HUMAN CARE 55200H Portable Ceiling Lift User Manual

June 16, 2024
HUMAN CARE

HUMAN CARE 55200H Portable Ceiling Lift

HUMAN-CARE-55200HPortable-Ceiling-Lift-PRODUCT

Product Information

Specifications

  • Lift Speed: Start/stop
  • Weight Capacity: 150 kg/330 lbs, 220 kg/484 lbs, 300 kg/661 lbs
  • Dimensions: 235x140x300mm
  • Weight: 7.4 kg
  • Battery: 2x12V, 2.3Ah
  • Power Source: Internally powered equipment
  • Water Resistance: IPX4
  • Waterproof Rating: IPX7

Product Description
The Altair is a lifting device designed to assist with lifting and transferring individuals with limited mobility. It is available in three weight capacity options: 150 kg/330 lbs, 220 kg/484 lbs, and 300 kg/661 lbs. The lift features a compact design with dimensions of 235x140x300mm and weighs 7.4 kg.

Product Usage Instructions

Suspending the Lift from the Rail
To suspend the lift from the rail system, follow these steps:

  1. Ensure that the rail system is securely installed and meets the necessary safety standards.
  2. Position the lift beneath the rail system.
  3. Engage the lifting mechanism and slowly raise the lift until it is securely suspended from the rail.

Hand Control
The hand control is used to operate the lift and adjust its position. Follow these instructions to use the hand control:

  1. Ensure that the hand control is properly connected to the lift.
  2. Use the buttons on the hand control to control the lifting and lowering of the lift.
  3. Adjust the position of the lift as needed by using the directional buttons on the hand control.

Indicator
The indicator on the lift provides important information about the status and operation of the lift. Refer to the user manual for a detailed explanation of the indicator’s different modes and meanings.

Charger
The charger is used to recharge the lift’s battery. Follow these steps to use the charger:

  1. Connect the charger to a power source.
  2. Connect the charger’s cable to the charging port on the lift.
  3. Allow the lift to charge for the recommended amount of time specified in the user manual.

Data Communication
The Altair lift is equipped with data communication capabilities. This allows for easy integration with other systems and devices. Refer to the user manual for detailed instructions on how to set up and use data communication features.

Hanger Bar
The hanger bar is used to support and secure the individual being lifted. Ensure that the hanger bar is properly attached to the lift and securely supports the weight of the individual.

Lifting Height
The lifting height of the Altair lift can be adjusted to accommodate different transfer needs. Use the hand control or other control mechanisms to adjust the lifting height as required.

Altair

Important!
Always read the user manual before using your product and keep this manual for future use.

WARNING!” This symbol is used when particular attention is requi-red

Lifting and moving a person always entails some risk. You should therefore read the user manuals for the lift and the lifting accessories carefully. Always make sure that the access-ories you are using are intended for use with the lift. As a caregiver, you are responsible for the client’s safety and you must know whether or not the client will be able to cope with the lifting process. Human Care’s lifts are designed only to raise and lower clients in a vertical direction, and never diagonally or through angles. The lift must always be positioned directly above the Client when lifting. The Client must always be positioned above the surface onto which they are to be lowered.

WARNING : LIFTING OR LOWERING A CLIENT DIAGONALLY OR IN AN ANGLED MOVEMENT MAY RESULT IN INJURY TO THE CLIENT OR THE CAREGIVER.

If the product is used in any way that is not re-commended by Human Care, then Human Care will not accept responsibility for any injuries or accidents that may occur.  The product is CE-marked in accordance with the EU Directive 93/42/EEC (MDD) and/or EU Regula-tion 2017/745 (MDR), as class I medical device. The CE mark is on the product. This product has been tested in accordance with ISO 10535.

  • If no other agreement or approval exists, the owner of the product is responsible for any combinations using accessories or products from suppliers other than Human Care.
  • Anyone using the equipment must have been given adequate tuition and training in use of the lift and its accessories
  • Clients must not use the lift without an adult present. Never leave a Client alone in the lift.
  • Always contact your supplier if you have any questions or if anything is unclear.
  • Human Care’s products are continuously being developed and updated and we reserve the right to make design changes without prior notice.
  • Notice to user/patient in case of serious incident
  • Any serious incident that has occurred in rela-tion to the product, should be reported to the local contact, who reports to the manufacturer, and the competent authority of the country in which the user/patient is established.

