NATIONAL INSTRUMENTS PXIe-4844 Optical Sensor Interrogator Scaling Instruction Manual
- June 16, 2024
- NATIONAL INSTRUMENTS
Table of Contents
- NATIONAL INSTRUMENTS PXIe-4844 Optical Sensor Interrogator Scaling
- Product Information
- Safety Guidelines
- Electromagnetic Compatibility Guidelines
- Unpacking
- What You Need to Get Started
- Installing the NI-OSI Driver Software
- Connecting Sensors
- Cleaning Sensors
- Using the NI-OSI Explorer and LabVIEW VIs
- Specifications
- Safety Standards
- Worldwide Support and Services
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
NATIONAL INSTRUMENTS PXIe-4844 Optical Sensor Interrogator Scaling
Product Information
Specifications
- Product: PXIe-4844
- Safety Guidelines:
- Do not enable the laser unless an optical connector or LC/APC connector cover is connected to the LC/APC connector port.
- Never look into the end of an optical cable attached to an optical output when the device is powered on.
- Do not modify the PXIe-4844 module.
- The protection provided by the PXIe-4844 can be impaired if it is used in a manner not described in this document.
- Electromagnetic Compatibility Guidelines:
- To ensure the specified EMC performance, operate this product only with shielded cables and accessories.
- To ensure the specified EMC performance when using the AUX port, install a snap-on ferrite bead (part number 781233-01) in accordance with the product installation instructions.
FAQ
- Q: What should I do if I notice loose components or damage on the hardware module?
- A: If you notice any loose components or damage on the hardware module, contact NI for further assistance. Do not install a damaged module into your system.
- Q: How should I transport the PXIe-4844 over long distances?
- A: When transporting the PXIe-4844 over long distances, remove the module from the chassis and place it in the original antistatic package and hard-shelled plastic case.
- Q: What LabVIEW versions are compatible with the NI-OSI driver software?
- A: The NI-OSI driver software is compatible with LabVIEW 2009 (32-bit), 2010 (32-bit), 2011 (32-bit), 2012 (32-bit), 2013 (32-bit), 2014 (32-bit), 2015 (32-bit), and 2016 (32-bit).
This document includes the installation instructions and specifications for the National Instruments PXIe-4844 Optical Sensor Interrogator (OSI) module. The PXIe-4844 is a dual-slot, 3U PXI Express data acquisition module for fiber Bragg grating (FBG) optical sensors. The PXIe-4844 provides four optical channels that are simultaneously sampled at 10 Hz, and is expandable to eight or 16 channels with an external optical multiplexer.
FBG fiber optic sensing provides many benefits over conventional electrical sensing because it is nonconductive, electrically passive, and immune to EMI. FBG technology also enables measurements over long distances without loss of signal accuracy and provides the ability to daisy-chain dozens of sensors along a single optical fiber. The PXIe-4844 requires no calibration as it continuously adjusts its measurements using an on-board NIST traceable wavelength reference.
Caution: This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash.
Safety Guidelines
Follow these guidelines when installing and using the PXIe-4844.
- Caution: Do not enable the laser unless an optical connector or LC/APC connector cover is connected to the LC/APC connector port. The laser enables when the device receives power.
- Caution: Never look into the end of an optical cable attached to an optical output when the device is powered on. The laser radiation is not visible to the human eye, but it can seriously damage your eyesight.
- Caution: Do not modify the PXIe-4844 module. This may result in hazardous radiation exposure from the laser source.
- Caution: The protection provided by the PXIe-4844 can be impaired if it is used in a manner not described in this document.
Electromagnetic Compatibility Guidelines
This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) as stated in the product specifications. These requirements and limits are designed to provide reasonable protection against harmful interference when the product is operated in its intended operational electromagnetic environment.
This product is intended for use in industrial locations. There is no guarantee that harmful interference will not occur in a particular installation, when the product is connected to a test object, or if the product is used in residential areas. To minimize the potential for the product to cause interference to radio and television reception or to experience unacceptable performance degradation, install and use this product in strict accordance with the instructions in the product documentation.
