Miele PG 8527 Laboratory Glassware Washer Instruction Manual
- June 16, 2024
- Miele
Table of Contents
- PG 8527 Laboratory Glassware Washer
- Product Information
- Specifications
- Product Usage Instructions
- Important Safety Instructions
- Use
- Safety Precautions
- Process Validation
- Routine Checks
- Cleaning the Filters in the Wash Cabinet
- Q: Can I use the Laboratory Glassware Washer PG 8527 / PG 8528
- Q: Can I install the machine myself?
- Q: What should I do if the machine is damaged?
PG 8527 Laboratory Glassware Washer
Product Information
Specifications
- Model: Laboratory Glassware Washer PG 8527 / PG 8528
- Country: US, CA
- Manual Link: https://manual-hub.com/
- M.-Nr: 07 207 090
Product Usage Instructions
Important Safety Instructions
This machine conforms to current safety requirements. However,
inappropriate use can lead to personal injury and property damage.
Read the operating instructions carefully before using this
machine.
Keep these instructions in a safe place and pass them on to any
future user.
Use
This machine is designed for commercial use and for specialized
applications only, as described in these Operating Instructions. Do
not use for purposes other than those for which it was designed, as
this could be dangerous.
This machine is intended for indoor use only.
The installation of this unit in non-stationary locations must
be performed by a qualified installer or service agency in strict
accordance with national and local safety regulations and
standards.
Safety Precautions
-
The machine should be commissioned and maintained only by a
Miele Service Technician. Repairs by unqualified persons could be
dangerous. -
Do not install the machine in an area where a danger of
explosion or freezing may be present. -
Be certain this appliance is properly installed and grounded by
an authorized technician. To guarantee the electrical safety of
this appliance, continuity must exist between the appliance and an
effective grounding system. It is imperative that this basic safety
requirement be met. If there is any doubt, have the electrical
system checked by a qualified technician. -
A damaged machine is dangerous. Turn off the machine
immediately at the power switch and call the Miele Service
Department. -
Personnel operating the machine should be trained regularly.
Children and untrained personnel must not be allowed access to the
machine or its controls. -
Take care when handling liquids such as detergents,
neutralizing agents, wetting agents, and rinse aids. These may
contain irritant or corrosive ingredients, acids, or alkalis. Never
use any organic solvents, as the danger of explosion exists. Wear
protective gloves and goggles. The manufacturer’s safety conditions
must be observed at all times. -
The water in the machine must not be used as drinking
water. -
Be careful when sorting items with sharp pointed ends. Position
them in the machine so that you do not hurt yourself or create a
danger for others.
Maintenance
Regular maintenance is important to ensure the proper
functioning and longevity of your Laboratory Glassware Washer PG
8527 / PG 8528. Follow these maintenance guidelines:
Process Validation
Perform process validation according to the instructions
provided in the manual to ensure the washer is operating correctly
and meeting required standards.
Routine Checks
Regularly perform routine checks on the washer to identify any
issues or abnormalities. Refer to the manual for detailed
instructions on what to check and how often.
Cleaning the Filters in the Wash Cabinet
Clean the filters in the wash cabinet regularly to prevent clogs
and maintain optimal performance. Follow the instructions in the
manual for proper cleaning procedures.
FAQ
Q: Can I use the Laboratory Glassware Washer PG 8527 / PG 8528
for residential purposes?
A: No, this machine is designed for commercial use only.
Q: Can I install the machine myself?
A: The installation of this unit in non-stationary locations
must be performed by a qualified installer or service agency in
strict accordance with national and local safety regulations and
standards.
Q: What should I do if the machine is damaged?
A: If the machine is damaged, turn off the machine immediately
at the power switch and call the Miele Service Department for
assistance.
Operating and Installation Instructions
Laboratory Glassware Washer PG 8527 / PG 8528
To prevent accidents and appliance damage read these instructions before installation or use.
en – US, CA
M.-Nr. 07 207 090
Contents
IMPORTANT SAFETY INSTRUCTIONS. . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . 4 Description of labwasher . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . 8 Guide to the labwasher . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . 9 Electronic controls . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . 12 Optional equipment . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . 14 Context menu . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . 15 Change the display
language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . 16 Change the operating level . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . 17 Delay start . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . 19 Change the heating type . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . 21 Current faults . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . 21 Loading package. . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Automatic mobile
unit recognition (AWK). . . . . . . . . . . . . . . . . . . . . . . . . . . .
. 22 Mobile unit coding . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . 22
Setting mobile unit coding . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . 23 Application . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Adding liquid detergents and neutralizers . . . . . . . . . . . . . . . . . .
. . . . . . . . . . 28 Dispensing systems. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Operation . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . 31 Main Switch – On/Off . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Turning On
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . 31 Starting a program . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Program sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . 34 End of program . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Turning off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . 34 Canceling a program . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . 34 Interrupting a program. . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . 35 Barcode scanner (optional) . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Barcode scan functions . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . 36 User scan load. . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
User scan unload. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . 37 Scan the load . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . 38 Loading packages . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . 41 Serial interface (data
transfer) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . 46
2
Contents
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . 47 Process validation . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . 48 Routine checks . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . 48 Cleaning the filters in the
wash cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
49
Cleaning the fine filter . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . 49 Cleaning the flat filter . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Cleaning the filter system for the circulation pump . . . . . . . . . . . . .
. . . . . . . . 50 Cleaning the spray arms . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . 51 Cleaning the control
panel and glass door (optional) . . . . . . . . . . . . . . . . . . . . . . 52
Cleaning the exterior . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . 52 Cleaning the wash cabinet . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Cleaning the door seal. . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . 52 Baskets and inserts . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Report printer (optional). . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . 54 Replacing the print paper. . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Replacing the ribbon cartridge. . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . 54 After sales service . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
INSTALLATION INSTRUCTIONS. . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . 57 Electrical connection . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Plumbing. . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . 59 Technical data . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 Caring
for the environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . 62
3
IMPORTANT SAFETY INSTRUCTIONS
This machine conforms to current safety requirements. However, inappropriate
use can lead to personal injury and property damage. Read the operating
instructions carefully before using this machine.
Keep these instructions in a safe place and pass them on to any future user.
Use
~ This machine is designed for
commercial use and for specialized applications only, as described in these
Operating Instructions. Do not use for purposes other than those for which it
was designed, as this could be dangerous.
~ This machine in intended for indoor
use only.