Compliances and Standards
Human Care is an ISO 13485:2016 certified Swedish medical device company. The Quality Management System is in compliance with US 21 CFR part 820.

Safety instructions

Before using for the first time, make sure that:

  • you have read and understood the user ma-nual for the lift and all of its accessories
  • the lift has been installed in accordance with the installation instructions
  • the lift accessories are intended for use with the lift and are correctly fitted
  • Before use: charge battery upon arrival. To ensure battery capacity is maintained, the new ceiling lift must be charged for 8 hours immediately upon arrival before it is put in use or stored. This is because the ceiling lift may have been stored for some time and during that period not been charged regularly.
  • Check to see if there is a distance strap suspended from the rail or if there is a hook fitted to the rail roller. The recommended minimum height from the floor to the distance strap is 1850 mm.
  • Remove the lift and charger from the packaging.
  • Remove the seal from the lift belt after first reading the user manual.
  • Plug the charger in at an easily accessible wall socket.
  • Plug the hand control into the charger (see page 8). If the indicator light on the charger is yellow, charge until the light turns green.

Important! The lift is designed exclusively for lifting people, and only one at a time.

Important! Maximum 220 kg weight capacity when using two-point suspension with Altair.

Before use, always ensure that:

  • personnel who use the equipment have been given adequate tuition and training in the use of the lift and its accessories
  • the lift belt is not twisted or worn and it can move freely in and out of the lift
  • lift accessories, e.g. sling, distance strap, and rail roller, are showing no signs of wear or other damage
  • the lift is correctly suspended in the distance strap, rail roller, and rail
  • the right types, sizes, materials, and designs of accessories have been selected concerning the client’s safety and needs
  • the lift accessories have been correctly installed.
  • The sling’s strap loops and suspension must be checked once the loops are fully extended when lifting but before the client is lifted from the underlying surface
  • It is important to always check that the loops are at the bottom of all hooks (see picture below), securely placed below the locking pin.

Human Care provides suitable training in the safe operation of the lift and other accessories. Further information about slings, rail systems, and other accessories can be found at the website: www.humancaregroup.com.

Important! To ensure that the lift belt will not be damaged, the lift must be balanced during moving, lifting, and lowering operations. Make sure the lift bar is balanced, the hooks are set at the same length and the sling is sus-pended correctly. Never hold or pull the lift belt when moving a Client. It is also important that the lift belt is tensioned as it is wound in.

Warning! Modification of the product is not permitted without the manufacturer’s consent!

Important! If the lift is being used with a 2-point suspension method, the lift belt should not be fully retracted. Leave 15 cm of the lift belt visible.

Specification

Lifting capacity, max.:

55200H| 150 kg/330 lbs
55210H| 220 kg/484 lbs
55220H| 300 kg/661 lbs
Dimensions (WxHxL):| 235x140x300mm/9.2×5.5×11.8”
Weight of lift:| 7,4 kg/16.3lbs
Batteries:| 2x12V, 2.3Ah
Emergency lowering:| Electronic and mechanical
Class:| Internally powered equipment
Type:| B
IP rating lift:| IPX4
IP rating hand control:| IPX7
Noise level:| <65dB (loaded and unloaded)
Lifting speed:| 1.8 m/min / 3.0 m/min (unloaded only)
Lifting height:| 2120 mm/83,5”, 1720 mm/67,7”
Start/stop:| soft
Indicator:

Standby| flashing GREEN
Active| constant GREEN
Error| constant RED
Service| flashing YELLOW/GREEN, YELLOW/RED
Charging| constant YELLOW
Charger| Input 100-240V AC 50-60 Hz, 1.6A; Output 27.8V, 0.8A DC
IP rating charger:| IPX0
ETL:| Yes

This product is designed for repeated use.

Product Description

HUMAN-CARE-55200HPortable-Ceiling-Lift-FIG- \(2\)

  1. Hand control
  2. Lift belt
  3. hanger bar
  4. Leg hook
  5. Secondary control panel
  6. Emergency lowering button
  7. Indicator
  8. Emergency stop

Intended use
A portable ceiling lift provides you with the ability to move a Client safely and easily. The lift can be easily hung up and taken down, which means the same lift can be used in several different rooms or wards. The lift is used for lifting and moving a Client, for example to or from a bed, wheelchair, or toilet, or from the floor. The apparatus can also be used for walking training, and with the help of some accessories, for balance training and weighing.

