CALIBRATION PROCEDURE
NI PXIe-6614
This document contains the performance verification and adjustment
procedures for the National Instruments NI PXIe-6614. For more information
about calibration, visit ni.com/calibration.
Software
Calibrating the NI PXIe-6614 requires NI-DAQmx version 9.8 or later on the
calibration system.
You can download NI-DAQmx from ni.com/downloads.
NI-DAQmx supports a number of application development environments (ADEs) and
programming languages, including LabVIEW, Lab Windows™ /CVI ™ , Microsoft
Visual C++, Microsoft Visual Basic. Please refer to NI-DAQ Readme file for
more details by selecting Start»All Programs “National Instruments» NI-DAQ.
When you install NI-DAQmx, you only need to install support for the ADE or
programming language you intend to use.
Documentation
Consult the following documents for information about the NI PXIe-6614, NI- DAQmx, and your application software. All documents are available on ni.com and help files install with the software.
| DAQ Getting Started Guide, Read Me First: NI-DAQmx, and DAQ Device
Installation Guide
Instructions for installation, hardware setup, and configuration of NI-DAQ
devices.
---|---
| NI 6614 User Manual
NI 6614 specific information.
| NI 6614 Specifications
NI 6614 specifications and calibration interval
| NI-DAQmx Readme
Operating system and application software support in NI-DAQmx
| NI DAQmx Help
Information about creating applications that use the NI-DAQmx driver
| LabVIEW Help
LabVIEW programming concepts and reference information about NI-DAQmx VIs and
functions
| NI-DAQmx C Reference Help
Reference information for NI-DAQmx C functions and NI-DAQmx C properties
| NI-DAQmx .NET Help Support for Visual Studio
Reference information for NI-DAQmx .NET methods and NI-DAQmx .NET properties,
key concepts, and a Centum to .Neptunium mapping table
Test Equipment
Table 1 lists the equipment recommended for the performance verification and
adjustment procedures. If the recommended equipment is not available, select a
substitute using the requirements listed in Table 1.
Table 1. Test Equipment
Equipment | Recommended Models | Requirements |
---|---|---|
10 MHz rubidium frequency standard clock or better* | Symmetrical 8040C or | |
Datum 8040 | ±0.75 ppb† accuracy or better | |
Shielded 68-conductor cable | SH68-68-D1 | Terminated with 68-pin female 0.050 |
series D-type connector
I/O connector block| BNC-2121, SCB-68A, or any 68-pin breakout box| —
BNC Male-to-BNC Male cable| —| 50 Ω
75 Ω BNC Male to 50 Ω BNC Female Adaptor| —| 50 to 75 Ω
- Allow adequate time for the clock to warm up prior to calibration.
† 1 ppb = 1/1000,000,000
Connecting the Clock Source to the NI PXIe-6614
- Connect the NI PXIe-6614 to the I/O connector block using the shielded 68-conductor cable.
- Connect the stable clock to PFI 39 on the NI PXIe-6614:
• If you are using the BNC-2121, attach the BNC adapter to PFI 39 on the BNC-2121 and then connect the stable clock source to it with the BNC cable.
• If you are using the SCB-68A, connect the stable clock to pin 2 (PFI 39) and then connect the ground of the stable clock to pin 36 (GND).
Test Conditions
The following setup and environmental conditions are required to ensure the NI PXIe-6614 meets published specifications.
- Keep connections to the NI PXIe-6614 as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
- Verify that all connections to the NI PXIe-6614 are secure.
- Use shielded coaxial cables or shielded copper wire for all cable connections to the NI PXIe-6614. Use twisted-pair wire to eliminate noise and thermal offsets.
- Maintain an ambient temperature of 23 °C ±5°C.
- Keep relative humidity between 5% to 95%, non-condensing.
- Allow a warm-up time of at least two hours to ensure the NI PXIe-6614 measurement circuitry is at a stable operating temperature.
Initial Setup
Refer to the DAQ Getting Started Guide, Read Me First: NI-DAQmx, and DAQ
Device Installation Guide for information about how to install the software
and hardware and how to configure the device in Measurement & Automation
Explorer (MAX).
Note When a device is configured in MAX, it is assigned a device name.
Each function call uses this device name to determine which device to
calibrate. This document uses Dev1 to refer to the device name. In the
following procedures, use the device name as it appears in MAX.
Complete the following steps to set up the NI PXIe-6614.
-
Install NI-DAQmx.
-
Power off the PXI Express chassis.
-
Install the NI PXIe-6614 into a PXIe compatible slot of the PXI Express chassis and then turn the chassis on.