~ The installation of this unit in non
stationary locations must be performed by a qualified installer or service
agency in strict accordance with national and local safety regulations and
standards.
Please pay attention to the following notes to maintain safe procedures.
~ The machine should be
commissioned and maintained only by a Miele Service Technician. Repairs by
unqualified persons could be dangerous.
~ Do not install the machine in an area
where a danger of explosion or freezing may be present.
~ Be certain this appliance is properly
installed and grounded by an authorized technician. To guarantee the
electrical safety of this appliance, continuity must exist between the
appliance and an effective grounding system. It is imperative that this basic
safety requirement be met. If there is any doubt, have the electrical system
checked by a qualified technician.
~ A damaged machine is dangerous.
Turn off the machine immediately at the power switch and call the Miele
Service Department.
~ Personnel operating the machine
should be trained regularly. Children and untrained personnel must not be
allowed access to the machine or its controls.
4
IMPORTANT SAFETY INSTRUCTIONS
~ Take care when handling liquids
such as detergents, neutralizing agents, wetting agents and rinse aids. These
may contain irritant or corrosive ingredients, acids or alkalis. Never use any
organic solvents, as the danger of explosion exists. Wear protective gloves
and goggles. The manufacturer’s safety conditions must be observed at all
times.
~ The water in the machine must not
be used as drinking water.
~ Be careful when sorting items with
sharp pointed ends. Position them in the machine so that you do not hurt
yourself or create a danger for others.
~ When using this machine regard the
high temperatures and be especially careful not to scald or burn yourself.
When opening the door bypassing the electrical lock, a danger of burning,
scalding and corrosion exist. Let mobile units, inserts and cleaned items cool
before touching them. Any water which may remain in containers will be very
hot and must be emptied into the wash cabinet.
~ If you are exposed to toxic vapors or
processing chemicals, consult the manufacturer’s material safety data sheets
for emergency procedures.
~ Always allow mobile units, modules,
inserts, and loads to cool before removing. Be sure to empty any water from
concave items into the wash cabinet before removing.
~ After drying with the drying unit
open the door to allow the items and inserts to cool.
~ If the pre-heater has been
programmed to “Pre-heater standby”, be particularly careful of hot water and
steam when opening the door. Danger of burning or scalding. The water inlet
pipe to the wash cabinet is located, as viewed from the loading side,
underneath the basket runner on the left hand side.
~ Steam heating is permissible up to a
pressure of 145 psi. This corresponds to a water steam temperature of 354°F
(179 °C).
~ Never clean the machine near or
with a water or high pressure hose.
~ Before servicing, disconnect the
power supply by either removing the fuse, unplugging the unit or manually
“tripping” the circuit breaker.
,Mount on non-combustible floors
only!
5
IMPORTANT SAFETY INSTRUCTIONS
The following points should be observed to assist in maintaining quality
standards for critical labware and to avoid damage to the loads being cleaned.
~ Only use process chemicals that are
approved by their manufacturer for the application involved. Any negative
effects on labware and the washer itself are the liability of the chemical
manufacturer.
~ If toxic chemicals may be present in
the wash cabinet, beware of this risk when interrupting a program and opening
the washer. The door seal and steam condensor must always be checked for
soundness.
~ Use only Miele approved cleaning
agents with this machine. Use of unsuitable cleaning agents could adversely
affect the components of the machine. Damages resulting from using unsuitable
cleaning agents are not covered by the warranty.
~ Pre-treating (e.g. with cleaning
agents), certain soiling and cleaning agents with a chemical interaction, can
cause foam. For pre-treatment and/or cleaning only use low-sudsing detergents
which have been approved by Miele. Suds can have an adverse effect on the
operation of the machine.
~ When a chemical additive is
recommended, the manufacturer of the machine takes no responsibility for the
effect of the chemical on the items being cleaned.
~ The machine must only be operated
with water and the recommended cleaning agents. Organic solvents must not be
used in the machine as there is the danger, under certain circumstances, of
explosion and damage to rubber and synthetic materials. Follow the dosage
recommendations of the cleaning agent’s manufacturer.
~ In critical applications where very
high requirements have to be met, it is strongly recommended that all the
relevant factors for the process, such as cleaning agent, quality of water,
etc. are discussed with the Miele Applications Specialists.
~ If the cleaning result is subject to
particularly stringent requirements (e.g. chemical analysis, specialized
processes), a regular quality control test should be carried out by the user
to ensure that the required standards of cleanliness are being achieved.
~ The mobile units and special inserts
should only be used for their specific application.
~ Empty any containers or utensils
before arranging them in the machine.
~ Do not allow any remains of acids,
solvents or corroding ferrous material, and in particular hydrochloric acid or
chloride solutions to get into the wash cabinet. Similarly avoid any materials
with a corrosive effect. The presence in compounds of any solvents should be
minimal, (especially those in hazard class A1).
6
IMPORTANT SAFETY INSTRUCTIONS
~ To avoid any corrosion damage
ensure that solutions or steam containing hydrochloric acid do not come in
contact with the stainless steel casing of the machine.
~ Please follow the installation advice
in these instructions and the separate Installation Instructions.
Using accessories
~ Only specific additional equipment
made by Miele should be connected to this machine. Consult a Miele Application
Specialist on the type and application of such equipment.
~ Only Miele mobile units, modules
and inserts should be used. If equipment from another manufacturer is used,
Miele cannot ensure the cleaning results. Damage or injury caused by this are
excluded from the warranty.
Disposal of an old appliance
~ When discarding a labwasher,
disconnect it from the power supply and cut off the power cord. For
environmental and safety reasons ensure the machine is completely drained of
any residual water and cleaning agent. (Observe safety regulations and wear
safety goggles and gloves). Make the door lock inoperative or remove the door
completely, so that children cannot accidentally shut themselves in. Make
appropriate arrangements for the safe disposal of the machine. Machines with a
tank system: remove the water from the tank before disposing the machine.
SAVE THESE
INSTRUCTIONS
7
Description of labwasher
The Miele PG 8527/ 8528 Laboratory Glassware Washer is designed for high-
throughput, centralized wash areas. This unit is ideal for cleaning both large
volumes of small items, and difficult to clean large items.
Typical applications include:
laboratory equipment for research and production,
laboratory equipment for analytical and preparation applications,
microbiology and biotechnology labware,
Labware can include anything from evaporation dishes to centrifuge tubes.
In these operating instructions, the general terms labware and glassware are
used to cover a range of laboratory equipment and medical items.
