REED SL-417 Sound Adapter Instruction Manual
- June 16, 2024
- REED
Table of Contents
SL-417 Sound Adapter
Instruction Manual
SL-417 Sound Adapter
Model SL-417 Sound Adapter
Instruction Manual
Features
- Works in cooperation with the SD-9300 Environment Meter to take professional Sound Level measurements
- 30 to 130 db, three ranges
- Frequency weighting networks meet IEC 61672 class 2 standards
- Fast time weighting dynamic characteristic modes
- Condenser microphone for high accuracy and long-term stability
- Over/Under range indicator
Specifications
Measuring Range:| Range1:30to80db;
Range.2:50to100db;
Range3:80to130db
---|---
Output:| 10mVper1db
Accuracy:| 31.5Hz:±3.5dB
63Hz:±2.5dB
125Hz:±2.0dB
250Hz:±1.9dB
500Hz:±1.9dB
1KHz:±1.4dB
2KHz:±2.6dB
4KHz:±3.6dB
8KHz:±5.6dB
Frequency:| 31.5Hzto8,000Hz
Microphone Type:| Electric condenser
Operating Temperature:| 0to50ºC(32to122ºF)
Operating Humidity:| >80%
Power Supply:| OneDC9Vbattery
Dimensions:| Microphone:12.7x58mm.(0.5×2.28”).
Instrument:107x53x29mm.(4.2.x.2.1.x.1.1”)
Weight:| 220g (0.49lb)
Instrument Description
- Microphone
- Calibration VR
- Power Indicator
- Over/Under Range Indicator
- Range/Power Off Switch
- Battery Cover/Compartment
- Output plug
Operating Instructions
- Connect the Output Plug to the Input Terminal on the SD-9300 Environment Meter.
- Determine the proper measuring range by using the Range Switch to minimize the tolerance of the readout. If the Over/Under Range Indicator is RED, the reading is over range and you need to select a higher range on the Range Switch. If the Over/Under Range Indicator is GREEN, the reading is under range and you need to select a lower range on the Range Switch.
- Point the microphone to the sound source, and the sound level will be displayed in dB on the SD-9300.
Note: Keep the Microphone dry and away from high vibrations
Battery Replacement
When the Power Indicator does not light up when you turn the meter on, you need to change the batteries.
- Open the Battery Cover and remove the battery.
- Replace the battery with a new 9V and replace the cover.
For service on this or any other REED product or information on other
REED products, contact REED Instruments at
info@reedinstruments.com
Notes _ ____
REED Instruments
1-877-849-2127
info@reedinstruments.com
www.reedinstruments.com