Note Installing the NI PXIe-6614 into the system timing slot automatically overdrives the PXI backplane clock. Therefore, performing adjustments in the system timing slot will cause the PXI backplane clock to change during the adjustment process. Refer to the Using the OXCO to Overdrive the PXI Backplane Clock section in the NI 6614 User Manual for more information. -
Reboot the host computer and launch Measurement & Automation Explorer (MAX).
-
Right-click the device name and select Self-Test to ensure that the module is working properly.
-
Connect the Clock source to the NI PXIe-6614. Refer to the Connecting the Clock Source to the NI PXIe-6614 sections for instructions.
Verification
The following performance verification procedure describes measuring the 100
MHz base clock on the NI PXIe-6614.
Clock Accuracy Verification
Verifying the performance of the oscillator on an NI PXIe-6614 requires a pair
of counters, which are Counters 0 and 1.
Counter 0 generates an accurate 40 second pulse using the external clock
source. The pulse is then routed to Counter 1, which counts the number of
edges of the internal time base. The number of clock edges is then used to
compute the internal time base frequency.
Figure 1. NI PXIe-6614 Verification Process
To conduct the measurement, configure the counters as stated below:
-
Use the DAQmx Create Channel (CO – Pulse Generation – Time) VI and configure Counter 0 as follows:
a. Set the counter to generate a continuous time pulse high time of 40 seconds, with 1 second low time.
b. Set the initial delay to 0, idle state to Low, and the pulse unit to Seconds. -
Configure the CO channel properties as indicated in Table 2 using a DAQmx Channel property node.
Table 2. Configuration Settings for CO Channel Property| Value
---|---
CO.Ctr.TimebaseSrc| PFI39
CO.Ctr.TimebaseRate| 10000000
CO.Pulse.Term| Ctr0InternalOutput -
Configure Counter 1 using DAQmx Create Channel (CI – Count Edges) VI. Set the count direction as Count Up, clock edge type as Rising, and initialize the initial count to 0.
-
Set the CI.CountEdges.Term to 100MHzTimeBase using a DAQmx Channel property node.
-
Configure the CI trigger properties as indicated in Table 3 using a DAQmx Trigger property node.
Table 3. Configuration Settings for CI Trigger Property| Value
---|---
Pause.TrigType| Digital Level
Pause.DigLvl.Src| Ctr0InternalOutput
Pause.DigLvl.When| Low -
Use DAQmx Read VI to count the number of clock edges within the 40 seconds pulse.
-
Calculate the frequency accuracy of the base clock using the following equation:
The frequency measurement has an uncertainty that is equal to the sum of the stable clock error and the quantization error caused by the counter. The stable clock accuracy must meet the specifications listed in the Test Equipment section of this document. The quantization error, which is the inherent uncertainty in digitizing an analog value, is 10 ns (equivalent to 1 clock edge miscount). Using a measurement duration of 40 seconds gives a quantization error as follows:
The sum of the stable clock error 0.75 ppb and the quantization error 0.25 ppb
gives a total measurement uncertainty of 1 ppb. You can reduce the measurement
uncertainty by using a more stable clock source, increasing the measurement
duration, or both.
Note The NI PXIe-6614 counters are of 32-bits resolution. Counter
overflow can occur when the generated pulse is more than 42.9 seconds.
Note The time needed for verification is slightly more than the gate
pulse duration.
Compare the measured clock frequency obtained with the acceptable frequency
range shown in Table 4.
Table 4. Initial Performance Verification Test Limits
Device | *Test Limit (Typical)** |
---|---|
NI PXIe-6614 | ±85 ppb† |
- This test limit is a typical specification; NI does not warrant that your initial measurements are in accordance with this limit. To ensure the performance of the equipment, NI recommends that you perform the adjustment procedure and reverify the equipment using the specifications in Table 5.
† ±85 ppb covers the ±75 ppb limit per year and the ±10 ppb OCXO retrace error.
Adjustment is necessary if the measured value is not within the calibration
measurement target frequency range listed in Table 5.
Note Test limits in Tables 4 and 5 are based upon the 374078A-01 edition
of the NI 6614 Specifications. Refer to the most recent NI 6614 Specifications
online at ni.com/manuals.
Adjustment
The following performance adjustment procedures describes the sequence of
operation required to adjust the NI PXIe-6614. The adjustment process for the
NI PXIe-6614 takes approximately 8 minutes or less to complete.
Clock Accuracy Adjustment
The adjustment process iteratively measures the 100 MHz time base and adjusts
the hardware until the measured frequency is within the target limits listed
in Table 5.
Table 5. Adjustment Target Frequency Range
Device| Calibration Measurement Accuracy| Calibration Measurement
Target Frequency Range—As Left
---|---|---
Low Limit| High Limit
NI PXIe-6614| ±4.25 ppb| 99,999,999.575 Hz| 100,000,000.425 Hz
To adjust the frequency accuracy of the 100 MHz Clock, do the following:
Note Throughout the procedure, refer to the NI-DAQmx function call
parameters for the LabVIEW input values.