For standardized results, it is preferable to machine-wash lab instruments and
equipment.
The cleaning parameters should always be optimally matched to the type of
soiling and labware involved. The processing chemicals used also should be
matched to the particular cleaning needs / analysis / analytical methods
involved.
8
Loading side
Guide to the labwasher
a Housing unit for Drying System DS and/or Steam Condensor SC
b “Profitronic” electronic control system (see “Programming manual”)
c Multiport
d Main Switch “On/Off” e Barcode scanner mount (optional) f Service panel g
Docking station for transfer cart
MF 27/28 h Lift door (closed)
9
Guide to the labwasher
Loading side
a Lift door (open) b Connection for transfer cart and
modules c Lower spray arm d Filter combination
e Container for dispensing systems DOS 1 – DOS 4
10
Unloading side (PG 8528 only)
Guide to the labwasher
a Lift door (closed) b Service panel with docking station
for transfer cart MF 27/28 c Multiport d Barcode scanner mount (optional)
e Printer (optional) f “Door” button g Display
11
Guide to the labwasher
Electronic controls
12
Guide to the labwasher
a On/Off button s turn washer on and off
b Door button a open and close the door
c Display If the washer is not in use, the display automatically switches off
after approx. 10 min.
^ To turn the display back on, touch any button.
Fault messages are shown in the display, together with a corresponding number.
See the “Programming manual” for more information.
d Optical Interface # used by Miele Service Technicians
e Start button h start a program
f Clear button +C return to the previous level discard entry cancel a
program
g and i Selection buttons – e shift position of marking in display changed
marked value scroll retrieve menu display user-defined operating
parameters during program operation
h OK Button – confirm marked menu item or selected value, – erase fault
messages, – erase dialogue messages, – access A0-Graph during program
operation.
13
Optional equipment
Optional modules:
Drying unit Steam condensor Pre-heater Conductivity meter Report
printer Barcode scanner connection up to 3 additional internal dispensing
systems Glass doors and wash cabinet
lighting combination steam/electric wash
cabinet heating
14
Context menu
Certain functions are available through the context menu at any time,
regardless of the operating level: change display language change
operating level, enter Delay Start for convertible washers:
change heating type, loading package to arrange barcode scanner. The current
fault messages can be shown in operating level D (if available).
^ To enter the context menu, press and hold the – e buttons at the same time.
The context menu appears in the display:
Context menu
B
Loading package
Change display language F
Change operating level
Use the buttons – e to move through the menu items.
15
Context menu
Change the display language
You can use the context menu to change the language used in the display. This
setting is stored until the washer is switched off with the s button or the
main switch.
The language set under “Settings – Language J” does not change. All reports
and print-outs will continue to be completed in the system language.
^ To select the menu item Change display language use – e , confirm with OK.
^ Select the desired language and confirm with OK.
The selected language appears in the display.
^ To leave the context menu, select +C.
The J symbol indicates the Change display language setting. If you accidently
select an unfamiliar language, simply select the menu item with J and select
the desired language.
16
Context menu
Change the operating level
Four operating levels are available for selection.
Operating levels A and B
C
D
Authorized access for :
selecting from a list of approved programs. automatic program assignment by
mobile unit code. open program selection through the Program overview,
Programming and Settings J menus.
The context menu can be used to switch between operating levels. To help prevent unauthorized access to the settings, a code is required to change operating levels.
To change the codes for existing users, or to enter new users, you must be a registered administrator. The Administrator can be registered by Miele.
^ Press and hold the – e buttons for at least 3 seconds.
The context menu appears in the display:
Context menu
B
Loading package
Change display language F
Change operating level
17
Context menu
^ Select the menu item Change operating level, confirm with OK.
Select the user
B
User ABC
Administrator
Technical Service
^ Select the user group and confirm with OK. ^ Enter the required code.
Change operating level
B
0 FFF
Enter code
^ Select – e to change the numbers. ^ Confirm with OK.
Repeat same process for each number.
Once complete the display will switch to the selected operating level. ^
Select +C to exit the Context menu.
If an incorrect entry is made, Code invalid will appear in the display. ^ The
message can be cancelled with OK.
Repeat the process to change the operating level.
18
Context menu
Delay start
The start time for a program can be pre-selected. ^ Select Delay start and
confirm. ^ To activate the Delay start option, go to the menu item
Activate and select Yes .
Delay start
B
Activate
Yes
Delay start
^ To use this feature, select Delay start and confirm. The start time menu will appear.
Delay start
B
5:48
Enter time of day
^ Select – e to enter the desired start time, then confirm.
^ Press and hold the +C until the Main menu appears in the display.
^ Select Program overview to select a program.
Program overview
B
CHANGE DS-FINE
LAB-STANDARD
LAB-UNIVERSAL
^ Select a program. ^ Start with h.
19
Context menu
The selected program, the delay start time, the current time of day, and the time remaining until the program starts are displayed.
LAB-STANDARD Start time 5:15
B 3:20
1:55 Remaining time
h
The selected program automatically starts at the delay start time.
^ To cancel the automatic program start, press +C.
Yes
Cancel start time? No
^ Select Yes and confirm.
The automatic program start is canceled. The display reverts to the program
overview. The Delay Start option is canceled. If you select No, the Delay
Start option will continue.
Once the program has run through, the function Delay Start is automatically
deactivated.
The function Delay Start can also be deactivated independent of a program
start. To do so, go to the Delay Start menu, and set the parameter Activate to
No.
20
Context menu
Change the heating type
Loading package
In convertible labwashers, you can use the context menu to change the heating
type – e. g. from electric to steam heating.
^ Select Steam/Electric, confirm with OK to activate.
^ Change the heating type as desired, e. g., Steam + Electric, confirm with
OK.
If a barcode scanner is activated and installed to the machine, loading
packages can be arranged in the context menu under Loading package.
See “Barcode scanner” for more information.
The new heating type is saved.
^ To leave the context menu, use +C.
Current faults
The current fault messages can be shown in operating level D.
^ Select the menu item and confirm with OK.
The current fault messages are displayed.
^ Use OK to scroll through the fault messages.
^ To leave the context menu, select +C.
21
Automatic mobile unit recognition (AWK)
In operating level C the AWK (optional) attributes a fixed program place to a
mobile unit with valid coding. Program places 1-15 are reserved for the AWK
automatic mobile unit recognition.
The unit coding (on the mobile unit) and the program place with the
corresponding program (in the “Profitronic” electronic control unit) must
match.