Furthermore, any changes or modifications to the product not expressly approved by National Instruments could void your authority to operate it under your local regulatory rules.
Caution: To ensure the specified EMC performance, operate this product
only with shielded cables and accessories.
Caution: To ensure the specified EMC performance when using the AUX port,
install a snap-on, ferrite bead (National Instruments part number 781233-01)
in accordance with the product installation instructions.
Unpacking
The product shipping kit includes the PXIe-4844 hardware module and the NI-OSI driver software DVD. The PXIe-4844 module ships in an antistatic package to prevent damage from electrostatic discharge (ESD), and the package is inside a hard-shelled plastic case. Before removing the hardware module from the antistatic package, touch the antistatic package to a metal part of your computer chassis to discharge any static electricity. Ground yourself using a grounding strap or by touching a grounded metal object.
Remove the hardware module from the package and inspect it for loose components or any signs of damage. Contact NI if the hardware module appears damaged in any way. Do not install a damaged module into your system. Store the module in the antistatic package and the hard-shelled plastic case when not in use.
Note: When transporting the PXIe-4844 over long distances, remove the module from the chassis and place the module in the original antistatic package and hard-shelled plastic case.
What You Need to Get Started
- LabVIEW 2009 (32-bit), 2010 (32-bit), 2011 (32-bit), 2012 (32-bit), 2013 (32-bit), 2014 (32-bit), 2015 (32-bit), or 2016 (32-bit)
- PXI Express chassis with
- controller, or
- MXI-Express (card or built-in)
- FBG sensors
- Multiplexer for additional channels (optional)
Installing the NI-OSI Driver Software
Install the NI-OSI driver software on the host computer before installing the PXIe-4844 module. Refer to readme_OSI.html, located in the National Instruments\OSI Explorer directory, for software installation instructions, LabVIEW compatibility information, and system requirements.
Installing the PXIe-4844
This section contains installation instructions for the PXIe-4844. Refer to your PXI Express chassis user manual for chassis instructions and warnings.
- Plug in your chassis before installing the PXIe-4844. The power cord grounds the chassis and protects it from electrical damage while you install the module.
- Make sure the chassis power switch is turned off.
- Caution: To protect both yourself and the chassis from electrical hazards, leave the chassis powered off until you finish installing the PXIe-4844 module.
- Touch a metal part on the chassis to discharge any static electricity that might be on your clothes or body.
- Remove the protective plastic covers from the four front panel mounting screws on the module, as shown in Figure 1.
- Make sure the PXIe-4844 injector/ejector handle is in its downward position.
- Identify two side-by-side empty chassis slots, except for a PXI Express System Controller slot. Of these two slots, the left-most slot must be one of the following PXI Express slots:
- PXI Express Peripheral Slot—a PXI Express slot marked with a solid circle containing the slot number.
- PXI Express Hybrid Peripheral Slot—a PXI Express hybrid slot marked with the letter “H” and a solid circle containing the slot number.
- PXI Express System Timing Slot—a PXI Express timing slot marked with a square surrounding a solid circle containing the slot number.
- Remove the filler panels covering the selected slots.
- Align the PXIe-4844 with the card guides on the top and bottom of the selected slots.
- Hold the injector/ejector handle down as you slowly slide the module into the chassis until the handle catches on the injector/ejector rail, as shown in Figure 2.
- Caution: When installing the module, make sure both edges are positioned inside the guides and that the module components do not come into contact with adjacent modules.
- Raise the injector/ejector handle to latch the module into the chassis. The front panel of the PXIe-4844 should be even with the front panel of the chassis.
- Tighten the four front panel mounting screws to 0.31 N · m (2.7 lb · in.) on the top and bottom of the module front panel to secure the PXIe-4844 to the chassis.
- Power on the chassis.
Removing the PXIe-4844 from the PXIe Chassis
To remove the PXIe-4844 from the PXI Express chassis, complete the following steps:
- Power off the chassis.
- Remove any cables or sensors attached to the PXIe-4844.
- Loosen the four front panel mounting screws on the module.