Refer to the NI-DAQmx Help or the NI-DAQmx C Reference Help for more
information on NI-DAQmx function calls.
- Open a calibration session on your device using the DAQmx Initialize External Calibration VI.
LabVIEW Block Diagram| NI-DAQmx Function Call
---|---
| Call DAQmxInitExtCal with the following parameters: device Name: dev1
password: The password required to open an external calibration session. The default password is NI.
Cal Handle: A reference to the calibration session - Perform an external calibration adjustment using the DAQmx Adjust TIO Time base Calibration VI.
LabVIEW Block Diagram| NI-DAQmx Function Call
---|---
| Call DAQmxAdjustTIOTimebaseCal with the following parameters:
Cal Handle: The session handle returned from DAQmxInitExtCal reference Frequency: 10000000.00
Note An error message is returned if DAQmx is unable to perform the
adjustment.
In this situation, verify that the external stable clock source is operating
normally and is properly connected to the NI PXIe-6614, and repeat the
adjustment process. If the error persists, contact National Instruments. Refer
to the Worldwide Support and Services section for contact information.
- Close the calibration session using the DAQmx Close External Calibration VI. This VI writes the calibration constants, date, time, and temperature to the internal calibration memory.
LabVIEW Block Diagram
| NI-DAQmx Function Call
---|---
|
Call DAQmxCloseExtCal with the following parameters:
Cal Handle: The session handle returned from DAQmxAdjustTIOTimebaseCal
action: Use DAQmx_VAL_EXT_CAL_COMMIT to update the internal calibration
memory. If you want to abort the operation without updating the calibration
memory, use DAQmx_VAL_EXT_CAL_CANCEL.
Note In LabVIEW, if the error in indicates that an error has occurred during adjustment, DAQmx does not update the internal calibration memory.
EEPROM Update
When an adjustment procedure successfully completes, the NI PXIe-6614 internal calibration memory (EEPROM) is immediately updated.
Reverification
Repeat the Verification section to determine the As-Left status of the device.
Note The reverification must be done immediately after adjustment without
power-cycling the device to ensure that no retrace error from OCXO is added to
the clock accuracy.
Note If any test fails Reverification after performing an adjustment,
verify that you have met the Test Conditions before returning your device to
NI. Refer to Worldwide Support and Services for assistance in returning the
device to NI.
Worldwide Support and Services
The National Instruments website is your complete resource for technical
support. At ni.com/support you have access to
everything from troubleshooting and application development self-help
resources to email and phone assistance from NI Application Engineers.
Visit ni.com/services for NI Factory Installation
Services, repairs, extended warranty, and other services.
Visit ni.com/register to register your National
Instruments product. Product registration facilitates technical support and
ensures that you receive important information updates from NI. National
Instruments corporate headquarters is located at 11500 North Mopar
Expressway, Austin, Texas, 78759-3504. National Instruments also has offices
located around the world. For telephone support in the United States, create
your service request at ni.com/support or dial 512
795 8248. For telephone support outside the United States, visit the Worldwide
Offices section of ni.com/niglobal to access the
branch office websites, which provide up-to-date contact information, support
phone numbers, email addresses, and current events.
Refer to the NI Trademarks and Logo Guidelines at
ni.com/trademarks for more information on National
Instruments trademarks. Other product and company names mentioned herein are
trademarks or trade names of their respective companies. For patents covering
National Instruments products/technology, refer to the appropriate location:
Help “Patents in your software, the patents.txt file on your media, or the
National Instruments Patents Notice at
ni.com/patents. You can find information about end-
user license agreements (EULAs) and third-party legal notices in the readme
file for your NI product. Refer to the Export Compliance Information at
ni.com/legal/export-compliance for
the National Instruments global trade compliance policy and how to obtain
relevant HTS codes, ECCNs, and other import/export data.
© 2013 National Instruments. All rights reserved.
1-800-915-6216
www.apexwaves.com
sales@apexwaves.com
All trademarks, brands, and brand names
are the property of their respective owners.
References
- Test and Measurement Systems, a part of Emerson - NI
- Test and Measurement Systems, a part of Emerson - NI
- Software and Driver Downloads - NI
- Product Documentation - NI
- Contact Us - NI
- Log In - National Instruments
- NI Services - NI
- Support - NI
- NI Trademarks and Logo Guidelines - NI
- Contact Us - NI
- Log In - National Instruments
- NI Services - NI
- Support - NI
- SCB-68A National Instruments Shielded I/O Connector Block | Apex Waves
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>