GREENLEE TG-2000 Infrared and Contact Thermometer
Read and understand all of the instructions and safety information in this manual before operating or servicing this tool.
Register this product at www.greenlee.com/ Registre este producto en www.greenlee.com
Description
The Greenlee TG-2000 Infrared and Contact Thermometer is a hand-held surface
temperature sensing device. The non-intrusive measuring capability of this
unit is safer and quicker than other methods, and is well suited to checking
the surface temperature of circuit breakers, motors, HVAC/R equipment,
engines, etc.
This device measures surface temperature by gathering three forms of infrared
energy — emitted, transmitted, and reflected — from an object and translating
that energy into a temperature.
This temperature instrument includes the following features:
- Dual laser sighting. The two lasers indicate the approximate diameter of the measurement area.
- User-adjustable high and low temperature alarms.
- Automatic recording of maximum, minimum, difference, and average temperatures while making measurements.
- Probe jack. Allows the thermometer to measure contact temperature using any K-type thermocouple with a standard mini connector.
- Lock mode. Allows long-term measurements without holding the trigger. The lasers are turned off in Lock mode unless the trigger is held down.
Safety
Safety is essential in the use and maintenance of Greenlee tools and
equipment. This instruction manual and any markings on the tool provide
information for avoiding hazards and unsafe practices related to the use of
this tool. Observe all of the safety information provided.
Purpose of this Manual
This instruction manual is intended to familiarize all personnel with the safe
operation and maintenance procedures for the Greenlee TG-2000 Infrared and
Contact Thermometer.
Keep this manual available to all personnel.
Replacement manuals are available upon request at no charge at
www.greenlee.com
Do not discard this product or throw away! For recycling information, go to www.greenlee.com
All specifications are nominal and may change as design improvements occur. Greenlee Tools, Inc. shall not be liable for damages resulting from misapplication or misuse of its products.
Important Safety Information
- Immediate hazards which, if not avoided, WILL result in severe injury or death.
- Hazards which, if not avoided, COULD result in severe injury or death.
- Hazards or unsafe practices which, if not avoided, MAY result in injury or property damage.
- Read and understand this material before operating or servicing this equipment. Failure to understand how to safely operate this tool could result in an accident causing serious injury or death.
- Electric shock hazard:
Do not use a thermocouple probe to measure the temperature of electrically energized objects such as terminals, fuses, or bare wires.
Contact with live circuits could result in severe injury or death.
WARNING: Electric shock and fire hazard:
- Do not expose this unit to rain or moisture.
- Do not use the unit if it is wet or damaged.
- Use this unit for the manufacturer’s intended purpose only, as described in this manual. Any other use can impair the protection provided by the unit.
- Do not open the case, except for the battery compartment.
- Before opening the case, release the trigger, unplug the thermocouple probe from the unit, and shut off the unit.
- Avoid laser exposure.
- Do not point laser directly at eye or at reflective objects. Failure to observe these precautions may result in injury.
- Avoid contact with the probe until it has cooled. The thermocouple probe may remain hot for several minutes after making a temperature measurement.
- Do not exceed the temperature rating of the thermocouple probe. Consult the manufacturer’s specifications for the selected probe.
- Do not attempt to repair this unit. It contains no user-serviceable parts.
- Do not expose the unit to extremes in temperature or high humidity. Refer to “Specifications.” Failure to observe these precautions may result in injury and can damage the unit.
- Temperature readings will be inaccurate on highly reflective objects.
- Temperature readings will be inaccurate on objects if emissivity is not adjusted properly. Follow masking and emissivity adjustment procedures.
- Allow 30 minutes for the unit to adjust to large changes in ambient temperature; inaccurate readings can result.
- Clean lens periodically; inaccurate readings can result. Refer to the “Maintenance” section. Failure to observe these precautions may result in injury and can damage the unit.
IMPORTANT
- Using this unit near equipment that generates electromagnetic interference can result in unstable or inaccurate readings.
- Protect the unit against static electricity.