The ceiling lift has been designed and tested for use indoors and is classified as an IPX4 product.

Important! The product must not be immersed in water.

The climatic conditions should be as follows: ambient temperature from 0 °C to 40 °C, relative ambient humidity from 30% to 80%, and ambient pressure from 790 hPa to 1060 hPa.

Product label
The product label contains information about the lift. This label is located underneath the lift.HUMAN-CARE-55200HPortable-Ceiling-Lift-FIG-
\(3\)

Wireless transmission equipment, mobile telephones, etc. must not be used close to the device as it may affect its function. Particular caution must be observed when using strong sources of interference such as diathermy and such, to ensure that no diathermy cables are placed on or near the device. If in doubt, ask the person in charge of the equipment or the supplier.

Symbols on the unit and product label

HUMAN-CARE-55200HPortable-Ceiling-Lift-FIG- \(4\) HUMAN-CARE-55200HPortable-Ceiling-Lift-FIG- \(5\)

Functions

Emergency stop
In an emergency, the emergency stop can be activated by pressing the RED button. To reset the emergency stop, turn the RED button clockwise.

Emergency lowering
In the event of an emergency, the emergency lowering function can be used. Press the emergency lowering button and down button for at least 3 seconds simultaneously to activate the emergency lowering. Older lifts (before July 2016) may double-keyable-key commands and delays. An audio signal indicates that the lift is lowering. Always ensure that emergency lowering can be performed as safely as possible for the client. The emergency lowering button only works when the lift is switched on and the emergency stop has been deactivated. When the emergency lowering is active, the end position switches are inactive. If the button is pushed until the lift is in its lowest position, the lift will start to go up again. Always stop at the lowest position.

The emergency lowering button must only be used in emergencies.

Safety catch
The lift is equipped with a safety catch. This prevents involuntary falls. If the lift belt is pulled out too quickly, the safety catch will engage, thus preventing the fall. Slacken the lift belt to unlock the catch. With the safety catch engaged, the lever (10) can never be turned, as this blocks the lowering function of the lift!

Mechanical emergency lowering
During power loss and the hoist is loaded, a user in the hoist, the mechanical emergency lowering is activated and the hoist is lowered safely and reliably.

Overheating
The lift has overheating protection which stops the motor if it gets too hot. Overheating can occur if the motor is overloaded or operated continuously for quite a long time. The lift’s work cycle (work: pause) is 15:85. Maximum continuous operation is two minutes. If this is exceeded, there is a risk of the motor overheating.

Belt tension guard
The lift is equipped with a belt tension guard that prevents unintentional discharging of the lift belt. The lift must be loaded to be able to pull the belt out. This means that when the lift is lowered to a surface, the lift belt is only fed out if it is pulled upwards while the DOWN button on the hand control or the control panel is held pressed in. The purpose of this is to pull out enough of the lift belt to be able to hook the lift onto the suspension hook or to unhook it.

Two Speeds
When the lift is loaded with less than approx. 40 kg it can be run at two speeds. If the Up or Down button is pressed for longer than 3 seconds, the lift will run in the faster mode. If the lift is loaded with more than approx. 40 kg it will only run in the slower mode.

Suspending the lift from the rail

  • Place the lift on a suitable surface. Feed out enough of the belt by pulling gently on the hook on the lift belt (2) while at the same time keeping the Down button on the hand control or the control panel of the lift pressed in. Hook onto the hook on the distance strap hanging from the rail roller diagonally from above (see fig)
  • Switch on at the mains and the lift is ready for use.HUMAN-CARE-55200HPortable-Ceiling-Lift-FIG- \(6\)

Hand control

HUMAN-CARE-55200HPortable-Ceiling-Lift-FIG- \(7\)

  • The hand control is equipped with an Up button (1) and a Down button (2) for raising and lowering the lift.
  • When the lift is loaded with less than approx. 40 kg it can be run at two speeds. If the Up or Down button is pressed for longer than 3 seconds, the lift will run in the faster mode. If the lift is loaded with more than approx. 40 kg it will only run in the slower mode.
  • On the short side of the hand control, there is an input for charging the lift and an input (USB-C) for communicating with the lift.
Secondary control panel

HUMAN-CARE-55200HPortable-Ceiling-Lift-FIG- \(8\)

  1. The lift is equipped with a secondary control panel located on the lift which can be used as an alternative to the hand control. It is equipped with an Up button (2) and a Down button (3) for raising and lowering, an OFF/ON button (1) for activation/deactivation of the lift, and an emergency lowering button (4) for quick lowering of the lift.
  2. Press the emergency lowering button and down button for at least 3 seconds simultaneously to activate the emergency lowering. Older lifts (before July 2016) may not have double-key commands and delays.

