Lazada TC-8200 Air Quality Dust Particle Counter User Manual
- June 16, 2024
- Lazada
Table of Contents
- TC-8200 Air Quality Dust Particle Counter
- Overview:
- Safety and maintenance
- structure appearance
- Button function
- Operation description:
- Counting and weighing mode conversion:
- Record data interval time setting:
- Steps to enter the function parameter setting:
- Technical specifications conventional
- Common problem analysis
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
TC-8200 Air Quality Dust Particle Counter
User Manual
TC-8200 Air Quality Dust Particle Counter
Dust Particle Counter Manual
Overview:
TC-8200 This product adopts the principle of laser scattering, can detect the
number and quality of 0.3um, 0.5um, 1.0um, 2.5um, 5.0um, 10um particle size
dust, the detection of clean room and below clean room levels, temperature,
humidity and AQI Grade detector.
It has 999 groups of data records, can view the data curve, with 3.2-inch TFT
full-color display, automatic measurement, real-time date and time,
rechargeable lithium battery or separate external USB power supply, stable
data, etc.
Applied to:
Class 100 clean room and below clean room inspection
Industrial production of electronics, optics, chemicals, food, medicine and
health, biological products, aerospace, etc.
Filter seal inspection
Particle pollution source detection
Safety and maintenance
- Do not use the instrument in a dusty or corrosive gas environment, so as not to reduce the life of the instrument or damage it
- When the battery icon on the display is empty or red, please charge it in time, and remove the battery if it is not used for a long time.
- Do not store or use the meter in a high temperature, high humidity, flammable, explosive and strong electromagnetic field environment.
- For maintenance, use a soft cloth and neutral detergent to clean the case. Do not use abrasives and solutions to prevent the case from corroding and damaging the instrument.
structure appearance
Function:
- Display screen
- Operation buttons
- USB interface
- Air outlet
- Battery compartment
- Air intake
- Temperature and humidity sensor
- Support frame nut
Note: 1. Do not block the air inlet with objects during the test
Button function
- Power on/off button: short press to power on or off
- OK button: 1) Historical trend interface press this button to exit to the main interface
- The history record interface can enter the page number view
- The function parameter interface can enter the parameter modification
- The alarm value setting can be exited to the main interface
- Mode button: 1) Short press to turn on the weighing mode or counting mode
- Shift page position of history record interface
- Parameters can be modified in the function parameter interface
- Displacement value of alarm value setting interface
- Setting button: 1) Short press the history record → function parameter → main interface
- Shift page position of history record interface
- Parameters can be modified in the function parameter interface
- Displacement value of alarm value setting interface
- Alarm value setting button: 1) Short press to enter the alarm value setting, press this button again to increase the alarm value
- The number of pages that can be turned up in the historical trend interface
- The number of pages that can be turned up in the history record interface
- The function parameter interface can go up to select parameter items 2)
- Temperature unit conversion button: 1) Short press to turn on the ℃ or ℉ unit
- The number of pages that can be turned down in the historical trend
interface
3)The number of pages that can be turned up in the history record interface
4)Function parameter interface can select parameter items down
- Historical trend button: 1) Short press to enter the historical trend interface
- Sound alarm button: 1) Short press to turn on or off the alarm sound
Operation description:
Power on or off:
Short press Button to turn on or off
Counting and weighing mode conversion:
Short press Button to switch between ug/m3 or Piece/L mode
Temperature unit conversion
Short press Button switch ℃/ ℉
Record data interval time setting:
Press during automatic measurement Press the button twice to enter the function parameter interface, Press Button to confirm to enter the modification, Press at this time Button to select [recording interval] item Press or Button to enter time selection Press Button to select the required interval time After the selection is complete, press Button twice Press again Button to return to the main interface。
History query:
Press Button once Press Go to
Steps to enter the function parameter setting:
Press Button twice and press Enter to
Backlight: Button to switch brightness25%-50%-75%-100%
Data Clear: Button to enter the data clearing menu, press again Button to
confirm clear, Button exit clear
Time Rev : Button to enter the time setting menu, press again The button
can be shifted, The button can modify the time, press Button to confirm the
modification, press again Button to exit modification , once again Button
and return to the main interface
Auto shutdown: Button to select OFF (no automatic shutdown) / 15 minutes
/ 30 minutes / 45 minutes / 1 hour / 2 hours / 4 hours / 8 hours.
Record interval : Button to enter the recording interval setting, press
Button to select the time recording interval, press when finished Button to
exit the interval interface, press again Button, then press Button to return
to the main interface
language settings: Button to switch Chinese/English
Factory settings: Button to enter to restore the original factory
parameter settings Press again Button to confirm reset ,
After reset, all will be restored to factory parameters
Technical specifications conventional
Particle detection
range| 0.3~10gm| unit
---|---|---
Six-channel particle size| 0.3, 0.5, 1.0, 2.5, 5.0, burn| urn
PM2.5 Range| 0-.400014/m’| µg/m³
PM2.5 Precision| 0-100gg/m³: ±10m/m³ 100-1000gg/m³: ±10%| µg/m³
Concentration resolution| 1gg/m³| µg/m³
Number of particles output range| 5 2000000 Granule/L| Granule/L
flow| 1 liter/min (2.83L=1L2.83)| L/min
Cleanliness standard| meets the ISO 14644-11505 level-1S09 level| level
Response time| 1| second
---|---|---
Boot stabilization time| 10| second
Temperature measurement range| -10^-60°C| °C
Temperature accuracy| ±2°C| °C
Temperature resolution| 0.01°C| °C
Humidity measurement range| 0^-100%| %
Humidity accuracy| ±3%| %
Humidity resolution| 0.01%| %
Power supply| 3.7VRemovable 18650 lithium battery| V
Charging voltage| SV/USB| V
length of work| 2000mAhCan work continuously> 9 hours| hours
data record| 999group| group
weight| 320g| g
size| 22280*46mm| mm
Common problem analysis
-
There is no data to show the number and concentration of particulate matter in the air
Analysis 1: The battery power is too low to start the sensor work -
Doesn’t boot
Analysis 1: If the battery is dead or damaged, use the USB port to supply power. If it can be turned on and there is a change in the power grid in the upper right corner, the battery is dead. If there is no change, only the OK icon will be displayed.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>