Dynalabs DYN-C-3000-SI Triaxial Capacitive Accelerometers User Manual
- June 16, 2024
- Dynalabs
Table of Contents
Dynalabs DYN-C-3000-SI Triaxial Capacitive Accelerometers
Model DYN-C-3000-SI Range [g]: 3, 5
Product Manual
Product Support
If at any time you have questions or problems with the DYN-C-3000-SI sensors, please contact a Dynalabs engineer at:
Phone: +90 312 386 21 89 (9 a.m. to 5 p.m., UTC +3)
E-mail: info@dynalabs.com.tr
Warranty
Our products are warranted against defective materials and workmanship for one year. Defects arising from user errors are not covered by the warranty.
Copyright
All copyrights of this manual belonging to Dynalabs products are reserved. It
cannot be reproduced without written consent.
Disclaimer
Dynalabs Ltd. provides this publication “as is” without warranty of any kind,
express or implied, including but not limited to, the implied warranties of
merchantability or fitness for a particular purpose. This document is subject
to change without notice, and should not be construed as a commitment or
representation by Dynalabs Ltd.
This publication may contain inaccuracies or typographical errors. Dynalabs
Ltd. will periodically update the material for inclusion in new editions.
Changes and improvements to the product described in this manual may be made
at any time.
Introduction
Capacitive accelerometers are based on proven micro-electro-mechanical systems
(MEMS) technology. These capacitive accelerometers are reliable and long-term
stable. They have a DC response. The advantage of these sensors is their
outstanding temperature stability, high frequency response and low noise-high
resolution. These sensors have a reliable aluminum housing with IP68
protection class.
Dynalabs 3000SI series triaxial accelerometers provide an ultra-low noise
performance from 0.7 to 1.2 μg/√Hz. These accelerometers provide excellent
bias and scale factor stability and a wide frequency range (±3dB) from 550 Hz
to 700 Hz.
DYN-C-3000-SI sensors offer the following options;
- Custom Cable Length (5m standard cable)
- Custom Housing Material
- Custom Connector
- Base plate
General Information
Unpacking and Inspection
Dynalabs products provide adequate protection for undamaged products to be
transported. Document the damages that occur indirectly during the transport
and contact the customer representative.
System Components
The DYN-C-3000-SI has the following components:
- MEMS Sensor
- Calibration Certificate
- Product Manual
Specifications
Table 1: Specifications datasheet
Full-scale acceleration| (g)| 3003SI ± 3| 3005SI ±
5
---|---|---|---
White Noise| (μg/√Hz)| 0.7| 1.2
Noise (Integrated over 0.1Hz to 100Hz)| (μg)| 8| 13
Dynamic range (0.1Hz to 100Hz)| (dB)| 108.5| 108.5
Scale Factor Sensitivity| (mV/g)| 900| 540
Bandwidth (±3dB)| (Hz)| 550| 700
Operating power consumption| (mW)| 90| 90
Environmental
Table 2 Environmental specifications datasheet
Protection Level | IP 68 |
---|---|
Operating Voltage | 6 V – 40 V |
Operating Temperature | -40 °C to +100 °C |
Isolation | Case isolated |
Physical
Table 3 Physical specifications datasheet
Sensing Element | MEMS Capacitive |
---|---|
Housing Material | Aluminum or Steel |
Connector (Optional) | D-Sub 9 or 15 pin, Lemo, Binder |
Mounting | Adhesive or screw mount |
Base plate (Optional) | Aluminum or Steel |
Weight (without cable) | 28 g (aluminum) |
55 g (steel)
Outline Drawing
The dimensional properties of DYN-C-3000-SI sensors are given below.
Technical Drawings
Operation and Installation
General
The general sensor connector configuration is given below; Cable Code/Pin
Configuration:
- Red : V + Power supply voltage +6 to +40 VDC
- Black : Ground Power GND
- X-Axis: Yellow : Signal (+) Positive, analog output voltage signal for differential mode
- Purple : Signal (-) Negative, analog output voltage signal for differential mode
- Y-Axis: Blue : Signal (+) Positive, analog output voltage signal for differential mode
- Green : Signal (-) Negative, analog output voltage signal for differential mode
- Z-Axis: White : Signal (+) Positive, analog output voltage signal for differential mode
- Orange : Signal (-) Negative, analog output voltage signal for differential mode
WARNING
- Never connect the power supply and/or the power ground to yellow, purple, blue, green, white and/or orange cables.
- Never connect the power supply to the power ground. Always use a clean power source and check the voltage range.
Sensor Static Calibration Verification
- Using gravity, voltage values are measured in the + and – gravity directions, providing a value of ±1 g. The measurement should be made as follows;
- When the acceleration value of 3000SI series sensors is entered into the data acquisition system, the sensor shows +1 g with the effect of gravity, which is in the direction of the axis to be calibrated.
- When the sensor is positioned in the opposite direction to the axis to be calibrated, the arrow shows -1g as shown below under the effect of gravity.
- Using gravity, the voltage values that provide 1 g in the + and – directions are measured and compared with the catalog value. The calibration value should be close to the catalog value with 10% tolerance. Sensor catalog sensitivity values are given in Table 1.
Declaration of Conformity
This declaration of conformity is issued under the sole responsibility of the manufacturer. The product(s) are developed, produced and tested according to following EC- directives:
- 2014/35/EU – Low Voltage Directive (LVD)
- 2006/42/EU – Machinery Safety Directive
- 2015/863/EU – RoHS Directive
Applied standards
- EN 61010-1:2010
- EN ISO 12100:2010
- MIL-STD-810-H-2019 (Test Methods: 501.7- High Temperature, 502.7- Low
Temperature, 514.8- Vibration, 516.8 – Shock)
DYNALABS MÜHENDİSLİK SANAYİ TİCARET LİMİTED ŞİRKETİ declares that above
mentioned products meet all the requirements of the above mentioned standards
and regulations.
Canan Karadeniz, General Manager
Ankara, 15.07.2021
Read User Manual Online (PDF format)