^ Each mobile unit must be coded before being used for the first time. See
“Mobile unit coding”.
^ The wash program through which the mobile unit is coded has to be assigned
to the relevant program place.
Mobile unit coding
The automatic mobile unit recognition feature assigns a program place to a
mobile unit. The mobile units must be coded with a magnetic strip, via a bit
combination. In operating level C the only program available for a coded
mobile unit is the one assigned to the corresponding program place.
When a coded mobile unit is inserted in the machine and the door closes, the
automatic mobile unit recognition selects the assigned program. Be sure no
small items connect to the magnetic strip.
^ Press 6 to start the program.
For program place changes, see the Programming Manual.
Important! Before starting the program with the “Start” button 6, check that the program needed for this mobile unit is displayed. Otherwise insufficient wash results may occur! Always check that programs for mobile units with AWK (on fixed program places) are not exchanged.
The coding is made up of 5 bits:
Bits 1 to 4 define the mobile unit code,
Bit 5 serves as a control (Parity-Bit).
Bit 6 controls the volume of water intake and the circulation pump for mobile
units with side coupling. This setting cannot be changed! Mobile units without
side coupling are coded with tracks which do not contain Bit 6!
22
Automatic mobile unit recognition (AWK)
Mobile unit coding strip
15 different codes can be set. They are assigned to program places 1 to 15.
Under “System function – Selector switch organization” the matching programs
have to be put into the first 15 program places.
Setting mobile unit coding
To set or alter the coding of a mobile unit with automatic recognition (AWK)
proceed as follows:
Program place
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Bit 1
0 l 0 l 0 l 0 l 0 l 0 l 0 l 0 l
Bit 2
0 0 l l 0 0 l l 0 0 l l 0 0 l l
Bit 3
0 0 0 0 l l l l 0 0 0 0 l l l l
Bit 4 Bit 5 ParityBit
0
0
0
l
0
l
0
0
0
l
0
0
0
0
0
l
l
l
l
0
l
0
l
l
l
0
l
l
l
l
l
0
^ Unscrew the track with an AWK (allen wrench) and remove it from the retainer.
The coded total must be an even number. Bit 6 is not included in this
calculation.
If the coded total equals an odd number the message CHECK MOBILE UNIT
RECOGNITION will appear. If the mobile unit code equals 0, the message NO
MOBILE UNIT RECOGNITION will appear. In both cases the program cannot be
started. The mobile unit recognition must be reset.
^ Remove the magnetic strip from the track.
23
Automatic mobile unit recognition (AWK)
^ Set the program place coding.
^ Place the track in the holder of the mobile unit and screw firmly in place.
Bit 6 is not an integral part of the modifiable magnetic strip. Make sure that
you code mobile units with side coupling to a track where Bit 6 is set to I.
Mobile units without side coupling must be coded with a track that does not
have Bit 6.
The magnetic strips for the PG 8527 / PG 8527 are black.
^ Put the magnetic strip back in the track.
Important: The magnetic strip must be placed in the track so that the Bit
coding, set according to the chart, is visible through the round windows of
the track.
24
The labwasher can be fitted with a variety of mobile units, baskets and inserts, depending on the cleaning application desired.
Application
Check before cleaning
Are the items to clean correctly sorted, loaded and connected?
Are the spray arms clean and can they rotate freely?
Is the filter combination free of coarse soiling? Clean if necessary.
Is the adapter to the water inlet in place? The spray arms correctly
connected?
Are the detergent and neutralizer containers full?
Check after cleaning
Check the cleaning results visually.
Perform protein analytical checks, e.g. with the Miele test kit.
Are the jets and connections securely connected with the mobile unit /
inserts?
25
Application
Loading tips
Select mobile units, baskets and inserts appropriate for the application.
^ Load the items to be washed so that water can access all surfaces. This
ensures that they will be properly cleaned.
^ Do not place items to be cleaned inside other pieces where they maybe
concealed.
^ Hollow vessels such as beakers, measuring cylinders, flasks, etc. should be
inverted and placed in the correct inserts so that water can flow in and out
unrestricted. A cover net can be used to reduce the risk of movement during
the wash process.
^ Use a cover net (e.g. an A 2) or mesh tray to secure small and/or
lightweight items to prevent them from blocking the spray arms or being caught
in the magnetic strip on the automatic mobile unit recognition system.
^ Deep based items should be placed at an angle to allow water to run off
easily.
^ Tall, narrow pieces should be placed in the center of the baskets. This
ensures good water coverage.
^ Mobile units with an adapter must engage correctly.
^ Engage inserts correctly into the module carts.
^ The spray arms must not be blocked by tall items or items which hang down in
their path. If necessary, manually rotate the arms to test.
^ Petri dishes and the similar should be placed in the correct insert with the
soiled side facing the center.
^ Insert pipettes with the pointed end downwards.
^ Quarter inserts should be placed as close to the middle of the unit as
possible.
^ Glassware should not stick out of the door area.
26
Application
Laboratory glassware (LG)
Wide-necked glassware, such as wide necked erlenmeyer flasks and petri dishes
or glassware with a cylindrical form, e.g. test tubes, can be cleaned inside
and out with the rotating spray arm. General glassware such as beakers,
conical flasks, petri dishes, test tubes, etc. can be arranged in full-size,
half or quarter inserts in an empty basket with spray arms. Mobile injector
units are available for direct injection spraying of narrow necked flasks,
pipettes, etc.
The following instructions relate only to basic preparation and loading of
glassware.
Loading the machine
Removing excess soils
^ Empty all glassware before loading into the machine. Take any hygienic
measures necessary to avoid infection.
,Ensure that no acid or solvent
residues, especially hydrochloric acid or chlorides get into the wash cabinet.
^ Remove all agar residues from petri dishes.
^ Remove blood clots and residues from test tubes, etc.
^ Remove all stoppers, corks, labels, sealing wax residues, etc.
^ Small parts such as stoppers and taps should be secured in suitable basket
inserts.
E 940 Injector mobile unit
Contains two levels, for narrow necked laboratory glassware (can also be used
without the upper level).
27
Adding liquid detergents and neutralizers
,Use only detergents and
neutralizers specially designed for labwashers. Always observe the
manufacturer’s recommendations for use.
The labwasher can be fitted with a maximum of 5 internal dispensing systems. They are color-coded to match the dispensing pumps to the corresponding siphon tubes.