Indicator
The indicator light has the following indicators, which are only accessible when the lift is switched on.

HUMAN-CARE-55200HPortable-Ceiling-Lift-FIG- \(9\)HUMAN-CARE-55200HPortable-Ceiling-Lift-FIG- \(11\)

Charger
The indicator on the lift flashes red when the batteries need recharging. Plug the charger cable into the hand control or put the hand control into the charger cradle and charging will start. When the lift is charging the indicator glows yellow.

Important! The lift can only be used with a charger recommended by Human Care, article number 50880.

  • Charge the lift regularly, preferably every night
  • The charger must be disconnected when you use the lift
  • The lift’s function is blocked during charging. Disconnect the charger to use the lift
  • The unit charges regardless of the position of the main switch
  • If the emergency stop is activated the lift cannot be charged. Check that the emergency stop is not activated when charging.
  • The charger must only be connected to an earthed socket.

Important! When the batteries reach a critically low battery level, it will no longer be possible to perform more lifts (up or down). However, there is still enough battery power for emergency lowering. See the separate manual for the charger, 99599.

Detaching the lift from the rail system

  • Lower the lift so that it stands solidly on a surface. The lift will stop automatically when it reaches the surface. It may be useful to use Human Care’s cart (Art. no. 50346). Feed more belts out by pulling gently on the hook on the lift belt while at the same time pressing the Down button on the hand control or control panel.
  • When the belt is slack it can be easily unhooked from the distance strap by lifting the belt loop upwards.
  • Press the Up button on the lift to move the belt back. Make sure the belt is kept taut and gets rolled up in the lift straight, without folding.

Important! When the lift is taken down, the belt must be kept taut while it is fed into the lift.

Using the lift
Press the Up button on the hand control or control panel on the lift to lift the Client. Press the Down button on the hand control or control panel to lower the Patient. Pressing the button activates a soft start, and releasing it causes a soft stop.

Data communication

  • Altair can communicate with a standard PC (Win 7/Win 8). This communication makes it possible to obtain statistics and other information on the lift.
  • To communicate with the ceiling lift you need the Human Care Data and Service Application, software that can be downloaded from the Human Care website, www.humancaregroup.com.
  • Connect the lift to your PC with a standard USB-C cable. Start the Human Care Data and Service Application and follow the instructions. The lift must be switched on for communication to be possible. The lift is deactivated while communication takes place.

Hanger bar

The following information can be read from the lift

  • Article number, serial number, and production date
  • Max. user weight
  • Software version
  • Service dates
  • Periodic usage statistics
  • Error log

Connected computer equipment must fulfill the requirements of IEC 60950-1 or IEC 60601-1 The latest software for the lift should always be used. You can find the latest version on our website. The hanger bar is available as an accessory on lifts, part no: 55751 Hanger Bar Altair/Roomer-S 2P 45 cm 55764 Hanger Bar Altair/ Roomer-S 2P 55 cm

Important! Do not hold on to the bar or the hooks when performing a lift.

Lifting height
The lifting height corresponds to the difference between the highest and lowest position the lift can reach. This corresponds to the length of the lift belt. The length of the lift belt is 1720 mm/66.1”(300 kg), 2120 mm (150/220 kg). If you want to extend the lift belt there are distance straps in various lengths available as extras. The lifting height of the lift is not affected by the length of the distance strap. The total height of the system consists of the rail in which the lift is mounted, the rail roller in the rail, a distance strap if relevant, and the lift.

Accessories
If the maximum weight for any of the accessories, e.g. slings, lift bars, and any other accessories is NOT the same as for the lift, then the lowest max. weight given for any one product is the one that shall apply. Always check the label on the lift and lifting accessories or contact Human Care if you have any questions or if anything is unclear.
Human Care recommends using only Human Care slings with Human Care lifts. See our website for all accessories.