CE Certification
This instrument conforms to the following standards:
• EN50081-1:1992, Electromagnetic Emissions
• EN50082-1:1992, Electromagnetic Susceptibility
Readings may be affected if the unit is operated with a radio frequency electromagnetic field of approximately 3 volts per meter, but the performance of the instrument will not be permanently affected. From 200 to 600 MHz, the maximum error is 10 °C (18 °F).
CFR 21 1040.10 and 1040.11
Performance Standards for Light-Emitting Products
Identification
-
Display
-
Decrease button and °C/°F selector
-
Mode selector
-
Increase button and Lock button
-
Probe jack
-
Trigger
-
Battery compartment door
-
Lens
-
Lasers
-
EMPTY
Display Icons
-
Lock mode indicator
-
Full battery indicator
12A. Low battery indicator
12B. Dead battery indicator -
Laser ON indicator
-
Backlight ON indicator
-
High/low alarm ON indicator
-
Scan mode indicator; blinks when reading is updated
-
Celsius or Fahrenheit indicator
-
Hold mode indicator
-
Temperature indicator
-
Mode value indicator
-
Mode indicator
Symbols on the Unit
Using the Features
Basic Features
- To turn the laser on or off: Press and hold the trigger, and then momentarily press the button.
- To turn the backlight on or off: Press and hold the trigger, and then momentarily press the button.
- To change between Celsius and Fahrenheit: If the LCD is blank, press and release the trigger. Momentarily press the °C/°F button.
- To lock the trigger for continuous measurement: If the LCD is blank, press and release the trigger. Momentarily press the Lock button. Momentarily press the Lock button again to return to normal measurement. The unit will automatically exit Lock mode after 60 minutes.
Advanced Features
Press the Mode button to access the advanced features. They appear in this
order: Emissivity Display (E), Emissivity Setting (iEh), Maximum (MAX),
Minimum (MIN), Difference (dIF), Average (AVG), High Alarm Limit (HAL), Low
Alarm Limit (LAL), and Probe (PRB).
Setting Alarm Points
After selecting the HAL or LAL mode, use the or button to set the alarm
point. A tone will sound and the alarm indicator will flash if the temperature
is higher than or lower than the corresponding set point. “HI” or “LOW” will
appear with the alarm indicator.
Temperature Probe
To use a K-type thermocouple temperature probe (purchased separately), connect
the probe to the probe jack on the side of the thermometer. Select PRB mode.
The probe temperature will appear in the lower right corner of the display.
The temperature detected by the infrared sensor will continue to appear in the
center of the display. You may still use the Mode button to access the other
advanced features, but the probe temperature is displayed only when PRB mode
is selected.
Note: The probe will not trip the high alarm and low alarm.
Measurement Review
After the unit shuts off, press Mode to review the measurement:
- MAX — the highest recorded temperature among all of the readings.
- MIN — the lowest recorded temperature among all of the readings.
- dIF — the difference between the maximum and minimum readings.
- AVG — a weighted rolling average of all of the readings.
Note: Values for MAX, MIN, dIF and AVG are reset each time the trigger is pressed.
Note: The AVG feature acts as a filter to eliminate fluctuations from the temperature readings. It is not an arithmetic mean, but rather a rolling weighted average of the readings that smoothes the data and excludes sudden spikes.
Distance from Target
Thermometers measure the surface temperature of an object by sensing its
emitted, reflected, and transmitted energy. The sensor collects this energy
from the area being measured (spot size). The spot size varies as the distance
to the target changes. The farther the distance from the area being targeted,
the larger the spot size.
When determining the appropriate distance from the object, the spot size must
be somewhat smaller than the target area. When accuracy is critical, the spot
size should be half as large as the target. If the spot size is larger than
the target size, the thermometer will pick up the temperature of the back-
ground, resulting in inaccurate readings.
Object 1 is the correct distance from the unit. Note the target is slightly
larger than the spot size being measured.
Object 2 is too far away. The thermometer also measures the background because
the target is smaller than the instrument’s spot size. This can result in an
inaccurate reading.
Distance to Spot Size
The distance to spot size is a ratio that provides the appropriate distance to
the target.
The nominal distance-to-spot-size (D:S) ratio for TG-2000 is 16:1.
Operation
-
Refer to “Typical Measurements” for specific measuring instructions.
-
Test the unit on an object with a known temperature.
- If the unit does not function as expected on an object of known temperature, replace the batteries.
- If the unit still does not function as expected, call Greenlee for technical assistance at 800-435-0786.
-
To measure the temperature of a selected area: Press and hold the trigger and point the ther-mometer at the object to be measured. The dual lasers indicate the approximate diameter of the measured area. “SCAN,” the measured temperature, and advanced feature data will appear on the display. The scan icon will blink. Release the trigger after completing the measurement. “HOLD,” the final temperature measurement, and advanced feature data will remain on the display for
1 minute. -
To take more than one temperature measurement, or to locate a hot spot: Press and hold the trigger, and scan the objects to be measured. The unit will record the maximum, minimum, differ-ence, and average temperatures.
To perform other functions, refer to “Advanced Features” in the “Using the Features” section.
Limitations
- The thermometer cannot measure through transparent objects, such as glass or plastic. It will measure the surface temperature of the glass or plastic.
- Steam, dust, smoke, or other particles can prevent accurate measurement by obstructing the unit’s infrared sensor.
Locating a Hot Spot
To find a hot spot, aim the thermometer outside the area of interest, and then
scan across with an up and down motion.
Troubleshooting
Code | Problem | Action |
---|---|---|
Er2 | Thermometer has been exposed to rapid temperature change. | Allow at least |
30 minutes for thermometer to stabilize at new ambient temperature.
Er3| Ambient temperature is outside operating limits (refer to
“Specifications”).| Operate thermometer within limits in the “Specifications”
section.
Er5 ~ 9
| ****
Other errors.
| Reset thermometer by removing batteries for at least 1 minute. Reinstall
batteries and check operation.
HI| Measured temperature is above range of thermometer.| Use thermometer
within its specified range.
LO| Measured temperature is below range of the thermometer.| Use thermometer
within its specified range.
Typical Measurements
Emissivity
Emissivity is the ability of an object to emit infrared energy. Most objects or surfaces have an emissivity value of about 0.95. Some objects, such as polished or shiny metallic surfaces, may have much lower emissivity. Using an infrared thermometer to measure the temperature of such objects will produce inaccurate results.
• Temperature readings will be inaccurate on highly reflective objects.
• Temperature readings will be inaccurate on objects if emissivity is not
adjusted properly. Follow masking and emissivity adjustment procedures.
Masking a Surface to Determine Emissivity
To determine the effect of emissivity on the measurement, cover the surface
with masking
tape or flat black paint. (Be sure to use tape or paint that can withstand the
anticipated temperature.) The emissivity of both these materials is about
0.95. Allow time for the tape or paint to reach the tem-perature of the
material underneath. Measure the covered area and a nearby area. The readings
should be about the same. If they are not, the emissivity may affect your
measurements.
Compensating for Emissivity
- If the temperature of the surface is known, adjust the emissivity until the unit agrees with the known temperature.
- If the material of the surface is known, adjust the emissivity according to the Common Emissivities table on the following page.
- Otherwise, mask the surface as described above. Measure temperature of the masked area with emissivity set at 0.95. Measure temperature of unmasked surface and adjust emissivity until tem-peratures agree. Write down emissivity value for future readings on this surface.
Common Emissivities
Material | Emissivity |
---|---|
Aluminum | 0.30 |
Asbestos | 0.95 |
Asphalt | 0.95 |
Basalt | 0.70 |
Brass* | 0.50 |
Brick | 0.90 |
Carbon | 0.85 |
Ceramic | 0.95 |
Concrete | 0.95 |
Copper* | 0.95 |
Dirt | 0.94 |
Food (frozen) | 0.90 |
Food (hot) | 0.93 |
Glass (plate) | 0.85 |
Ice | 0.98 |
Material | Emissivity |
--- | --- |
Iron* | 0.70 |
Lead* | 0.50 |
Limestone | 0.98 |
Oil | 0.94 |
Paint | 0.93 |
Paper | 0.95 |
Plastic** | 0.95 |
Rubber | 0.95 |
Sand | 0.90 |
Snow | 0.90 |
Steel* | 0.80 |
Textiles | 0.94 |
Water | 0.93 |
Wood*** | 0.94 |
Accuracy
Accuracy Table for Infrared Sensor
Target Temperature | Ambient Temperature | Accuracy |
---|---|---|
15 °C to 35 °C | 25 °C | ± 1.0 °C |
-60 °C to 0 °C | **** |
20 °C to 26 °C
| ± (2.0–0.05 x reading) °C
0 °C to 625 °C| ± 2.0% of reading or 2 °C, whichever is greater
59 °F to 95 °F| 77 °F| ± 1.8 °F
-76 °F to 32 °F| ****
73 °F to 79 °F
| ± (5.2–0.05 x reading) °F
32 °F to 1157 °F| ± 2.0% of reading or 4 °F, whichever is greater
Accuracy for K-Type Thermocouple Measurements
Instrument error is ± 1.0% of reading or 1 °C (1.8 °F), whichever is greater.
Specification applies at ambient temperature of 17 °C to 29 °C (63 °F to 84
°F). Thermocouple range or accuracy not included. Consult thermocouple probe
data sheet for usable temperature range and accuracy specification.
Specifications
-
Display: 4–digit LCD (999.9 or 9999 maximum display)
-
Response Time: 1 second, 90% response
-
Temperature Range:
Infrared Sensor: -60 °C to 625 °C (-76 °F to 1157 °F)
K-Type Thermocouple Probe: -64 °C to 1400 °C (-83.2 °F to 2552 °F) -
Spectral Response: 8 to 14 µm
-
Repeatability: ± 0.5% of reading or ± 1 °C (± 2 °F), whichever is greater
-
Emissivity: 0.10 to 1.0, digitally adjustable
-
Ambient Operating Temperature Range: 0 °C to 50 °C (32 °F to 122 °F)*
-
Relative Humidity: 10% to 90% (non-condensing) at < 30 °C (86 °F) ambient temperature Altitude: 2000 m (6500′) maximum
-
Storage Conditions: -10 °C to 60 °C (14 °F to 140 °F)*.
Remove batteries. -
Battery: Two 1.5 volt batteries (AAA, NEDA 24A or IEC LR03)
-
For improved accuracy, allow unit 30 minutes to adjust to large change in temperature.
Statement of Conformity
Greenlee Tools, Inc. is certified in accordance with ISO 9001 (2000) for our
Quality Management Systems.
The instrument enclosed has been checked and/or calibrated using equipment
that is traceable to the National Institute for Standards and Technology
(NIST).
Maintenance
- Electric shock hazard:
- Do not attempt to repair this unit. It contains no user-serviceable parts.
- Failure to observe this precaution may result in injury and can damage the unit.
Battery Replacement
1. Pivot the top of the door down and away from the handle.
2. Replace the batteries (observe polarity).
3. Close the battery compartment door.
Cleaning Lens
- Blow off loose particles using clean compressed air.
- Gently brush any remaining debris away with a camel hair brush or soft cloth.
- Carefully wipe the surface with a water-moistened cotton swab.
Note: Do not use solvents!
Housing
Periodically wipe the housing with a damp cloth and mild detergent; do not use
abrasives or solvents.
Authorized Representative:
- Gustav Klauke GmbH
- Auf dem Knapp 46
- 42855 Remscheid
- Germany
www.greenlee.com
4455 Boeing Drive • Rockford, IL 61109-2988 • USA •
815-397-7070 An ISO 9001 Company •
©2021 Greenlee Tools, Inc.
USA
Tel: 800-435-0786
Fax: 800-451-2632
Canada
Tel: 800-435-0786
Fax: 800-524-2853
International
Tel: +1-815-397-7070
Fax: +1-815-397-9247
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>