Dispensing system
1 2 3 4 5
Color coding
blue white red green yellow
The washer comes equipped, as standard, with two internal dispensing systems:
Dispensing system DOS 1 for use with alkaline products, e.g., liquid
detergents. The dosing rate is 200 ml/min.
Dispensing system DOS 3 for use with acidic products, e.g., neutralizers.
The dosing rate is 105 ml/min.
Up to three additional dispensing systems can be installed optionally:
Dosing rate 465 ml/min 200 ml/min 105 ml/min
Product(s) alkaline alkaline acidic
Depending on the application(s) chosen for this machine, the appropriate amounts of detergent and neutralizer are dispensed through these systems.
Instead of the internal dispensing system, up to five external dispensing
systems can be connected to the washer.
For special applications, the external dispensing systems can be activated, in
addition to the internal systems.
28
Adding liquid detergents and neutralizers
Dispensing systems
Four 10 l containers of processing chemicals can be placed behind the service
panel in the washer. Any additional containers will need to be set up outside
the washer.
,Use caution when handling
processing chemicals. These products can cause irritation and burning. Always
follow all applicable safety procedures and observe the manufacturer’s
instructions. Use protective eye wear and gloves.
Message “Fill DOS [X] container”
^ Refill or replace the container indicated in the display.
Adding liquid detergents and neutralizers
^ Open the service panel, then lift it up and out at the bottom.
Fill the container(s) when you are prompted in the display, e.g., Fill DOS1 container. This will prevent containers from becoming completely empty and needing to be primed (cleared of air).
29
Adding liquid detergents and neutralizers
Message “Check dispensing system [X]”
The currently running program pauses.
^ Check the container(s) and dispensing hoses indicated in the display. Refill
or replace as needed.
^ Remove the container from the washer.
^ Unscrew the siphon tube and remove it.
^ Fill the container with the required product.
Containers and dispensing systems that are not in use can be shut off, to
avoid fault messages in the display (see “Machine function – Container
inquiry” in the Programming manual).
Priming the dispensing system
Whenever a container has been allowed to completely empty, it must be primed
(cleared of air) after refilling.
^ Select the corresponding service program, e. g., DOS1-FILL.
^ Press the h start button.
^ Insert the siphon tube into the container opening and screw it into place.
Once the containers have been filled, the message clears from the display.
30
Operation
Main Switch – On/Off
The main switch disconnects the labwasher from the power supply
^ Turn the main switch to I-ON.
Once the start procedure is completed, the labwasher is ready to for use.
The following message may appear: Fault 2218: “New power frequency 50 Hz Flow
rate automatically adjusted” or “New power frequency 60 Hz Flow rate
automatically adjusted”. During the start procedure, the Profitronic system
detected a different power frequency, and adjusted the flow rate of the
dispensing pump accordingly.
Before operation, you will need to check if the change in power frequency was
caused by a fault in the power supply. If so, incorrect amounts of processing
chemicals would be dispensed during the program.
The fault message can only be confirmed in operating level D and in the
service mode.
Turning On
^ Press and hold the s button .
The activating time for the s button can be set to a maximum of 10 seconds (see “System functions On/Off button activation time” in the Programming Manual).
Depending on the set operating level, the following will appear in the display:
Operating levels A and B
C
D
Displayed
A list of approved programs. Instructions for Automatic Mobile Unit
recognition. Select from: Program Overview Programming Settings J.
If the washer is not in use, the display automatically switches off after approx. 10 min. To turn the display back on, touch any button.
31
Operation
Opening and closing the door
^ Press the a button.
The door opens or closes.
To open the door on the clean side (PG 8528), note the following:
programs with a process requirement must have been completed without faults
according to the program parameters set,
or
the gate function is activated, i.e., door mechanism is set to one of the
following parameters: – Sluice, – Door clean + Sluice, – Door unclean +
Sluice. (see “Machine functions / Door mechanism” in the Programming manual).
,If the pre-heater has been
programmed for heating, beware of hot water or steam exiting the open door.
Looking from the unclean side, the inlet pipe for the pre-heater is located
beneath the right basket guide rail.
Starting a program
You can find detailed and important information regarding the Miele default
programs in the Program charts in the Appendix of the Programming Manual.
If a barcode scanner is connected to the labwasher and the user or load scan
is activated, the selected program cannot be started until scanning has been
completed. For more information see “Barcode scanner”.
32
Operation
Operating levels A and B
A list of all selectable programs appears in the display. ^ Select the desired
program with – e,
confirm with OK. ^ Press the h start button.
The program begins.
Operating level D ^ Select the menu item Program
overview, confirm with OK. ^ Select the desired program with – e,
confirm with OK. ^ Press the h start button.
The program begins.
Operating level C
,Make sure that no metallic parts
or pieces are stuck to the magnetic strip or its underside. These objects can
cause the coding to be misread or unreadable.
^ Push the coded basket into place in the washer.
^ Close the door a and press the start button h.
^ The program begins.
33
Operation
Program sequence
Canceling a program
Once started, the program proceeds automatically. The program steps are shown in the display.
A program can only be interrupted or canceled in operating levels B and D.
Detailed information on the program sequences is available in the “Programming
Manual”.
,Do not change the printer paper
roll or ribbon cartridge during a running program.
End of program
^ Once the program ends, End of program appears in the display.
Turning off
^ Press the s button.
^ Press the +C button.
The program is interrupted. The following prompts appear in the display:
Program Cancel (OK) or Continue (Clear)?
^ Confirm the program cancellation with the OK button.
Program canceled or Water drainage appears in the display.
After the water has drained away, the program sequence appears in the display.
If the machine is to be left off for several days, be sure to turn the main switch to 0-OFF.
34
Operation
Interrupting a program
A program can only be interrupted or cancelled in operating levels B and D.
Only interrupt a program if the door must be opened for urgent reasons, e.g.,
if items are moving around too much, or the cleaning performance needs to be
checked.
If a program with a defined process requirement is interrupted then continued,
make sure to check the display upon program completion. If the display reads
Process parameter not achieved, the door was opened after process parameter
monitoring began, and so the defined requirements were not achieved. If
necessary, repeat the program.
In operating levels B or D
^ Press the +C button.
The program is interrupted. The following prompts appear in the display:
Program Cancel (OK) or Continue (Clear)?
^ Open the door a
Note PG 8528: Which door can be opened depends on which setting has been
programmed under “Machine Function – Automatic door” (see the Programming
Manual).
The door on the unclean side can always be opened, no matter which parameter
has been set.
The door on the clean side can only be opened if one of the following
parameters are set: – Sluice, – Door clean + Sluice, – Door unclean + Sluice.
,Caution! Items in the machine
may be hot. There is the danger of burning or scalding.
^ Arrange items securely. ^ Close the door a. ^ Press the +C button.
The program continues.
35
Barcode scanner (optional)
Barcode scanners can be connected to the labwasher via the multiport on the
unclean side and, if available, on the clean side as well.
The configuration of the serial interface for the barcode scanner is described
in the section “PC / Print functions – Serial Interface configuration” in the
Programming Manual.
Barcode scan functions
All barcode scanner functions are available at every operational level:
User scan load
User scan unload
Scan the load
The scanned bar codes are transmitted to the record of users and/or loads.
User scan and load scan can be activated independently of each other.
Activation is described in the section “System functions – Barcode” in the
Programming Manual.
36
Barcode scanner (optional)
User scan load
When User scan load is activated, after program selection the following
message is displayed (immediately in the case of automatic mobile unit
recognition on operating level C):
Scan user
Once the user barcode has been scanned, the program can be started.
When Scan the load is also activated, after the user scan you are prompted to
scan the load.
Cancelling User scan load
User scan load can be cancelled using the +C key. The program overview is
displayed again.
User scan unload
If User scan unload is activated, you will be prompted to scan the user
barcode once a program has ended:
End of program Scan user
The door cannot be opened until the user barcode has been scanned. Depending
on the settings in the Menu Machine functions / Automatic door the door will
open automatically.
after cancelling a program with +C
If User scan unload is activated, the following message is displayed after
cancelling the program:
Program canceled Water drain
After the water has drained, the Scan user prompt will appear:
Scan user
Skip user scan for unloading
On operating levels B and D, you can skip User scan unload by pressing OK.
At the end of the wash protocol, “Skip user scan” is printed.
37
Barcode scanner (optional)
Scan the load
When load scan is switched on, the following message prompts the operators to
scan the load:
Scan the load
A
:
Scan the load
^ Scan the first barcode of the load. In the menu Scan the load the barcode numbers of the scanned load are listed.
Scan the load
A
0987664
^ Scan further bar codes if applicable.
If a scanned barcode already exists, the following query appears:
Load scanned, accept?
A maximum of 99 load bar codes can be scanned. Then the following message
appears:
Maximum number of entries reached
The last scanned barcode can be deleted from the list with +C. To prevent
unintentional removal from the list you will be prompted to confirm the
deletion:
Delete last entry?
38
Barcode scanner (optional)
Cancel load scan
If no load was scanned, the load scan can be cancelled using +C. The program
overview is displayed again.
If load scans already exist, they must be deleted prior to cancellation. To do
this, answer the query Delete last entry with Yes for each scanned load.
Skip load scan
On operating levels B and D, you can skip the load scan by pressing OK.
If you answer the query Skip load scan? with No, the program overview will be
displayed.
If you answer the query with Yes, the program can be started by pressing h.
Finish load scan
^ After all the load bar codes have been scanned, finish the process by
pressing OK.
h flashes and the program can be started.
Using loading packages
When load scan is on, prior to program start the Profitronic verifies that at
least one loading package has been saved.
At the beginning of the wash protocol, “User scan skipped” is printed.
39
Barcode scanner (optional)
If this is the case, the following message is displayed:
Yes
Use loading package?
^ Acknowledge the message by selecting Yes. All loading packages are
displayed:
Select loading package 1_14.02.2007_08:23 2_14.02.2007_09:11
A No
A
^ Select a loading package and confirm with OK. h flashes and the program can
be started.
At the end of the program, the message Process parameters achieved appears and
the completed program package is deleted.
The last processed loading package is not deleted:
if the program is cancelled or if an error occurs during the program
(Process parameters not achieved),
with programs in whose program header the parameter Process information is
not defined.
The next time the program is started the loading package will be processed
again. It can be manually deleted or moved on the list.
If none of the saved loading packages are to be used: ^ Select No and confirm.
^ The message Scan the load is displayed. ^ Scan the load manually as
described in “Scan the load”.
40
Barcode scanner (optional)
Loading packages
Loading packages can be created to optimize the function sequences when
loading the labwasher. To do this, the load scan must be activated, see
“System functions – Barcode” in the Programming Manual).
Context menu
B
Loading package
Change display language F
Change operating level
After the context menu is opened and the menu item Loading package confirmed,
the following functions are available:
Add loading package Display loading package Delete loading package Move
loading package
41
Barcode scanner (optional)
Add loading package
To create a loading package, open the Context menu, then select and confirm Loading package.
Loading package
B
Move loading package
Add loading package
Display loading package
^ Select and confirm Add loading package.
The following message appears:
Scan the load
^ Scan the individual components of the load.
^ Press OK to confirm the complete loading package.
Repeat the steps for further loading packages. The loading packages are stored in a list (see “Display loading packages”). The last scanned loading package is added to the end of the list.
A maximum of 50 loading packages can be stored. If the max. amount has been reached, loading packages must be processed or deleted before further packages can be added.
42
Barcode scanner (optional)
Display loading package
To display the loading packages, open the context menu, then select and
confirm Loading package.
^ Select and confirm the menu item Display loading package.
The list with all the loading packages is displayed. Each loading package is identified by a number, date and time.
Display loading package
A
1_14.02.2007_08:23
2_14.02.2007_09:11
^ Select and confirm the desired loading package. The components of the loading package are listed in the menu Scan the load.
Scan the load
A
0987664
0983558
0991724
^ Press +C to return to the menu Display loading package. ^ Press +C to return to the menu Loading package.
43
Barcode scanner (optional)
Delete loading package
To delete a loading package, open the context menu, then select and confirm
the menu item Loading package.
^ Select and confirm Delete loading package.
All existing loading packages are displayed with number, date and time.
Delete loading package
A
1_14.02.2007_08:23
2_14.02.2007_09:11
^ Select and confirm the desired loading package. The following message
appears:
Delete loading package?
^ Confirm the request. The loading package is deleted. ^ Press +C to return to
the menu Loading package.
44
Barcode scanner (optional)
Move loading package
To move a loading package, open the context menu, then select and confirm the
Loading package.
^ Select and confirm the menu item Move loading package.
All existing loading packages are displayed with number, date and time.