Troubleshooting

Description Indicator Action

Nothing happens when you press the Up button or the Down button on the hand control or the control panel| Not lit| 1)  Check that the main switch is on.

2)  The batteries are fully discharged. Charge the lift immediately

3)  The emergency stop is pushed in. Deactivate by turning the red button clockwise.

Nothing happens when you press the Up button or the Down button on the hand control| Flashing green Flashing yellow Flashing red| Check the hand control. If the hand control is defective it needs to be replaced. Contact a service technician.
Nothing happens when you press the Up button or the Down button on the hand control or the control panel.| Constant red| 1) The lift motor is overheated. Wait a while and try again.

2)  Lift is overloaded. Lower using the emergency lowering button

Nothing happens when you press the Up button or the Down button on the hand control or the control panel.| Constant yellow| Charging in progress. Interrupt charging.
Nothing happens when you press the Up button on the hand control or the control panel.| Flashing green Flashing yellow Flashing red| The lift is in its maximum upper position. Press the Down button on the hand control or the control panel to lower the lift.
Nothing happens when you press the Down button on the hand control or the control panel.| Flashing green Flashing yellow Flashing red| The lift is in its maximum lower position. Press the Up button on the hand control or the control panel to raise the lift.
The belt is not fed out when the lift is on a surface.| Flashing green Constant green Flashing yellow Flashing red| The lift belt has to be pulled upwards at the same time as the Down button on the hand control or the control panel is held pressed in.
The indicator switches from flashing green or yellow to flashing red.| Flashing red| The batteries need charging but you can still lift the Patient a few times. The lift should be set to charge as soon as possible.
The indicator switches from flashing green or yellow to flashing red.| Flashing red| The batteries need charging but you can still lift the client a few times. The lift should be set to charge as soon as possible.

Care instructions for the lift and accessories

Care instructions for the lift

  • The lift should be recharged every night
  • Wipe the ceiling lift using a slightly dampened drying cloth. Washing up liquid can be used
  • Never wash the ceiling lift using water or any other fluids.
  • NEVER use strong detergents on the ceiling lift.
  • If the lift needs to be disinfected, DuPontTM Virkon® can be used
  • The lift belt must be washed and disinfected as instructed above
  • Activate the emergency stop if the ceiling lift is not to be used for a few days. Otherwise, the batteries will run down. The batteries must be charged at least every 12 months.
  • The lift must always be stored or moved lying horizontally

Annual inspection
The lift must be inspected at least once a year by someone with Human Care qualifications and by Human Care’s directive. Pay particular attention to parts that show wear. The lift’s service indicator will flash yellow when a service needs to be performed.

Daily inspection

  • Always check the lift belt and sling before use. There must be no loose seams or other signs of wear to lift belts, belts, and fabric. If there are, replace the sling or lift belt.
  • Check the ceiling lift for visible signs of damage
  • Check that the lift is correctly mounted in the distance strap or rail roller

Monthly inspection
Check the lift belts for damage or wear and tear. Feed out enough of the lift belts to allow their entire length to be checked. Contact a retailer if the belt is damaged. The lift belt should be replaced every second year.

Service
Installation, service, and maintenance must only be performed by personnel with Human Care qualifications and by Human Care’s directive. Only original spare parts may be used.

Service agreements
Human Care invites you to sign a service agreement for annual inspection and test loading.

Transport and storage
During transport, or when the lift is not going to be used for some time, the mains switch must be off or the emergency stop button pushed in, otherwise, the batteries will discharge. The lift must be transported and stored lying horizontally. The climatic conditions should be as follows: ambient temperature from 0 °C to 40 °C, relative ambient humidity from 30% to 80%, and ambient pressure from 790 hPa to 1060 hPa.

Product Lifetime
The product´s expected lifetime is ten (10) years or 10.000 lifts if the product is used as intended and maintained according to the manufacturer´s instructions, depending on the intensity of use and maximum load applied during use. If the product label is no longer legible, the product should be discarded.