- Press the injector/ejector handle down.
- Slide the module out of the chassis.
- Place the PXIe-4844 in its original antistatic bag. Store the module within its hard-shelled plastic case.
Connecting Sensors
The PXIe-4844 has four simplex singlemode LC/APC connector ports to connect sensors. If your sensor does not have an LC/APC connector, you need an adapter to connect the sensor to the LC/APC connector port.
Caution: Connecting damaged or dirty sensors to the module can damage the
LC/APC connector ports on the module. Always clean optical connectors before
connecting to the module.
Caution: Never force an optical connector into an LC/APC connector port.
A ferrule may break and damage the module.
Connect the LC/APC connector on the sensor to an available LC/APC connector port on the module.
Caution: When connecting any accessory to the AUX port, install a ferrite bead (NI part number 781233-01) on the connecting cable as close to the AUX port as possible.
Note: The AUX port is an 8-pin mini-DIN connector you can use to connect third-party multiplexers to the PXIe-4844.
Cleaning Sensors
To clean the optical sensors, use a ferrule cleaner or follow the standard guidelines:
- Fold a lint-free wipe into a compress.
- Moisten the compress with isopropyl alcohol.
- Remove the protective cover from the sensor connector.
- Press the connector end face firmly to the moistened section of the compress, then forcefully wipe the connector with a twisting motion toward the edge of the compress, finishing in a clean, dry section of the compress. Do not reuse dirty sections of the compress.
- Discard the used compress.
Calibration
The PXIe-4844 has continuous on-board calibration using epoxy-free Telcordia- qualified optical referencing components and continuous on-board NIST traceable wavelength reference components to ensure that sensor wavelength measurements remain within specifications over the life of the product.
Using the NI-OSI Explorer and LabVIEW VIs
The NI-OSI driver software installs the NI-OSI Explorer and the NI-OSI LabVIEW VIs. Use the NI-OSI Explorer to identify and configure the optical sensors connected to the PXIe-4844. Use the Optical Measurement VIs in LabVIEW to perform optical sensing measurements. To get started:
- Select Start»All Programs»National Instruments»NI-OSI Explorer»NI-OSI Explorer.
- Read the Welcome dialog. Follow the links provided for more information on configuring sensors and performing measurements.
Specifications
The following specifications are typical for the PXIe-4844 operating at 25 °C unless otherwise noted.
Bus Interface
- Form factor ……………………………………………….x4 PXI Express, v1.0 compliant
Laser
- Type ………………………………………………………….Fiber laser
- Class …………………………………………………………1
- Output power (continuous wave)
- Min …………………………………………………… 0.06 mW
- Max ………………………………………………….. 0.25 mW
- Beam diameter …………………………………………..9 mm (0.35 in.)
- Numerical aperture …………………………………….0.1
- Optical Input
- Number of channels …………………………………… 4
- Wavelength range ……………………………………… 1510 nm to 1590 nm
- Sample rate ………………………………………………. 10 Hz ± 0.1 Hz
- Optical dynamic range ……………………………….. 40 dB
- FBG Wavelength Detection
- Accuracy ………………………………………………….. 1 pm
- Stability (0 °C to 55 °C) …………………………….. 1 pm
- Repeatability …………………………………………….. 1 pm
Physical Characteristics
- If you need to clean the PXIe-4844, use a soft, non-metallic brush.
- Make sure that the device is completely dry and free from contaminants before returning it to the PXI Express chassis.
Note: For two-dimensional drawings and three-dimensional models of the PXIe-4844 module and connectors, visit ni.com/dimensions and search by module number.
- Dimensions (without connectors) ………………… 13.1 cm × 21.4 cm × 4.1 cm
- (5.1 in. × 8.4 in. × 1.6 in.)
- Weight ……………………………………………………… 213 g (7.5 oz)
- I/O connector ……………………………………………. LC/APC
- Slot requirements ………………………………………. Two side-by-side chassis slots, other than a PXI Express System Controller slot. The left-most slot must be a PXI Express, PXI Express Hybrid, or PXI Express System Timing slot.