Move loading package
A
1_14.02.2007_08:23
2_14.02.2007_09:11
3_14.02.2007_10:31
^ Select the desired loading package. ^ Press e – to move the loading package, and confirm. The list will be saved in the new sequence. ^ Press +C to return to the menu Loading package.
45
Serial interface (data transfer)
Each washer comes supplied with a 16.5 ft. (5 m) interface cable and RJ45
plug, for data transfer between the Profitronic system and an external report
printer or PC.
The interface cable is connected to the Ethernet interface in the washer.
Pin position of the 9-pole Sub-D plug
(serial interface):
5
GND ground
3
TXD
transmit
2
RXD receive
7
CTS
clean to send
8
RTS
ready to send
1-4-6-9 empty
The supplied cable can be used to connect the Ethernet connection of the
washer to suitable network components (e. g., a hub or a switch). For a direct
Ethernet connection between the washer and an external device, you will need a
crossover cable.
The supplied cable can also be used to connect the washer to an external
device via the RS 232 interface. An RJ45 adapter / Sub-D plug is also
supplied.
Only a Miele Technician can connect a different interface.
Pin position of the RJ45 plug (Ethernet):
1
Transmit Data +
2
Transmit Data –
3
Receive Data +
6
Receive Data –
4-5-7-8 empty
The interface configuration is described in the section “PC/Print functions”
in the Programming manual.
Please note the following when connecting a printer or PC:
Only use an industry-standard PC or printer.
When installing the machine take the dimensions of the printer/PC into
account.
Any extension cable from the serial interface to the printer or PC must not
exceed 32′ 10″ (10 m); from the Ethernet interface the maximum extension cable
length is 328ft. (100 m).
For external printers, only HP LaserJets or compatible devices are acceptable.
Settings for the external printer functions are described in the section “PC/Print functions” in the programming manual.
46
Maintenance
Labwashers should undergo routine maintenance by the Miele Technical Service
every 1000 operating hours or at least once every 6 months. Maintenance covers
the following: Electrical safety Door mechanism and door seal Screws and
connections in the wash
cabinet Water inlet and drain Internal and external dispensing
systems Spray arms Triple filter system Drain pump and non-return valve
Steam condensor All mobile units, baskets, and inserts If applicable:
Drying unit Connected printer
Also during maintenance, the following function checks are conducted:
a test program is run
wash temperatures are measured
ensure a water tight door seal
all safety-relevant measuring systems are checked (fault display).
47
Maintenance
Process validation
As a rule, it is the responsibility of the user to ensure that items cleaned
in the labwasher meet the required standards.
Routine checks
The user must complete these checks every day before using the washer. A
checklist is supplied with the washer.
Check the following points:
the filters in the base of the wash cabinet
the spray arms in the washer and in the mobile units and inserts
the wash cabinet and the door seal and
all mobile units, baskets, and inserts
48
Maintenance
Cleaning the filters in the wash cabinet
The filters in the base of the wash cabinet prevent large particles from
entering the system. The filters can become clogged, they need to be checked
regularly and cleaned as needed.
,Do not use the washer unless all
the filters are in place.
Cleaning the fine filter
,Beware of glass splinters,
needles, etc. danger of injury.
^ Remove the fine filter and clean if necessary.
Cleaning the flat filter
^ Remove the flat filter and clean if necessary.
49
Maintenance
Cleaning the filter system for the circulation pump To protect the circulation
pump, there are two additional filters under the flat filter.
^ Pull the filters up and out of the holder to remove. Clean if necessary.
^ Replace the filter combination in reverse order. After replacing, check that
all filters are properly in place.
50
Maintenance
Cleaning the spray arms
The spray arms can become clogged. Check daily and clean if needed. ^ Use a
sharp, pointed object to push
any particles into the spray arm jets, then rinse throughly under running
water.
To do so, remove the spray arms as follows: ^ Remove the basket.
Labwasher spray arms
Spray arms on basket / inserts ^ Use a wrench to loosen the nut on
the spray arm guide, then pull the spray arm downward.
If you notice wear and tear on the bearings or the gasket, please contact
Miele.
^ After cleaning, put the spray arms back in place and tighten the nut. Make
sure the nut is in the proper position.
After replacing the spray arms make sure that they rotate freely.
,Make sure that no metallic parts
or pieces become attached to the magnets at the ends of the spray arms. These
objects can cause the spray arm rotations to be miscalculated.
^ Loosen the nut with a wrench and pull the spray arm either upward or downward.
51
Maintenance
Cleaning the control panel and glass door (optional)
^ Press s to turn off the washer.
^ Clean the control panel and the glass door with a damp cloth or glass
cleaner. To disinfect, use an approved cleaning agent.
,Do not use abrasive or
all-purpose cleaners. The chemicals contained in these products can seriously
damage the surface of the glass.
Cleaning the exterior
^ To clean the stainless steel exterior, use a damp cloth and dish soap, or a
non-abrasive stainless steel cleaner.
^ To help prevent re-soiling (fingerprints, etc.), a stainless steel
conditioner can be used after cleaning.
,Do not use thinner or
ammonium-based cleaners. They can damage the surface.
,Do not hose down the washer or
the immediate vicinity, e.g., with a water hose or pressure washer.
Cleaning the wash cabinet
The wash cabinet is mostly self-cleaning. If you notice a build-up of
deposits, contact Miele.
Cleaning the door seal
To remove soiling, clean the door seal regularly with a damp cloth.
Damaged or leaky door seals should be replaced by Miele Technical Service.
52
Maintenance
Baskets and inserts
If there are spray arms:
To ensure that the mobile units, baskets and inserts function properly, they
must be checked routinely. A checklist is supplied with the labwasher.
Check the following points:
Are the rollers in proper condition, and are they securely attached to the
basket / insert?
Are the counter-nuts on the guide rails firmly secured?
Is the basket connection correctly set and firmly screwed on?
Do the spray arms rotate freely?
Are the spray jets clogged? See the section on “Cleaning and Care / Cleaning
the spray arms”.
Are the screws in the magnet rails for Automatic Mobile Unit Recognition
tightly secured?
Are the magnets in place on the spray arm ends?
Are the magnets on the spray arms free of attached metal parts?
For baskets in the modular system, are the caps in the module connection working properly?
Are all spray jets, spray sleeves, and hose adapters securely attached to the basket / insert?
Are all spray jets, spray sleeves, and hose adapters unclogged so that wash water can flow through?
Are all caps, covers, and fasteners securely attached to the spray sleeves?