Warranty and support
If you need information or support, please contact your Human Care office,

Guidance and manufacturer’s declaration electromagnetic emissions

The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the user of the Human Care lift should ensure that it is used in such an environment.
Emission test| Compliance| Electromagnetic environment – guidance
RF emissions CISPR 11| Group 1| The Human Care lift uses RF energy only because its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The Human Care lift is suitable for use in all establishments including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations/flicker emissions IEC

61000-3-3

| Complies
Guidance and manufacturer’s declaration electromagnetic immunity

The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the user of the Human Care lift should ensure that it is used in such an environment.
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic environment guidance
Electrostatic discharge (ESD) IEC 61000-4-2| +/- 6 kV contact

+/- 8 kV air

| +/- 6 kV contact

+/- 8 kV air

| Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material,

the relative humidity should be at least 30 %.

Electrical fast transient / Burst

IEC 61000-4-4

| +/- 2 kV for power supply lines

+/- 1 kV for input/output lines

| +/- 2 kV for power supply lines

n/a. for input/output lines

| Mains power quality should be that of a typical commercial or hospital environment.
Surge

IEC 61000-4-5

| +/- 1 kV differential mode

+/- 2 kV common mode

| +/- 1 kV differential mode

n/a. for common mode

| Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions

and voltage variations on power supply input lines IEC 61000-4-11

| <5 % UT

(>95 % dip in UT) for 0,5 cycle

40 % UT

(60 % dip in UT) for 5 cycles

70 % UT

(30 % dip in UT) for 25 cycles

<5 % UT (>95 % dip in UT))

for 5 sec

| <5 % UT

(>95 % dip in UT) for 0,5 cycle

40 % UT

(60 % dip in UT) for 5 cycles

70 % UT

(30 % dip in UT) for 25 cycles

<5 % UT

(>95 % dip in UT)) for 5 sec

| Mains power quality should be that of a typical commercial or hospital environment. If the user of the Human Care lift requires continued operation during power mains interruptions, it is recommended that the lift be powered from an uninterruptible power supply or battery.
Power frequency (50/60 Hz) magnetic field

IEC 61000-4-8

| 3 A/m| 3 A/m| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment
NOTE UT is the a.c. mains voltage before application of the test level.
Guidance and manufacturer’s declaration electromagnetic immunity

The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the user of the Human Care lift should ensure that it is used in such an environment.
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic environment guidance
Conducted RF IEC 61000-4-6

Radiated RF IEC 61000-4-3

| 3 Vrms

150 kHz to 80 MHz

3 V/m

80MHz to 2,5GHz

10 V/m

800MHz to 2,5GHz

| 3 Vrms

3 V/m

10V/m

| Portable and mobile RF communications equipment should be used no closer to any part of the Human Care lifts, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distance

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. b

Interference may occur in the vicinity of equipment marked with the following symbol.

  • NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
  • NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcasts, and TV broadcasts cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Human Care lifts are used exceeds the applicable RF compliance level above, the Human Care lifts should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Human Care lifts.

Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

Recommended separation distances between portable and mobile RF communications equipment and the Human Care lifts.

The Human Care lifts are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Human Care lifts can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Human Care lifts as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of the transmitter

W

| Separation distance according to frequency of transmitter m
150 kHz to 80 MHz d = 1,2√P| 80 MHz to 800 MHz d = 1,2√P| 800 MHz to 2.5 GHz d = 0,7√P
0.01| 0.12| 0.12| 0.07
0.1| 0,37| 0,37| 0,22
1| 1,16| 1,16| 0,7
10| 3,67| 3,67| 2,21
100| 11.6| 11.6| 7
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

Recycling
Once the product has reached the end of its life, it should be disposed of according to your country’s rules, especially for the separate collection of electrical and electronic products. Correct disposal helps prevent negative consequences for the environment and human health.

99701, Rev15, ML, User Manual, Altair, Nov23

FAQ’s

What is the weight capacity of the Altair lift?

The Altair lift is available in three weight capacity options: 150 kg/330 lbs, 220 kg/484 lbs, and 300 kg/661 lbs.

How do I charge the Altair lift?

To charge the Altair lift, connect the charger to a power source and then connect the charger's cable to the charging port on the lift. Allow the lift to charge for the recommended amount of time specified in the user manual.

Can the Altair lift be used with other systems or devices?

Yes, the Altair lift is equipped with data communication capabilities, allowing for easy integration with other systems and devices. Refer to the user manual for detailed instructions on how to set up and use data communication features.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals