intended purpose only, as described in this manual. Any other use can impair the protection provided by the unit.
Failure to observe these warnings could result in severe injury or death.
WARNING
Electric shock hazard:
- Do not apply more than the rated voltage between any two input terminals, or between any input terminal and earth ground.
- Keep hands and fingers below the barriers on the test leads and the clamp meter body.
Failure to observe these warnings could result in severe injury or death.
WARNING
Electric shock hazard:
- Do not operate with the case open.
- Before opening the case, remove the test leads from the circuit and shut off the unit.
Failure to observe these warnings could result in severe injury or death.
WARNING
Electric shock hazard:
-
Unless measuring voltage, current, or frequency, shut off and lock out power. Make sure that all capacitors are discharged. Voltage must not be present.
-
Set the selector and connect the test leads so that they correspond to the intended measurement.
Incorrect settings or connections can result in incorrect measurements or damage to the unit. -
Using this unit near equipment that generates electromagnetic interference can result in unstable or inaccurate readings.
Failure to observe these warnings could result in severe injury or death.
CAUTION
Electric shock hazard:
- Do not change the measurement function while the test leads are connected to a component or circuit.
- Do not clamp the jaw around a conductor carrying a frequency greater than 400 Hz. The magnetic circuitry may reach a hazardous temperature if this frequency is exceeded.
Failure to observe these precautions may result in injury and can damage the unit.
CAUTION
Electric shock hazard:
- Do not attempt to repair this unit. It contains no user-serviceable parts.
- Do not expose the unit to extremes in temperature or high humidity. Refer to “Specifications.”
Failure to observe these precautions may result in injury and can damage the unit.
Identification
- Antenna for non-contact EF detection
- Jaw center indicator for best current accuracy; polarity indicator for DC amps
- Selects a function or turns power off
- LCD
- Input jack for phase and motor rotation functions only
- Positive input terminal; L2 input for motor and phase rotation
- Negative, common, or ground input terminal; L1 input for motor and phase rotation
- Refer to explanation under “Using the Features.”
- Jaw lever
- Keep hand or fingers below the dashed line during operation.
- Jaw
- Center indicator for AmpTip™ function
Symbols on the Unit
Warning—Read the instruction manual
Warning—Risk of electric shock
Earth ground
Double insulation
AC (alternating current)
DC (direct current)
Application around or removal from hazardous live conductors is permitted.
Display Icons
- CCrest
- R Record function is enabled.
- Diode
- Continuity
- PEAK-RMS mode is enabled.
- Phase rotation main supply
- Motor rotation
- 8.8.8.8 Numeric display
- µ Micro (10-6)
- F Farad
- A Ampere
- V Volt
- Hz Hertz (frequency in cycles per second)
- Ω Ohm
- k Kilo (103
- AmpTip™ function is enabled.
- ∆ Relative zero function is enabled.
- – Polarity indicator
- VFD AC measurement with filter for accuracy on variable frequency drives
- **** AC measurement is selected.
- + AC + DC measurement is selected.
- **** DC measurement is selected.
- Low battery
- H Hold function is selected.
- MAX Maximum value being displayed
- MIN Minimum value being displayed
- AVG Average value being displayed
- AUTO Automatic ranging is enabled.
Using the Features
All Models
-
Intelligent Auto Power Off (APO): The meter automatically turns off after approximately 32 minutes of inactivity to extend battery life. Active meter function is specified as: selector switch is rotated or pushbuttons are pressed; readings above 8.5% of the range; non-OL readings for resistance, continuity, or diode function; non-zero readings for Hz function; phase rotation segment moving. To wake up the meter from APO, press the SELECT button momentarily and release or turn the rotary switch OFF and then back on. Always turn the rotary switch to the OFF position when the meter is not in use.
-
: Press the button momentarily to toggle the LCD backlight. The backlight automatically turns off after 32 seconds to extend battery life.
-
: Press the button momentarily to toggle the flashlight. The flashlight automatically turns off after 32 seconds to extend battery life.
-
C CREST: Press for 1 second to activate. LCD will C indicate when mode is active. This mode will capture current or voltage peaks in duration of as small as 5 ms. Press CREST to toggle between MIN and MAX. Add 250 digits to accuracy for changes > 5 ms.
-
H HOLD: Press momentarily to hold the present value on the display. Press again to exit this mode.
-
R REC: Press momentarily to activate MAX/MIN/AVG recording mode. R and “MAX MIN AVG” appear on the display. The meter beeps when new maximum or minimum reading is updated. The average reading is calculated over time. Press the button momentarily to read the MAX, MIN, and AVG readings in sequence. Press the button for 1 second to exit MAX/MIN/AVG recording mode. APO is disabled automatically in this mode.
-
SELECT: Press momentarily to toggle between functions, measurement modes, or display modes. The last used setting becomes the default when that function is selected again. To change the default, select a new setting. The setting will be stored in nonvolatile memory.
-
∆: Press the ∆ button for 1 second to toggle between zero and normal mode. In zero mode the value on the LCD when the button is pressed becomes the offset. As the input value changes, the offset is subtracted from the actual value. This modified value is shown on the LCD.
Note: ∆ operation when DCA or DC + ACA functions are selected offsets any magnetic hysteresis in the jaw. If DCA or DC + ACA reading before measuring is not zero, press the ∆ button for 1 second. The LCD will display “dc_0”. When this clears, the meter is ready for operation. When measuring more than 5 A, this mode is disabled. As an indication, the meter will beep three times. CM-1560, CMH-1000, and CMI-1000 Only -
PEAK-RMS: Press the PEAK-RMS button for 1 second to activate PEAK-RMS mode to capture inrush current or voltage in duration as short as 80 ms. P-RMS appears on the display. Press the button for 1 second to exit PEAK-RMS mode. APO is disabled automatically in this mode.
CMH-1000 and CMI-1000 Only -
and : If desired, press and hold the REC button while turning the meter ON to enable the Beeper feature. If the segments indicate a clockwise movement, the beeper sounds a single long beep per segment cycle. If the segments indicate a counterclockwise movement, the beeper sounds three short beeps per segment cycle.
-
: Ensure the motor is electrically isolated from its power source. Press the SELECT button momentarily to select Motor function. Connect the meter to the motor as described in the “Settings Table” section of this manual. From the perspective of looking down the shaft of the motor, spin it clockwise. If the meter indicates a clockwise movement, the motor leads connected to L1, L2, and L3 of the meter are T1, T2, and T3, respectively. If the meter indicates a counterclockwise movement, swap any two connects between the meter and motor. Then retest.
-
: Press the SELECT button momentarily to select Phase function. Connect the meter to the phases as described in the “Settings Table” section of this manual. If the meter indicates a clockwise movement, the phases connected to L1, L2, and L3 of the meter are L1, L2, and L3, respectively. If the meter indicates a counterclockwise movement, swap any two connects between the meter and phases. Then retest.
Using the Test Leads
WARNING
Electric shock hazard:
The test leads supplied with this product comply with safety standards that
limit the exposed length of the probe tip to 4 mm for measurement categories
III and IV. These test leads include a cap that must be in place when used in
measurement category III or IV applications.
Failure to observe this warning could result in severe injury or death. AC
Measurement
AC measurements are usually displayed as RMS (root mean square) values. The
RMS value is equal to the value of a DC waveform, which would deliver the same
power if it replaced the time-varying waveform.
Two AC measurement methods are average-responding RMS calibrated and true RMS-
reading.
The average-responding RMS calibrated method takes the average value of the
input signal after full wave rectification, multiplies it by 1.11, and
displays the result. This method is accurate if the input signal is a pure
sine wave.
The true RMS-reading method uses internal circuitry to read the true RMS
value. This method is accurate, within the specified crest factor limitations,
whether the input signal is a pure sine wave, square wave, triangle wave, half
wave, or signal with harmonics. The ability to read true RMS provides much
more measurement versatility. The Greenlee CM-1360, CM-1560, CMH-1000, and
CMI-1000 are true RMS meters.
The Waveforms and Crest Factors table shows some typical AC signals and their
RMS values.
Waveforms and Crest Factors
Waveform| | | |
---|---|---|---|---
RMS Value| 100| 100| 100| 100
Average Value| 90| 100| 87| 64
Crest Factor* | 1.414| 1| 1.73| 2
-
The crest factor is the ratio of the peak value to the RMS value; it is represented by the Greek letter .
AC + DC True RMS
AC + DC true RMS calculates both of the AC and DC components given by the expression when making measurements and responds accurately to the total effective RMS value regardless of the waveform. Distorted waveforms with the presence of DC components and harmonics may cause:- Transformers, generators, and motors to overheat
- Circuit breakers to trip prematurely
- Fuses to blow
- Neutrals to overheat due to the triplen harmonics present on the neutral
- Bus bars and electrical panels to vibrate
CM-1560 and CMI-1000 can be set to display AC true RMS or AC + DC true RMS.