53
Maintenance
Report printer (optional)
Replacing the print paper
When the red indicator light on the front panel of the printer lights, the
print paper needs to be replaced.
^ Open the front panel of the printer by the upper edge, and pull down.
^ Remove the paper roll and spindle together from the holder, fit the new roll
onto the spindle, and install them back into the printer.
^ Guide the paper up and over the paper transport roller (slit behind the
ribbon cartridge). Press the green paper transport button until the paper re-
emerges above the ribbon cartridge.
^ Guide the paper through the slit in the front panel. Shut the panel.
Replacement paper rolls (2 ¼” (58 mm) wide / outer diameter approx. 2″ (50 mm)
can be obtained from Miele.
Replacing the ribbon cartridge
^ Open the front panel of the printer by the upper edge, and pull down.
^ The ribbon cartridge is above the paper roll. Pull it forward and out of the
holder and replace it with a new one. The paper must be guided between the
ribbon and the cartridge housing.
^ On the right, turn the small wheel for manual ribbon transport clockwise
until the ribbon is taut.
^ Guide the paper through the slit in the front panel. Shut the panel.
Replacement ribbon cartridges can be obtained from Miele.
54
In the event of a fault which you cannot easily fix yourself, please contact
the Miele Technical Service Department at the address on the back of this
booklet.
When contacting Technical Service please quote the serial number and model of
the appliance.
After sales service
55
56
INSTALLATION INSTRUCTIONS
Electrical connection
All electrical work must be carried out by a qualified electrician in
accordance with local and national safety regulations.
Installation, repairs and other work by unqualified persons could be
dangerous.
The machine must only be operated with the voltage, frequency and fusing shown
on the data plate.
^ For technical data see the data plate or wiring diagram supplied.
The data plate is on the cover plate behind the service panel of the unclean
side.
The wiring diagram is secured to the inner side of the service panel.
This unit is convertible for use with steam or electric heating.
Hook up requirements are as follows:
1. Electric heat only: 208 V 3 phase 60 HZ 60 amp
2. Steam heat only: without drying option 208 V 3 phase 60 HZ 20 amp with
drying option 208 V 3 phase 60 HZ 30 amp
WARNING: THIS APPLIANCE MUST BE GROUNDED
58
Plumbing
Water connection
This machine must be connected to the water supply in accordance with all
national and local plumbing codes.
,Water in the machine is not
suitable for drinking.
The washer is constructed so that it may be connected to a water supply
without an extra non-return valve, unless required by code.
Water and drain connections should be situated as close to the machine as
possible, and be easily accessible.
Water supply pressure Hot + Cold water connection =
min. 22 psi – max.145 psi DI water connection =
min. 9 psi – max. 145 psi
Recommended water pressure Hot + Cold water connection =
over 37 psi – max.145 psi DI water connection =
over 22 psi – max. 145 psi to avoid long intake times
The machine cannot be operated with an inflow pressure of less than 10 psi.
The machine is supplied with 3 connection hoses 5 ft. (1.5 m) long with 3/4″
(19 mm) female hose thread ends.
Large area filters, supplied with the machine, are to be installed in the hose
attached to the water supply. See “Cleaning and Care – Cleaning the water
inlet filters”.
,Do not shorten or damage the
hoses! Do not remove the filters in the hoses.
59
Plumbing
Requirements
3. DI water connection
(refer to the Installation Diagram for more information)
1. Cold water connection
One ½” (13 mm) cold water supply line with a standard 3/4″ (19 mm) male hose
thread faucet is needed. The cold water supply hose is marked with blue tape.
One ½” (13 mm) deionized water supply line with a standard 3/4″ (19 mm) male hose thread faucet is needed. The machine cannot be operated conventionally with a flow pressure of less than 10 psi. In this case the installation of an optional DI pump is required. The DI water supply hose is marked with green tape.
2. Hot water connection
One ½” (13 mm) hot water supply line with a standard 3/4″ male hose thread
faucet is needed. The water temperature must not exceed 158°F / 70°C
(recommended: 140°F / 60°C). The hot water supply hose is marked with red
tape. If no hot water line is available, the red marked hose must be connected
to the cold water line.
If the water pressure is below 29 psi, the intake time will increase.
,If the machine will not be
connected to DI water, the programs must be changed by Miele Technical
Service. The intake hose stays at the rear of the machine.
DI water connection (pressure less) 1.25 – 9 psi (optional)
For connection to 1.25 – 9 psi water pressure the machine must be converted if it was not ordered from the factory as a special model. A feed pump must be installed by Miele’s Technical Service Dept.
The discharge nipple of the DI water container (pressure less) must be at least at the same height as the machine’s upper edge.
60
Technical data
Height Width Depth Weight (net) Operating weight Voltage (volts) Connected load (watts) Current (amps) Air pressure connection Steam connection: with Electric TA with Steam TA Water pressure (flow rate) Cold, hot, and deionized (DI) water Room temperature Altitude above sea level !mark
98 7/16″ 250 cm (minimum room height) 45 1/4″ 115 cm 34 1/4″ 87 cm approx.
1210 lbs 550 kg approx. 1760 lbs 800 kg see data plate see data plate see data
plate 85 psi 600 kPa (required for steam operation)
36 – 145 psi / 284 – 356°F (140 – 180 °C) 36 – 87 psi 1) / 284 – 327°F (140 –
164 °C) 30 – 145 psi over pressure up to max. 158°F (70 °C)
41 – 104°F (5 °C to 40 °C) up to 4921 ft. (1500 m)2) MPG Guidelines –
93/42/EEC, Class IIa
- At a steam pressure of 36 – 43 psi (250 – 300 kPa), the drying unit reaches a maximum drying temperature of 194 – 212°F (90 – 100°C).
- If the machine is installed above 4921 ft. (1500 m), the wash water will boil at a lower temperature. This may require you to change the program parameters depending on the wash and rinse temperatures selected.
61
Caring for the environment
Disposal of the packing materials
The cardboard box and packing materials protect the appliance during shipping.
They have been designed to be biodegradable and recyclable.
Ensure that any plastic wrappings, bags, etc. are disposed of safely and kept
out of the reach of children. Danger of suffocation! Please recycle.
Disposal of an old appliance
Old appliances may contain materials that can be recycled. Please contact your
local authorities about recycling in your area.
Ensure that the appliance presents no danger to children while being stored
for disposal. See “Important Safety Instructions”.
62
63
Alteration rights reserved / 2609
M.-Nr. 07 207 090 / 00
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>