- Slot compatibility ……………………………………… x4, x8, and x16 PXI Express or PXI Express Hybrid slots
Safety Standards
This product meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Note For UL and other safety certifications, refer to the product label or the Online Product Certification section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
- EN 63126-1 (IEC 61326-1): Class A emissions, Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- AS/NZS CISPR 11: Group 1, Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
Note
- For EMC declarations and certifications, refer to the Online Product Certification section.
Laser Compliance
This product meets the requirements of the following laser compliance standards for electrical equipment for measurement, control, and laboratory use:
- IEC 60825-1, ED 2.0, 2007-03; US CDRH 21 CFR Subchapter J
CE Compliance
This product meets the essential requirements of applicable European Directives as follows:
- 2014/35/EU; Low-Voltage Directive (safety)
- 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
Online Product Certification
To obtain product certifications and the Declaration of Conformity (DoC) for this product, visit ni.com/certification, search by module number or product line, and click the appropriate link in the Certification column.
Shock and Vibration
- Mechanical shock
- Operating
- (IEC 60068-2-7 Annex A, section A.4, Table A.1) ………………………..15 g peak, half-sine, 11 ms pulse
- Non-operating (IEC 60068-2-7) ……………25 g peak, half-sine, 11 ms pulse
- Random vibration
- Operating (ETSI 300 019-2-3) ………………0.15 grms, 5 Hz to 100 Hz
- Non-operating (IEC 60068-2-64) ………….0.8 grms, 10 Hz to 150 Hz
Environmental
This device is intended for indoor use only.
Caution: Do not exceed the operating temperature, even when using the module in a chassis with a higher temperature range.
- Operating temperature
- (IEC 60068-2-1, IEC 60068-2-2) ………………… 0 °C to 55 °C
- Storage temperature
- (IEC 60068-2-1, IEC 60068-2-2) ………………… -40 °C to 70 °C
- Operating humidity (IEC 60068-2-56) …………. 10% to 90%, non-condensing
- Storage humidity (IEC 60068-2-56) …………….. 5% to 95%, non-condensing
- Maximum altitude ……………………………………… 2,000 m
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of the product life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers, National Instruments WEEE initiatives, and compliance with WEEE Directive 2002/96/EC on Waste and Electronic Equipment, visit ni.com/environment/weee.
RoHS
National Instruments
- (RoHS) National Instruments RoHS ni.com/environment/rohs_china
- (For information about China RoHS compliance, go to ni.com/environment/rohs_china.)
Worldwide Support and Services
The NI website is your complete resource for technical support. At ni.com/support you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers.
Visit ni.com/services for NI Factory Installation
Services, repairs, extended warranty, and other services.
Visit ni.com/register to register your NI product.
Product registration facilitates technical support and ensures that you
receive important information updates from NI.
A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European Communities using the manufacturer’s declaration of conformity. This system affords the user protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC for your product by visiting ni.com/certification. If your product supports calibration, you can obtain the calibration certificate for your product at ni.com/calibration.
NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. NI also has offices located around the world. For telephone support in the United States, create your service request at ni.com/support or dial 1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the Worldwide Offices section of ni.com/niglobal to access the branch office websites, which provide up-to-date contact information, support phone numbers, email addresses, and current events.
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patents Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product.
Refer to the Export Compliance Information at ni.com/legal/export- compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2010–2017 National Instruments. All rights reserved.
References
- Test and Measurement Systems, a part of Emerson - NI
- Test and Measurement Systems, a part of Emerson - NI
- Product Certifications - NI
- Dimensional Drawings - NI
- Engineering a Healthy Planet - NI
- Contact Us - NI
- Log In - National Instruments
- NI Services - NI
- Support - NI
- NI Trademarks and Logo Guidelines - NI
- Product Certifications - NI
- Dimensional Drawings - NI
- Engineering a Healthy Planet - NI
- Product Take-Back Program and Recycling - NI
- Contact Us - NI
- Log In - National Instruments
- NI Services - NI
- Support - NI
- Manual-Hub.com – Free PDF manuals!
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>