AC Bandwidth
AC bandwidth is the range of frequencies over which AC measurements can be
made within the specified accuracy. It is the frequency response of the AC
functions—not of the frequency measurement functions. A meter cannot
accurately measure the AC value with frequency spectrums outside its
bandwidth. Complex waveforms, noise, and distorted waveforms contain frequency
components that are much higher than the fundamental; for example, high
frequency noise on a 50/60 Hz power line.
Operation
WARNING
Electric shock hazard:
Contact with live circuits could result in severe injury or death.
Operating Procedure
-
Refer to the “Settings Table.” Set the selector to the proper setting, press SELECT (when instructed to do so), and connect the test leads to the meter.
-
Refer to “Typical Measurements” for specific measurement instructions.
-
Test the unit on a known functioning circuit or component.
• If the unit does not function as expected on a known functioning circuit, replace the batteries.
• If the unit still does not function as expected, call Greenlee for technical assistance at 800-435-0786. -
Take the reading from the circuit or component to be tested.
Settings Table
To measure this value …| Set the selector to this symbol …| These icons appear
on the
display …| Connect the red lead to …| Connect the black lead to …
---|---|---|---|---
ALL MODELS
Voltage (AC)| | V | +| COM
Voltage (DC)| | V | +| COM
Current, AmpTip™ (AC)| | | N/A| COM
Current (AC)| | | N/A| COM
Continuity| | Ω| +| COM
Resistance| Ω| kΩ| +| COM
Diode| | V| +| COM
Frequency| Hz| Hz| +| COM
Electric Field (EF)| EF| EF| N/A| COM
Capacitance| | µF| +| COM
CM-1560 and CMI-1000 Only
Voltage (AC + DC) | | +| COM
Current, AmpTip™ (DC)| | | N/A| N/A
Current, AmpTip™ (AC + DC)| | | N/A| N/A
Current (DC)| | | N/A| N/A
Current (AC + DC)| | | N/A| N/A
CMH-1000 and CMI-1000 Only
Temperature| Temp| C or F|
Current (µDC)| | µA ****| +| COM
Phase Rotation†| | | +| COM
Motor Rotation†| | | +| COM
- Use the top side of the stationary jaw, or for more precise measurements use a probe in the COM terminal.
** Discharge capacitor before measurement. Refer to “Typical Measurements” regarding polarized capacitors.
*** Thermocouple + connects to +, thermocouple – connects to COM.
† Yellow lead connects to L3.
Typical Measurements and Accuracy
Refer to the “Specifications” section for operating conditions and temperature
coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a fixed
amount) at 23 °C ± 5 °C (73.4 °F ± 9 °F).
Frequency must be within the specified bandwidth for non-sinusoidal waveforms.
Crest factors, unless otherwise specified, are as follows:
- Crest factor < 2.5:1 at full scale
- Crest factor < 5.0:1 at half scale
Frequency
Function | **Sensitivity (Sine RMS)*** | Range |
---|---|---|
600 V, 1000 V | 50 V | 5.00 Hz ~ 999.9 Hz |
60.0 A AmpTip™ | 40 A | 40.00 Hz ~ 400.0 Hz |
60.00 A, 600.0 A, 1000 A | 40 A | 40.00 Hz ~ 400.0 Hz |
- DC bias must be less than 50% of the Sine RMS value.
Accuracy of Frequency Range
Display Range | Accuracy |
---|---|
xx.xx | ± (1.0% + 0.05 Hz) |
xxx.x | ± (1.0% + 0.5 Hz) |
Voltage
Measurement AC Voltage
Range| Accuracy at 20 Hz to 50 Hz| Accuracy at 50 Hz to 60 Hz| Accuracy at 60
Hz to 200 Hz| Accuracy at 200 Hz to 400 Hz
---|---|---|---|---
600.0 V| ± (1.5% + 0.5 V)| ± (0.8% + 0.5 V)| ± (1.5% + 0.5 V)| ± (10% + 0.5 V)
1000 V| ± (1.5% + 5 V)| ± (0.8% + 5 V)| ± (1.5% + 5 V)| ± (10% + 5 V)
DC Voltage
Range | Accuracy | Range | Accuracy |
---|---|---|---|
600.0 V | ± (0.8% + 0.5 V | 1000 V | ± (0.8% + 5 V) |
AC + DC Voltage (CM-1560 and CMI-1000 only)
Range| Accuracy at
40 Hz to 50 Hz| Accuracy at DC,
50 Hz to 60 Hz| Accuracy at
60 Hz to 200 Hz| Accuracy at
200 Hz to 400 Hz
---|---|---|---|---
600.0 V| ± (1.8% + 0.7 V)| ± (1.0% + 0.7 V)| ± (1.8% + 0.7 V)| ± (12% + 0.7 V)
1000 V| ± (1.8% + 7 V)| ± (1.0% + 0.7 V)| ± (1.8% + 7 V)| ± (12% + 0.7 V)
Input Impedance: 10 MΩ, 100 pF nominal
Current Measurement
Typical Measurements and Accuracy (cont’d)
µA DC Current
μDC Current
(CMH-1000 and CMI-1000 only)
Range | Accuracy | Burden Voltage |
---|---|---|
200.0 μA | ± (1.0% + 0.5 μA) | 3.5 mV/μA |
200.0 μA | ± (1.0% + 5 μA) |
AmpTip™ Function AmpTip™ AC Current
Range| Accuracy Range| Accuracy at 40 Hz to 100 Hz†|
Accuracy at 100 Hz to 400 Hz†
---|---|---|---
60.00 A| 0.00 A to 3.99 A| ± (1.5% + 0.15 A)| ± (2.0% + 0.15 A)
4.00 A to 60.00 A| ± (1.5% + 0.15 A)| ± (2.0% + 0.05 A)
† Induced error for adjacent current carrying conductors is < 0.02 A/A.
AmpTip™ DC Current (CM-1560 and CMI-1000 only)
Range | Accuracy Range | Accuracy† |
---|---|---|
60.00 A | 0.00 A to 3.99 A | ± (1.5% + 0.15 A) |
4.00 A to 60.00 A | ± (1.5% + 0.05 A) |
† Induced error for adjacent current carrying conductors is < 0.02 A/A.
AmpTip™ DC + AC Current (CM-1560 and CMI-1000 only)
Range| Accuracy Range| Accuracy at DC, 40 Hz to 100 Hz†|
Accuracy at 100 Hz to 400 Hz†
---|---|---|---
60.00 A| 0.00 A to 3.99 A| ± (2.0% + 0.17 A)| ± (2.2% + 0.17 A)
4.00 A to 60.00 A| ± (2.0% + 0.07 A)| ± (2.2% + 0.07 A)
† Induced error for adjacent current carrying conductors is < 0.08 A/A.
Note: Use Δ if there is an offset before the measurement is taken.
Typical Measurements and Accuracy (cont’d)
Current Measurement (cont’d)
AC and DC Clamp AC Current (CM-1560 and CMI-1000 only)
Range| Accuracy Range| Accuracy at
40 Hz to 100 Hz†| Accuracy at
100 Hz to 400 Hz†
---|---|---|---
60.00 A| 0.00 A to 8.99 A| ± (1.8% + 0.15 A)| ± (2.2% + 0.15 A)
9.00 A to 60.00 A| ± (1.8% + 0.05 A)| ± (2.2% + 0.15 A)
600.0 A| 60.0 A to 600.0 A| ± (1.8% + 0.5 A)| ± (2.2% + 0.5 A)
1000 A| 600 A to 1000 A| ± (1.8% + 5 A)| ± (2.2% + 5 A)
- Maximum crest factor < 1.4:1 at full scale and < 2.8:1 at half scale.
† Induced error for adjacent current carrying conductors is < 0.02 A/A.
AC Current (CM-1360 and CMI-1360 only)
Range| Accuracy Range| Accuracy at 40 Hz to 100 Hz†|
Accuracy at 100 Hz to 400 Hz†^
---|---|---|---
60.00 A| 0.00 A to 5.99 A| ± (1.8% + 0.15 A)| ± (2.2% + 0.15 A)
6.00 A to 60.00 A| ± (1.8% + 0.05 A)| ± (2.2% + 0.05 A)
60.00 A| 6.00 A to 60.00 A| ± (1.8% + 0.5 A)| ± (2.2% + 0.5 A)
1000 A| 600 A to 1000 A| ± (1.8% + 5 A)| ± (2.2% + 5 A)
- Maximum crest factor < 1.4:1 at full scale and < 2.8:1 at half scale.
† Induced error for adjacent current carrying conductors is < 0.05 A/A.
^ For any measurement, when the conductor is not positioned on the center of the jaws, add 2.0% to the accuracy for position error.
DC Current (CM-1560 and CMI-1000 only)
Range | Accuracy Range | Accuracy† |
---|---|---|
60.00 A | 0.00 A to 8.99 A | ± (1.8% + 0.15 A) |
9.00 A to 60.00 A | ± (1.8% + 0.05 A) | |
600.0 A | 60.0 A to 600.0 A | ± (1.8% + 0.5 A) |
1000 A | 600 A to 1000 A | ± (1.8% + 5 A) |
† Induced error for adjacent current carrying conductors is < 0.02 A/A.
AC + DC Current (CM-1560 and CMI-1000 only)
Range| Accuracy Range| Accuracy at DC, 40 Hz to 100 Hz†| Accuracy at 100 Hz
to 400 Hz†
---|---|---|---
60.00 A| 0.00 A to 8.99 A| ± (2.2% + 0.17 A)| ± (2.5% + 0.17 A)
4.00 A to 60.00 A| ± (2.2% + 0.07 A)| ± (2.5% + 0.07 A)
600.0 A| 60.0 A to 600.0 A| ± (2.2% + 0.7 A)| ± (2.5% + 0.7 A)
1000 A| 600 A to 1000 A| ± (2.2% + 7 A)| ± (2.5% + 7 A)
- Maximum crest factor < 1.4:1 at full scale and < 2.8:1 at half scale.
† Induced error for adjacent current carrying conductors is < 0.08 A/A.
Note: Use Δ if there is an offset before the measurement is taken.
Typical Measurements and Accuracy (cont’d) Resistance
Range | Accuracy |
---|---|
600.0 Ω | ± (1.0% + 0.5 Ω) |
6.000 kΩ | ± (1.0% + 0.005 kΩ) |
60.00 kΩ | ± (1.0% + 0.05 kΩ) |
Continuity
Tone Threshold: Between 10 Ω and 250 Ω
Response Time: Approximately 32 ms Capacitance
Range | Accuracy* |
---|---|
200.0 μF | ± (2.0% + 0.4 μF) |
2500 μF | ± (2.0% + 4 μF) |
Diode Test
Measuring Range: 2.000 V
Accuracy: ± (1.5% + 0.005 V)
Test Current (typical): 0.3 mA
Open Circuit Voltage (typical): < 3.5 VDC
Typical Measurements and Accuracy (cont’d)
Temperature (CMH-1000 and CMI-1000 only)
Range | Accuracy* |
---|---|
-40.0 °C ~ 99.9 °C | ± (1.0% + 0.8 °C) |
100 °C ~ 400 °C | ± (1.0% + 1 °C) |
-40.0 °F ~ 211.8 °F | ± (1.0% + 1.5 °F) |
212 °F ~ 752 °F | ± (1.0% + 2 °F) |
- Thermocouples supplied with meter meet standard JIS C 1602-1981.B
Non-contact EF Detection
Typical Voltage | Range at 50 Hz to 60 Hz | LCD Indication |
---|---|---|
20 V | 10 V ~ 36 V | – |
55 V | 23 V ~ 83 V | – – |
110 V | 59 V ~ 165 V | – – – |
220 V | 124 V ~ 330 V | – – – — |
440 V | 250 V ~ 1000 V | – – – – – – |
Specifications
Display: LCD, 3-5/6 digits 6000 counts
Polarity: Automatic
Update Rate: 5 per second nominal
Operating Conditions:
-10 °C to 10 °C (14 °F to 50 °F), relative humidity non-condensing
11 °C to 30 °C (51.8 °F to 86 °F), relative humidity ≤ 90%
30 °C to 40 °C (86 °F to 104 °F), relative humidity ≤ 75%
40 °C to 50 °C (104 °F to 122 °F), relative humidity ≤ 45%
Pollution degree: 2
Indoor use only
Altitude:
Operating: Below 2000 m (6500 ft)
Storage: 3000 m (9843 ft)
Storage Temperature: -20 °C to 60 °C, relative humidity < 80%, with batteries
removed
Temperature Coefficient: Nominal 0.10 x (Accuracy) per °C below 18 °C from -10
°C to 18 °C (14 °F to 64.4 °F) and above 28 °C from 28 °C to 50 °C (82.4 °F to
122 °F) or otherwise specified
Sensing: True RMS
E.M.C.:
Meets EN61326-1:2006 (EN55022, EN61000-3-2, EN61000-3-3, EN61000-4-2,
EN61000-4-3,
EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11)
In an RF field of 1 V/m, resistance modes all models and DC µA (CMH-1000 and
CMI-1000):
Total Accuracy = Specified Accuracy + 80 digits
In an RF field of 1 V/m at frequencies 200 MHz ~ 300 MHz, DCA, DC+AC A and ACA
mode:
Total Accuracy = Specified Accuracy + 60 digits
In an RF field of 3V/m, all other modes and models not specified above:
Total Accuracy = Specified Accuracy + 20 digits
Safety: Double insulation per UL/IEC/EN61010-1 Ed. 3, IEC/EN61010-2-033 Ed. 1,
CAN/CSA C22.2 No.
61010-1 Ed. 3, IEC/EN61010-2-032 Ed. 2 & IEC/EN61010-031 Ed. 1.1 to CAT III
600V and CAT IV 300V AC and DC
Overload Protections:
Jaws:
Current and Frequency Function: 1000 A DC/AC RMS at ≤ 400 Hz
Terminals:
Voltage and Phase Rotation Function: 1100 V DC/AC RMS
All Other Modes and Functions: 1000 V DC/AC RMS
Transient Protection: 8.0 kV (1.2/50 μs surge)
Power Supply: Two 1.5 V AA batteries (IEC LR06)
Low Battery:
Capacitance and Frequency Functions: Approximately 2.85 V
All other Functions: Approximately 2.5 V
Current Consumption:
CM-1560 and CMI-1000: Typically 13 mA
CM-1360 and CMH-1000: Typically 4.3 mA
APO Timing: Idle for 32 minutes
APO Consumption: 5 μA typical
Dimensions: 258 mm x 94 mm x 44 mm (10.1″ x 3.7″ x 1.7″)
Jaw Opening and Conductor Diameter: 51 mm max (2.0″)
Statement of Conformity
Greenlee Textron Inc. is certified in accordance with ISO 9001 (2000) for our
Quality Management Systems.
The instrument enclosed has been checked and/or calibrated using equipment
that is traceable to the National Institute for Standards and Technology
(NIST).
Measurement Categories
These definitions were derived from the international safety standard for
insulation coordination as it applies to measurement, control, and laboratory
equipment. These measurement categories are explained in more detail by the
International Electrotechnical Commission; refer to either of their
publications: IEC 61010-1 or IEC 60664.
Measurement Category II
Local level. Appliances, portable equipment, and the circuits they are plugged
into. Some examples include light fixtures, televisions, and long branch
circuits.
Measurement Category III
Distribution level. Permanently installed machines and the circuits they are
hard-wired to. Some examples include conveyor systems and the main circuit
breaker panels of a building’s electrical system.
Measurement Category IV
Primary supply level. Overhead lines and other cable systems. Some examples
include cables, meters, transformers, and other exterior equipment owned by
the power utility.
Maintenance
WARNING
Electric shock hazard:
Before opening the case, remove the test leads from the circuit and shut off
the unit.
Failure to observe this warning could result in severe injury or death.
Replacing the Batteries
-
Disconnect the unit from the circuit.
Turn the unit OFF. -
Remove the two screws from the battery access door and remove it.
-
Replace the batteries, making sure to observe the polarity.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use
abrasives or solvents.
If the meter will not be used for periods longer than 60 days, remove the
batteries and store them separately.
Lifetime Limited Warranty
Greenlee Textron Inc. warrants to the original purchaser of these goods for
use that these products will be free from defects in workmanship and material
for their useful life, excepting normal wear and abuse. This warranty is
subject to the same terms and conditions contained in Greenlee Textron Inc.’s
standard one-year limited warranty.
For all Test Instrument repairs, contact Customer Service at
800-435-0786 and request a Return
Authorization.
For items not covered under warranty (such as items dropped, abused, etc.), a
repair cost quote is available upon request.
Note: Prior to returning any test instrument, please check replaceable
batteries or make sure the battery is at full charge.
Register this product at www.greenlee.com
4455 Boeing Drive • Rockford, IL 61109-2988 • USA •
815-397-7070
An ISO 9001 Company • Greenlee Textron Inc. is a subsidiary of Textron Inc.
www.greenlee.com
USA
Tel: 800-435-0786
Fax: 800-451-2632
Canada
Tel: 800-435-0786
Fax: 800-524-2853
International
Tel: +1-815-397-7070
Fax: +1-815-397-9247
References
- Manual-Hub.com - Free PDF manuals!
- Professional. Nothing Less.™ | Greenlee
- Manual-Hub.com – Free PDF manuals!